Advertisement for orthosearch.org.uk
Results 1 - 20 of 38
Results per page:

Aims. The aim of this study was to compare the mid-term patient-reported outcome, bone remodelling, and migration of a short stem (Collum Femoris Preserving; CFP) with a conventional uncemented stem (Corail). Methods. Of 81 patients who were initially enrolled, 71 were available at five years’ follow-up. The outcomes at two years have previously been reported. The primary outcome measure was the clinical result assessed using the Oxford Hip Score (OHS). Secondary outcomes were the migration of the stem, measured using radiostereometric analysis (RSA), change of bone mineral density (BMD) around the stem, the development of radiolucent lines, and additional patient-reported outcome measures (PROMs). Results. There were no statistically significant differences between the groups regarding PROMs (median OHS (CFP 45 (interquartile range (IQR) 35 to 48); Corail 45 (IQR 40 to 48); p = 0.568). RSA showed stable stems in both groups, with little or no further subsidence between two and five years. Resorption of the femoral neck was evident in nine patients in the CFP group and in none of the 15 Corail stems with a collar that could be studied. Dual X-ray absorbiometry showed a significantly higher loss of BMD in the proximal Gruen zones in the CFP group (mean changes in BMD: Gruen zone 1, CFP -9.5 (95% confidence interval (CI) -14.8 to -4.2), Corail 1.0 (95% CI 3.4 to 5.4); Gruen zone 7, CFP -23.0 (95% CI -29.4 to -16.6), Corail -7.2 (95% CI -15.9 to 1.4). Two CFP stems were revised before two years’ follow-up due to loosening, and one Corail stem was revised after two years due to chronic infection. Conclusion. The CFP stem has a similar clinical outcome and subsidence pattern when compared with the Corail stem. More pronounced proximal stress-shielding was seen with the CFP stem, suggesting diaphyseal fixation, and questioning its femoral neck-sparing properties in the long term. Cite this article: Bone Joint J 2022;104-B(5):581–588


Bone & Joint 360
Vol. 11, Issue 2 | Pages 52 - 54
1 Apr 2022
Evans JT Evans JP Whitehouse MR


The Bone & Joint Journal
Vol. 104-B, Issue 1 | Pages 19 - 26
1 Jan 2022
Sevaldsen K Schnell Husby O Lian ØB Farran KM Schnell Husby V

Aims. Highly polished stems with force-closed design have shown satisfactory clinical results despite being related to relatively high early migration. It has been suggested that the minimal thickness of cement mantles surrounding the femoral stem should be 2 mm to 4 mm to avoid aseptic loosening. The line-to-line cementing technique of the femoral stem, designed to achieve stem press-fit, challenges this opinion. We compared the migration of a highly polished stem with force-closed design by standard and line-to-line cementing to investigate whether differences in early migration of the stems occur in a clinical study. Methods. In this single-blind, randomized controlled, clinical radiostereometric analysis (RSA) study, the migration pattern of the cemented Corail hip stem was compared between line-to-line and standard cementing in 48 arthroplasties. The primary outcome measure was femoral stem migration in terms of rotation and translation around and along with the X-, Y-, and Z- axes measured using model-based RSA at three, 12, and 24 months. A linear mixed-effects model was used for statistical analysis. Results. Results from mixed model analyses revealed a lower mean retroversion for line-to-line (0.72° (95% confidence interval (CI) 0.38° to 1.07°; p < 0.001), but no significant differences in subsidence between the techniques (-0.15 mm (95% CI -0.53 to 0.227; p = 0.429) at 24 months. Radiolucent lines measuring < 2 mm wide were found in three and five arthroplasties cemented by the standard and line-to-line method, respectively. Conclusion. The cemented Corail stem with a force-closed design seems to settle earlier and better with the line-to-line cementing method, although for subsidence the difference was not significant. However, the lower rate of migration into retroversion may reduce the wear and cement deformation, contributing to good long-term fixation and implant survival. Cite this article: Bone Joint J 2022;104-B(1):19–26


The Bone & Joint Journal
Vol. 99-B, Issue 11 | Pages 1435 - 1441
1 Nov 2017
Buttaro MA Oñativia JI Slullitel PA Andreoli M Comba F Zanotti G Piccaluga F

