Aims. This study aimed to analyze the accuracy and
Aims. Navigation devices are designed to improve a surgeon’s accuracy in positioning the acetabular and femoral components in total hip arthroplasty (THA). The purpose of this study was to both evaluate the accuracy of an optical computer-assisted surgery (CAS) navigation system and determine whether preoperative spinopelvic mobility (categorized as hypermobile, normal, or stiff) increased the risk of acetabular component placement
Aims. Traditionally, total hip arthroplasty (THA) templating has been performed on anteroposterior (AP) pelvis radiographs. Recently, additional AP hip radiographs have been recommended for accurate measurement of the femoral offset (FO). To verify this claim, this study aimed to establish quantitative data of the measurement
This study used CT analysis to determine the rotational alignment of 39 painful and 26 painless fixed-bearing total knee replacements (TKRs) from a cohort of 740 NexGen Legacy posterior-stabilised and cruciate-retaining prostheses implanted between May 1996 and August 2003. The mean rotation of the tibial component was 4.3° of internal rotation (25.4° internal to 13.9° external rotation) in the painful group and 2.2° of external rotation (8.5° internal to 18.2° external rotation) in the painfree group (p = 0.024). In the painful group 17 tibial components were internally rotated more than 9° compared with none in the painfree group (p <
0.001). Additionally, six femoral components in the painful group were internally rotated more than 6° compared with none in the painfree group (p = 0.017). External rotational
We performed a CT-based computer simulation study
to determine how the relationship between any inbuilt posterior
slope in the proximal tibial osteotomy and cutting jig rotational
orientation
Reimers’ hip migration percentage is commonly used to document the extent of subluxation of the hip in children with spasticity. In this study, two measurers, with six months paediatric orthopaedic experience, measured the migration percentage on 44 pelvic radiographs of children with cerebral palsy, aged between two and eight years. Unknown to the measurers, each radiograph was duplicated, giving 22 non-identical radiographs (44 hips) which were measured twice at time 0 and twice six weeks later. The intra-measurer, intra-sessional absolute differences between the first and second measurements ranged from 0% to 23%, with median values of 2.5% to 3.6%. The intra-sessional median absolute differences were not statistically different between the two measurers and measuring sessions (p = 0.42, Kruskal-Wallis test). The inter-sessional absolute differences for measurements made by the same measurers ranged from 0% to 18% with a median absolute difference of 1.7% to 3.2%. Overall, only 5% of the intra-measurer measurement differences, within and between sessions, were above 13%. Repeated measurements by one measurer over time must, therefore, vary by more than 13% in order to be 95% confident of a true change. The inter-measurer
We present an audit comparing our level I major trauma centre’s
data for a cohort of patients with hip fractures in the National
Hip Fracture Database (NHFD) with locally held data on these patients. A total of 2036 records for episodes between July 2009 and June
2014 were reviewed. Aims
Patients and Methods
A study was undertaken to assess the degree of inter-observer
Wrong-level surgery is a unique pitfall in spinal
surgery and is part of the wider field of wrong-site surgery. Wrong-site
surgery affects both patients and surgeons and has received much
media attention. We performed this systematic review to determine
the incidence and prevalence of wrong-level procedures in spinal
surgery and to identify effective prevention strategies. We retrieved
12 studies reporting the incidence or prevalence of wrong-site surgery
and that provided information about prevention strategies. Of these,
ten studies were performed on patients undergoing lumbar spine surgery
and two on patients undergoing lumbar, thoracic or cervical spine procedures.
A higher frequency of wrong-level surgery in lumbar procedures than
in cervical procedures was found. Only one study assessed preventative
strategies for wrong-site surgery, demonstrating that current site-verification protocols
did not prevent about one-third of the cases. The current literature
does not provide a definitive estimate of the occurrence of wrong-site
spinal surgery, and there is no published evidence to support the
effectiveness of site-verification protocols. Further prevention
strategies need to be developed to reduce the risk of wrong-site surgery.
Aims. We aimed to assess the reliability and validity of OpenPose, a posture estimation algorithm, for measurement of knee range of motion after total knee arthroplasty (TKA), in comparison to radiography and goniometry. Methods. In this prospective observational study, we analyzed 35 primary TKAs (24 patients) for knee osteoarthritis. We measured the knee angles in flexion and extension using OpenPose, radiography, and goniometry. We assessed the test-retest reliability of each method using intraclass correlation coefficient (1,1). We evaluated the ability to estimate other measurement values from the OpenPose value using linear regression analysis. We used intraclass correlation coefficients (2,1) and Bland–Altman analyses to evaluate the agreement and
Aims. The aim of this study was to evaluate the reliability and validity of a patient-specific algorithm which we developed for predicting changes in sagittal pelvic tilt after total hip arthroplasty (THA). Methods. This retrospective study included 143 patients who underwent 171 THAs between April 2019 and October 2020 and had full-body lateral radiographs preoperatively and at one year postoperatively. We measured the pelvic incidence (PI), the sagittal vertical axis (SVA), pelvic tilt, sacral slope (SS), lumbar lordosis (LL), and thoracic kyphosis to classify patients into types A, B1, B2, B3, and C. The change of pelvic tilt was predicted according to the normal range of SVA (0 mm to 50 mm) for types A, B1, B2, and B3, and based on the absolute value of one-third of the PI-LL mismatch for type C patients. The reliability of the classification of the patients and the prediction of the change of pelvic tilt were assessed using kappa values and intraclass correlation coefficients (ICCs), respectively. Validity was assessed using the overall mean
Aims. Achieving accurate implant positioning and restoring native hip biomechanics are key surgeon-controlled technical objectives in total hip arthroplasty (THA). The primary objective of this study was to compare the reproducibility of the planned preoperative centre of hip rotation (COR) in patients undergoing robotic arm-assisted THA versus conventional THA. Methods. This prospective randomized controlled trial (RCT) included 60 patients with symptomatic hip osteoarthritis undergoing conventional THA (CO THA) versus robotic arm-assisted THA (RO THA). Patients in both arms underwent pre- and postoperative CT scans, and a patient-specific plan was created using the robotic software. The COR, combined offset, acetabular orientation, and leg length discrepancy were measured on the pre- and postoperative CT scanogram at six weeks following surgery. Results. There were no significant differences for any of the baseline characteristics including spinopelvic mobility. The absolute
Aims. This study aimed to investigate the relationship between changes in patellar height and clinical outcomes at a mean follow-up of 7.7 years (5 to 10) after fixed-bearing posterior-stabilized total knee arthroplasty (PS-TKA). Methods. We retrospectively evaluated knee radiographs of 165 knees, which underwent fixed-bearing PS-TKA with patella resurfacing. The incidence of patella baja and changes in patellar height over a minimum of five years of follow-up were determined using Insall-Salvati ratio (ISR) measurement. We examined whether patella baja (ISR < 0.8) at final follow-up affected clinical outcomes, knee joint range of motion (ROM), and Knee Society Score (KSS). We also assessed inter- and intrarater reliability of ISR measurements and focused on the relationship between patellar height reduction beyond measurement
Aims. Cross-table lateral (CTL) radiographs are commonly used to measure acetabular component anteversion after total hip arthroplasty (THA). The CTL measurements may differ by > 10° from CT scan measurements but the reasons for this discrepancy are poorly understood. Anteversion measurements from CTL radiographs and CT scans are compared to identify spinopelvic parameters predictive of inaccuracy. Methods. THA patients (n = 47; 27 males, 20 females; mean age 62.9 years (SD 6.95)) with preoperative spinopelvic mobility, radiological analysis, and postoperative CT scans were retrospectively reviewed. Acetabular component anteversion was measured on postoperative CTL radiographs and CT scans using 3D reconstructions of the pelvis. Two cohorts were identified based on a CTL-CT
The April 2023 Hip & Pelvis Roundup. 360. looks at: Do technical
