Advertisement for orthosearch.org.uk
Results 161 - 165 of 165
Results per page:
The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 3 | Pages 315 - 320
1 Mar 2011
Hart AJ Ilo K Underwood R Cann P Henckel J Lewis A Cobb J Skinner J

We measured the orientation of the acetabular and femoral components in 45 patients (33 men, 12 women) with a mean age of 53.4 years (30 to 74) who had undergone revision of metal-on-metal hip resurfacings. Three-dimensional CT was used to measure the inclination and version of the acetabular component, femoral version and the horizontal femoral offset, and the linear wear of the removed acetabular components was measured using a roundness machine.

We found that acetabular version and combined version of the acetabular and femoral components were weakly positively correlated with the rate of wear. The acetabular inclination angle was strongly positively correlated with the rate of wear. Femoral version was weakly negatively correlated with the rate of wear. Application of a threshold of > 5 μm/year for the rate of wear in order to separate the revisions into low or high wearing groups showed that more high wearing components were implanted outside Lewinnek’s safe zone, but that this was mainly due to the inclination of the acetabular component, which was the only parameter that significantly differed between the groups.

We were unable to show that excess version of the acetabular component alone or combined with femoral version was associated with an increase in the rate of wear based on our assessment of version using CT.


Bone & Joint 360
Vol. 3, Issue 1 | Pages 40 - 41
1 Feb 2014
Ivory J

Metal-on-metal (MoM) hip resurfacing was developed in the 1990s by surgeons in Birmingham, UK, as a surgical solution to the problem of osteoarthritis in younger, more active patients. Early results were promising and the procedure gained in popularity. However, adverse reports of soft-tissue reaction and failure started to appear from 2008 onwards. Surgeons may be asked to write medico-legal reports on the surgical aspects of an individual case for claimant lawyers or in defence for the NHSLA or indemnity insurers. The purpose of this article is to cover some of the aspects of the operation that may be considered in such medico-legal reports.


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 3 | Pages 356 - 361
1 Mar 2010
Kwon Y- Glyn-Jones S Simpson DJ Kamali A McLardy-Smith P Gill HS Murray DW

The presence of pseudotumours, which are soft-tissue masses relating to the hip, after metal-on-metal hip resurfacing arthroplasty has been associated with elevated levels of metal ions in serum, suggesting that pseudotumours occur when there is increased wear. We aimed to quantify the wear in vivo of implants revised for pseudotumours (eight) and of a control group of implants (22) revised for other reasons of failure. We found that the implant group with pseudotumours had a significantly higher rate of median linear wear of the femoral component at 8.1 μm/year (2.75 to 25.4) than the 1.79 μm/year (0.82 to 4.15; p = 0.002) of the non-pseudotumour group. For the acetabular component a significantly higher rate of median linear wear of 7.36 μm/year (1.61 to 24.9) was observed in the pseudotumour group compared with 1.28 μm/year (0.81 to 3.33, p = 0.001) in the other group. Wear of the acetabular component in the pseudotumour group always involved the edge of the implant, indicating that edge-loading had occurred. Our findings are the first direct evidence that pseudotumour is associated with increased wear at the metal-on-metal articulation. Furthermore, edge-loading with the loss of fluid-film lubrication may be an important mechanism of generation of wear in patients with a pseudotumour


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 6 | Pages 738 - 744
1 Jun 2009
Hart AJ Sabah S Henckel J Lewis A Cobb J Sampson B Mitchell A Skinner JA

We carried out metal artefact-reduction MRI, three-dimensional CT measurement of the position of the component and inductively-coupled plasma mass spectrometry analysis of cobalt and chromium levels in whole blood on 26 patients with unexplained pain following metal-on-metal resurfacing arthroplasty.

MRI showed periprosthetic lesions around 16 hips, with 14 collections of fluid and two soft-tissue masses. The lesions were seen in both men and women and in symptomatic and asymptomatic hips. Using three-dimensional CT, the median inclination of the acetabular component was found to be 55° and its positioning was outside the Lewinnek safe zone in 13 of 16 cases. Using inductively-coupled plasma mass spectrometry, the levels of blood metal ions tended to be higher in painful compared with well-functioning metal-on-metal hips.

These three clinically useful investigations can help to determine the cause of failure of the implant, predict the need for future revision and aid the choice of revision prostheses.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 7 | Pages 989 - 990
1 Jul 2007
DANIEL J ZIAEE H PRADHAN C PYNSENT PB McMINN DJW