Advertisement for orthosearch.org.uk
Results 121 - 140 of 148
Results per page:
The Bone & Joint Journal
Vol. 95-B, Issue 9 | Pages 1204 - 1208
1 Sep 2013
Kerens B Boonen B Schotanus MGM Lacroix H Emans PJ Kort NP

Although it has been suggested that the outcome after revision of a unicondylar knee replacement (UKR) to total knee replacement (TKR) is better when the mechanism of failure is understood, a comparative study on this subject has not been undertaken.

A total of 30 patients (30 knees) who underwent revision of their unsatisfactory UKR to TKR were included in the study: 15 patients with unexplained pain comprised group A and 15 patients with a defined cause for pain formed group B. The Oxford knee score (OKS), visual analogue scale for pain (VAS) and patient satisfaction were assessed before revision and at one year after revision, and compared between the groups.

The mean OKS improved from 19 (10 to 30) to 25 (11 to 41) in group A and from 23 (11 to 45) to 38 (20 to 48) in group B. The mean VAS improved from 7.7 (5 to 10) to 5.4 (1 to 8) in group A and from 7.4 (2 to 9) to 1.7 (0 to 8) in group B. There was a statistically significant difference between the mean improvements in each group for both OKS (p = 0.022) and VAS (p = 0.002). Subgroup analysis in group A, performed in order to define a patient factor that predicts outcome of revision surgery in patients with unexplained pain, showed no pre-operative differences between both subgroups.

These results may be used to inform patients about what to expect from revision surgery, highlighting that revision of UKR to TKR for unexplained pain generally results in a less favourable outcome than revision for a known cause of pain.

Cite this article: Bone Joint J 2013;95-B:1204–8.


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 1 | Pages 35 - 39
1 Jan 2006
Beaulé PE Campbell PA Hoke R Dorey F

During hip resurfacing arthroplasty, excessive valgus positioning or surgical technique can result in notching of the femoral neck. Although mechanical weakening and subsequent fracture of the femoral neck are well described, the potential damage to the retinacular vessels leading to an ischaemic event is relatively unknown. Using laser Doppler flowmetry, we measured the blood flow in 14 osteoarthritic femoral heads during routine total hip replacement surgery, before and after notching of the femoral neck. In ten hips there was a reduction in blood flow of more than 50% from the baseline value after simulated notching of the femoral neck. Our results suggest that femoral head vascularity in the osteoarthritic state is similar to the non-arthritic state, where damage to the extraosseous vessels can predispose to avascular necrosis. Surgeons who perform resurfacing arthroplasty of the hip should pay careful attention to these vessels by avoiding excessive dissection around the femoral neck and/or notching.


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 5 | Pages 741 - 744
1 May 2005
Beaulé PE Krismer M Mayrhofer P Wanner S Le Duff M Mattesich M Stoeckl B Amstutz HC Biedermann R

Studies on the migration of an implant may be the only way of monitoring the early performance of metal-on-metal prostheses. The Ein Bild Roentgen Analyse - femoral component analysis (EBRA-FCA) method was adapted to measure migration of the femoral component in a metal-on-metal surface arthroplasty of the hip using standard antero-posterior radiographs. In order to determine the accuracy and precision of this method a prosthesis was implanted into cadaver bones. Eleven series of radiographs were used to perform a zero-migration study. After adjustment of the femoral component to simulate migration of 3 mm the radiographs were repeated. All were measured independently by three different observers.

The accuracy of the method was found to be ± 1.6 mm for the x-direction and ± 2 mm for the y-direction (95% percentile). The method was validated using 28 hips with a minimum follow-up of 3.5 years after arthroplasty. Seventeen were sound, but 11 had failed because of loosening of the femoral component. The normal (control) group had a different pattern of migration compared with that of the loose group. At 29.2 months, the control group showed a mean migration of 1.62 mm and 1.05 mm compared with 4.39 mm and 4.05 mm in the failed group, for the centre of the head and the tip of the stem, respectively (p = 0.001). In the failed group, the mean time to migration greater than 2 mm was earlier than the onset of clinical symptoms or radiological evidence of failure, 19.1 versus 32.2 months (p = 0.001) and 24.8 months (p = 0.012), respectively.

EBRA-FCA is a reliable and valid tool for measuring migration of the femoral component after surface arthroplasty and can be used to predict early failure of the implant. It may be of value in determining the long-term performance of surface arthroplasty.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 7 | Pages 890 - 896
1 Jul 2011
Bajwa AS Villar RN

Arthroscopy of the native hip is an established diagnostic and therapeutic procedure. Its application in the symptomatic replaced hip is still being explored. We describe the use of arthroscopy of the hip in 24 symptomatic patients following total hip replacement, resurfacing arthroplasty of the hip and partial resurfacing (study group), and compared it with arthroscopy of the native hip in 24 patients (control group). A diagnosis was made or confirmed at arthroscopy in 23 of the study group and a therapeutic arthroscopic intervention resulted in relief of symptoms in ten of these. In a further seven patients it led to revision hip replacement. In contrast, arthroscopy in the control group was diagnostic in all 24 patients and the resulting arthroscopic therapeutic intervention provided symptomatic relief in 21.

The mean operative time in the study group (59.7 minutes (35 to 93)) was less than in the control group (71 minutes (40 to 100), p = 0.04) but the arthroscopic approach was more difficult in the arthroplasty group. We suggest that arthroscopy has a role in the management of patients with a symptomatic arthroplasty when other investigations have failed to provide a diagnosis.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 9 | Pages 1169 - 1177
1 Sep 2011
Underwood R Matthies A Cann P Skinner JA Hart AJ

The Articular Surface Replacement (ASR) hip resurfacing arthroplasty has a failure rate of 12.0% at five years, compared with 4.3% for the Birmingham Hip Resurfacing (BHR). We analysed 66 ASR and 64 BHR explanted metal-on-metal hip replacements with the aim of understanding their mechanisms of failure. We measured the linear wear rates of the acetabular and femoral components and analysed the clinical cause of failure, pre-revision blood metal ion levels and orientation of the acetabular component.

There was no significant difference in metal ion levels (chromium, p = 0.82; cobalt, p = 0.40) or head wear rate (p = 0.14) between the two groups. The ASR had a significantly increased rate of wear of the acetabular component (p = 0.03) and a significantly increased occurrence of edge loading (p < 0.005), which can be attributed to differences in design between the ASR and BHR. The effects of differences in design on the in vivo wear rates are discussed: these may provide an explanation as to why the ASR is more sensitive to suboptimal positioning than the BHR.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 3 | Pages 405 - 408
1 Mar 2009
Monk AP Keys GW Murray DW

We describe a technique for the diagnosis of loosening of the femoral component of the Oxford Unicompartmental Knee Replacement using accurately aligned lateral radiographs in extension and flexion. If gaps are present between the component and cement on one radiograph and not on the other, the component is loose.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 8 | Pages 1071 - 1076
1 Aug 2012
Lim H Bae J Song S Kim S

Medium-term survivorship of the Oxford phase 3 unicompartmental knee replacement (UKR) has not yet been established in an Asian population. We prospectively evaluated the outcome of 400 phase 3 Oxford UKRs in 320 Korean patients with a mean age at the time of operation of 69 years (48 to 82). The mean follow-up was 5.2 years (1 to 10). Clinical and radiological assessment was carried out pre- and post-operatively. At five years, the mean Knee Society knee and functional scores had increased significantly from 56.2 (30 to 91) pre-operatively to 87.2 (59 to 98) (p = 0.034) and from 59.2 (30 to 93) to 88.3 (50 to 100) (p = 0.021), respectively. The Oxford knee score increased from a mean of 25.8 (12 to 39) pre-operatively to 39.8 (25 to 58) at five years (p = 0.038). The ten-year survival rate was 94% (95% confidence interval 90.1 to 98.0). A total of 14 UKRs (3.5%) required revision. The most common reason for revision was dislocation of the bearing in 12 (3%). Conversion to a total knee replacement was required in two patients who developed osteoarthritis of the lateral compartment.

This is the largest published series of UKR in Korean patients. It shows that the mid-term results after a minimally invasive Oxford phase 3 UKR can yield satisfactory clinical and functional results in this group of patients.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 10 | Pages 1321 - 1329
1 Oct 2012
Sarmah SS Patel S Hossain FS Haddad FS

Radiological assessment of total and unicompartmental knee replacement remains an essential part of routine care and follow-up. Appreciation of the various measurements that can be identified radiologically is important. It is likely that routine plain radiographs will continue to be used, although there has been a trend towards using newer technologies such as CT, especially in a failing knee, where it provides more detailed information, albeit with a higher radiation exposure.

The purpose of this paper is to outline the radiological parameters used to evaluate knee replacements, describe how these are measured or classified, and review the current literature to determine their efficacy where possible.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 6 | Pages 762 - 767
1 Jun 2012
Sternheim A Rogers BA Kuzyk PR Safir OA Backstein D Gross AE

The treatment of substantial proximal femoral bone loss in young patients with developmental dysplasia of the hip (DDH) is challenging. We retrospectively analysed the outcome of 28 patients (30 hips) with DDH who underwent revision total hip replacement (THR) in the presence of a deficient proximal femur, which was reconstructed with an allograft prosthetic composite. The mean follow-up was 15 years (8.5 to 25.5). The mean number of previous THRs was three (1 to 8). The mean age at primary THR and at the index reconstruction was 41 years (18 to 61) and 58.1 years (32 to 72), respectively. The indication for revision included mechanical loosening in 24 hips, infection in three and peri-prosthetic fracture in three.

Six patients required removal and replacement of the allograft prosthetic composite, five for mechanical loosening and one for infection. The survivorship at ten, 15 and 20 years was 93% (95% confidence interval (CI) 91 to 100), 75.5% (95% CI 60 to 95) and 75.5% (95% CI 60 to 95), respectively, with 25, eight, and four patients at risk, respectively. Additionally, two junctional nonunions between the allograft and host femur required bone grafting and plating.

An allograft prosthetic composite affords a good long-term outcome in the management of proximal femoral bone loss in revision THR in patients with DDH, while preserving distal host bone.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 6 | Pages 720 - 724
1 Jun 2009
Ollivere B Darrah C Brankin RC Donnell ST Walton NP

We report the follow-up at 12 years of the use of the Elite Plus total hip replacement (THR). We have previously reported the results at a mean of 6.4 years. Of the 217 patients (234 THRs), 83 had died and nine had been lost to follow-up. The patients were reviewed radiologically and clinically using the Oxford hip score.

Of the 234 THRs, 19 (8.1%) had required a revision by the final follow-up in all but one for aseptic loosening. Survivorship analysis for revision showed a survival of 93.9% (95% confidence interval (CI) 89.2 to 96.5) at ten years, and of 88.0% (95% CI 81.8 to 92.3) at 12 years. At the final follow-up survival analysis showed that 37% (95% CI 37.3 to 44.7) of the prostheses had either failed radiologically or had been revised. Patients with a radiologically loose femoral component had a significantly poorer Oxford hip score than those with a well-fixed component (p = 0.03). Radiological loosening at 6.4 years was predictive of failure at 12 years.

Medium-term radiographs and clinical scores should be included in the surveillance of THR to give an early indication of the performance of specific implants.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 2 | Pages 158 - 162
1 Feb 2012
Sternheim A Backstein D Kuzyk PRT Goshua G Berkovich Y Safir O Gross AE

We report the use of porous metal acetabular revision shells in the treatment of contained bone loss. The outcomes of 53 patients with 50% acetabular bleeding host bone contact were compared with a control group of 49 patients with > 50% to 85% bleeding host bone contact. All patients were treated with the same type of trabecular metal acetabular revision shell. The mean age at revision was 62.4 years (42 to 80) and the mean follow-up of both groups was 72.4 months (60 to 102). Clinical, radiological and functional outcomes were assessed. There were four (7.5%) mechanical failures in the 50% host bone contact group and no failures in the > 50% host bone contact group (p = 0.068). Out of both groups combined there were four infections (3.9%) and five recurrent dislocations (4.9%) with a stable acetabular component construct that were revised to a constrained liner. Given the complexity of the reconstructive challenge, porous metal revision acetabular shells show acceptable failure rates at five to ten years’ follow-up in the setting of significant contained bone defects. This favourable outcome might be due to the improved initial stability achieved by a high coefficient of friction between the acetabular implant and the host bone, and the high porosity, which affords good bone ingrowth.


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 3 | Pages 265 - 271
1 Mar 2008
Mandalia V Eyres K Schranz P Toms AD

Evaluation of patients with painful total knee replacement requires a thorough clinical examination and relevant investigations in order to reach a diagnosis. Awareness of the common and uncommon problems leading to painful total knee replacement is useful in the diagnostic approach. This review article aims to act as a guide to the evaluation of patients with painful total knee replacement.


