In an interdisciplinary project involving electronic
engineers and clinicians, a telemetric system was developed to measure
the bending load in a titanium internal femoral fixator. As this
was a new device, the main question posed was: what clinically relevant
information could be drawn from its application? As a first clinical
investigation, 27 patients (24 men, three women) with a mean age
of 38.4 years (19 to 66) with femoral nonunions were treated using the
system. The mean duration of the nonunion was 15.4 months (5 to
69). The elasticity of the plate-callus system was measured telemetrically
until union. Conventional radiographs and a CT scan at 12 weeks
were performed routinely, and healing was staged according to the
CT scans. All nonunions healed at a mean of 21.5 weeks (13 to 37).
Well before any radiological signs of healing could be detected,
a substantial decrease in elasticity was recorded. The relative
elasticity decreased to 50% at a mean of 7.8 weeks (3.5 to 13) and
to 10% at a mean of 19.3 weeks (4.5 to 37). At 12 weeks the mean
relative elasticity was 28.1% (0% to 56%). The relative elasticity
was significantly different between the different healing stages
as determined by the CT scans. Incorporating load measuring electronics into implants is a promising
option for the assessment of bone healing. Future application might
lead to a reduction in the need for exposure to ionising radiation
to monitor fracture healing.
Objectives. The monitoring of fracture healing is a complex process. Typically, successive radiographs are performed and an emerging calcification of the fracture area is evaluated. The aim of this study was to investigate whether different bone healing patterns can be distinguished using a telemetric instrumented femoral internal plate fixator. Materials and Methods. An electronic telemetric system was developed to assess bone healing mechanically. The system consists of a telemetry module which is applied to an internal locking plate fixator, an external reader device, a sensor for measuring externally applied load and a laptop computer with processing software. By correlation between externally applied load and load measured in the implant, the elasticity of the osteosynthesis is calculated. The elasticity decreases with ongoing consolidation of a fracture or nonunion and is an appropriate parameter for the course of bone healing. At our centre, clinical application has been performed in 56 patients suffering nonunion or fracture of the femur. Results. A total of 39 cases of clinical application were reviewed for this study. In total, four different types of healing curves were observed: fast healing; slow healing; plateau followed by healing; and non-healing. Conclusion. The electronically instrumented
Aims. The anterior pelvic
Femoral periprosthetic fractures are rising in incidence. Their management is complex and carries a high associated mortality. Unlike native hip fractures, there are no guidelines advising on time to theatre in this group. We aim to determine whether delaying surgical intervention influences morbidity or mortality in femoral periprosthetic fractures. We identified all periprosthetic fractures around a hip or knee arthroplasty from our prospectively collated database between 2012 and 2021. Patients were categorized into early or delayed intervention based on time from admission to surgery (early = ≤ 36 hours, delayed > 36 hours). Patient demographics, existing implants, Unified Classification System fracture subtype, acute medical issues on admission, preoperative haemoglobin, blood transfusion requirement, and length of hospital stay were identified for all patients. Complication and mortality rates were compared between groups.Aims
Methods
To evaluate whether low-intensity pulsed ultrasound (LIPUS) accelerates bone healing at osteotomy sites and promotes functional recovery after open-wedge high tibial osteotomy (OWHTO). Overall, 90 patients who underwent OWHTO without bone grafting were enrolled in this nonrandomized retrospective study, and 45 patients treated with LIPUS were compared with 45 patients without LIPUS treatment in terms of bone healing and functional recovery postoperatively. Clinical evaluations, including the pain visual analogue scale (VAS) and Japanese Orthopaedic Association (JOA) score, were performed preoperatively as well as six weeks and three, six, and 12 months postoperatively. The progression rate of gap filling was evaluated using anteroposterior radiographs at six weeks and three, six, and 12 months postoperatively.Aims
Methods
This study investigated vancomycin-microbubbles (Vm-MBs) and meropenem (Mp)-MBs with ultrasound-targeted microbubble destruction (UTMD) to disrupt biofilms and improve bactericidal efficiency, providing a new and promising strategy for the treatment of device-related infections (DRIs). A film hydration method was used to prepare Vm-MBs and Mp-MBs and examine their characterization. Biofilms of methicillin-resistant Aims
Methods
Most hip fractures treated with modern internal
fixation techniques will heal. However, failures occasionally occur and
require revision procedures. Salvage strategies employed during
revision are based on whether the fixation failure occurs in the
femoral neck, or in the intertrochanteric region. Patient age and
remaining bone stock also influence decision making. For fractures
in young patients, efforts are generally focused on preserving the
native femoral head via osteotomies and repeat internal fixation.
