Polyethylene wear debris can cause osteolysis
and the failure of total hip arthroplasty. We present the five-year
wear rates of a highly cross-linked polyethylene (X3) bearing surface
when used in conjunction with a 36 mm ceramic femoral head. This was a prospective study of a cohort of 100 THAs in 93 patients.
Pain and activity scores were measured pre- and post-operatively.
Femoral head penetration was measured at two months, one year, two
years and at five years using validated edge-detecting software
(PolyWare Auto). At a mean of 5.08 years (3.93 to 6.01), 85 hips in 78 patients
were available for study. The mean age of these patients was 59.08
years (42 to 73, the mean age of males (n = 34) was 59.15 years,
and females (n = 44) was 59.02 years). All patients had significant
improvement in their functional scores (p <
0.001). The steady
state two-dimensional linear wear rate was 0.109 mm/year. The steady
state volumetric wear rate was 29.61 mm3/year. No significant
correlation was found between rate of wear and age (p = 0.34), acetabular
component size (p = 0.12) or clinical score (p = 0.74). Our study shows low steady state wear rates at five years in
X3 highly cross-linked polyethylene in conjunction with a 36 mm
ceramic femoral head. The linear wear rate was almost identical
to the osteolysis threshold of 0.1 mm/year recommended in the literature. Cite this article:
Eleven patients were reviewed an average of 23 years after they had been treated by excision of a symptomatic calcaneonavicular bar in 16 of their feet. Of these feet 69% (11 feet) had a good or excellent result. Of the five failures, three feet had good results after subsequent triple arthrodesis, but two treated by repeated excision of the bar were still unsatisfactory. Beaking of the talus seen before operation correlated with poor results.
A retrospective review of medial compartment arthroplasty in 22 patients (22 knees) is reported. The operations were performed between 1973 and 1978. Eighty-six per cent were rated good or excellent using the knee rating system devised by the Hospital for Special Surgery, New York. Excellent or good results were achieved in six patients who had previously had a high tibial osteotomy. At the time of follow-up significant loosening had not occurred, although progression of patellofemoral disease was noted. This study supports the promising results reported for unicompartmental resurfacing arthroplasty in the elderly.