Instability is the reason for revision of a primary
total knee replacement (TKR) in 20% of patients. To date, the diagnosis
of instability has been based on the patient’s symptoms and a subjective
clinical assessment. We assessed whether a measured standardised
forced leg extension could be used to quantify instability. A total of 25 patients (11 male/14 female, mean age 70 years;
49 to 85) who were to undergo a revision TKR for instability of
a primary implant were assessed with a Nottingham rig pre-operatively
and then at six and 26 weeks post-operatively. Output was quantified
(in revolutions per minute (rpm)) by accelerating a stationary flywheel.
A control group of 183 patients (71 male/112 female, mean age 69
years) who had undergone primary TKR were evaluated for comparison. Pre-operatively, all 25 patients with instability exhibited a
distinctive pattern of reduction in ‘mid-push’ speed. The mean reduction
was 55 rpm ( Cite this article:
As part of the national initiative to reduce
waiting times for joint replacement surgery in Wales, the Cardiff
and Vale NHS Trust referred 224 patients to the NHS Treatment Centre
in Weston-Super-Mare for total knee replacement (TKR). A total of
258 Kinemax TKRs were performed between November 2004 and August
2006. Of these, a total of 199 patients (232 TKRs, 90%) have been
followed up for five years. This cohort was compared with 258 consecutive
TKRs in 250 patients, performed at Cardiff and Vale Orthopaedic
Centre (CAVOC) over a similar time period. The five year cumulative
survival rate was 80.6% (95% confidence interval (CI) 74.0 to 86.0)
in the Weston-Super-Mare cohort and 95.0% (95% CI 90.2 to 98.2)
in the CAVOC cohort with revision for any reason as the endpoint.
The relative risk for revision at Weston-Super-Mare compared with
CAVOC was 3.88 (p <
0.001). For implants surviving five years,
the mean Oxford knee scores (OKS) and mean EuroQol (EQ-5D) scores
were similar (OKS: Weston-Super-Mare 29 (2 to 47) The results show a higher revision rate for patients operated
at Weston-Super-Mare Treatment Centre, with a reduction in functional
outcome and quality of life after revision. This further confirms
that patients moved from one area to another for joint replacement
surgery fare poorly.
The April 2014 Oncology Roundup360 looks at: Eyeball as good as microscope for tumour margins; when is best to stabilise femoral metastases?; fluorine does not cause bone tumours; whether giant cell tumour of the proximal femur ever successfully managed; extraskeletal osteosarcoma; modular lower limb tumour reconstruction; and observational studies the basis for most bone tumour treatment.
Mechanical alignment has been a fundamental tenet of total knee arthroplasty (TKA) since modern knee replacement surgery was developed in the 1970s. The objective of mechanical alignment was to infer the greatest biomechanical advantage to the implant to prevent early loosening and failure. Over the last 40 years a great deal of innovation in TKA technology has been focusing on how to more accurately achieve mechanical alignment. Recently the concept of mechanical alignment has been challenged, and other alignment philosophies are being explored with the intention of trying to improve patient outcomes following TKA. This article examines the evolution of the mechanical alignment concept and whether there are any viable alternatives.