For failures in older patients, some kind of hip replacement is
usually selected. Disuse osteopenia, deformity, bone loss, and stress-risers
from previous
In contrast to operations performed for other fractures, there is a high incidence rate of surgical site infection (SSI) post-open reduction and internal fixation (ORIF) done for tibial plateau fractures (TPFs). This study investigates the effect of induced membrane technique combined with internal fixation for managing SSI in TPF patients who underwent ORIF. From April 2013 to May 2017, 46 consecutive patients with SSI post-ORIF for TPFs were managed in our centre with an induced membrane technique. Of these, 35 patients were included for this study, with data analyzed in a retrospective manner.Aims
Methods
The proximal humerus internal locking system is an
Techniques are described by which metal implants can be designed and produced to fit precisely a bony site at a subsequent operation. CT scans and solid modelling were used to produce an accurate three-dimensional representation of the surface of the bone. These techniques were applied to the production of an
Surgical reconstruction of deformed Charcot feet carries a high risk of nonunion, metalwork failure, and deformity recurrence. The primary aim of this study was to identify the factors contributing to these complications following hindfoot Charcot reconstructions. We retrospectively analyzed patients who underwent hindfoot Charcot reconstruction with an intramedullary nail between January 2007 and December 2019 in our unit. Patient demographic details, comorbidities, weightbearing status, and postoperative complications were noted. Metalwork breakage, nonunion, deformity recurrence, concurrent midfoot reconstruction, and the measurements related to intramedullary nail were also recorded.Aims
Methods
In our unit, we adopt a two-stage surgical reconstruction approach using internal fixation for the management of infected Charcot foot deformity. We evaluate our experience with this functional limb salvage method. We conducted a retrospective analysis of prospectively collected data of all patients with infected Charcot foot deformity who underwent two-stage reconstruction with internal fixation between July 2011 and November 2019, with a minimum of 12 months’ follow-up.Aims
Methods
We have studied the intervertebral discs adjacent to fractured vertebral bodies using MRI in 63 patients at a minimum of 18 months after injury. There were 75 thoracolumbar fractures of which 26 were treated conservatively and 37 by posterior reduction and fusion with an AO
We have prospectively studied the outcome of infections associated with implants which were retained and treated using a standardised antimicrobial protocol. Over a period of four years, we studied 24 consecutive patients who had symptoms of infection for less than one year, a stable implant, no sinus tract and a known pathogen which was susceptible to recommended antimicrobial agents. The infections involved hip prostheses (14), knee prostheses (5), an
A recent report from France suggested an association between the use of cobalt-chrome (CoCr) femoral heads in total hip arthroplasties (THAs) and an increased risk of dilated cardiomyopathy and heart failure. CoCr is a commonly used material in orthopaedic implants. If the reported association is causal, the consequences would be significant given the millions of joint arthroplasties and other orthopaedic procedures in which CoCr is used annually. We examined whether CoCr-containing THAs were associated with an increased risk of all-cause mortality, heart outcomes, cancer, and neurodegenerative disorders in a large national database. Data from the National Joint Registry was linked to NHS English hospital inpatient episodes for 374,359 primary THAs with up to 14.5 years' follow-up. We excluded any patients with bilateral THAs, knee arthroplasties, indications other than osteoarthritis, aged under 55 years, and diagnosis of one or more outcome of interest before THA. Implants were grouped as either containing CoCr or not containing CoCr. The association between implant construct and the risk of all-cause mortality and incident heart failure, cancer, and neurodegenerative disorders was examined.Aims
Methods
An experimental study is reported of fracture healing in the femora of 36 Beagle dogs, comparing the results of using stainless steel plates with those of using less rigid titanium alloy plates. The alloy plates led to the appearance of a small amount of periosteal callus without any histological evidence of fracture instability, thus allowing the radiological assessment of fracture union. This also produced less bone loss during the remodelling phase. Radiological measurements 24 weeks after osteotomy showed cortical thickness to be reduced by six per cent under titanium alloy and by 19 per cent under stainless steel, while histological measurements showed a total bone loss of 3.7 per cent under titanium alloy and of 11 per cent under stainless steel plates. Removal of the titanium alloy plates after eight weeks followed by a recovery period of 16 weeks produced an increase of cortical thickness of 69 per cent and a gain in total bone mass of 30 per cent. Titanium alloy plates also produced less soft-tissue reaction than stainless steel plates. It is concluded that this alloy is a promising material for
We performed a retrospective review of the case notes of 84 consecutive patients who had suffered a severe (Gustilo IIIb or IIIc) open fracture of the tibia after blunt trauma between 1990 and 1998. All had been treated by a radical protocol which included early soft-tissue cover with a muscle flap by a combined orthopaedic and plastic surgery service. Our ideal management is a radical debridement of the wound outside the zone of injury, skeletal stabilisation and early soft-tissue cover with a vascularised muscle flap. All patients were followed clinically and radiologically to union or for one year. After exclusion of four patients (one unrelated death and three patients lost to follow-up), we reviewed 80 patients with 84 fractures. There were 67 men and 13 women with a mean age of 37 years (3 to 89). Five injuries were grade IIIc and 79 grade IIIb; 12 were site 41, 43 were site 42 and 29 were site 43. Debridement and stabilisation of the fracture were invariably performed immediately. In 33 cases the soft-tissue reconstruction was also completed in a single stage, while in a further 30 it was achieved within 72 hours. In the remaining 21 there was a delay beyond 72 hours, often for critical reasons unrelated to the limb injury. All grade-IIIc injuries underwent immediate vascular reconstruction, with an immediate cover by a flap in two. All were salvaged. There were four amputations, one early, one mid-term and two late, giving a final rate of limb salvage of 95%. Overall, nine pedicled and 75 free muscle flaps were used; the rate of flap failure was 3.5%. Stabilisation of the fracture was achieved with 19 external and 65
Post-traumatic elbow stiffness is a disabling condition that remains challenging for upper limb surgeons. Open elbow arthrolysis is commonly used for the treatment of stiff elbow when conservative therapy has failed. Multiple questions commonly arise from surgeons who deal with this disease. These include whether the patient has post-traumatic stiff elbow, how to evaluate the problem, when surgery is appropriate, how to perform an excellent arthrolysis, what the optimal postoperative rehabilitation is, and how to prevent or reduce the incidence of complications. Following these questions, this review provides an update and overview of post-traumatic elbow stiffness with respect to the diagnosis, preoperative evaluation, arthrolysis strategies, postoperative rehabilitation, and prevention of complications, aiming to provide a complete diagnosis and treatment path. Cite this article:
Tibiotalocalcaneal (TTC) fusion is used to treat a variety of conditions affecting the ankle and subtalar joint, including osteoarthritis (OA), Charcot arthropathy, avascular necrosis (AVN) of the talus, failed total ankle arthroplasty, and severe deformity. The prevalence of postoperative complications remains high due to the complexity of hindfoot disease seen in these patients. The aim of this study was to analyze the relationship between preoperative conditions and postoperative complications in order to predict the outcome following primary TTC fusion. We retrospectively reviewed the medical records of 101 patients who underwent TTC fusion at the same institution between 2011 and 2019. Risk ratios (RRs) associated with age, sex, diabetes, cardiovascular disease, smoking, preoperative ankle deformity, and the use of bone graft during surgery were related to the postoperative complications. We determined from these data which pre- and perioperative factors significantly affected the outcome.Aims
Methods
The aim of this study was to describe the technique of distraction
osteogenesis followed by arthrodesis using internal fixation to
manage complex conditions of the ankle, and to present the results
of this technique. Between 2008 and 2014, distraction osteogenesis followed by arthrodesis
using internal fixation was performed in 12 patients with complex
conditions of the ankle due to trauma or infection. There were eight
men and four women: their mean age was 35 years (23 to 51) at the
time of surgery. Bone healing and functional recovery were evaluated
according to the criteria described by Paley. Function was assessed
using the ankle-hindfoot scale of the American Orthopedic Foot and
Ankle Society (AOFAS).Aims
Patients and Methods
The best method of treating unstable pelvic fractures that involve
the obturator ring is still a matter for debate. This study compared
three methods of treatment: nonoperative, isolated posterior fixation
and combined anteroposterior stabilization. The study used data from the German Pelvic Trauma Registry and
compared patients undergoing conservative management (n = 2394),
surgical treatment (n = 1345) and transpubic surgery, including
posterior stabilization (n = 730) with isolated posterior osteosynthesis
(n = 405) in non-complex Type B and C fractures that only involved the