Aims. The Corail stem has good long-term results. After four years of using this stem, we have detected a small group of patients who have presented with symptomatic metaphyseal debonding. The aim of this study was to quantify the incidence of this complication, to delineate the characteristics of patients presenting with this complication and to compare these patients with asymptomatic controls to determine any important predisposing factors. Patients and Methods. Of 855 Corail collarless cementless stems implanted for osteoarthritis, 18 presented with symptomatic metaphyseal debonding. A control group of 74 randomly selected patients was assembled. Clinical and radiological parameters were measured and a logistic regression model was created to evaluate factors associated with metaphyseal debonding. Results. The prevalence of this complication was 2.1% in our series. In the multivariable model, the presence of a Dorr B-type proximal femur was associated with metaphyseal debonding (odds ratio (OR) 10.73, 95% confidence interval (CI) 2.31 to 49.97, p = 0.002), as was a body mass index > 25 kg/m. 2. (OR 6.85, 95% CI 1.06 to 44.28, p = 0.04). Smaller stems and the use of a polyethylene acetabular liner appeared to be protective when compared with metal and ceramic setting hard-on-hard bearings. Conclusion. We have described an uncommon but important mode of failure of the Corail stem. Surgeons should be aware of this phenomenon; overweight patients with Dorr B-type femurs and in whom hard bearings are used appear to be particularly at risk. Cite this article: Bone Joint J 2017;99-B:1435–41


The Bone & Joint Journal
Vol. 100-B, Issue 12 | Pages 1559 - 1564
1 Dec 2018
McConnell JS Saunders PRJ Young SK

Aims. Cementless femoral stems must be correctly sized and well-seated to obtain satisfactory biological fixation. The change in sound that occurs during impaction of the femoral broach is said to indicate good fit, but this has not been widely studied. We set out to find whether the presence or absence of these sound changes could predict correct sizing. Patients and Methods. We recorded the sound generated during femoral broaching for 105 cementless total hip arthroplasties using the Corail stem. Four cases were excluded, leaving 101 recordings for analysis. There were 36 male patients and 65 female patients, with a mean age of 69.9 years (. sd. 12.3) and median body mass index (BMI) of 29 kg/m. 2. (interquartile range (IQR) 26 to 32). The recordings were analyzed to identify the frequencies of the sounds produced during impaction of the femoral broach. Results. The emergence of a low-frequency band of sound in the 1 kHz range, during the final femoral broaching, was a strong predictor of a well-sized implant stem. The frequency was related to femoral length, supporting our hypothesis that the sound arose from the bone itself. Conclusion. The low-frequency sound generated during femoral broaching can be monitored spectrographically, its frequency can be predicted from femoral length, and it is a good predictor of appropriate stem sizing


The Bone & Joint Journal
Vol. 97-B, Issue 6 | Pages 741 - 748
1 Jun 2015
Bonnin MP Neto CC Aitsiselmi T Murphy CG Bossard N Roche S

The aim of this study was to investigate the relationship between the geometry of the proximal femur and the incidence of intra-operative fracture during uncemented total hip arthroplasty (THA). We studied the pre-operative CT scans of 100 patients undergoing THA with an uncemented femoral component. We measured the anteroposterior and mediolateral dimensions at the level of division of the femoral neck to calculate the aspect ratio of the femur. Wide variations in the shape of the femur were observed, from round, to very narrow elliptic. The femurs of women were narrower than those of men (p < 0.0001) and small femurs were also narrower than large ones. Patients with an intra-operative fracture of the calcar had smaller and narrower femurs than those without a fracture (p < 0.05) and the implanted Corail stems were smaller in those with a fracture (mean size 9 vs 12, p < 0.0001). The variability of the shape of the femoral neck at the level of division contributes to the understanding of the causation of intra-operative fractures in uncemented THA. Cite this article: Bone Joint J 2015;97-B:741–8


The Bone & Joint Journal
Vol. 96-B, Issue 3 | Pages 299 - 305
1 Mar 2014
Bell KR Clement ND Jenkins PJ Keating JF

We performed a case–control study to compare the rates of further surgery, revision and complications, operating time and survival in patients who were treated with either an uncemented hydroxyapatite-coated Corail bipolar femoral stem or a cemented Exeter stem for a displaced intracapsular fracture of the hip. The mean age of the patients in the uncemented group was 82.5 years (53 to 97) and in the cemented group was 82.7 years (51 to 99) We used propensity score matching, adjusting for age, gender and the presence or absence of dementia and comorbidities, to produce a matched cohort receiving an Exeter stem (n = 69) with which to compare the outcome of patients receiving a Corail stem (n = 69). The Corail had a significantly lower all-cause rate of further surgery (p = 0.016; odds ratio (OR) 0.18, 95% CI 0.04 to 0.84) and number of hips undergoing major further surgery (p = 0.029; OR 0.13, 95% CI 0.01 to 1.09). The mean operating time was significantly less for the Corail group than for the cemented Exeter group (59 min [12 to 136] vs 70 min [40 to 175], p = 0.001). The Corail group also had a lower risk of a peri-prosthetic fracture (p = 0.042; OR 0.19, 95% CI 0.01 to 1.42) . There was no difference in the mortality rate between the groups. There were significantly fewer complications in the uncemented group, suggesting that the use of this stem would result in a decreased rate of morbidity in these frail patients. Whether this relates to an improved functional outcome remains unknown. Cite this article: Bone Joint J 2014;96-B:299–305


The Bone & Joint Journal
Vol. 106-B, Issue 6 | Pages 548 - 554
1 Jun 2024
Ohyama Y Minoda Y Masuda S Sugama R Ohta Y Nakamura H

Aims

The aim of this study was to compare the pattern of initial fixation and changes in periprosthetic bone mineral density (BMD) between patients who underwent total hip arthroplasty (THA) using a traditional fully hydroxyapatite (HA)-coated stem (T-HA group) and those with a newly introduced fully HA-coated stem (N-HA group).

Methods

The study included 36 patients with T-HA stems and 30 with N-HA stems. Dual-energy X-ray absorptiometry was used to measure the change in periprosthetic BMD, one and two years postoperatively. The 3D contact between the stem and femoral cortical bone was evaluated using a density-mapping system, and clinical assessment, including patient-reported outcome measurements, was recorded.


Bone & Joint 360
Vol. 13, Issue 1 | Pages 13 - 16
1 Feb 2024

The February 2024 Hip & Pelvis Roundup360 looks at: Trial of vancomycin and cefazolin as surgical prophylaxis in arthroplasty; Is preoperative posterior femoral neck tilt a risk factor for fixation failure? Cemented versus uncemented hemiarthroplasty for displaced intracapsular fractures of the hip; Periprosthetic fractures in larger hydroxyapatite-coated stems: are collared stems a better alternative for total hip arthroplasty?; Postoperative periprosthetic fracture following hip arthroplasty with a polished taper slip versus composite beam stem; Is oral tranexamic acid as good as intravenous?; Stem design and the risk of early periprosthetic femur fractures following THA in elderly patients; Does powered femoral broaching compromise patient safety in total hip arthroplasty?


Bone & Joint Open
Vol. 5, Issue 11 | Pages 971 - 976
5 Nov 2024
Baker G Hill J O'Neill F McChesney J Stevenson M Beverland D

Aims

In 2015, we published the results of our ceramic-on-metal (CoM) total hip arthroplasties (THAs) performed between October 2007 and July 2009 with a mean follow-up of 34 months (23 to 45) and a revision rate of 3.1%. The aim of this paper is to present the longer-term outcomes.

Methods

A total of 264 patients were reviewed at a mean of 5.8 years (4.6 to 7.2) and 10.1 years (9.2 to 10.6) to determine revision rate, pain, outcome scores, radiological analysis, and blood ion levels. Those who were unwilling or unable to travel were contacted by telephone.


The Bone & Joint Journal
Vol. 106-B, Issue 9 | Pages 879 - 883
1 Sep 2024
Kayani B Staats K Haddad FS


The Bone & Joint Journal
Vol. 106-B, Issue 3 Supple A | Pages 110 - 114
1 Mar 2024
Yee AHF Chan VWK Fu H Chan P Chiu KY

Aims

The aim of this study was to evaluate the survival of a collarless, straight, hydroxyapatite-coated femoral stem in total hip arthroplasty (THA) at a minimum follow-up of 20 years.