We are entering a new era with governmental bodies
taking an increasingly guiding role, gaining control of registries,
demanding direct access with release of open public information
for quality comparisons between hospitals. This review is written
by physicians and scientists who have worked with the Swedish Knee
Arthroplasty Register (SKAR) periodically since it began. It reviews
the history of the register and describes the methods used and lessons
learned. Cite this article:
We describe our technique and rationale using
hybrid fixation for primary total hip arthroplasty (THA) at the Hospital
for Special Surgery. Modern uncemented acetabular components have
few screw holes, or no holes, polished inner surfaces, improved
locking mechanisms, and maximised thickness and shell-liner conformity. Uncemented
sockets can be combined with highly cross-linked polyethylene liners,
which have demonstrated very low wear and osteolysis rates after
ten to 15 years of implantation. The results of cement fixation
with a smooth or polished surface finished stem have been excellent,
virtually eliminating complications seen with cementless fixation
like peri-operative femoral fractures and thigh pain. Although mid-term
results of modern cementless stems are encouraging, the long-term
data do not show reduced revision rates for cementless stems compared
with cemented smooth stems. In this paper we review the conduct
of a hybrid THA, with emphasis on pre-operative planning, surgical
technique, hypotensive epidural anaesthesia, and intra-operative
physiology. Cite this article:
We investigated the static and cyclical strength of parallel and angulated locking plate screws using rigid polyurethane foam (0.32 g/cm3) and bovine cancellous bone blocks. Custom-made stainless steel plates with two conically threaded screw holes with different angulations (parallel, 10° and 20° divergent) and 5 mm self-tapping locking screws underwent pull-out and cyclical pull and bending tests. The bovine cancellous blocks were only subjected to static pull-out testing. We also performed finite element analysis for the static pull-out test of the parallel and 20° configurations. In both the foam model and the bovine cancellous bone we found the significantly highest pull-out force for the parallel constructs. In the finite element analysis there was a 47% more damage in the 20° divergent constructs than in the parallel configuration. Under cyclical loading, the mean number of cycles to failure was significantly higher for the parallel group, followed by the 10° and 20° divergent configurations. In our laboratory setting we clearly showed the biomechanical disadvantage of a diverging locking screw angle under static and cyclical loading.
We reviewed 87 patients who underwent revision reconstruction of the anterior cruciate ligament. The incidence of meniscal tears and degenerative change was assessed and related to the interval between failure of the primary graft and revision reconstruction. Patients were divided into two groups: early revision surgery within six months of graft failure, and delayed revision. Degenerative change was scored using the French Society of Arthroscopy system. There was a significantly higher incidence of articular cartilage degeneration in the delayed group (Mann-Whitney U-test, 53.2% We conclude that revision reconstruction should be carried out within six months of primary graft failure, in order to minimise the risk of degenerative change.
Isolated patellofemoral arthritis is a common
condition and there are varying opinions on the most effective treatments.
Non-operative and operative treatments have failed to demonstrate
effective long-term treatment for those in an advanced stage of
the condition. Newer designs and increased technology in patellofemoral replacement
(PFR) have produced more consistent outcomes. This has led to a
renewed enthusiasm for this procedure. Newer PFR prostheses have
addressed the patellar maltracking issues plaguing some of the older designs.
Short-term results with contemporary prostheses and new technology
are described here. Cite this article:
Total hip replacement (THR) is a very common
procedure undertaken in up to 285 000 Americans each year. Patient
satisfaction with THR is very high, with improvements in general
health, quality of life, and function while at the same time very
cost effective. Although the majority of patients have a high degree
of satisfaction with their THR, 27% experience some discomfort,
and up to 6% experience severe chronic pain. Although it can be
difficult to diagnose the cause of the pain in these patients, this
clinical issue should be approached systematically and thoroughly.
