We conducted a meta-analysis, including randomised
controlled trials (RCTs) and cohort studies, to examine the effect
of patient-specific instruments (PSI) on radiological outcomes after
total knee replacement (TKR) including: mechanical axis alignment
and malalignment of the femoral and tibial components in the coronal,
sagittal and axial planes, at a threshold of >
3º from neutral.
Relative risks (RR) for malalignment were determined for all studies
and for RCTs and cohort studies separately. Of 325 studies initially identified, 16 met the eligibility criteria,
including eight RCTs and eight cohort studies. There was no significant
difference in the likelihood of mechanical axis malalignment with
PSI We conclude that PSI does not improve the accuracy of alignment
of the components in TKR compared with conventional instrumentation. Cite this article:
This systematic review and meta-analysis was conducted to determine
the mid- to long-term clinical outcomes for a medial-pivot total
knee replacement (TKR) system. The objectives were to synthesise
available survivorship, Knee Society Scores (KSS), and reasons for
revision for this system. A systematic search was conducted of two online databases to
identify sources of survivorship, KSS, and reasons for revision.
Survivorship results were compared with values in the National Joint
Registry of England, Wales, and Northern Ireland (NJR).Objectives
Methods
There are many types of treatment used to manage the frozen shoulder, but there is no consensus on how best to manage patients with this painful and debilitating condition. We conducted a review of the evidence of the effectiveness of interventions used to manage primary frozen shoulder using the Cochrane Database of Systematic Reviews, the Database of Abstracts of Reviews of Effects, the Physiotherapy Evidence Database, MEDLINE and EMBASE without language or date restrictions up to April 2009. Two authors independently applied selection criteria and assessed the quality of systematic reviews using the Assessment of Multiple Systematic Reviews (AMSTAR) tool. Data were synthesised narratively, with emphasis placed on assessing the quality of evidence. In total, 758 titles and abstracts were identified and screened, which resulted in the inclusion of 11 systematic reviews. Although these met most of the AMSTAR quality criteria, there was insufficient evidence to draw firm conclusions about the effectiveness of treatments commonly used to manage a frozen shoulder. This was mostly due to poor methodological quality and small sample size in primary studies included in the reviews. We found no reviews evaluating surgical interventions. More rigorous randomised trials are needed to evaluate the treatments used for frozen shoulder.
Fractures of the femoral neck in children are
rare, high-energy injuries with high complication rates. Their treatment has
become more interventional but evidence of the efficacy of such
measures is limited. We performed a systematic review of studies
examining different types of treatment and their outcomes, including
avascular necrosis (AVN), nonunion, coxa vara, premature physeal
closure (PPC), and Ratliff’s clinical criteria. A total of 30 studies
were included, comprising 935 patients. Operative treatment and
open reduction were associated with higher rates of AVN. Delbet
types I and II fractures were most likely to undergo open reduction
and internal fixation. Coxa vara was reduced in the operative group,
whereas nonunion and PPC were not related to surgical intervention. Nonunion
and coxa vara were unaffected by the method of reduction. Capsular
decompression had no effect on AVN. Although surgery allows a more
anatomical union, it is uncertain whether operative treatment or
the type of reduction affects the rate of AVN, nonunion or PPC,
because more severe fractures were operated upon more frequently.
A delay in treatment beyond 24 hours was associated with a higher
incidence of AVN. Cite this article:
The increasing prevalence of osteoporosis in
an ageing population has contributed to older patients becoming
the fastest-growing group presenting with acetabular fractures.
We performed a systematic review of the literature involving a number
of databases to identify studies that included the treatment outcome
of acetabular fractures in patients aged >
55 years. An initial
search identified 61 studies; after exclusion by two independent
reviewers, 15 studies were considered to meet the inclusion criteria.
All were case series. The mean Coleman score for methodological
quality assessment was 37 (25 to 49). There were 415 fractures in
414 patients. Pooled analysis revealed a mean age of 71.8 years
(55 to 96) and a mean follow-up of 47.3 months (1 to 210). In seven
studies the results of open reduction and internal fixation (ORIF)
were presented: this was combined with simultaneous hip replacement
(THR) in four, and one study had a mixture of these strategies.
The results of percutaneous fixation were presented in two studies,
and a single study revealed the results of non-operative treatment. With fixation of the fracture, the overall mean rate of conversion
to THR was 23.1% (0% to 45.5%). The mean rate of non-fatal complications
was 39.8% (0% to 64%), and the mean mortality rate was 19.1% (5%
to 50%) at a mean of 64 months (95% confidence interval 59.4 to
68.6; range 12 to 143). Further data dealing with the classification
of the fracture, the surgical approach used, operative time, blood
loss, functional and radiological outcomes were also analysed. This study highlights that, of the many forms of treatment available
for this group of patients, there is a trend to higher complication
rates and the need for further surgery compared with the results
of the treatment of acetabular fractures in younger patients. Cite this article:
We report a systematic review and meta-analysis of published randomised controlled trials evaluating the efficacy of tranexamic acid (TXA) in reducing blood loss and transfusion in total hip replacement (THR). The data were evaluated using the generic evaluation tool designed by the Cochrane Bone, Joint and Muscle Trauma Group. We identified 11 clinical trials which were suitable for detailed extraction of data. There were no trials that used TXA in revision THR. A total of seven studies (comprising 350 patients) were eligible for the blood loss outcome data. The use of TXA reduced intra-operative blood loss by a mean of 104 ml (95% confidence interval (CI) −164 to −44, p = 0.0006, heterogeneity I2 0%), postoperative blood loss by a mean of 172 ml (95% CI −263 to −81, p = 0.0002, heterogeneity I2 63%) and total blood loss by a mean of 289 ml (95% CI −440 to −138, p <
0.0002, heterogeneity I2 54%). TXA led to a significant reduction in the proportion of patients requiring allogeneic blood transfusion (risk difference −0.20, 95% CI −0.29 to −0.11, p <
0.00001, I2 15%). There were no significant differences in deep-vein thrombosis, pulmonary embolism, infection rates or other complications among the study groups.
Total hip replacement (THR) has been shown to
be a cost-effective procedure. However, it is not risk-free. Certain conditions,
such as diabetes mellitus, are thought to increase the risk of complications.
In this study we have evaluated the prevalence of diabetes mellitus
in patients undergoing THR and the associated risk of adverse operative
outcomes. A meta-analysis and systematic review were conducted according
to the guidelines of the meta-analysis of observational studies
in epidemiology. Inclusion criteria were observational studies reporting
the prevalence of diabetes in the study population, accompanied
by reports of at least one of the following outcomes: venous thromboembolic
events; acute coronary events; infections of the urinary tract,
lower respiratory tract or surgical site; or requirement for revision
arthroplasty. Altman and Bland’s methods were used to calculate differences
in relative risks. The prevalence of diabetes mellitus was found
to be 5.0% among patients undergoing THR, and was associated with
an increased risk of established surgical site infection (odds ratio
(OR) 2.04 (95% confidence interval (CI) 1.52 to 2.76)), urinary
infection (OR 1.43 (95% CI 1.33 to 1.55)) and lower respiratory
tract infections (OR 1.95 (95% CI 1.61 to 2.26)). Diabetes mellitus
is a relatively common comorbidity encountered in THR. Diabetic
patients have a higher rate of developing both surgical site and
non-surgical site infections following THR. Cite this article:
The purpose of this study was to undertake a
meta-analysis to determine whether there is lower polyethylene wear and
longer survival when using mobile-bearing implants in total knee
replacement when compared with fixed-bearing implants. Of 975 papers
identified, 34 trials were eligible for data extraction and meta-analysis
comprising 4754 patients (6861 knees). We found no statistically
significant differences between the two designs in terms of the incidence
of radiolucent lines, osteolysis, aseptic loosening or survival.
There is thus currently no evidence to suggest that the use of mobile-bearing
designs reduce polyethylene wear and prolong survival after total
knee replacement. Cite this article:
We report a systematic review and meta-analysis
of published randomised and quasi-randomised trials evaluating the
efficacy of pre-operative skin antisepsis and cleansing techniques
in reducing foot and ankle skin flora. The post-preparation culture
number (Post-PCN) was the primary outcome. The data were evaluated
using a modified version of the Cochrane Collaboration’s tool. We
identified eight trials (560 participants, 716 feet) that met the inclusion
criteria. There was a significant difference in the proportions
of Post-PCN between hallux nailfold (HNF) and toe web spaces (TWS)
sites: 0.47 Meta-analyses showed that alcoholic chlorhexidine had better
efficacy than alcoholic povidone-iodine (PI) at HNF sites (risk
difference 0.19 (95% CI 0.08 to 0.30); p = 0.0005); a two-step intervention
using PI scrub and paint (S&
P) followed by alcohol showed significantly
better efficacy over PI (S&
P) alone at TWS sites (risk difference
0.13 (95% CI 0.02 to 0.24); p = 0.0169); and a two-step intervention
using chlorhexidine scrub followed by alcohol showed significantly
better efficacy over PI (S&
P) alone at the combined (HNF with
TWS) sites (risk difference 0.27 (95% CI 0.13 to 0.40); p <
0.0001).
No significant difference was found between cleansing techniques. Cite this article:
We conducted a systematic review and meta-analysis
of randomised controlled trials evaluating the effect of tranexamic
acid (TXA) upon blood loss and transfusion in primary total knee
replacement. The review used the generic evaluation tool designed
by the Cochrane Bone, Joint and Muscle Trauma Group. A total of
19 trials were eligible: 18 used intravenous administration, one
also evaluated oral dosing and one trial evaluated topical use.
TXA led to a significant reduction in the proportion of patients
requiring blood transfusion (risk ratio (RR) 2.56, 95% confidence
interval (CI) 2.1 to 3.1, p <
0.001; heterogeneity I2 =
75%; 14 trials, 824 patients). Using TXA also reduced total blood
loss by a mean of 591 ml (95% CI 536 to 647, p <
0.001; I2 =
78%; nine trials, 763 patients). The clinical interpretation of
these findings is limited by substantial heterogeneity. However,
subgroup analysis of high-dose (>
4 g) TXA showed a plausible consistent
reduction in blood transfusion requirements (RR 5.33; 95% CI 2.44
to 11.65, p <
0.001; I2 = 0%), a finding that should
be confirmed by a further well-designed trial. The current evidence
from trials does not support an increased risk of deep-vein thrombosis
(13 trials, 801 patients) or pulmonary embolism (18 trials, 971
patients) due to TXA administration.
We undertook a retrospective case-control study
to assess the clinical variables associated with infections in open fractures.
A total of 1492 open fractures were retrieved; these were Gustilo
and Anderson grade I in 663 (44.4%), grade II in 370 (24.8%), grade
III in 310 (20.8%) and unclassifiable in 149 (10.0%). The median
duration of prophylaxis was three days (interquartile range (IQR)
1 to 3), and the median number of surgical interventions was two
(1 to 9). We identified 54 infections (3.6%) occurring at a median
of ten days (IQR 5 to 20) after trauma. Pathogens intrinsically
resistant to the empirical antibiotic regimen used (enterococci, Infection in open fractures is related to the extent of tissue
damage but not to the duration of prophylactic antibiotic therapy.
Even for grade III fractures, a one-day course of prophylactic antibiotics
might be as effective as prolonged prophylaxis. Cite this article:
The aims of this piece of work were to: 1) record the background
concentrations of blood chromium (Cr) and cobalt (Co) concentrations
in a large group of subjects; 2) to compare blood/serum Cr and Co
concentrations with retrieved metal-on-metal (MoM) hip resurfacings;
3) to examine the distribution of Co and Cr in the serum and whole
blood of patients with MoM hip arthroplasties; and 4) to further
understand the partitioning of metal ions between the serum and
whole blood fractions. A total of 3042 blood samples donated to the local transfusion
centre were analysed to record Co and Cr concentrations. Also, 91
hip resurfacing devices from patients who had given pre-revision
blood/serum samples for metal ion analysis underwent volumetric
wear assessment using a coordinate measuring machine. Linear regression analysis
was carried out and receiver operating characteristic curves were
constructed to assess the reliability of metal ions to identify
abnormally wearing implants. The relationship between serum and
whole blood concentrations of Cr and Co in 1048 patients was analysed
using Bland-Altman charts. This relationship was further investigated
in an Objectives
Methods
We systematically reviewed all the evidence published
in the English language on proximal interphalangeal joint (PIPJ)
replacement, to determine its effectiveness on the function of the
hand and the associated post-operative complications. Original studies were selected if they reported clinical outcome
with a minimum of one year’s follow-up. Quality was assessed using
the Cowley systematic review criteria modified for finger-joint
replacements. Of 319 articles identified, only five were adequately
reported according to our quality criteria; there were no randomised
controlled trials. PIPJ replacements had a substantial effect size
on hand pain of -23.2 (95% confidence interval (CI) -27.3 to -19.1)
and grip strength 1.2 (95% CI -10.7 to 13.1), and a small effect
on range of movement 0.2 (95% CI -0.4 to 0.8). A dorsal approach
was most successful. Post-operative loosening occurred in 10% (95%
CI 3 to 30) of ceramic and 12.5% (95% CI 7 to 21) of pyrocarbon
replacements. Post-operative complications occurred in 27.8% (95%
CI 20 to 37). We conclude that the effectiveness of PIPJ replacement has not
been established. Small observational case studies and short-term
follow-up, together with insufficient reporting of patient data,
functional outcomes and complications, limit the value of current
evidence. We recommend that a defined core set of patients, surgical and
outcome data for this intervention be routinely and systematically
collected within the framework of a joint registry.
Plasma levels of cobalt and chromium ions and
Metal Artefact Reduction Sequence (MARS)-MRI scans were performed
on patients with 209 consecutive, unilateral, symptomatic metal-on-metal
(MoM) hip arthroplasties. There was wide variation in plasma cobalt
and chromium levels, and MARS-MRI scans were positive for adverse reaction
to metal debris (ARMD) in 84 hips (40%). There was a significant
difference in the median plasma cobalt and chromium levels between
those with positive and negative MARS-MRI scans (p <
0.001).
Compared with MARS-MRI as the potential reference standard for the
diagnosis of ARMD, the sensitivity of metal ion analysis for cobalt
or chromium with a cut-off of >
7 µg/l was 57%. The specificity was
65%, positive predictive value was 52% and the negative predictive
value was 69% in symptomatic patients. A lowered threshold of >
3.5 µg/l for cobalt and chromium ion levels improved the sensitivity
and negative predictive value to 86% and 74% but at the expense
of specificity (27%) and positive predictive value (44%). Metal ion analysis is not recommended as a sole indirect screening
test in the surveillance of symptomatic patients with a MoM arthroplasty.
The investigating clinicians should have a low threshold for obtaining
cross-sectional imaging in these patients, even in the presence
of low plasma metal ion levels.
The aim of this study was to determine whether there is any significant
difference in temporal measurements of pain, function and rates
of re-tear for arthroscopic rotator cuff repair (RCR) patients compared
with those patients undergoing open RCR. This study compared questionnaire- and clinical examination-based
outcomes over two years or longer for two series of patients who
met the inclusion criteria: 200 open RCR and 200 arthroscopic RCR
patients. All surgery was performed by a single surgeon. Objectives
Methods
We aimed first to summarise minimal clinically important differences
(MCIDs) after total hip (THR) or knee replacement (TKR) in health-related
quality of life (HRQoL), measured using the Short-Form 36 (SF-36).
Secondly, we aimed to improve the precision of MCID estimates by
means of meta-analysis. We conducted a systematic review of English and non-English articles
using MEDLINE, the Cochrane Controlled Trials Register (1960–2011),
EMBASE (1991–2011), Web of Science, Academic Search Premier and
Science Direct. Bibliographies of included studies were searched
in order to find additional studies. Search terms included MCID
or minimal clinically important change, THR or TKR and Short-Form
36. We included longitudinal studies that estimated MCID of SF-36
after THR or TKR.Objectives
Methods
We present the clinical results and survivorship
of consecutive 100 Birmingham Hip Resurfacings in 90 patients at
a minimum follow-up of ten years. All procedures were carried out
by an independent surgeon who commenced a prospective study in 1998.
Patients were assessed clinically using the Western Ontario and
McMaster Universities osteoarthritis index, Short-Form 36, Harris
hip score and University of California, Los Angeles activity score. Radiological
analysis was performed by an independent observer and blood metal
ion levels concentrations were measured at ten years post-operatively
in 62 patients. The median acetabular component inclination was
46.2° (34° to 59°) and anteversion 11.0° (0° to 30°). The median
chromium concentration in the unilateral group was 1.74 (0.41 to
15.23) and for the bilateral group was 2.98 (1.57 to 18.01). The
equivalent values for cobalt were 1.67 (0.54 to 20.4) and 1.88 (1.38
to 19.32). In total there were eight failures giving an overall survival
at ten years of 92% (95% confidence interval (CI) 86.7 to 97.3).
BHR in male patients had an improved survivorship of 94.6% (95%
CI 89.4 to 100) compared with females at 84.6% (95% CI 70.7 to 98.5),
but this did not reach statistical significance (p = 0.119). Four
of the nine BHRs with a bearing diameter of 42 mm failed. The overall results were consistent with data produced from other
centres in that the clinical outcome of large male patients was
extremely encouraging, whereas the survival of the smaller joints
was less satisfactory.
We report a systematic review and meta-analysis
of the peer-reviewed literature focusing on metal sensitivity testing
in patients undergoing total joint replacement (TJR). Our purpose
was to assess the risk of developing metal hypersensitivity post-operatively
and its relationship with outcome and to investigate the advantages
of performing hypersensitivity testing. We undertook a comprehensive search of the citations quoted in
PubMed and EMBASE: 22 articles (comprising 3634 patients) met the
inclusion criteria. The frequency of positive tests increased after
TJR, especially in patients with implant failure or a metal-on-metal
coupling. The probability of developing a metal allergy was higher
post-operatively (odds ratio (OR) 1.52 (95% confidence interval
(CI) 1.06 to 2.31)), and the risk was further increased when failed
implants were compared with stable TJRs (OR 2.76 (95% CI 1.14 to
6.70)). Hypersensitivity testing was not able to discriminate between
stable and failed TJRs, as its predictive value was not statistically
proven. However, it is generally thought that hypersensitivity testing
should be performed in patients with a history of metal allergy
and in failed TJRs, especially with metal-on-metal implants and
when the cause of the loosening is doubtful.
The aim of this study was to investigate the
possible benefit of large-head metal-on-metal bearing on a stem
for primary hip replacement compared with a 28 mm diameter conventional
metal-on-polyethylene bearing in a prospective randomised controlled
trial. We investigated cemented stem behaviour between these two
different bearings using Einzel-Bild-Röntgen-Analyse, clinical and
patient reported measures (Harris hip score, Western Ontario and
McMaster Universities osteoarthritis index, Short Form-36 and satisfaction)
and whole blood metal ion levels at two years. A power study indicated
that 50 hips were needed in each group to detect subsidence of >
5 mm at two years with a
p-value of <
0.05. Significant improvement (p <
0.001) was found in the mean
clinical and patient reported outcomes at two years for both groups.
Comparison of outcomes between the groups at two years showed no
statistically significant difference for mean stem migration, clinical
and patient reported outcomes; except overall patient satisfaction which
was higher for metal-on-metal group (p = 0.05). Metal ion levels
were raised above the Medicines and Healthcare products Regulatory
Agency advised safety level (7 µg per litre) in 20% of the metal-on-metal
group and in one patient in metal-on-polyethylene group (who had
a metal-on-metal implant on the contralateral side). Two patients
in the metal-on-metal group were revised, one for pseudotumour and
one for peri-prosthetic fracture. Use of large modular heads is associated with a risk of raised
whole blood metal ion levels despite using a proven bearing from
resurfacing. The head-neck junction or excess stem micromotion are
possibly the weak links warranting further research.
A new generation of knee prostheses has been introduced with the intention of improving post-operative knee flexion. In order to evaluate whether this goal has been achieved we performed a systematic review and meta-analysis. Systematic literature searches were conducted on MEDLINE and EMBASE from their inception to December 2007, and proceedings of scientific meetings were also searched. Only randomised, clinical trials were included in the meta-analysis. The mean difference in the maximum post-operative flexion between the ‘high-flex’ and conventional types of prosthesis was defined as the primary outcome measure. A total of five relevant articles was identified. Analysis of these trials suggested that no clinically relevant or statistically significant improvement was obtained in flexion with the ‘high-flex’ prostheses. The weighted mean difference was 2.1° (95% confidence interval −0.2 to +4.3; p = 0.07).