Aims. The aim of this study was to create artificial intelligence (AI) software with the purpose of providing a second opinion to physicians to support distal radius fracture (DRF) detection, and to compare the accuracy of fracture detection of physicians with and without software support. Methods. The dataset consisted of 26,121 anonymized anterior-posterior (AP) and lateral standard view radiographs of the wrist, with and without DRF. The convolutional neural network (CNN) model was trained to detect the presence of a DRF by comparing the radiographs containing a fracture to the inconspicuous ones. A total of 11 physicians (six surgeons in training and five hand surgeons) assessed 200 pairs of randomly selected digital radiographs of the wrist (AP and lateral) for the presence of a DRF. The same images were first evaluated without, and then with, the support of the CNN model, and the diagnostic accuracy of the two methods was compared. Results. At the time of the study, the CNN model showed an area under the receiver operating curve of 0.97. AI assistance improved the physician’s sensitivity (correct fracture detection) from 80% to 87%, and the specificity (correct fracture exclusion) from 91% to 95%. The overall
The use of artificial intelligence (AI) is rapidly growing across many domains, of which the medical field is no exception. AI is an umbrella term defining the practical application of algorithms to generate useful output, without the need of human cognition. Owing to the expanding volume of patient information collected, known as ‘big data’, AI is showing promise as a useful tool in healthcare research and across all aspects of patient care pathways. Practical applications in orthopaedic surgery include: diagnostics, such as fracture recognition and tumour detection; predictive models of clinical and patient-reported outcome measures, such as calculating mortality rates and length of hospital stay; and real-time rehabilitation monitoring and surgical training. However, clinicians should remain cognizant of AI’s limitations, as the development of robust reporting and validation frameworks is of paramount importance to prevent avoidable
Aims. The aim of this study was to identify the minimal clinically important difference (MCID), minimal important change (MIC), minimal detectable change (MDC), and patient-acceptable symptom state (PASS) in the Forgotten Joint Score (FJS) according to patient satisfaction six months following total hip arthroplasty (THA) in a UK population. Methods. During a one-year period, 461 patients underwent a primary THA and completed preoperative and six-month FJS, with a mean age of 67.2 years (22 to 93). At six months, patient satisfaction was recorded as very satisfied, satisfied, neutral, dissatisfied, or very dissatisfied. The difference between patients recording neutral (n = 31) and satisfied (n = 101) was used to define the MCID. MIC for a cohort was defined as the change in the FJS for those patients declaring their outcome as satisfied, whereas receiver operating characteristic curve analysis was used to determine the MIC for an individual and the PASS. Distribution-based methodology was used to calculate the MDC. Results. Using satisfaction as the anchor, the MCID for the FJS was 8.1 (95% confidence interval (CI) 3.7 to 15.9; p = 0.040), which was affirmed when adjusting for confounding. The MIC for the FJS for a cohort of patients was 17.7 (95% CI 13.7 to 21.7) and for an individual patient was 18. The MDC90 for the FJS was eight, meaning that 90% of patients scoring more than this will have experienced a real change that is beyond measurement
Aims. To map the Oxford Knee Score (OKS) and High Activity Arthroplasty Score (HAAS) items to a common scale, and to investigate the psychometric properties of this new scale for the measurement of knee health. Methods. Patient-reported outcome measure (PROM) data measuring knee health were obtained from the NHS PROMs dataset and Total or Partial Knee Arthroplasty Trial (TOPKAT). Assumptions for common scale modelling were tested. A graded response model (fitted to OKS item responses in the NHS PROMs dataset) was used as an anchor to calibrate paired HAAS items from the TOPKAT dataset. Information curves for the combined OKS-HAAS model were plotted. Bland-Altman analysis was used to compare common scale scores derived from OKS and HAAS items. A conversion table was developed to map between HAAS, OKS, and the common scale. Results. We included 3,329 response sets from 528 patients undergoing knee arthroplasty. These generally met the assumptions of unidimensionality, monotonicity, local independence, and measurement invariance. The HAAS items provided more information than OKS items at high levels of knee health. Combining both instruments resulted in higher test-level information than either instrument alone. The mean
Aims. The Oxford Shoulder Score (OSS) is a 12-item measure commonly used for the assessment of shoulder surgeries. This study explores whether computerized adaptive testing (CAT) provides a shortened, individually tailored questionnaire while maintaining test accuracy. Methods. A total of 16,238 preoperative OSS were available in the National Joint Registry (NJR) for England, Wales, Northern Ireland, the Isle of Man, and the States of Guernsey dataset (April 2012 to April 2022). Prior to CAT, the foundational item response theory (IRT) assumptions of unidimensionality, monotonicity, and local independence were established. CAT compared sequential item selection with stopping criteria set at standard