A detailed history and clinical examination can often provide the
correct diagnosis and guide the appropriate selection of investigations, which
will then serve to confirm the clinical diagnosis made. Cite this article:
A modular femoral head–neck junction has practical
advantages in total hip replacement. Taper fretting and corrosion
have so far been an infrequent cause of revision. The role of design
and manufacturing variables continues to be debated. Over the past
decade several changes in technology and clinical practice might
result in an increase in clinically significant taper fretting and
corrosion. Those factors include an increased usage of large diameter
(36 mm) heads, reduced femoral neck and taper dimensions, greater
variability in taper assembly with smaller incision surgery, and
higher taper stresses due to increased patient weight and/or physical
activity. Additional studies are needed to determine the role of
taper assembly compared with design, manufacturing and other implant
variables. Cite this article:
This study compared the primary stability of two commercially
available acetabular components from the same manufacturer, which
differ only in geometry; a hemispherical and a peripherally enhanced
design (peripheral self-locking (PSL)). The objective was to determine
whether altered geometry resulted in better primary stability. Acetabular components were seated with 0.8 mm to 2 mm interference
fits in reamed polyethylene bone substrate of two different densities
(0.22 g/cm3 and 0.45 g/cm3). The primary stability
of each component design was investigated by measuring the peak
failure load during uniaxial pull-out and tangential lever-out tests.Objective
Methods
We report on the long-term results of 163 bicruciate-retaining
Hermes 2C total knee replacements in 130 patients at a mean follow-up
of 22.4 years (20.3 to 23.5). Even when the anterior cruciate ligament
had a partially degenerative appearance it was preserved as long
as the knee had a normal anterior drawer and Lachman’s test pre-operatively. The
description and surgical technique of this minimally constrained
prosthesis were published in 1983 and the ten-year clinical results
in 1999. A total of 12% of the knees (20 of 163) in this study were revised
because of wear of the polyethylene tibial insert. Excellent stability
was achieved and the incidence of aseptic component loosening was
4.3% (seven of 163). The survival rate using revision for any reason
as the endpoint was 82% (95% confidence interval 76.2 to 88.0). Although this series included a relatively small number of replacements,
it demonstrated that the anterior cruciate ligament, even when partially
degenerated at the time of TKR, remained functional and provided
adequate stability at a long-term follow-up. Cite this article:
We performed a systematic review and meta-analysis
of modern total ankle replacements (TARs) to determine the survivorship,
outcome, complications, radiological findings and range of movement,
in patients with end-stage osteoarthritis (OA) of the ankle who
undergo this procedure. We used the methodology of the Cochrane Collaboration,
which uses risk of bias profiling to assess the quality of papers
in favour of a domain-based approach. Continuous outcome scores
were pooled across studies using the generic inverse variance method
and the random-effects model was used to incorporate clinical and
methodological heterogeneity. We included 58 papers (7942 TARs)
with an interobserver reliability (Kappa) for selection, performance,
attrition, detection and reporting bias of between 0.83 and 0.98.
The overall survivorship was 89% at ten years with an annual failure
rate of 1.2% (95% confidence interval (CI) 0.7 to 1.6). The mean
American Orthopaedic Foot and Ankle Society score changed from 40 (95%
CI 36 to 43) pre-operatively to 80 (95% CI 76 to 84) at a mean follow-up
of 8.2 years (7 to 10) (p <
0.01). Radiolucencies were identified
in up to 23% of TARs after a mean of 4.4 years (2.3 to 9.6). The
mean total range of movement improved from 23° (95% CI 19 to 26)
to 34° (95% CI 26 to 41) (p = 0.01). Our study demonstrates that TAR has a positive impact on patients’
lives, with benefits lasting ten years, as judged by improvement
in pain and function, as well as improved gait and increased range
of movement. However, the quality of evidence is weak and fraught
with biases and high quality randomised controlled trials are required
to compare TAR with other forms of treatment such as fusion. Cite this article:
The purpose of this study was twofold: first,
to determine whether the five-year results of hip resurfacing arthroplasty
(HRA) in Canada justified the continued use of HRA; and second,
to identify whether greater refinement of patient selection was
warranted. This was a retrospective cohort study that involved a review
of 2773 HRAs performed between January 2001 and December 2008 at
11 Canadian centres. Cox’s proportional hazards models were used
to analyse the predictors of failure of HRA. Kaplan–Meier survival
analysis was performed to predict the cumulative survival rate at
five years. The factors analysed included age, gender, body mass
index, pre-operative hip pathology, surgeon’s experience, surgical
approach, implant sizes and implant types. The most common modes
of failure were also analysed. The 2773 HRAs were undertaken in 2450 patients: 2127 in men and
646 in women. The mean age at operation was 50.5 years ( The failure rates of HRA at five years justify the ongoing use
of this technique in men. Female gender is an independent predictor
of failure, and a higher failure rate at five years in women leads
the authors to recommend this technique only in exceptional circumstances
for women. Cite this article:
We undertook a retrospective review of 33 patients who underwent total femoral endoprosthetic replacement as limb salvage following excision of a malignant bone tumour. In 22 patients this was performed as a primary procedure following total femoral resection for malignant disease. Revision to a total femoral replacement was required in 11 patients following failed segmental endoprosthetic or allograft reconstruction. There were 33 patients with primary malignant tumours, and three had metastatic lesions. The mean age of the patients was 31 years (5 to 68). The mean follow-up was 4.2 years (9 months to 16.4 years). At five years the survival of the implants was 100%, with removal as the endpoint and 56% where the endpoint was another surgical intervention. At five years the patient survival was 32%. Complications included dislocation of the hip in six patients (18%), local recurrence in three (9%), peri-prosthetic fracture in two and infection in one. One patient subsequently developed pulmonary metastases. There were no cases of aseptic loosening or amputation. Four patients required a change of bushings. The mean Musculoskeletal Tumour Society functional outcome score was 67%, the mean Harris Hip Score was 70, and the mean Oxford Knee Score was 34. Total femoral endoprosthetic replacement can provide good functional outcome without compromising patient survival, and in selected cases provides an effective alternative to amputation.
In this study we present our experience with
four generations of uncemented total knee arthroplasty (TKA) from Smith
&
Nephew: Tricon M, Tricon LS, Tricon II and Profix, focusing
on the failure rates correlating with each design change. Beginning
in 1984, 380 Tricon M, 435 Tricon LS, 305 Tricon 2 and 588 Profix
were implanted by the senior author. The rate of revision for loosening
was 1.1% for the Tricon M, 1.1% for the Tricon LS, 0.5% for the
Tricon 2 with a HA coated tibial component, and 1.3% for the Profix
TKA. No loosening of the femoral component was seen with the Tricon
M, Tricon LS or Tricon 2, with no loosening seen of the tibial component
with the Profix TKA. Regarding revision for wear, the incidence
was 13.1% for the Tricon M, 6.6% for the Tricon LS, 2.3% for the
Tricon 2, and 0% for the Profix. These results demonstrate that
improvements in the design of uncemented components, including increased
polyethylene thickness, improved polyethylene quality, and the introduction
of hydroxyapatite coating, has improved the outcomes of uncemented
TKA over time.
Operative fixation is the treatment of choice for a rupture of the distal tendon of biceps. A variety of techniques have been described including transosseous sutures and suture anchors. The poor quality of the bone of the radial tuberosity might affect the load to failure of the tendon repair in early rehabilitation. The aim of this study was to determine the loads to failure of different techniques of fixation and to investigate their association with the bone mineral density of the radial tuberosity. Peripheral quantitative computed tomography was carried out to measure the trabecular and cortical bone mineral density of the radial tuberosity in 40 cadaver specimens. The loads to failure in four different techniques of fixation were determined. The Endobutton-based method showed the highest failure load at 270 N ( The transosseous technique is an easy and cost-saving procedure for fixation of the distal biceps tendon. TwinFix-QuickT 5.0 mm had significantly lower failure loads, which might affect early rehabilitation, particularly in older patients.
An understanding of the remodelling of tendon is crucial for the development of scientific methods of treatment and rehabilitation. This study tested the hypothesis that tendon adapts structurally in response to changes in functional loading. A novel model allowed manipulation of the mechanical environment of the patellar tendon in the presence of normal joint movement via the application of an adjustable external fixator mechanism between the patella and the tibia in sheep, while avoiding exposure of the patellar tendon itself. Stress shielding caused a significant reduction in the structural and material properties of stiffness (79%), ultimate load (69%), energy absorbed (61%), elastic modulus (76%) and ultimate stress (72%) of the tendon compared with controls. Compared with the material properties the structural properties exhibited better recovery after re-stressing with stiffness 97%, ultimate load 92%, energy absorbed 96%, elastic modulus 79% and ultimate stress 80%. The cross-sectional area of the re-stressed tendons was significantly greater than that of stress-shielded tendons. The remodelling phenomena exhibited in this study are consistent with a putative feedback mechanism under strain control. This study provides a basis from which to explore the interactions of tendon remodelling and mechanical environment.