Advertisement for orthosearch.org.uk
Results 1781 - 1800 of 2331
Results per page:
The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 12 | Pages 1666 - 1669
1 Dec 2012
Gulotta LV Choi D Marinello P Wright T Cordasco FA Craig EV Warren RF

Reverse total shoulder replacement (RTSR) depends on adequate deltoid function for a successful outcome. However, the anterior deltoid and/or axillary nerve may be damaged due to prior procedures or injury. The purpose of this study was to determine the compensatory muscle forces required for scapular plane elevation following RTSR when the anterior deltoid is deficient. The soft tissues were removed from six cadaver shoulders, except for tendon attachments. After implantation of the RTSR, the shoulders were mounted on a custom-made shoulder simulator to determine the mean force in each muscle required to achieve 30° and 60° of scapular plane elevation. Two conditions were tested: 1) Control with an absent supraspinatus and infraspinatus; and 2) Control with anterior deltoid deficiency. Anterior deltoid deficiency resulted in a mean increase of 195% in subscapularis force at 30° when compared with the control (p = 0.02). At 60°, the subscapularis force increased a mean of 82% (p < 0.001) and the middle deltoid force increased a mean of 26% (p = 0.04).

Scapular plane elevation may still be possible following an RTSR in the setting of anterior deltoid deficiency. When the anterior deltoid is deficient, there is a compensatory increase in the force required by the subscapularis and middle deltoid. Attempts to preserve the subscapularis, if present, might maximise post-operative function.


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 8 | Pages 1025 - 1031
1 Aug 2008
Mizu-uchi H Matsuda S Miura H Okazaki K Akasaki Y Iwamoto Y

We compared the alignment of 39 total knee replacements implanted using the conventional alignment guide system with 37 implanted using a CT-based navigation system, performed by a single surgeon. The knees were evaluated using full-length weight-bearing anteroposterior radiographs, lateral radiographs and CT scans.

The mean hip-knee-ankle angle, coronal femoral component angle and coronal tibial component angle were 181.8° (174.2° to 188.3°), 88.5° (84.0° to 91.8°) and 89.7° (86.3° to 95.1°), respectively for the conventional group and 180.8° (178.2° to 185.1°), 89.3° (85.8° to 92.0°) and 89.9° (88.0° to 93.0°), respectively for the navigated group.

The mean sagittal femoral component angle was 85.5° (80.6° to 92.8°) for the conventional group and 89.6° (85.5° to 94.0°) for the navigated group.

The mean rotational femoral and tibial component angles were −0.7° (−8.8° to 9.8°) and −3.3° (−16.8° to 5.8°) for the conventional group and −0.6° (−3.5° to 3.0°) and 0.3° (−5.3° to 7.7°) for the navigated group.

The ideal angles of all alignments in the navigated group were obtained at significantly higher rates than in the conventional group. Our results demonstrated significant improvements in component positioning with a CT-based navigation system, especially with respect to rotational alignment.


The Bone & Joint Journal
Vol. 95-B, Issue 1 | Pages 106 - 110
1 Jan 2013
Jeyaseelan L Singh VK Ghosh S Sinisi M Fox M

We present our experience of managing patients with iatropathic brachial plexus injury after delayed fixation of a fracture of the clavicle. It is a retrospective cohort study of patients treated at our peripheral nerve injury unit and a single illustrative case report. We identified 21 patients in whom a brachial plexus injury occurred as a direct consequence of fixation of a fracture of the clavicle between September 2000 and September 2011.

The predominant injury involved the C5/C6 nerves, upper trunk, lateral cord and the suprascapular nerve. In all patients, the injured nerve was found to be tethered to the under surface of the clavicle by scar tissue at the site of the fracture and was usually associated with pathognomonic neuropathic pain and paralysis.

Delayed fixation of a fracture of the clavicle, especially between two and four weeks after injury, can result in iatropathic brachial plexus injury. The risk can be reduced by thorough release of the tissues from the inferior surface of the clavicle before mobilisation of the fracture fragments. If features of nerve damage appear post-operatively urgent specialist referral is recommended.

