Advertisement for orthosearch.org.uk
Results 141 - 160 of 161
Results per page:
The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 11 | Pages 1499 - 1505
1 Nov 2012
Kim J Kasat NS Bae J Kim S Oh S Lim H

The purpose of this study was to measure the radiological parameters of femoral component alignment of the Oxford Phase 3 unicompartmental knee replacement (UKR), and evaluate their effect on clinical outcome. Multiple regression analysis was used to examine the relative contributions of the radiological assessment of femoral component alignment in 189 consecutive UKRs performed by a single surgeon. The American Knee Society scores were compared between groups, defined as being within or outside recommended tolerances of the position of the femoral component. For the flexion/extension position 21 UKRs (11.1%) lay outside the recommended limits, and for posterior overhang of the femoral component nine (4.8%) lay outside the range. The pre-operative hip/knee/ankle (HKA) angle, narrowest canal distance from the distal femoral entry point of the alignment jig and coronal entry-point position had significant effects on the flexion/extension position. Pre-operative HKA angle had a significant influence on posterior overhang of the femoral component. However, there was no significant difference in American Knee Society scores relative to the position of the femoral component.


The Bone & Joint Journal
Vol. 95-B, Issue 4 | Pages 436 - 444
1 Apr 2013
Scott CEH Nutton RW Biant LC

The lateral compartment is predominantly affected in approximately 10% of patients with osteoarthritis of the knee. The anatomy, kinematics and loading during movement differ considerably between medial and lateral compartments of the knee. This in the main explains the relative protection of the lateral compartment compared with the medial compartment in the development of osteoarthritis. The aetiology of lateral compartment osteoarthritis can be idiopathic, usually affecting the femur, or secondary to trauma commonly affecting the tibia. Surgical management of lateral compartment osteoarthritis can include osteotomy, unicompartmental knee replacement and total knee replacement. This review discusses the biomechanics, pathogenesis and development of lateral compartment osteoarthritis and its management.

Cite this article: Bone Joint J 2013;95-B:436–44.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 8 | Pages 1058 - 1066
1 Aug 2012
Baker PN Deehan DJ Lees D Jameson S Avery PJ Gregg PJ Reed MR

Patient-reported outcome measures (PROMs) are increasingly being used to assess functional outcome and patient satisfaction. They provide a framework for comparisons between surgical units, and individual surgeons for benchmarking and financial remuneration. Better performance may bring the reward of more customers as patients and commissioners seek out high performers for their elective procedures. Using National Joint Registry (NJR) data linked to PROMs we identified 22 691 primary total knee replacements (TKRs) undertaken for osteoarthritis in England and Wales between August 2008 and February 2011, and identified the surgical factors that influenced the improvements in the Oxford knee score (OKS) and EuroQol-5D (EQ-5D) assessment using multiple regression analysis. After correction for patient factors the only surgical factors that influenced PROMs were implant brand and hospital type (both p < 0.001). However, the effects of surgical factors upon the PROMs were modest compared with patient factors. For both the OKS and the EQ-5D the most important factors influencing the improvement in PROMs were the corresponding pre-operative score and the patient’s general health status. Despite having only a small effect on PROMs, this study has shown that both implant brand and hospital type do influence reported subjective functional scores following TKR. In the current climate of financial austerity, proposed performance-based remuneration and wider patient choice, it would seem unwise to ignore these effects and the influence of a range of additional patient factors.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 8 | Pages 1054 - 1059
1 Aug 2011
van Jonbergen HPW Scholtes VAB van Kampen A Poolman RW

The efficacy of circumpatellar electrocautery in reducing the incidence of post-operative anterior knee pain is unknown. We conducted a single-centre, outcome-assessor and patient-blinded, parallel-group, randomised, controlled trial to compare circumpatellar electrocautery with no electrocautery in total knee replacement in the absence of patellar resurfacing. Patients requiring knee replacement for primary osteoarthritis were randomly assigned circumpatellar electrocautery (intervention group) or no electrocautery (control group). The primary outcome measure was the incidence of anterior knee pain. A secondary measure was the standardised clinical and patient-reported outcomes determined by the American Knee Society scores and the Western Ontario and McMaster Universities (WOMAC) osteoarthritis index. A total of 131 knees received circumpatellar electrocautery and 131 had no electrocautery.

The overall incidence of anterior knee pain at follow-up at one year was 26% (20% to 31%), with 19% (12% to 26%) in the intervention group and 32% (24% to 40%) in the control group (p = 0.02). The relative risk reduction from electrocautery was 40% (9% to 61%) and the number needed to treat was 7.7 (4.3 to 41.4). The intervention group had a better mean total WOMAC score at follow-up at one year compared with the control group (16.3 (0 to 77.7) versus 21.6 (0 to 76.7), p = 0.04). The mean post-operative American Knee Society knee scores and function scores were similar in the intervention and control groups (knee score: 92.4 (55 to 100) versus 90.4 (51 to 100), respectively (p = 0.14); function score: 86.5 (15 to 100) versus 84.5 (30 to 100), respectively (p = 0.49)).

Our study suggests that in the absence of patellar resurfacing electrocautery around the margin of the patella improves the outcome of total knee replacement.


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 11 | Pages 1533 - 1534
1 Nov 2008


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 9 | Pages 1223 - 1231
1 Sep 2011
Babazadeh S Dowsey MM Swan JD Stoney JD Choong PFM

The role of computer-assisted surgery in maintaining the level of the joint in primary knee joint replacement (TKR) has not been well defined. We undertook a blinded randomised controlled trial comparing joint-line maintenance, functional outcomes, and quality-of-life outcomes between patients undergoing computer-assisted and conventional TKR. A total of 115 patients were randomised (computer-assisted, n = 55; conventional, n = 60).

Two years post-operatively no significant correlation was found between computer-assisted and conventional surgery in terms of maintaining the joint line. Those TKRs where the joint line was depressed post-operatively improved the least in terms of functional scores. No difference was detected in terms of quality-of-life outcomes. Change in joint line was found to be related to change in alignment. Change in alignment significantly affects change in joint line and functional scores.


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 3 | Pages 374 - 379
1 Mar 2010
Kendrick BJL Rout R Bottomley NJ Pandit H Gill HS Price AJ Dodd CAF Murray DW

With medial unicompartmental osteoarthritis (OA) there is occasionally a full-thickness ulcer of the cartilage on the medial side of the lateral femoral condyle. It is not clear whether this should be considered a contraindication to unicompartmental knee replacement (UKR). The aim of this study was to determine why these ulcers occur, and whether they compromise the outcome of UKR.

Case studies of knees with medial OA suggest that cartilage lesions on the medial side of the lateral condyle are caused by impingement on the lateral tibial spine as a result of the varus deformity and tibial subluxation. Following UKR the varus and the subluxation are corrected, so that impingement is prevented and the damaged part of the lateral femoral condyle is not transmitting load. An illustrative case report is presented.

Out of 769 knees with OA of the medial compartment treated with the Oxford UKR, 59 (7.7%) had partial-thickness cartilage loss and 20 (2.6%) had a full-thickness cartilage deficit on the medial side of the lateral condyle. The mean Oxford Knee Score (OKS) at the last follow-up at a mean of four years was 41.9 (13 to 48) in those with partial-thickness cartilage loss and 41.0 (20 to 48) in those with full-thickness loss. In those with normal or superficially damaged cartilage the mean was 39.5 (5 to 48) and 39.7 (8 to 48), respectively. There were no statistically significant differences between the pre-operative OKS, the final review OKS or of change in the score in the various groups.

We conclude that in medial compartment OA, damage to the medial side of the lateral femoral condyle is caused by impingement on the tibial spine and should not be considered a contraindication to an Oxford UKR, even if there is extensive full-thickness ulceration of the cartilage.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 2 | Pages 198 - 204
1 Feb 2011
Pandit H Jenkins C Gill HS Barker K Dodd CAF Murray DW

This prospective study describes the outcome of the first 1000 phase 3 Oxford medial unicompartmental knee replacements (UKRs) implanted using a minimally invasive surgical approach for the recommended indications by two surgeons and followed up independently. The mean follow-up was 5.6 years (1 to 11) with 547 knees having a minimum follow-up of five years. At five years their mean Oxford knee score was 41.3 (sd 7.2), the mean American Knee Society Objective Score 86.4 (sd 13.4), mean American Knee Society Functional Score 86.1 (sd 16.6), mean Tegner activity score 2.8 (sd 1.1). For the entire cohort, the mean maximum flexion was 130° at the time of final review.

The incidence of implant-related re-operations was 2.9%; of these 29 re-operations two were revisions requiring revision knee replacement components with stems and wedges, 17 were conversions to a primary total knee replacement, six were open reductions for dislocation of the bearing, three were secondary lateral UKRs and one was revision of a tibial component. The most common reason for further surgical intervention was progression of arthritis in the lateral compartment (0.9%), followed by dislocation of the bearing (0.6%) and revision for unexplained pain (0.6%). If all implant-related re-operations are considered failures, the ten-year survival rate was 96% (95% confidence interval, 92.5 to 99.5). If only revisions requiring revision components are considered failures the ten-year survival rate is 99.8% (confidence interval 99 to 100).

This is the largest published series of UKRs implanted through a minimally invasive surgical approach and with ten-year survival data. The survival rates are similar to those obtained with a standard open approach whereas the function is better. This demonstrates the effectiveness and safety of a minimally invasive surgical approach for implanting the Oxford UKR.


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 5 | Pages 699 - 699
1 May 2006
Bannister G


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 7 | Pages 896 - 902
1 Jul 2009
Gulati A Chau R Pandit HG Gray H Price AJ Dodd CAF Murray DW

Narrow, well-defined radiolucent lines commonly observed at the bone-implant interface of unicompartmental knee replacement tibial components have been referred to as physiological radiolucencies. These should be distinguished from pathological radiolucencies, which are poorly defined, wide and progressive, and associated with loosening and infection. We studied the incidence and clinical significance of tibial radiolucent lines in 161 Oxford unicondylar knee replacements five years after surgery. All the radiographs were aligned with fluoroscopic control to obtain views parallel to the tibial tray to reveal the tibial bone-implant interface.

We found that 49 knees (30%) had complete, 52 (32%) had partial and 60 (37%) had no radiolucent lines. There was no relationship between the incidence of radiolucent lines and patient factors such as gender, body mass index and activity, or operative factors including the status of the anterior cruciate ligament and residual varus deformity. Nor was any statistical relationship established between the presence of radiolucent lines and clinical outcome, particularly pain, assessed by the Oxford Knee score and the American Knee Society score.

We conclude that radiolucent lines are common after Oxford unicompartmental knee replacement but that their aetiology remains unclear. Radiolucent lines were not a source of adverse symptoms or pain. Therefore, when attempting to identify a source of postoperative pain after Oxford unicompartmental knee replacement the presence of a physiological radiolucency should be ignored.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 6 | Pages 745 - 749
1 Jun 2009
Ritter MA

The long-term success of total knee replacement is multifactorial, including factors relating to the patient, the operation and the implant. The purpose of this study was to examine the 20-year survival of the cemented Anatomical Graduated Component (AGC) total knee replacement. Between 1983 and 2004, 7760 of these were carried out at our institution. Of these, 6726 knees which received the non-modular metal-backed tibial component with compression-moulded polyethylene and had a minimum two-year follow-up were available for study. In all, 36 knees were followed over 20 years with a survival of the tibial and femoral components together of 97.8% (95% confidence interval (CI) 0.9851 to 0.9677), with no implants being revised for polyethylene wear or osteolysis. Age > 70 was associated with increased survival (99.6%, 95% CI 99.0 to 99.8) (p < 0.0001) but pre-operative valgus alignment reduced survival (95.1%, 95% CI 90.0 to 97.6) (p = 0.0056). Age < 55 (p = 0.129), pre-operative varus alignment (p = 0.707), osteonecrosis (p = 0.06), rheumatoid arthritis (p = 0.247), and gender (p = 0.666) were not statistically associated with failure.

We attribute the success of the AGC implant to its relatively unconstrained articular geometry and the durability of a non-modular metal-backed tibial component with compression moulded polyethylene.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 4 | Pages 469 - 474
1 Apr 2009
Gulati A Pandit H Jenkins C Chau R Dodd CAF Murray DW

Varus malalignment after total knee replacement is associated with a poor outcome. Our aim was to determine whether the same was true for medial unicompartmental knee replacement (UKR). The anatomical leg alignment was measured prospectively using a long-arm goniometer in 160 knees with an Oxford UKR. Patients were then grouped according to their mechanical leg alignment as neutral (5° to 10° of valgus), mild varus (0° to 4° of valgus) and marked varus (> 0° of varus). The groups were compared at five years in terms of absolute and change in the Oxford Knee score, American Knee Society score and the incidence of radiolucent lines.

Post-operatively, 29 (18%) patients had mild varus and 13 (8%) had marked varus. The mean American Knee Society score worsened significantly (p < 0.001) with increasing varus. This difference disappeared if a three-point deduction for each degree of malalignment was removed. No other score deteriorated with increasing varus, and the frequency of occurrence of radiolucent lines was the same in each group.

We therefore conclude that after Oxford UKR, about 25% of patients have varus alignment, but that this does not compromise their clinical or radiological outcome. Following UKR the deductions for malalignment in the American Knee Society score are not justified.


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 9 | Pages 1121 - 1127
1 Sep 2008
Sikorski JM

The advent of computer-assisted knee replacement surgery has focused interest on the alignment of the components. However, there is confusion at times between the alignment of the limb as a whole and that of the components. The interaction between them is discussed in this article. Alignment is expressed relative to some reference axis or plane and measurements will vary depending on what is selected as the reference. The validity of different reference axes is discussed. Varying prosthetic alignment has direct implications for surrounding soft-tissue tension. In this context the interaction between alignment and soft-tissue balance is explored and the current knowledge of the relationship between alignment and outcome is summarised.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 3 | Pages 408 - 412
1 Mar 2007
Ma H Lu Y Kwok T Ho F Huang C Huang C

One of the most controversial issues in total knee replacement is whether or not to resurface the patella. In order to determine the effects of different designs of femoral component on the conformity of the patellofemoral joint, five different knee prostheses were investigated. These were Low Contact Stress, the Miller-Galante II, the NexGen, the Porous-Coated Anatomic, and the Total Condylar prostheses. Three-dimensional models of the prostheses and a native patella were developed and assessed by computer. The conformity of the curvature of the five different prosthetic femoral components to their corresponding patellar implants and to the native patella at different angles of flexion was assessed by measuring the angles of intersection of tangential lines.

The Total Condylar prosthesis had the lowest conformity with the native patella (mean 8.58°; 0.14° to 29.9°) and with its own patellar component (mean 11.36°; 0.55° to 39.19°). In the other four prostheses, the conformity was better (mean 2.25°; 0.02° to 10.52°) when articulated with the corresponding patellar component. The Porous-Coated Anatomic femoral component showed better conformity (mean 6.51°; 0.07° to 9.89°) than the Miller-Galante II prosthesis (mean 11.20°; 5.80° to 16.72°) when tested with the native patella. Although the Nexgen prosthesis had less conformity with the native patella at a low angle of flexion, this improved at mid (mean 3.57°; 1.40° to 4.56°) or high angles of flexion (mean 4.54°; 0.91° to 9.39°), respectively. The Low Contact Stress femoral component had the best conformity with the native patella (mean 2.39°; 0.04° to 4.56°). There was no significant difference (p > 0.208) between the conformity when tested with the native patella or its own patellar component at any angle of flexion.

The geometry of the anterior flange of a femoral component affects the conformity of the patellofemoral joint when articulating with the native patella. A more anatomical design of femoral component is preferable if the surgeon decides not to resurface the patella at the time of operation.


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 1 | Pages 44 - 48
1 Jan 2006
Keene G Simpson D Kalairajah Y

Twenty patients underwent simultaneous bilateral medial unicompartmental knee arthroplasty. Pre-operative hip-knee-ankle alignment and valgus stress radiographs were used to plan the desired post-operative alignment of the limb in accordance with established principles for unicompartmental arthroplasty. In each patient the planned alignment was the same for both knees. Overall, the mean planned post-operative alignment was to 2.3° of varus (0° to 5°).

The side and starting order of surgery were randomised, using conventional instrumentation for one knee and computer-assisted surgery for the opposite side.

The mean variation between the pre-operative plan and the achieved correction in the navigated and the non-navigated limb was 0.9° (sd 1.1; 0° to 4°) and 2.8° (sd 1.4; 1° to 7°), respectively. Using the Wilcoxon signed rank test, we found the difference in variation statistically significant (p < 0.001).

Assessment of lower limb alignment in the non-navigated group revealed that 12 (60%) were within ± 2° of the pre-operative plan, compared to 17 (87%) of the navigated cases.

Computer-assisted surgery significantly improves the post-operative alignment of medial unicompartmental knee arthroplasty compared to conventional techniques in patients undergoing bilateral simultaneous arthroplasty. Improved alignment after arthroplasty is associated with better function and increased longevity.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 12 | Pages 1602 - 1607
1 Dec 2007
Beard DJ Pandit H Ostlere S Jenkins C Dodd CAF Murray DW

Anterior knee pain and/or radiological evidence of degeneration of the patellofemoral joint are considered to be contraindications to unicompartmental knee replacement. The aim of this study was to determine whether this is the case.

Between January 2000 and September 2003, in 100 knees (91 patients) in which Oxford unicompartmental knee replacements were undertaken for anteromedial osteoarthritis, pre-operative anterior knee pain and the radiological status of the patellofemoral joint were defined using the Altman and Ahlback systems. Outcome was evaluated at two years with the Oxford knee score and the American Knee Society score.

Pre-operatively 54 knees (54%) had anterior knee pain. The clinical outcome was independent of the presence or absence of pre-operative anterior knee pain. Degenerative changes of the patellofemoral joint were seen in 54 patients (54%) on the skyline radiographs, including ten knees (10%) with joint space obliteration. Patients with medial patellofemoral degeneration had a similar outcome to those without. For some outcome measures patients with lateral patellofemoral degeneration had a worse score than those without, but these patients still had a good outcome, with a mean Oxford knee score of 37.6 (SD 9.5). These results show that neither anterior knee pain nor radiologically-demonstrated medial patellofemoral joint degeneration should be considered a contraindication to Oxford unicompartmental knee replacement. With lateral patellofemoral degeneration the situation is less well defined and caution should be observed.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 4 | Pages 471 - 476
1 Apr 2007
Kim Y Kim J Yoon S

Bilateral sequential total knee replacement was carried out under one anaesthetic in 100 patients. One knee was replaced using a CT-free computer-assisted navigation system and the other conventionally without navigation. The two methods were compared for accuracy of orientation and alignment of the components. There were 85 women and 15 men with a mean age of 67.6 years (54 to 83). Radiological and CT imaging was carried out to determine the alignment of the components. The mean follow-up was 2.3 years (2 to 3).

