Advertisement for orthosearch.org.uk
Results 1681 - 1700 of 1756
Results per page:
The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 12 | Pages 1638 - 1644
1 Dec 2007
Nath RK Lyons AB Melcher SE Paizi M

The medial rotation contracture caused by weak external rotation secondary to obstetric brachial plexus injury leads to deformation of the bones of the shoulder. Scapular hypoplasia, elevation and rotation deformity are accompanied by progressive dislocation of the humeral head. Between February and August 2005, 44 children underwent a new surgical procedure called the ‘triangle tilt’ operation to correct this bony shoulder deformity. Surgical levelling of the distal acromioclavicular triangle combined with tightening of the posterior glenohumeral capsule (capsulorrhaphy) improved shoulder function and corrected the glenohumeral axis in these patients. The posture of the arm at rest was improved and active external rotation increased by a mean of 53° (0° to 115°) in the 40 children who were followed up for more than one year.

There was a mean improvement of 4.9 points (1.7 to 8.3) of the Mallet shoulder function score after surgical correction of the bony deformity.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 4 | Pages 486 - 489
1 Apr 2007
Bassi RS Simmons D Ali F Nuttall D Birch A Trail IA Stanley JK

The Acclaim total elbow replacement is a modular system which allows implantation in both unlinked and linked modes. The results of the use of this implant in primary total elbow replacement in 36 patients, operated on between July 2000 and August 2002, are presented at a mean follow-up of 36 months (24 to 49). Only one patient did not have good relief of pain, but all had improved movement and function.

No implant showed clinical or radiological loosening, although one had a lucent area in three of seven humeral zones. The short-term results of the Acclaim total elbow replacement are encouraging. However, 11 patients (30.5%) suffered an intra-operative fracture of the humeral condyle. This did not affect the outcome, or the requirement for further surgery, except in one case where the fracture failed to unite. This problem has hopefully been addressed by redesigning the humeral resection guide. Other complications included three cases of ulnar neuropathy (8.3%) and one of deep infection (2.8%).


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 2 | Pages 133 - 139
1 Feb 2008
Schuind F Eslami S Ledoux P

Kienböck’s disease is a form of osteonecrosis affecting the lunate. Its aetiology remains unknown. Morphological variations, such as negative ulnar variance, high uncovering of the lunate, abnormal radial inclination and/or a trapezoidal shape of the lunate and the particular pattern of its vascularity may be predisposing factors. A history of trauma is common. The diagnosis is made on plain radiographs, but MRI can be helpful early in the disease. A CT scan is useful to demonstrate fracture or fragmentation of the lunate. Lichtman classified Kienböck disease into five stages. The natural history of the condition is not well known, and the symptoms do not correlate well with the changes in shape of the lunate and the degree of carpal collapse. There is no strong evidence to support any particular form of treatment. Many patients are improved by temporary immobilisation of the wrist, which does not stop the progression of carpal collapse. Radial shortening may be the treatment of choice in young symptomatic patients presenting with stages I to III-A of Kienböck’s disease and negative ulnar variance. Many other forms of surgical treatment have been described.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 2 | Pages 202 - 205
1 Feb 2007
Arya AP Kulshreshtha R Kakarala GK Singh R Compson JP

Disorders of the pisotriquetral joint are well recognised as the cause of pain on the ulnar side of the wrist. The joint is not usually examined during routine arthroscopy because it is assumed to have a separate joint cavity to the radiocarpal joint, although there is often a connection between the two.

We explored this connection during arthroscopy and in fresh-frozen cadaver wrists and found that in about half of the cases the pisotriquetral joint could be visualised through standard wrist portals. Four different types of connection were observed between the radiocarpal joint and the pisotriquetral joint. They ranged from a complete membrane separating the two, to no membrane at all, with various other types of connection in between.