Cite this article: Bone Joint J 2013;95-B:106–10.


The Bone & Joint Journal
Vol. 95-B, Issue 1 | Pages 70 - 74
1 Jan 2013
Dattani R Smith CD Patel VR

We investigated the incidence of and risk factors for venous thromboembolism (VTE) following surgery of the shoulder and elbow and assessed the role of thromboprophylaxis in upper limb surgery. All papers describing VTE after shoulder and elbow surgery published in the English language literature before 31 March 2012 were reviewed. A total of 14 papers were available for analysis, most of which were retrospective studies and case series. The incidence of VTE was 0.038% from 92 440 shoulder arthroscopic procedures, 0.52% from 42 261 shoulder replacements, and 0.64% from 4833 procedures for fractures of the proximal humerus (open reduction and internal fixation or hemiarthroplasty). The incidence following replacement of the elbow was 0.26% from 2701 procedures. Diabetes mellitus, rheumatoid arthritis and ischaemic heart disease were identified as the major risk factors.

The evidence that exists on thromboprophylaxis is based on level III and IV studies, and we therefore cannot make any recommendations on prophylaxis based on the current evidence. It seems reasonable to adopt a multimodal approach that involves all patients receiving mechanical prophylaxis, with chemical prophylaxis reserved for those who are at high risk for VTE.

Cite this article: Bone Joint J 2013;95-B:70–4.


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 10 | Pages 1332 - 1337
1 Oct 2010
Leung KS Tang N Cheung LWH Ng E

Fluoronavigation is an image-guided technology which uses intra-operative fluoroscopic images taken under a real-time tracking system and registration to guide surgical procedures. With the skeleton and the instrument registered, guidance under an optical tracking system is possible, allowing fixation of the fracture and insertion of an implant. This technology helps to minimise exposure to x-rays, providing multiplanar views for monitoring and accurate positioning of implants. It allows real-time interactive quantitative data for decision-making and expands the application of minimally invasive surgery. In orthopaedic trauma its use can be further enhanced by combining newer imaging technologies such as intra-operative three-dimensional fluoroscopy and optical image guidance, new advances in software for fracture reduction, and new tracking mechanisms using electromagnetic technology. The major obstacles for general and wider applications are the inability to track individual fracture fragments, no navigated real-time fracture reduction, and the lack of an objective assessment method for cost-effectiveness.

We believe that its application will go beyond the operating theatre and cover all aspects of patient management, from pre-operative planning to intra-operative guidance and postoperative rehabilitation.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 11_Supple_A | Pages 52 - 54
1 Nov 2012
Rodriguez JA Rathod PA

Large femoral heads have been used with increasing frequency over the last decade. The prime reason is likely the effect of large heads on stability. The larger head neck ratio, combined with the increased jump distance of larger heads result in a greater arc of impingement free motion, and greater resistance to dislocation in a provocative position. Multiple studies have demonstrated clear clinical efficacy in diminishing dislocation rates with the use of large femoral heads. With crosslinked polyethylene, wear has been shown to be equivalent between larger and smaller heads. However, the stability advantages of increasing diameter beyond 38 mm have not been clearly demonstrated. More importantly, recent data implicates large heads in the increasing prevalence of groin pain and psoas impingement. There are clear benefits with larger femoral head diameters, but the advantages of diameters beyond 38 mm have not yet been demonstrated clinically.


Bone & Joint 360
Vol. 1, Issue 6 | Pages 23 - 25
1 Dec 2012

The December 2012 Trauma Roundup360 looks at: whether tranexamic acid stops bleeding in trauma across the board; antibiotic beads and VAC; whether anaesthetic determines the outcome in surgery for distal radial fractures; high complications in surgery on bisphosphonate-hardened bone; better outcomes but more dislocations in femoral neck fractures; the mythical hip fracture; plate augmentation in nonunion surgery; and SIGN intramedullary nailing and infections.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 10 | Pages 1281 - 1286
1 Oct 2009
Olsen M Davis ET Chiu M Gamble P Tumia N Boyle RA Schemitsch EH

The computed neck-shaft angle and the size of the femoral component were recorded in 100 consecutive hip resurfacings using imageless computer-navigation and compared with the angle measured before operation and with actual component implanted. The reliability of the registration was further analysed using ten cadaver femora. The mean absolute difference between the measured and navigated neck-shaft angle was 16.3° (0° to 52°). Navigation underestimated the measured neck-shaft angle in 38 patients and the correct implant size in 11. Registration of the cadaver femora tended to overestimate the correct implant size and provided a low level of repeatability in computing the neck-shaft angle.

Prudent pre-operative planning is advisable for use in conjunction with imageless navigation since misleading information may be registered intraoperatively, which could lead to inappropriate sizing and positioning of the femoral component in hip resurfacing.


The Bone & Joint Journal
Vol. 95-B, Issue 2 | Pages 151 - 159
1 Feb 2013
Duckworth AD McQueen MM Ring D

Most fractures of the radial head are stable undisplaced or minimally displaced partial fractures without an associated fracture of the elbow or forearm or ligament injury, where stiffness following non-operative management is the primary concern. Displaced unstable fractures of the radial head are usually associated with other fractures or ligament injuries, and restoration of radiocapitellar contact by reconstruction or prosthetic replacement of the fractured head is necessary to prevent subluxation or dislocation of the elbow and forearm. In fractures with three or fewer fragments (two articular fragments and the neck) and little or no metaphyseal comminution, open reduction and internal fixation may give good results. However, fragmented unstable fractures of the radial head are prone to early failure of fixation and nonunion when fixed. Excision of the radial head is associated with good long-term results, but in patients with instability of the elbow or forearm, prosthetic replacement is preferred.

This review considers the characteristics of stable and unstable fractures of the radial head, as well as discussing the debatable aspects of management, in light of the current best evidence.

Cite this article: Bone Joint J 2013;95-B:151–9.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 7 | Pages 970 - 974
1 Jul 2011
Giannoudis PV Kanakaris NK Dimitriou R Mallina R Smith RM

Isolated fractures of the anterior column and anterior wall are a relatively rare subgroup of acetabular fractures. We report our experience of 30 consecutive cases treated over ten years. Open reduction and internal fixation through an ilioinguinal approach was performed for most of these cases (76.7%) and percutaneous techniques were used for the remainder. At a mean follow-up of four years (2 to 6), 26 were available for review. The radiological and functional outcomes were good or excellent in 23 of 30 patients (76.7%) and 22 of 26 patients (84.6%) according to Matta’s radiological criteria and the modified Merlé d’Aubigné score, respectively.

Complications of minor to moderate severity were seen in six of the 30 cases (20%) and none of the patients underwent secondary surgery or replacement of the hip.


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 10 | Pages 1392 - 1396
1 Oct 2010
Wang J Chen W Lin P Hsu C Wang C

Intra-articular resection of bone with soft-tissue balancing and total knee replacement (TKR) has been described for the treatment of patients with severe osteoarthritis of the knee associated with an ipsilateral malunited femoral fracture. However, the extent to which deformity in the sagittal plane can be corrected has not been addressed. We treated 12 patients with severe arthritis of the knee and an extra-articular malunion of the femur by TKR with intra-articular resection of bone and soft-tissue balancing. The femora had a mean varus deformity of 16° (8° to 23°) in the coronal plane. There were seven recurvatum deformities with a mean angulation of 11° (6° to 15°) and five antecurvatum deformities with a mean angulation of 12° (6° to 15°).

The mean follow-up was 93 months (30 to 155). The median Knee Society knee and function scores improved from 18.7 (0 to 49) and 24.5 (10 to 50) points pre-operatively to 93 (83 to 100) and 90 (70 to 100) points at the time of the last follow-up, respectively. The mean mechanical axis of the knee improved from 22.6° of varus (15° to 27° pre-operatively to 1.5° of varus (3° of varus to 2° of valgus) at the last follow-up. The recurvatum deformities improved from a mean of 11° (6° to 15°) pre-operatively to 3° (0° to 6°) at the last follow-up. The antecurvatum deformities in the sagittal plane improved from a mean of 12° (6° to 16°) pre-operatively to 4.4° (0° to 8°) at the last follow-up.

Apart from varus deformities, TKR with intra-articular bone resection effectively corrected the extra-articular deformity of the femur in the presence of antecurvatum of up to 16° and recurvatum of up to 15°.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 8 | Pages 1088 - 1092
1 Aug 2011
Lizaur A Sanz-Reig J Gonzalez-Parreño S

The purpose of this study was to review the long-term outcomes of a previously reported prospective series of 46 type III acromioclavicular dislocations. These were treated surgically with temporary fixation of the acromioclavicular joint with wires, repair of the acromioclavicular ligaments, and overlapped suture of the deltoid and trapezius muscles. Of the 46 patients, one had died, four could not be traced, and three declined to return for follow-up, leaving 38 patients in the study. There were 36 men and two women, with a mean age at follow-up of 57.3 years (41 to 71). The mean follow-up was 24.2 years (21 to 26).

Patients were evaluated using the Imatani and University of California, Los Angeles (UCLA) scoring systems. Their subjective status was assessed using the Disabilities of the Arm, Shoulder and Hand and Simple Shoulder Test questionnaires, and a visual analogue scale for patient satisfaction. The examination included radiographs of the shoulder.

At a follow-up of 21 years, the results were satisfactory in 35 (92.1%) patients and unsatisfactory in three (7.9%). In total, 35 patients (92.1%) reported no pain, one slight pain, and two moderate pain. All except two patients had a full range of shoulder movement compared with the opposite side. Unsatisfactory results were the result of early redisplacement in two patients, and osteoarthritis without redisplacement in one. According to the Imatani and UCLA scores, there was no difference between the operated shoulder and the opposite shoulder (p > 0.05). Given the same situation, 35 (92.1%) patients would opt for the same surgical treatment again.

Operative treatment of type III acromioclavicular joint injuries produces satisfactory long-term results.


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 11 | Pages 1574 - 1579
1 Nov 2010
Hariri A Mascard E Atlan F Germain MA Heming N Dubousset JF Wicart P

We describe a retrospective review of 38 cases of reconstruction following resection of the metaphysiodiaphysis of the lower limb for malignant bone tumours using free vascularised fibular grafts. The mean follow-up was for 7.6 years (0.4 to 18.4). The mean Musculoskeletal Tumor Society score was 27.2 (20 to 30). The score was significantly higher when the graft was carried out in a one-stage procedure after resection of the tumour rather than in two stages. Bony union was achieved in 89% of the cases. The overall mean time to union was 1.7 years (0.2 to 10.3).

Free vascularised fibular transfer is a major operation with frequent, but preventable, complications which allows salvage of the limb with satisfactory functional results.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 11_Supple_A | Pages 85 - 89
1 Nov 2012
Drexler M Dwyer T Marmor M Abolghasemian M Sternheim A Cameron HU

In this study we present our experience with four generations of uncemented total knee arthroplasty (TKA) from Smith & Nephew: Tricon M, Tricon LS, Tricon II and Profix, focusing on the failure rates correlating with each design change. Beginning in 1984, 380 Tricon M, 435 Tricon LS, 305 Tricon 2 and 588 Profix were implanted by the senior author. The rate of revision for loosening was 1.1% for the Tricon M, 1.1% for the Tricon LS, 0.5% for the Tricon 2 with a HA coated tibial component, and 1.3% for the Profix TKA. No loosening of the femoral component was seen with the Tricon M, Tricon LS or Tricon 2, with no loosening seen of the tibial component with the Profix TKA. Regarding revision for wear, the incidence was 13.1% for the Tricon M, 6.6% for the Tricon LS, 2.3% for the Tricon 2, and 0% for the Profix. These results demonstrate that improvements in the design of uncemented components, including increased polyethylene thickness, improved polyethylene quality, and the introduction of hydroxyapatite coating, has improved the outcomes of uncemented TKA over time.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 12 | Pages 1607 - 1611
1 Dec 2009
Stufkens SAS Knupp M Lampert C van Dijk CN Hintermann B