Methods

We reviewed the results of 165 THAs using the Omnifit HA system in 138 patients, performed between August 1993 and December 1999. The mean age of the patients at the time of surgery was 46 years (20 to 77). Avascular necrosis was the most common indication for THA, followed by ankylosing spondylitis and primary osteoarthritis. The mean follow-up was 22 years (20 to 31). At 20 and 25 years, 113 THAs in 91 patients and 63 THAs in 55 patients were available for review, respectively, while others died or were lost to follow-up. Kaplan-Meier analysis was performed to evaluate the survival of the stem. Radiographs were reviewed regularly, and the stability of the stem was evaluated using the Engh classification.


The Bone & Joint Journal
Vol. 105-B, Issue 10 | Pages 1045 - 1051
1 Oct 2023
Turgeon TR Righolt CH Burnell CD Gascoyne TC Hedden DR Bohm ER

Aims

The primary aim of this trial was to compare the subsidence of two similar hydroxyapatite-coated titanium femoral components from different manufacturers. Secondary aims were to compare rotational migration (anteversion/retroversion and varus/valgus tilt) and patient-reported outcome measures between both femoral components.

Methods

Patients were randomized to receive one of the two femoral components (Avenir or Corail) during their primary total hip arthroplasty between August 2018 and September 2020. Radiostereometric analysis examinations at six, 12, and 24 months were used to assess the migration of each implanted femoral component compared to a baseline assessment. Patient-reported outcome measures were also recorded for these same timepoints. Overall, 50 patients were enrolled (62% male (n = 31), with a mean age of 65.7 years (SD 7.3), and mean BMI of 30.2 kg/m2 (SD 5.2)).


Bone & Joint Open
Vol. 2, Issue 5 | Pages 293 - 300
3 May 2021
Lewis PM Khan FJ Feathers JR Lewis MH Morris KH Waddell JP

Aims

“Get It Right First Time” (GIRFT) and NHS England’s Best Practice Tariff (BPT) have published directives advising that patients over the ages of 65 (GIRFT) and 69 years (BPT) receiving total hip arthroplasty (THA) should receive cemented implants and have brought in financial penalties if this policy is not observed. Despite this, worldwide, uncemented component use has increased, a situation described as a ‘paradox’. GIRFT and BPT do, however, acknowledge more data are required to support this edict with current policies based on the National Joint Registry survivorship and implant costs.

Methods

This study compares THA outcomes for over 1,000 uncemented Corail/Pinnacle constructs used in all age groups/patient frailty, under one surgeon, with identical pre- and postoperative pathways over a nine-year period with mean follow-up of five years and two months (range: nine months to nine years and nine months). Implant information, survivorship, and regular postoperative Oxford Hip Scores (OHS) were collected and two comparisons undertaken: a comparison of those aged over 65 years with those 65 and under and a second comparison of those aged 70 years and over with those aged under 70.


The Bone & Joint Journal
Vol. 106-B, Issue 3 Supple A | Pages 130 - 136
1 Mar 2024
Morlock M Perka C Melsheimer O Kirschbaum SM

Aims

Despite higher rates of revision after total hip arthroplasty (THA) being reported for uncemented stems in patients aged > 75 years, they are frequently used in this age group. Increased mortality after cemented fixation is often used as a justification, but recent data do not confirm this association. The aim of this study was to investigate the influence of the design of the stem and the type of fixation on the rate of revision and immediate postoperative mortality, focusing on the age and sex of the patients.

Methods

A total of 333,144 patients with primary osteoarthritis (OA) of the hip who underwent elective THA between November 2012 and September 2022, using uncemented acetabular components without reconstruction shells, from the German arthroplasty registry were included in the study. The revision rates three years postoperatively for four types of stem (uncemented, uncemented with collar, uncemented short, and cemented) were compared within four age groups: < 60 years (Young), between 61 and 70 years (Mid-I), between 71 and 80 years (Mid-II), and aged > 80 years (Old). A noninferiority analysis was performed on the most frequently used designs of stem.


Bone & Joint Open
Vol. 2, Issue 9 | Pages 696 - 704
1 Sep 2021
Malhotra R Gautam D Gupta S Eachempati KK

Aims

Total hip arthroplasty (THA) in patients with post-polio residual paralysis (PPRP) is challenging. Despite relief in pain after THA, pre-existing muscle imbalance and altered gait may cause persistence of difficulty in walking. The associated soft tissue contractures not only imbalances the pelvis, but also poses the risk of dislocation, accelerated polyethylene liner wear, and early loosening.

