The aim of this study was to quantify the stability
of fracture-implant complex in fractures after fixation. A total
of 15 patients with an undisplaced fracture of the femoral neck,
treated with either a dynamic hip screw or three cannulated hip
screws, and 16 patients with an AO31-A2 trochanteric fracture treated
with a dynamic hip screw or a Gamma Nail, were included. Radiostereometric
analysis was used at six weeks, four months and 12 months post-operatively
to evaluate shortening and rotation. Migration could be assessed in ten patients with a fracture of
the femoral neck and seven with a trochanteric fracture. By four
months post-operatively, a mean shortening of 5.4 mm (-0.04 to 16.1)
had occurred in the fracture of the femoral neck group and 5.0 mm
(-0.13 to 12.9) in the trochanteric fracture group. A wide range
of rotation occurred in both types of fracture. Right-sided trochanteric
fractures seem more rotationally stable than left-sided fractures. This prospective study shows that migration at the fracture site
occurs continuously during the first four post-operative months,
after which stabilisation occurs. This information may allow the
early recognition of patients at risk of failure of fixation. Cite this article:
Fractures of the proximal femur are a common clinical problem, and a number of orthopaedic devices are available for the treatment of such fractures. The objective of this study was to assess the rotational stability, a common failure predictor, of three different rotational control design philosophies: a screw, a helical blade and a deployable crucifix. Devices were compared in terms of the mechanical work (W) required to rotate the implant by 6° in a bone substitute material. The substitute material used was Sawbones polyurethane foam of three different densities (0.08 g/cm3, 0.16 g/cm3 and 0.24 g/cm3). Each torsion test comprised a steady ramp of 1°/minute up to an angular displacement of 10°.Objectives
Methods
This study aimed to determine the diagnostic performance of radiographic
criteria to detect aseptic acetabular loosening after revision total
hip arthroplasty (THA). Secondary aims were to determine the predictive
values of different thresholds of migration and to determine the
predictive values of radiolucency criteria. Acetabular component migration to re-revision was measured retrospectively
using Ein-Bild-Rontgen-Analyse (EBRA-Cup) and manual measurements
(Sutherland method) in two groups: Group A, 52 components (48 patients) found
not loose at re-revision and Group B, 42 components (36 patients)
found loose at re-revision between 1980 and 2015. The presence and
extent of radiolucent lines was also assessed.Aims
Patients and Methods
We carried out a further study of the long-term results of the
cemented Exeter femoral component in patients under the age of 40
with a mean follow-up of 13.6 years (10 to 20). We reviewed our original cohort of 104 cemented Exeter stems
in 78 consecutive patients with a mean age of 31 years (16 to 39).
Only one patient was lost to radiological follow-up.Aims
Patients and Methods
Femoral impaction bone grafting was first developed in 1987 using
morselised cancellous bone graft impacted into the femoral canal
in combination with a cemented, tapered, polished stem. We describe
the evolution of this technique and instrumentation since that time. Between 1987 and 2005, 705 revision total hip arthroplasties
(56 bilateral) were performed with femoral impaction grafting using
a cemented femoral stem. All surviving patients were prospectively
followed for a mean of 14.7 years (9.8 to 28.3) with no loss to
follow-up. By the time of the final review, 404 patients had died.Aims
Patients and Methods
The accuracy and precision of two new methods of model-based
radiostereometric analysis (RSA) were hypothesised to be superior
to a plain radiograph method in the assessment of polyethylene (PE)
wear. A phantom device was constructed to simulate three-dimensional
(3D) PE wear. Images were obtained consecutively for each simulated
wear position for each modality. Three commercially available packages
were evaluated: model-based RSA using laser-scanned cup models (MB-RSA),
model-based RSA using computer-generated elementary geometrical
shape models (EGS-RSA), and PolyWare. Precision (95% repeatability
limits) and accuracy (Root Mean Square Errors) for two-dimensional
(2D) and 3D wear measurements were assessed.Objectives
Methods
As the number of younger and more active patients
treated with total knee arthroplasty (TKA) continues to increase,
consideration of better fixation as a means of improving implant
longevity is required. Cemented TKA remains the reference standard
with the largest body of evidence and the longest follow-up to support
its use. However, cementless TKA, may offer the opportunity of a
more bone-sparing procedure with long lasting biological fixation
to the bone. We undertook a review of the literature examining advances
of cementless TKA and the reported results. Cite this article:
The purpose of the present study was to examine the long-term
fixation of a cemented fixed-bearing polished titanium tibial baseplate
(Genesis ll). Patients enrolled in a previous two-year prospective trial (n
= 35) were recalled at ten years. Available patients (n = 15) underwent
radiostereometric analysis (RSA) imaging in a supine position using
a conventional RSA protocol. Migration of the tibial component in
all planes was compared between initial and ten-year follow-up.