We have compared the results at a mean follow-up of 13 years (11 to 14) of two groups of supination-external rotation type-4 fractures of the ankle, in one of which there was a fracture of the medial malleolus and in the other the medial deltoid ligament had been partially or completely ruptured.

Of 66 patients treated operatively between 1993 and 1997, 36 were available for follow-up. Arthroscopy had been performed in all patients pre-operatively to assess the extent of the intra-articular lesions. The American Orthopaedic Foot and Ankle Society hind-foot score was used for clinical evaluation and showed a significant difference in both the total and the functional scores (p < 0.05), but not in those for pain or alignment, in favour of the group with a damaged deltoid ligament (p < 0.05). The only significant difference between the groups on the short-form 36 quality-of-life score was for bodily pain, again in favour of the group with a damaged deltoid ligament. There was no significant difference between the groups in the subjective visual analogue scores or in the modified Kannus radiological score.

Arthroscopically, there was a significant difference with an increased risk of loose bodies in the group with an intact deltoid ligament (p < 0.005), although there was no significant increased risk of deep cartilage lesions in the two groups.

At a mean follow-up of 13 years after operative treatment of a supination-external rotation type-4 ankle fracture patients with partial or complete rupture of the medial deltoid ligament tended to have a better result than those with a medial malleolar fracture.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 2 | Pages 162 - 169
1 Feb 2009
Bardakos NV Villar RN

Although the association between femoroacetabular impingement and osteoarthritis is established, it is not yet clear which hips have the greatest likelihood to progress rapidly to end-stage disease. We investigated the effect of several radiological parameters, each indicative of a structural aspect of the hip joint, on the progression of osteoarthritis. Pairs of plain anteroposterior pelvic radiographs, taken at least ten years apart, of 43 patients (43 hips) with a pistol-grip deformity of the femur and mild (Tönnis grade 1) or moderate (Tönnis grade 2) osteoarthritis were reviewed. Of the 43 hips, 28 showed evidence of progression of osteoarthritis. There was no significant difference in the prevalence of progression between hips with initial Tönnis grade 1 or grade 2 osteoarthritis (p = 0.31). Comparison of the hips with and without progression of arthritis revealed a significant difference in the mean medial proximal femoral angle (81° vs 87°, p = 0.004) and the presence of the posterior wall sign (39% vs 7%, p = 0.02) only. A logistic regression model was constructed to predict the influence of these two variables in the development of osteoarthritis.

Mild to moderate osteoarthritis in hips with a pistol-grip deformity will not progress rapidly in all patients. In one-third, progression will take more than ten years to manifest, if ever. The individual geometry of the proximal femur and acetabulum partly influences this phenomenon. A hip with cam impingement is not always destined for end-stage arthritic degeneration.


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 8 | Pages 1096 - 1099
1 Aug 2010
Sutherland AG Cooper K Alexander LA Nicol M Smith FW Scotland TR

We identified a series of 128 patients who had unilateral open reconstruction of the anterior cruciate ligament (ACL) by a single surgeon between 1993 and 2000. In all, 79 patients were reviewed clinically and radiologically eight to 15 years after surgery. Assessment included measurement of the Lysholm and Tegner scores, the ACL quality-of-life score and the Short Form-12 score, as well as the International Knee Documentation Committee clinical assessment, measurement of laxity by the KT-1000 arthrometer, a single-leg hop test and standardised radiography of both knees using the uninjured knee as a control.

Of the injured knees, 46 (57%) had definite radiological evidence of osteoarthritis (Kellgren-Lawrence grade 2 or 3), with a mean difference between the injured and non-injured knees of 1.2 grades. The median ACL quality-of-life score was 80 (interquartile range (IQR) 60 to 90), the Lysholm score 84 (IQR 74 to 95), the Short Form-12 physical component score 54 (IQR 49 to 56) and the mean Hop Index 0.94 (0.52 to 1.52). In total 58 patients were graded as normal, 20 as nearly normal and one as abnormal on the KT-1000 assessment and pivot-shift testing. Taking the worst-case scenario of assuming all non-attenders (n = 48), two septic failures and one identified unstable knee found at review to be failures, the failure rate was 40%. Only two of the patients reviewed stated that they would not have similar surgery again.

Open reconstruction of the ACL gives good, durable functional results, but with a high rate of radiologically evident osteoarthritis.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 3 | Pages 310 - 315
1 Mar 2009
Olsen M Davis ET Waddell JP Schemitsch EH

We have investigated the accuracy of placement of the femoral component using imageless navigation in 100 consecutive Birmingham Hip Resurfacings. Pre-operative templating determined the native neck-shaft angle and planned stem-shaft angle of the implant. The latter were verified post-operatively using digital anteroposterior unilateral radiographs of the hip.

The mean neck-shaft angle determined before operation was 132.7° (118° to 160°). The mean planned stem-shaft angle was a relative valgus alignment of 9.7° (sd 2.6). The stem-shaft angle after operation differed from that planned by a mean of 2.8° (sd 2.0) and in 86% of cases the final angle measured within ± 5° of that planned. We had no instances of notching of the neck or varus alignment of the implant in our series. A learning curve was observed in the time taken for navigation, but not for accurate placement of the implant.

Navigation in hip resurfacing may afford the surgeon a reliable and accurate method of placement of the femoral component.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 8 | Pages 1098 - 1103
1 Aug 2011
Ruggieri P Mavrogenis AF Guerra G Mercuri M

We retrospectively studied 14 patients with proximal and diaphyseal tumours and disappearing bone (Gorham’s) disease of the humerus treated with wide resection and reconstruction using an allograft-resurfacing composite (ARC). There were ten women and four men, with a mean age of 35 years (8 to 69). At a mean follow-up of 25 months (10 to 89), two patients had a fracture of the allograft. In one of these it was revised with a similar ARC and in the other with an intercalary prosthesis. A further patient had an infection and a fracture of the allograft that was revised with a megaprosthesis. In all patients with an ARC, healing of the ARC-host bone interface was observed. One patient had failure of the locking mechanism of the total elbow replacement. The mean post-operative Musculoskeletal Tumor Society score for the upper extremity was 77% (46.7% to 86.7%), which represents good and excellent results; one patient had a poor result (46.7%).

In the short term ARC effectively relieves pain and restores shoulder function in patients with wide resection of the proximal humerus. Fracture and infection remain significant complications.


The Bone & Joint Journal
Vol. 95-B, Issue 5 | Pages 583 - 597
1 May 2013
Kurien T Pearson RG Scammell BE

We reviewed 59 bone graft substitutes marketed by 17 companies currently available for implantation in the United Kingdom, with the aim of assessing the peer-reviewed literature to facilitate informed decision-making regarding their use in clinical practice. After critical analysis of the literature, only 22 products (37%) had any clinical data. Norian SRS (Synthes), Vitoss (Orthovita), Cortoss (Orthovita) and Alpha-BSM (Etex) had Level I evidence. We question the need for so many different products, especially with limited published clinical evidence for their efficacy, and conclude that there is a considerable need for further prospective randomised trials to facilitate informed decision-making with regard to the use of current and future bone graft substitutes in clinical practice.

Cite this article: Bone Joint J 2013;95-B:583–97.