Methods

In all, ten hips in ten patients with PPRP with fixed pelvic obliquity who underwent THA as per an algorithmic approach in two centres from January 2014 to March 2018 were followed-up for a minimum of two years (2 to 6). All patients required one or more additional soft tissue procedures in a pre-determined sequence to correct the pelvic obliquity. All were invited for the latest clinical and radiological assessment.


Bone & Joint Research
Vol. 11, Issue 3 | Pages 180 - 188
1 Mar 2022
Rajpura A Asle SG Ait Si Selmi T Board T

Aims

Hip arthroplasty aims to accurately recreate joint biomechanics. Considerable attention has been paid to vertical and horizontal offset, but femoral head centre in the anteroposterior (AP) plane has received little attention. This study investigates the accuracy of restoration of joint centre of rotation in the AP plane.

Methods

Postoperative CT scans of 40 patients who underwent unilateral uncemented total hip arthroplasty were analyzed. Anteroposterior offset (APO) and femoral anteversion were measured on both the operated and non-operated sides. Sagittal tilt of the femoral stem was also measured. APO measured on axial slices was defined as the perpendicular distance between a line drawn from the anterior most point of the proximal femur (anterior reference line) to the centre of the femoral head. The anterior reference line was made parallel to the posterior condylar axis of the knee to correct for rotation.


The Bone & Joint Journal
Vol. 102-B, Issue 12 | Pages 1654 - 1661
1 Dec 2020
Perelgut ME Polus JS Lanting BA Teeter MG

Aims

The direct anterior (DA) approach has been associated with rapid patient recovery after total hip arthroplasty (THA) but may be associated with more frequent femoral complications including implant loosening. The objective of this study was to determine whether the addition of a collar to the femoral stem affects implant migration, patient activity, and patient function following primary THA using the DA approach.

Methods

Patients were randomized to either a collared (n = 23) or collarless (n = 26) cementless femoral stem implanted using the DA approach. Canal fill ratio (CFR) was measured on the first postoperative radiographs. Patients underwent a supine radiostereometric analysis (RSA) exam postoperatively on the day of surgery and at two, four, six, 12, 26, and 52 weeks postoperatively. Patient-reported outcome measures (Western Ontario and McMaster Universities Osteoarthritis (WOMAC) Index, the 12-item Short Form Health Survey Mental and Physical Score, and University of California, Los Angeles (UCLA) Activity Score) were measured preoperatively and at each post-surgery clinic visit. Activity and function were also measured as the weekly average step count recorded by an activity tracker, and an instrumented timed up-and-go (TUG) test in clinic, respectively.


The Bone & Joint Journal
Vol. 100-B, Issue 8 | Pages 1010 - 1017
1 Aug 2018
Jain S Magra M Dube B Veysi VT Whitwell GS Aderinto JB Emerton ME Stone MH Pandit HG

Aims

This study aimed to evaluate implant survival of reverse hybrid total hip arthroplasty (THA) at medium-term follow-up.

Patients and Methods

A consecutive series of 1082 THAs in 982 patients with mean follow-up of 7.9 years (5 to 11.3) is presented. Mean age was 69.2 years (21 to 94). Of these, 194 (17.9%) were in patients under 60 years, 663 (61.3%) in female patients and 348 (32.2%) performed by a trainee. Head size was 28 mm in 953 hips (88.1%) or 32 mm in 129 hips (11.9%). Survival analysis was performed and subgroups compared using log rank tests.


The Bone & Joint Journal
Vol. 101-B, Issue 7 | Pages 779 - 786
1 Jul 2019
Lamb JN Baetz J Messer-Hannemann P Adekanmbi I van Duren BH Redmond A West RM Morlock MM Pandit HG

Aims

The aim of this study was to estimate the 90-day risk of revision for periprosthetic femoral fracture associated with design features of cementless femoral stems, and to investigate the effect of a collar on this risk using a biomechanical in vitro model.

Materials and Methods

A total of 337 647 primary total hip arthroplasties (THAs) from the United Kingdom National Joint Registry (NJR) were included in a multivariable survival and regression analysis to identify the adjusted hazard of revision for periprosthetic fracture following primary THA using a cementless stem. The effect of a collar in cementless THA on this risk was evaluated in an in vitro model using paired fresh frozen cadaveric femora.