Outcome scores including the Knee Society Score, Western Ontario
and McMaster Universities Arthritis Index, 12-item Short Form Health
Survey, Forgotten Joint Score, and University of California, Los
Angeles Activity Score were recorded.Aims
Patients and Methods
The December 2015 Hip &
Pelvis Roundup360 looks at: Vitamin E infusion helpful in polyethylene; Hip replacement in fracture and arthritis; Non-surgical treatment for arthritis; Cost and approach in hip surgery; Who does well in FAI surgery?; AAOS Thromboembolism guidelines; Thromboprophylaxis and periprosthetic joint infection; Fluid collections not limited to metal-on-metal THR
Throughout the 20th Century, it has been postulated that the knee moves on the basis of a four-bar link mechanism composed of the cruciate ligaments, the femur and the tibia. As a consequence, the femur has been thought to roll back with flexion, and total knee arthroplasty (TKA) prostheses have been designed on this basis. Recent work, however, has proposed that at a position of between 0° and 120° the medial femoral condyle does not move anteroposteriorly whereas the lateral femoral condyle tends, but is not obliged, to roll back – a combination of movements which equates to tibial internal/ femoral external rotation with flexion. The aim of this paper was to assess if the articular geometry of the GMK Sphere TKA could recreate the natural knee movements The pattern of knee movement was studied in 15 patients (six male: nine female; one male with bilateral TKAs) with 16 GMK Sphere implants, at a mean age of 66 years (53 to 76) with a mean BMI of 30 kg/m2 (20 to 35). The motions of all 16 knees were observed using pulsed fluoroscopy during a number of weight-bearing and non-weight-bearing static and dynamic activities.Objectives
Methods
Progressive retroversion of a cemented stem is
predictive of early loosening and failure. We assessed the relationship
between direct post-operative stem anteversion, measured with CT,
and the resulting rotational stability, measured with repeated radiostereometric
analysis over ten years. The study comprised 60 cemented total hip
replacements using one of two types of matt collared stem with a
rounded cross-section. The patients were divided into three groups
depending on their measured post-operative anteversion (<
10°,
10° to 25°, >
25°). There was a strong correlation between direct
post-operative anteversion and later posterior rotation. At one
year the <
10° group showed significantly more progressive retroversion
together with distal migration, and this persisted to the ten-year
follow-up. In the <
10° group four of ten stems (40%) had been
revised at ten years, and an additional two stems (20%) were radiologically
loose. In the ‘normal’ (10° to 25°) anteversion group there was
one revised (3%) and one loose stem (3%) of a total of 30 stems,
and in the >
25° group one stem (5%) was revised and another loose (5%)
out of 20 stems. This poor outcome is partly dependent on the design
of this prosthesis, but the results strongly suggest that the initial
rotational position of cemented stems during surgery affects the
subsequent progressive retroversion, subsidence and eventual loosening.
The degree of retroversion may be sensitive to prosthetic design
and stem size, but <
10° of anteversion appears deleterious to
the long-term outcome for cemented hip prosthetic stems. Cite this article:
The aim of this randomised controlled study was
to compare functional and radiological outcomes between modern cemented
and uncemented hydroxyapatite coated stems after one year in patients
treated surgically for a fracture of the femoral neck. A total of
141 patients aged >
65 years were included. Patients were randomised
to be treated with a cemented Exeter stem or an uncemented Bimetric
stem. The patients were reviewed at four and 12 months. The cemented group performed better than the uncemented group
for the Harris hip score (78 In conclusion, our data do not support the use of an uncemented
hydroxyapatite coated stem for the treatment of displaced fractures
of the femoral neck in the elderly. Cite this article:
This randomised trial evaluated the outcome of
a single design of unicompartmental arthroplasty of the knee (UKA) with
either a cemented all-polyethylene or a metal-backed modular tibial
component. A total of 63 knees in 45 patients (17 male, 28 female)
were included, 27 in the all-polyethylene group and 36 in the metal-backed
group. The mean age was 57.9 years (39.6 to 76.9). At a mean follow-up
of 6.4 years (5 to 9.9), 11 all-polyethylene components (41%) were
revised (at a mean of 5.8 years; 1.4 to 8.0) post-operatively and
two metal-backed components were revised (at one and five years).
One revision in both groups was for unexplained pain, one in the
metal-backed group was for progression of osteoarthritis. The others
in the all-polyethylene group were for aseptic loosening. The survivorship
at seven years calculated by the Kaplan–Meier method for the all-polyethylene
group was 56.5% (95% CI 31.9 to 75.2, number at risk 7) and for
the metal-backed group was 93.8% (95% CI 77.3 to 98.4, number at
risk 16) This difference was statistically significant (p <
0.001).
At the most recent follow-up, significantly better mean Western
Ontario and McMaster Universities Arthritis Index Scores were found
in the all-polyethylene group (13.4 This randomised study demonstrates that all-polyethylene components
in this design of fixed bearing UKA had unsatisfactory results with
significantly higher rates of failure before ten years compared
with the metal-back components. Cite this article:
This prospective study reports the 15-year survival and ten-year
functional outcome of a consecutive series of 1000 minimally invasive
Phase 3 Oxford medial UKAs (818 patients, 393 men, 48%, 425 women,
52%, mean age 66 years; 32 to 88). These were implanted by two surgeons
involved with the design of the prosthesis to treat anteromedial
osteoarthritis and spontaneous osteonecrosis of the knee, which
are recommended indications. Patients were prospectively identified
and followed up independently for a mean of 10.3 years (5.3 to 16.6). At ten years, the mean Oxford Knee Score was 40 (standard deviation
( This is the only large series of minimally invasive UKAs with
15-year survival data. The results support the continued use of
minimally invasive UKA for the recommended indications. Cite this article:
Oxidised zirconium (OxZi) has been developed
as an alternative bearing surface for femoral heads in total hip arthroplasty
(THA). This study has investigated polyethylene wear, functional
outcomes and complications, comparing OxZi and cobalt–chrome (CoCr)
as part of a three-arm, multicentre randomised controlled trial.
Patients undergoing THA from four institutions were prospectively
randomised into three groups. Group A received a CoCr femoral head
and highly cross-linked polyethylene (XLPE) liner; Group B received
an OxZi femoral head and XLPE liner; Group C received an OxZi femoral
head and ultra-high molecular weight polyethylene (UHMWPE) liner.
At five years, 368 patients had no statistically significant differences
in short-form-36 (p = 0.176 mental, p = 0.756 physical), Western
Ontario and McMaster Universities Osteoarthritis Index (p = 0.847),
pain scores
(p = 0.458) or complications. The mean rate of linear wear was 0.028
mm/year (standard deviation ( Cite this article: