A 66-year-old man presented with an infected resurfacing hip replacement in the immediate post-operative period following an uneventful transurethral resection of the prostate. Prophylactic antibiotics had been administered on induction of anaesthesia. The hip prosthesis had been inserted seven years previously and had been hitherto asymptomatic. The hip was washed out and microbiological examination identified Despite current recommendations, clinicians undertaking invasive procedures which can lead to bacteraemia in patients with prosthetic joint replacements should be aware of the risk of haematogenous seeding in such prostheses, which although low, has potentially disastrous consequences.
We describe three cases of fracture of the titanium JRI-Furlong hydroxyapatite-ceramic (HAC)-coated femoral component. We have examined previous case reports of failure of this stem and conclude that fracture may occur in two places, namely at the neck-shoulder junction and at the conical-distal cylindrical junction. These breakages are the result of fatigue in a metallurgically-proven normal femoral component. All the cases of failure of the femoral component have occurred in patients with a body mass index of more than 25 in whom a small component, either size 9 or 10, had been used. In patients with a body mass index above normal size 9 components should be avoided and the femoral canal should be reamed sufficiently to accept a large femoral component to ensure that there is adequate metaphyseal fixation.