obturator ring anteriorly. Calculated odds ratios were adjusted
for potential confounders. Outcome criteria were intraoperative
and general short-term complications, the incidence of nerve injuries,
and mortality.Aims
Patients and Methods
Computer hexapod assisted orthopaedic surgery (CHAOS), is a method
to achieve the intra-operative correction of long bone deformities
using a hexapod external fixator before definitive internal fixation
with minimally invasive stabilisation techniques. The aims of this study were to determine the reliability of this
method in a consecutive case series of patients undergoing femoral
deformity correction, with a minimum six-month follow-up, to assess
the complications and to define the ideal group of patients for
whom this treatment is appropriate. The medical records and radiographs of all patients who underwent
CHAOS for femoral deformity at our institution between 2005 and
2011 were retrospectively reviewed. Records were available for all
55 consecutive procedures undertaken in 49 patients with a mean
age of 35.6 years (10.9 to 75.3) at the time of surgery.Aims
Patients and Methods
Despite worldwide clinical use of bio-absorbable devices for internal fixation in orthopaedic surgery, the degradation behaviour and tissue replacement of these implants are not fully understood. In a long-term experimental study, we have determined the patterns of tissue restoration 36 and 54 months after implantation of polyglycolic acid and poly-laevo-lactic acid screws in the distal femur of the rabbit. After 36 months in the polyglycolic acid group the specimens showed no remaining polymer and loose connective tissue occupied 80% of the screw track. Tissue restoration remained poor at 54 months, the amounts of trabecular bone and haematopoietic elements being significantly lower than those in the intact control group. The amount of trabecular bone within the screw track at 54 months in the polyglycolic acid group was less than in the empty drill holes (p = 0.04). In the poly-laevo-lactic acid group, polymeric material was present in abundance after 54 months, occupying 60% of the cross-section of the core area of the screw track. When using absorbable internal fixation implants we should recognise that the degradation of the devices will probably not be accompanied by the restoration of normal trabecular bone.
We report the outcome of 32 patients (37 knees) who underwent hemicallostasis with a dynamic external fixator for osteoarthritis of the medial compartment of the knee. There were 16 men (19 knees) and 16 women (18 knees) with a mean age at operation of 54.6 years (27 to 72). The aim was to achieve a valgus overcorrection of 2° to 8° or mechanical axis at 62.5% (± 12.5%). At a mean follow-up of 62.8 months (51 to 81) there was no change in the mean range of movement, and no statistically significant difference in the Insall-Salvati index or tibial slope (p = 0.11 and p = 0.15, respectively). The mean hip-knee-ankle angle changed from 190.6 (183° to 197°) to 176.0° (171° to 181°), with a mean final position of the mechanical axis of 58.5% (35.1% to 71.2%). The desired alignment was attained in 31 of 37 (84%) knees. There were 21 excellent, 13 good, two fair and one poor result according to the Oxford knee score with no correlation between age and final score. This score was at its best at one year with a statistically significant deterioration at two years (p = 0.001) followed by a small but not statistically significant deterioration until the final follow-up (p = 0.17). All the knees with Ahlback grade 1 osteoarthritis had excellent or good results. Complications included pin tract infections involving 16.4% of all pins used, delayed union in two, knee stiffness in four, fracture of the lateral cortex in one and ring sequestrum in one. In conclusion, hemicallostasis provides precision in attaining the desired alignment without interfering with tibial slope or patellar height, and is relatively free of serious complications.
We studied the effects of coating titanium implants with teicoplanin and clindamycin in 30 New Zealand White rabbits which were randomly assigned to three groups. The intramedullary canal of the left tibia of each rabbit was inoculated with 500 colony forming units of Staphylococcus aureus. Teicoplanin-coated implants were implanted into rabbits in group 1, clindamycin-coated implants into rabbits in group 2, and uncoated implants into those in group 3. All the rabbits were killed one week later. The implants were removed and cultured together with pieces of tibial bone and wound swabs. The rate of colonisation of the organisms in the three groups was compared. Organisms were cultured from no rabbits in group 1, one in group 2 but from all in group 3. There was no significant difference between groups 1 and 2 (p = 1.000). There were significant differences between groups 1 and 3 and groups 2 and 3 (p <
0.001). Significant protection against bacterial colonisation and infection was found with teicoplanin- and clindamycin-coated implants in this experimental model.