We recommend that inspection of the pisotriquetral joint should be a part of the protocol for routine arthroscopy of the wrist.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 11 | Pages 1498 - 1503
1 Nov 2007
Hanna SA David LA Aston WJS Gikas PD Blunn GW Cannon SR Briggs TWR

Between 1988 and 2006, 18 patients had a custom-made endoprosthetic replacement of the distal humerus for bone tumours at our institution. There were 11 primary malignant neoplasms, six secondary deposits, and one benign aggressive tumour. The mean follow-up was for 4.4 years (1 to 18.2). Complications occurred in nine patients and included aseptic loosening in three (16.6%), local recurrence in two (11%), infection in two (11%), neuropraxia of the radial nerve in one (5.5%) and a peri-prosthetic fracture in one (5.5%). Excision was inadequate in four patients (22%), all of which developed local recurrence and/or metastases. There were seven deaths from the primary disease after a mean of 2.3 years (1 to 5), one of whom had an above-elbow amputation for local recurrence seven months before death. The remaining six had satisfactory elbow function at their last follow-up. The 11 living patients were evaluated using the Musculoskeletal Tumour Society and Toronto Extremity Salvage scoring systems. The mean scores achieved were 76% (67% to 87%) and 73% (59% to 79%), respectively. Overall, 17 of 18 patients had significant improvement in the degree of their pain following operation.

Custom-made endoprosthetic reconstruction of the elbow for bone tumours is a viable treatment in carefully selected patients. It maintains satisfactory function and provides good pain relief.


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 3 | Pages 308 - 313
1 Mar 2008
Murtha PE Hafez MA Jaramaz B DiGioia AM

Three-dimensional surface models of the normal hemipelvis derived from volumetric CT data on 42 patients were used to determine the radius, depth and orientation of the native acetabulum. A sphere fitted to the lunate surface and a plane matched to the acetabular rim were used to calculate the radius, depth and anatomical orientation of the acetabulum. For the 22 females the mean acetabular abduction, anteversion, radius and normalised depth were 57.1° (50.7° to 66.8°), 24.1° (14.0° to 33.3°), 25 mm (21.7 to 30.3) and 0.79 mm (0.56 to 1.04), respectively. The same parameters for the 20 males were 55.5° (47.7° to 65.9°), 19.3° (8.5° to 32.3°), 26.7 mm (24.5 to 28.7) and 0.85 mm (0.65 to 0.99), respectively.

The orientation of the native acetabulum did not match the safe zone for acetabular component placement described by Lewinnek. During total hip replacement surgeons should be aware that the average abduction angle of the native acetabulum exceeds that of the safe zone angle. If the concept of the safe zone angle is followed, abduction of the acetabular component should be less than the abduction of the native acetabulum by approximately 10°.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 5 | Pages 615 - 619
1 May 2007
Smith R Wood PLR

A consecutive series of 23 patients (25 ankles) with osteoarthritis of the ankle and severe varus or valgus deformity were treated by open arthrodesis using compression screws. Primary union was achieved in 24 ankles one required further surgery to obtain a solid fusion. The high level of satisfaction in this group of patients reinforces the view that open arthrodesis, as opposed to ankle replacement or arthroscopic arthrodesis, continues to be the treatment of choice when there is severe varus or valgus deformity associated with the arthritis.


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 8 | Pages 1059 - 1065
1 Aug 2008
Elhassan B Ozbaydar M Massimini D Diller D Higgins L Warner JJP

Transfer of pectoralis major has evolved as the most favoured option for the management of the difficult problem of irreparable tears of subscapularis. We describe our experience with this technique in 30 patients divided into three groups. Group I comprised 11 patients with a failed procedure for instability of the shoulder, group II included eight with a failed shoulder replacement and group III, 11 with a massive tear of the rotator cuff. All underwent transfer of the sternal head of pectoralis major to restore the function of subscapularis.

At the latest follow-up pain had improved in seven of the 11 patients in groups I and III, but in only one of eight in group II. The subjective shoulder score improved in seven patients in group I, in one in group II and in six in group III. The mean Constant score improved from 40.9 points (28 to 50) in group I, 32.9 (17 to 47) in group II and 28.7 (20 to 42) in group III pre-operatively to 60.8 (28 to 89), 41.9 (24 to 73) and 52.3 (24 to 78), respectively.

Failure of the tendon transfer was highest in group II and was associated with pre-operative anterior subluxation of the humeral head. We conclude that in patients with irreparable rupture of subscapularis after shoulder replacement there is a high risk of failure of transfer of p?ctoralis major, particularly if there is pre-operative anterior subluxation of the humeral head.


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 2 | Pages 166 - 171
1 Feb 2008
Lundblad H Kreicbergs A Jansson K

We suggest that different mechanisms underlie joint pain at rest and on movement in osteoarthritis and that separate assessment of these two features with a visual analogue scale (VAS) offers better information about the likely effect of a total knee replacement (TKR) on pain. The risk of persistent pain after TKR may relate to the degree of central sensitisation before surgery, which might be assessed by determining the pain threshold to an electrical stimulus created by a special tool, the Pain Matcher. Assessments were performed in 69 patients scheduled for TKR. At 18 months after operation, separate assessment of pain at rest and with movement was again carried out using a VAS in order to enable comparison of pre- and post-operative measurements. A less favourable outcome in terms of pain relief was observed for patients with a high pre-operative VAS score for pain at rest and a low pain threshold, both features which may reflect a central sensitisation mechanism.


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 2 | Pages 172 - 179
1 Feb 2008
Pinczewski LA Salmon LJ Jackson WFM von Bormann RBP Haslam PG Tashiro S

There is little evidence examining the relationship between anatomical landmarks, radiological placement of the tunnels and long-term clinical outcomes following anterior cruciate ligament (ACL) reconstruction. The aim of this study was to investigate the reproducibility of intra-operative landmarks for placement of the tunnels in single-bundle reconstruction of the ACL using four-strand hamstring tendon autografts.

Isolated reconstruction of the ACL was performed in 200 patients, who were followed prospectively for seven years with use of the International Knee Documentation Committee forms and radiographs. Taking 0% as the anterior and 100% as the posterior extent, the femoral tunnel was a mean of 86% (sd 5) along Blumensaat’s line and the tibial tunnel was 48% (sd 5) along the tibial plateau. Taking 0% as the medial and 100% as the lateral extent, the tibial tunnel was 46% (sd 3) across the tibial plateau and the mean inclination of the graft in the coronal plane was 19° (sd 5.5).

The use of intra-operative landmarks resulted in reproducible placement of the tunnels and an excellent clinical outcome seven years after operation. Vertical inclination was associated with increased rotational instability and degenerative radiological changes, while rupture of the graft was associated with posterior placement of the tibial tunnel. If the osseous tunnels are correctly placed, single-bundle reconstruction of the ACL adequately controls both anteroposterior and rotational instability.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 7 | Pages 971 - 976
1 Jul 2007
Kampa RJ Prasthofer A Lawrence-Watt DJ Pattison RM

In order to determine the potential for an internervous safe zone, 20 hips from human cadavers were dissected to map out the precise pattern of innervation of the hip capsule. The results were illustrated in the form of a clock face. The reference point for measurement was the inferior acetabular notch, representing six o’clock. Capsular branches from between five and seven nerves contributed to each hip joint, and were found to innervate the capsule in a relatively constant pattern. An internervous safe zone was identified anterosuperiorly in an arc of 45° between the positions of one o’clock and half past two.

Our study shows that there is an internervous zone that could be safely used in a capsule-retaining anterior, anterolateral or lateral approach to the hip, or during portal placement in hip arthroscopy.


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 12 | Pages 1580 - 1583
1 Dec 2006
Ling ZX Kumar VP

We investigated the relationship of the inferior gluteal nerve to gluteus maximus by dissecting the muscle in 12 fresh-frozen and formalin-treated cadavers. The anatomy was recorded using still digital photography. The course of the inferior gluteal nerve was carefully traced and was noted to enter the deep surface of gluteus maximus approximately 5 cm from the tip of the greater trochanter of the femur. The susceptibility of the nerve to injury during a posterior approach to the hip may be explained by its close relationship to the deep surface of gluteus maximus. It is easily damaged before it has been seen if the muscle is split and parted more than 5 cm from the tip of the greater trochanter of the femur. We suggest that a modified posterior approach be used to expose the hip to avoid damage to this nerve.


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 12 | Pages 1585 - 1593
1 Dec 2008
Henricson A Linder L Nilsson KG

We compared the performance of uncemented trabecular metal tibial components in total knee replacement with that of cemented tibial components in patients younger than 60 years over two years using radiostereophotogrammetric analysis (RSA). A total of 22 consecutive patients (mean age 53 years, 33 to 59, 26 knees) received an uncemented NexGen trabecular metal cruciate-retaining monobloc tibial component and 19 (mean 53 years, 44 to 59, 21 knees) a cemented NexGen Option cruciate-retaining modular tibial component.

All the trabecular metal components migrated during the initial three months and then stabilised. The exception was external rotation, which did not stabilise until 12 months. Unlike conventional metal-backed implants which displayed a tilting migration comprising subsidence and lift-off from the tibial tray, most of the trabecular metal components showed subsidence only, probably due to the elasticity of the implant.

This pattern of subsidence is regarded as being beneficial for uncemented fixation.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 11 | Pages 1545 - 1550
1 Nov 2007
Koslowsky TC Mader K Dargel J Koebke J Hellmich M Pennig D

We have evaluated four different fixation techniques for the reconstruction of a standard Mason type-III fracture of the radial head in a sawbone model. The outcome measurements were the quality of the reduction, and stability.

A total of 96 fractures was created. Six surgeons were involved in the study and each reconstructed 16 fractures with 1.6 mm fine-threaded wires (Fragment Fixation System (FFS)), T-miniplates, 2 mm miniscrews and 2 mm Kirschner (K-) wires; four fractures being allocated to each method using a standard reconstruction procedure.

The quality of the reduction was measured after definitive fixation. Biomechanical testing was performed using a transverse plane shear load in two directions to the implants (parallel and perpendicular) with respect to ultimate failure load and displacement at 50 N.

A significantly better quality of reduction was achieved using the FFS wires (Tukey’s post hoc tests, p < 0.001) than with the other devices with a mean step in the articular surface and the radial neck of 1.04 mm (sd 0.96) for the FFS, 4.25 mm (sd 1.29) for the miniplates, 2.21 mm (sd 1.06) for the miniscrews and 2.54 mm (sd 0.98) for the K-wires. The quality of reduction was similar for K-wires and miniscrews, but poor for miniplates.

The ultimate failure load was similar for the FFS wires (parallel, 196.8 N (sd 46.8), perpendicular, 212.5 N (sd 25.6)), miniscrews (parallel, 211.8 N (sd 47.9), perpendicular, 208.0 N (sd 65.9)) and K-wires (parallel, 200.4 N (sd 54.5), perpendicular, 165.2 N (sd 37.9)), but significantly worse (Tukey’s post hoc tests, p < 0.001) for the miniplates (parallel, 101.6 N (sd 43.1), perpendicular, 122.7 N (sd 40.7)). There was a significant difference in the displacement at 50 N for the miniplate (parallel, 4.8 mm (sd 2.8), perpendicular, 4.8 mm (sd 1.7)) vs FFS (parallel, 2.1 mm (sd 0.8), perpendicular, 1.9 mm (sd 0.7)), miniscrews (parallel, 1.8 mm (sd 0.5), perpendicular, 2.3 mm (sd 0.8)) and K-wires (parallel, 2.2 mm (sd 1.8), perpendicular, 2.4 mm (sd 0.7; Tukey’s post hoc tests, p < 0.001)).

The fixation of a standard Mason type-III fracture in a sawbone model using the FFS system provides a better quality of reduction than that when using conventional techniques. There was a significantly better stability using FFS implants, miniscrews and K-wires than when using miniplates.


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 8 | Pages 1166 - 1166
1 Aug 2005


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 2 | Pages 206 - 209
1 Feb 2007
Houshian S Chikkamuniyappa C Schroeder H

We present the outcome of the treatment of chronic post-traumatic contractures of the proximal interphalangeal joint by gradual distraction correction using an external fixator. A total of 30 consecutive patients with a mean age of 34 years (17 to 54) had distraction for a mean of 16 days (10 to 22). The fixator was removed after a mean of 29 days (16 to 40). Assessment at a mean of 34 months (18 to 54) after completion of treatment showed that the mean active range of movement had significantly increased by 63° (30° to 90°; p < 0.001). The mean active extension gained was 47° (30° to 75°).

Patients aged less than 40 years fared slightly better with a mean gain in active range of movement of 65° (30° to 90°) compared with those aged more than 40 years, who had a mean gain in active range of movement of 55° (30° to 70°) but the difference was not statistically significant (p = 0.148).

The use of joint distraction to correct chronic flexion contracture of the proximal interphalangeal joint is a minimally-invasive and effective method of treatment.


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 12 | Pages 1703 - 1703
1 Dec 2005
Dowd GSE


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 1 | Pages 89 - 93
1 Jan 2007
Herscovici D Scaduto JM Infante A

Between 1992 and 2000, 57 patients with 57 isolated fractures of the medial malleolus were treated conservatively by immobilisation in a cast. The results were assessed by examination, radiography and completion of the short form-36 questionnaire and American Orthopaedic Foot and Ankle Society ankle-hindfoot score.

Of the 57 fractures 55 healed without further treatment. The mean combined dorsi- and plantar flexion was 52.3° (25° to 82°) and the mean short form-36 and American Orthopaedic Foot and Ankle Society scores 48.1 (28 to 60) and 89.8 (69 to 100), respectively. At review there was no evidence of medial instability, dermatological complications, malalignment of the mortise or of post-traumatic arthritis.

Isolated fractures of the medial malleolus can obtain high rates of union and good functional results with conservative treatment. Operation should be reserved for bi- or trimalleolar fractures, open fractures, injuries which compromise the skin or those involving the plafond or for patients who develop painful nonunion.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 6 | Pages 794 - 798
1 Jun 2007
Strauss EJ Frank JB Walsh M Koval KJ Egol KA

Many orthopaedic surgeons believe that obese patients have a higher rate of peri-operative complications and a worse functional outcome than non-obese patients. There is, however, inconsistency in the literature supporting this notion.

This study was performed to evaluate the effect of body mass index (BMI) on injury characteristics, the incidence of complications, and the functional outcome after the operative management of unstable ankle fractures.

We retrospectively reviewed 279 patients (99 obese (BMI ≥ 30) and 180 non-obese (BMI < 30) patients who underwent surgical fixation of an unstable fracture of the ankle. We found that obese patients had a higher number of medical co-morbidities, and more Orthopaedic Trauma Association type B and C fracture types than non-obese patients. At two years from the time of injury, however, the presence of obesity did not affect the incidence of complications, the time to fracture union or the level of function.

These findings suggest that obese patients should be treated in line with standard procedures, keeping in mind any known associated medical co-morbidities.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 6 | Pages 785 - 789
1 Jun 2007
Ross R Mirza AH Norris HE Khatri M

Between January 1990 and December 2000 we carried out 226 SB Charité III disc replacements for lumbar disc degeneration in 160 patients. They were reviewed at a mean follow-up of 79 months (31 to 161) to determine the clinical and radiological outcome. The clinical results were collected by an independent observer, who was not involved in patient selection, treatment or follow-up, using a combination of outcome measures, including the Oswestry Disability Index. Pain was recorded using a visual analogue score, and the most recent radiographs were reviewed.

Survival of the device was analysed by the Kaplan-Meier method and showed a cumulative survival of 35% at 156 months when radiological failure was taken as the endpoint. The mean improvement in the Oswestry disability index scores after disc replacement was 14% (6% to 21%) and the mean improvement in the pain score was 1.6 (0.46 to 2.73), both falling below the clinically significant threshold. Removal of the implant was required in 12 patients, four because of implant failure.

These poor results indicate that further use of this implant is not justified.