The operating and tourniquet times were significantly longer in the navigation group (p < 0.001). There were no significant pre- or post-operative differences between the knee scores of the two groups (p = 0.288 and p = 0.429, respectively). The results of imaging and the number of outliers for all radiological parameters were not statistically different (p = 0.109 to p = 0.920).

In this series computer-assisted navigated total knee replacement did not result in more accurate orientation and alignment of the components than that achieved by conventional total knee replacement.


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 1 | Pages 31 - 36
1 Jan 2008
Yoo JH Kang YG Chang CB Seong SC Kim TK

We examined the placement of the stem in relation to the medial tibial cortex when using total knee replacements (TKRs) with medially-offset tibial stems in Korean patients. Measurements were performed on the pre- and post-operative radiographs of 246 osteoarthritic knees replaced between January 2005 and May 2006 using the Genesis II or E-motion TKR with a medially-offset stem. Pre-operatively, we measured the distance between the mechanical axis and that of the tibial shaft and post-operatively, that between the midline of the tibial stem and the axis of the shaft.

Knees were identified in which there was radiological contact between the tip of the stem and the medial tibial cortex. The mechanical axis was located medial to the axis of the shaft in 203 knees (82.5%). Post-operatively, the midline of the tibial stem was located medial to the tibial shaft axis in 196 knees (79.7%). In 16 knees (6.5%) there was radiological contact between the tibial stem or cement mantle and the medial tibial cortex.

Our study has shown that the medially-offset stem in the tibial component may not be a good option for knees undergoing replacement for advanced osteoarthritis in some Korean patients.


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 6 | Pages 804 - 808
1 Jun 2005
Matsuda Y Ishii Y Noguchi H Ishii R

We performed a randomised, prospective study of 80 mobile-bearing total knee arthroplasties (80 knees) in order to measure the effects of varus-valgus laxity and balance on the range of movement (ROM) one year after operation. Forty knees had a posterior-cruciate-ligament (PCL)-retaining prosthesis and the other 40 a PCL-sacrificing prosthesis. In the balanced group (69 knees) in which the difference between varus and valgus was less than 2°, the mean ROM improved significantly from 107.6° to 117.7° (p < 0.0001). By contrast, in the 11 knees which were unbalanced and in which the difference between varus and valgus laxity exceeded 2°, the ROM decreased from a mean of 121.0° to 112.7° (p = 0.0061). We conclude that coronal laxity, especially balanced laxity, is important for achieving an improved ROM in mobile-bearing total knee arthroplasty.


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 1 | Pages 43 - 49
1 Jan 2008
Smith AJ Wood DJ Li M

We have examined the differences in clinical outcome of total knee replacement (TKR) with and without patellar resurfacing in a prospective, randomised study of 181 osteoarthritic knees in 142 patients using the Profix total knee system which has a femoral component with features considered to be anatomical and a domed patellar implant.

The procedures were carried out between February 1998 and November 2002. A total of 159 TKRs in 142 patients were available for review at a mean of four years (3 to 7). The patients and the clinical evaluator were blinded in this prospective study. Evaluation was undertaken annually by an independent observer using the knee pain scale and the Knee Society clinical rating system. Specific evaluation of anterior knee pain, stair-climbing and rising from a seated to a standing position was also undertaken.

No benefit was shown of TKR with patellar resurfacing over that without resurfacing with respect to any of the measured outcomes. In 22 of 73 knees (30.1%) with and 18 of 86 knees (20.9%) without patellar resurfacing there was some degree of anterior knee pain (p = 0.183). No revisions related to the patellofemoral joint were performed in either group. Only one TKR in each group underwent a re-operation related to the patellofemoral joint. A significant association between knee flexion contracture and anterior knee pain was observed in those knees with patellar resurfacing (p = 0.006).