We report the outcomes of 20 patients (12 men,
8 women, 21 feet) with Charcot neuro-arthropathy who underwent correction
of deformities of the ankle and hindfoot using retrograde intramedullary
nail arthrodesis. The mean age of the patients was 62.6 years (46
to 83); their mean BMI was 32.7 (15 to 47) and their median American
Society of Anaesthetists score was 3 (2 to 4). All presented with
severe deformities and 15 had chronic ulceration. All were treated
with reconstructive surgery and seven underwent simultaneous midfoot
fusion using a bolt, locking plate or a combination of both. At
a mean follow-up of 26 months (8 to 54), limb salvage was achieved
in all patients and 12 patients (80%) with ulceration achieved healing
and all but one patient regained independent mobilisation. There was
failure of fixation with a broken nail requiring revision surgery
in one patient. Migration of distal locking screws occurred only
when standard screws had been used but not with hydroxyapatite-coated
screws. The mean American Academy of Orthopaedic Surgeons Foot and
Ankle (AAOS-FAO) score improved from 50.7 (17 to 88) to 65.2 (22
to 88), (p = 0.015). The mean Short Form (SF)-36 Health Survey Physical
Component Score improved from 25.2 (16.4 to 42.8) to 29.8 (17.7
to 44.2), (p = 0.003) and the mean Euroqol EQ‑5D‑5L score improved
from 0.63 (0.51 to 0.78) to 0.67 (0.57 to 0.84), (p = 0.012). Single-stage correction of deformity using an intramedullary
hindfoot arthrodesis nail is a good form of treatment for patients
with severe Charcot hindfoot deformity, ulceration and instability
provided a multidisciplinary care plan is delivered. Cite this article:
Despite advances in the prevention and treatment of osteoporotic fractures, their prevalence continues to increase. Their operative treatment remains a challenge for the surgeon, often with unpredictable outcomes. This review highlights the current aspects of management of these fractures and focuses on advances in implant design and surgical technique.
We investigated the clinical outcome of internal
fixation for pathological fracture of the femur after primary excision of
a soft-tissue sarcoma that had been treated with adjuvant radiotherapy. A review of our database identified 22 radiation-induced fractures
of the femur in 22 patients (seven men, 15 women). We noted the
mechanism of injury, fracture pattern and any complications after
internal fixation, including nonunion, hardware failure, secondary
fracture or deep infection. The mean age of the patients at primary excision of the tumour
was 58.3 years (39 to 86). The mean time from primary excision to
fracture was 73.2 months (2 to 195). The mean follow-up after fracture
fixation was 65.9 months (12 to 205). Complications occurred in
19 patients (86%). Nonunion developed in 18 patients (82%), of whom
11 had a radiological nonunion at 12 months, five a nonunion and
hardware failure and two an infected nonunion. One patient developed
a second radiation-associated fracture of the femur after internal
fixation and union of the initial fracture. A total of 13 patients
(59%) underwent 24 revision operations. Internal fixation of a pathological fracture of the femur after
radiotherapy for a soft-tissue sarcoma has an extremely high rate
of complication and requires specialist attention. Cite this article:
The increasing prevalence of osteoporosis in
an ageing population has contributed to older patients becoming
the fastest-growing group presenting with acetabular fractures.
We performed a systematic review of the literature involving a number
of databases to identify studies that included the treatment outcome
of acetabular fractures in patients aged >
55 years. An initial
search identified 61 studies; after exclusion by two independent
reviewers, 15 studies were considered to meet the inclusion criteria.
All were case series. The mean Coleman score for methodological
quality assessment was 37 (25 to 49). There were 415 fractures in
414 patients. Pooled analysis revealed a mean age of 71.8 years
(55 to 96) and a mean follow-up of 47.3 months (1 to 210). In seven
studies the results of open reduction and internal fixation (ORIF)
were presented: this was combined with simultaneous hip replacement
(THR) in four, and one study had a mixture of these strategies.
The results of percutaneous fixation were presented in two studies,
and a single study revealed the results of non-operative treatment. With fixation of the fracture, the overall mean rate of conversion
to THR was 23.1% (0% to 45.5%). The mean rate of non-fatal complications
was 39.8% (0% to 64%), and the mean mortality rate was 19.1% (5%
to 50%) at a mean of 64 months (95% confidence interval 59.4 to
68.6; range 12 to 143). Further data dealing with the classification
of the fracture, the surgical approach used, operative time, blood
loss, functional and radiological outcomes were also analysed. This study highlights that, of the many forms of treatment available
for this group of patients, there is a trend to higher complication
rates and the need for further surgery compared with the results
of the treatment of acetabular fractures in younger patients. Cite this article:
Hip arthrodesis remains a viable surgical technique
in well selected patients, typically the young manual labourer with
isolated unilateral hip disease. Despite this, its popularity with
patients and surgeons has decreased due to the evolution of hip
replacement, and is seldom chosen by young adult patients today.
The surgeon is more likely to encounter a patient who requests conversion
to total hip replacement (THR). The most common indications are
a painful pseudarthrosis, back pain, ipsilateral knee pain or contralateral
hip pain. Occasionally the patient will request conversion because
of difficulty with activities of daily living, body image and perceived
cosmesis. The technique of conversion and a discussion of the results
are presented. Cite this article:
We determined the midterm survival, incidence
of peri-prosthetic fracture and the enhancement of the width of
the femur when combining struts and impacted bone allografts in
24 patients (25 hips) with severe femoral bone loss who underwent
revision hip surgery. The pre-operative diagnosis was aseptic loosening
in 16 hips, second-stage reconstruction in seven, peri-prosthetic
fracture in one and stem fracture in one hip. A total of 14 hips
presented with an Endoklinik grade 4 defect and 11 hips a grade
3 defect. The mean pre-operative Merle D’Aubigné and Postel score
was 5.5 points (1 to 8). The survivorship was 96% (95% confidence interval 72 to 98) at
a mean of 54.5 months (36 to 109). The mean functional score was
17.3 points (16 to 18). One patient in which the strut did not completely
bypass the femoral defect was further revised using a long cemented
stem due to peri-prosthetic fracture at six months post-operatively.
The mean subsidence of the stem was 1.6 mm (1 to 3). There was no
evidence of osteolysis, resorption or radiolucencies during follow-up
in any hip. Femoral width was enhanced by a mean of 41% (19% to
82%). A total of 24 hips had partial or complete bridging of the
strut allografts. This combined biological method was associated with a favourable
survivorship, a low incidence of peri-prosthetic fracture and enhancement
of the width of the femur in revision total hip replacement in patients
with severe proximal femoral bone loss.
We investigated the static and cyclical strength of parallel and angulated locking plate screws using rigid polyurethane foam (0.32 g/cm3) and bovine cancellous bone blocks. Custom-made stainless steel plates with two conically threaded screw holes with different angulations (parallel, 10° and 20° divergent) and 5 mm self-tapping locking screws underwent pull-out and cyclical pull and bending tests. The bovine cancellous blocks were only subjected to static pull-out testing. We also performed finite element analysis for the static pull-out test of the parallel and 20° configurations. In both the foam model and the bovine cancellous bone we found the significantly highest pull-out force for the parallel constructs. In the finite element analysis there was a 47% more damage in the 20° divergent constructs than in the parallel configuration. Under cyclical loading, the mean number of cycles to failure was significantly higher for the parallel group, followed by the 10° and 20° divergent configurations. In our laboratory setting we clearly showed the biomechanical disadvantage of a diverging locking screw angle under static and cyclical loading.
Percutaneous stabilisation of tibial fractures by locking plates has become an accepted form of osteosynthesis. A potential disadvantage of this technique is the risk of damage to the neurovascular bundles in the anterior and peroneal compartments. Our aim in this anatomical study was to examine the relationship of the deep peroneal nerve to a percutaneously-inserted Less Invasive Stabilisation System tibial plate in the lower limbs of 18 cadavers. Screws were inserted through stab incisions. The neurovascular bundle was dissected to reveal its relationship to the plate and screws. In all cases, the deep peroneal nerve was in direct contact with the plate between the 11th and the 13th holes. In ten specimens the nerve crossed superficial to the plate, in six it was interposed between the plate and the bone and in the remaining two specimens it coursed at the edge of the plate. Percutaneous insertion of plates with more than ten holes is not recommended because of the risk of injury to the neurovascular structures. When longer plates are required we suggest distal exposure so that the neurovascular bundle may be displayed and protected.
The treatment of infected exposed implants which have been used for internal fixation usually involves debridement and removal of the implant. This can result in an unstable fracture or spinal column. Muscle flaps may be used to salvage these implants since they provide soft-tissue cover and fresh vascularity. However, there have been few reports concerning their use and these have concentrated on the eradication of the infection and successful soft-tissue cover as the endpoint. There is no information on the factors which may influence the successful salvage of the implant using muscle flaps. We studied the results and factors affecting outcome in nine pedicled muscle flaps used in the treatment of exposed metal internal fixation with salvage of the implant as the primary endpoint. This was achieved in four cases. Factors predicting success were age <
30 years, the absence of comorbid conditions and a favourable microbiological profile. The growth of multiple organisms, a history of smoking and the presence of methicillin-resistant Staphylococcus aureus on wound cultures indicated a poor outcome. The use of antibiotic beads, vacuum-assisted closure and dressing, the surgical site, the type of flap performed and the time from primary surgery to flap cover were not predictive of outcome.
Between November 1994 and June 1999, 35 patients referred to our Problem Fracture Service with chronic diaphyseal osteomyelitis were treated using a closed double-lumen suction irrigation system after reaming and arthroscopic debridement of the intramedullary canal. This is a modified system based on that of Lautenbach. Between June and July 2007 the patients were reviewed by postal questionnaire and telephone and from the case notes. At a mean follow-up of 101 months (2 to 150), 26 had no evidence of recurrence and four had died from unrelated causes with no evidence of recurrent infection. One had been lost to follow-up at two months and was therefore excluded. Four had persisting problems with sinus discharge and one had his limb amputated for recurrent metaplastic change. Our results represent a clearance of infection of 85.3% (29 of 34), with recurrence in 11.8% (4 of 34). They are comparable to the results of the Papineau and Belfast techniques, but with considerably less surgical insult to the patient.
We report the outcome at ten to 15 years of two-stage revision for hip infection in 99 patients using the Prostalac articulated hip spacer system. All the patients were contacted to determine their current functional and infection status using the Oxford-12, Short form-12, and Western Ontario and McMaster University Osteoarthritis Index questionnaires. A total of 11 of the 99 patients had a further infection, of whom seven responded to repeat surgery with no further sequelae. The mean interval between the stages was five months (1 to 36). We were able to review 48 living patients, with a mean age of 72 years (46 to 86), 34 (71%) of whom provided health-related quality-of-life outcome scores. The mean follow-up was 12 years (10 to 15). The long-term success rate was 89% and with additional surgery this rose to 96%. The mean global Western Ontario and McMaster University Osteoarthritis Index score was 80.6 ( Two-stage revision for hip infection using a Prostalac interim spacer offers a predictable and lasting solution for patients with this difficult problem.
Conventional treatment of mild slipped capital femoral epiphysis consists of fixation Between October 1984 and December 1995 we treated 16 patients for unilateral mild slipped capital femoral epiphysis by fixation While our clinical data favours conventional treatment, our radiological findings are in support of restoring the anatomy of the proximal femur to avoid or delay the development of femoroacetabular impingement following mild slipped capital femoral epiphysis.
We have updated our previous randomised controlled trial comparing release of chromium (Cr) and cobalt (Co) ions and included levels of titanium (Ti) ions. We have compared the findings from 28 mm metal-on-metal total hip replacement, performed using titanium CLS/Spotorno femoral components and titanium AlloFit acetabular components with Metasul bearings, with Durom hip resurfacing using a Metasul articulation or bearing and a titanium plasma-sprayed coating for fixation of the acetabular component. Although significantly higher blood ion levels of Cr and Co were observed at three months in the resurfaced group than in total hip replacement, no significant difference was found at two years post-operatively for Cr, 1.58 μg/L and 1.62 μg/L respectively (p = 0.819) and for Co, 0.67 μg/L and 0.94 μg/L respectively (p = 0.207). A steady state was reached at one year in the resurfaced group and after three months in the total hip replacement group. Interestingly, Ti, which is not part of the bearing surfaces with its release resulting from metal corrosion, had significantly elevated ion levels after implantation in both groups. The hip resurfacing group had significantly higher Ti levels than the total hip replacement group for all periods of follow-up. At two years the mean blood levels of Ti ions were 1.87 μg/L in hip resurfacing and and 1.30 μg/L in total hip replacement (p = 0.001). The study confirms even with different bearing diameters and clearances, hip replacement and 28 mm metal-on-metal total hip replacement produced similar Cr and Co metal ion levels in this randomised controlled trial study design, but apart from wear on bearing surfaces, passive corrosion of exposed metallic surfaces is a factor which influences ion concentrations. Ti plasma spray coating the acetabular components for hip resurfacing produces significantly higher release of Ti than Ti grit-blasted surfaces in total hip replacement.
New developments in osteotomy techniques and methods of fixation have caused a revival of interest of osteotomies around the knee. The current consensus on the indications, patient selection and the factors influencing the outcome after high tibial osteotomy is presented. This paper highlights recent research aimed at joint pressure redistribution, fixation stability and bone healing that has led to improved surgical techniques and a decrease of post-operative time to full weight-bearing.
Failure of fixation is a common problem in the treatment of osteoporotic fractures around the hip. The reinforcement of bone stock or of fixation of the implant may be a solution. Our study assesses the existing evidence for the use of bone substitutes in the management of these fractures in osteoporotic patients. Relevant publications were retrieved through Medline research and further scrutinised. Of 411 studies identified, 22 met the inclusion criteria, comprising 12 experimental and ten clinical reports. The clinical studies were evaluated with regard to their level of evidence. Only four were prospective and randomised. Polymethylmethacrylate and calcium-phosphate cements increased the primary stability of the implant-bone construct in all experimental and clinical studies, although there was considerable variation in the design of the studies. In randomised, controlled studies, augmentation of intracapsular fractures of the neck of the femur with calcium-phosphate cement was associated with poor long-term results. There was a lack of data on the long-term outcome for trochanteric fractures. Because there were only a few, randomised, controlled studies, there is currently poor evidence for the use of bone cement in the treatment of fractures of the hip.
The treatment of fractures of the proximal tibia is complex and makes great demands on the implants used. Our study aimed to identify what levels of primary stability could be achieved with various forms of osteosynthesis in the treatment of diaphyseal fractures of the proximal tibia. Pairs of human tibiae were investigated. An unstable fracture was simulated by creating a defect at the metaphyseal-diaphyseal junction. Six implants were tested in a uniaxial testing device (Instron) using the quasi-static and displacement-controlled modes and the force-displacement curve was recorded. The movements of each fragment and of the implant were recorded video-optically (MacReflex, Qualysis). Axial deviations were evaluated at 300 N. The results show that the nailing systems tolerated the highest forces. The lowest axial deviations in varus and valgus were also found for the nailing systems; the highest axial deviations were recorded for the buttress plate and the less invasive stabilising system (LISS). In terms of rotational displacement the LISS was better than the buttress plate. In summary, it was found that higher loads were better tolerated by centrally placed load carriers than by eccentrically placed ones. In the case of the latter, it appears advantageous to use additive procedures for medial buttressing in the early phase.
The results of meta-analysis show a revision rate of 33% for internal fixation of displaced fractures of the femoral neck, mostly because of nonunion. Osteopenia and osteoporosis are highly prevalent in elderly patients. Bone density has been shown to correlate with the intrinsic stability of the fixation of the fracture in cadaver and retrospective studies. We aimed to confirm or refute this finding in a clinical setting. We performed a prospective, multicentre study of 111 active patients over 60 years of age with a displaced fracture of the femoral neck which was eligible for internal fixation. The bone density of the femoral neck was measured pre-operatively by dual-energy x-ray absorptiometry (DEXA). The patients were divided into two groups namely, those with osteopenia (66%, mean T-score −1.6) and those with osteoporosis (34%, mean T-score −3.0). Age (p = 0.47), gender (p = 0.67), delay to surgery (p = 0.07), the angle of the fracture (p = 0.33) and the type of implant (p = 0.48) were similar in both groups. Revision to arthroplasty was performed in 41% of osteopenic and 42% of osteoporotic patients (p = 0.87). Morbidity (p = 0.60) and mortality were similar in both groups (p = 0.65). Our findings show that the clinical outcome of internal fixation for displaced fractures of the femoral neck does not depend on bone density and that pre-operative DEXA is not useful.