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 6 | Pages 730 - 733
1 Jun 2006
Quinlan JF O’Shea K Doyle F Brady OH

The in-cement technique for revision hip arthroplasty involves retaining the original cement-bone interface. This has been proven to be a biomechanically stronger method than recementing after complete removal of the original cement mantle.

This study reviewed a series of 54 consecutive revision hip arthroplasty procedures, using the in-cement technique, between November 1999 and November 2003. Clinical and radiological follow-up included functional assessment.

There were 54 procedures performed in 51 patients, whose mean age at surgery was 70.3 years (45 to 85). A total of 42 were available at a mean follow-up of 29.2 months (6 to 51). There was no radiological evidence of loosening. Functional assessments were available for 40 patients who had a mean Harris hip score of 85.2 (51.9 to 98.5), a mean Oxford hip score of 19.6 (12 to 41), a mean UCLA activity profile score of 5.9 (3 to 8) and a mean SF-36 score of 78.0 (31.6 to 100).

The in-cement technique provides consistent, high functional outcomes and should be considered in appropriately selected cases.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 1 | Pages 16 - 20
1 Jan 2007
Gill HS Campbell PA Murray DW De Smet KA

Resurfacing arthroplasty of the hip is being used increasingly as an alternative to total hip replacement, especially for young active patients. There is concern about necrosis of the femoral head after resurfacing which can result in fracture and loosening. Most systems use a cemented femoral component, with the potential for thermal necrosis of the cancellous bone of the reamed femoral head. We used thermal probes to record temperatures close to the cement-bone interface during resurfacing arthroplasty.

The maximum temperature recorded at the cement-bone interface in four cases was approximately 68°C which was higher than that reported to kill osteocytes. A modified surgical technique using insertion of a suction cannula into the lesser trochanter, generous pulsed lavage and early reduction of the joint significantly reduced the maximum recorded cancellous bone temperature to approximately 36°C in five cases (p = 0.014).

We recommend the modified technique since it significantly reduces temperatures at the cement-bone interface.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 1 | Pages 21 - 25
1 Jan 2007
Khan A Yates P Lovering A Bannister GC Spencer RF

We determined the effect of the surgical approach on perfusion of the femoral head during hip resurfacing arthroplasty by measuring the concentration of cefuroxime in bone samples from the femoral head. A total of 20 operations were performed through either a transgluteal or an extended posterolateral approach.

The concentration of cefuroxime in bone was significantly greater when using the transgluteal approach (mean 15.7 mg/kg; 95% confidence interval 12.3 to 19.1) compared with that using the posterolateral approach (mean 5.6 mg/kg; 95% confidence interval 3.5 to 7.8; p < 0.001). In one patient, who had the operation through a posterolateral approach, cefuroxime was undetectable.

Using cefuroxime as an indirect measure of blood flow, the posterolateral approach was found to be associated with a significant reduction in the blood supply to the femoral head during resurfacing arthroplasty compared with the transgluteal approach.


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 2 | Pages 167 - 170
1 Feb 2005
Treacy RBC McBryde CW Pynsent PB

We report the survival at five years of 144 consecutive metal-on-metal resurfacings of the hip implanted between August 1997 and May 1998. Failure was defined as revision of either the acetabular or femoral component for any reason during the study period. The survival at the end of five years was 98% overall and 99% for aseptic revisions only. The mean age of the patients at implantation was 52.1 years.

Three femoral components failed during the first two years, two were infected and one fractured. A single stage revision was carried out in each case. No other revisions were performed or are impending. No patients were lost to follow-up. Four died from unrelated causes during the study period.

This study confirms that hip resurfacing using a metal-on-metal bearing of known provenance can provide a solution in the medium term for the younger more active adult who requires surgical intervention for hip disease.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 7 | Pages 879 - 886
1 Jul 2007
Howie DW Wimhurst JA McGee MA Carbone TA Badaruddin BS

We investigated 219 revisions of total hip replacement (THR) in 211 patients using a collarless double-taper cemented femoral component. The mean age of the patients was 72 years (30 to 90). The 137 long and 82 standard length stems were analysed separately. The mean follow-up was six years (2 to 18), and no patient was lost to follow-up.

Survival of the long stems to re-revision for aseptic loosening at nine years was 98% (95% confidence interval (CI) 94 to 100), and for the standard stems was 93% (95% CI 85 to 100). At five years, one long stem was definitely loose radiologically and one standard stem was probably loose. Pre-operative femoral bone deficiency did not influence the results for the long stems, and corrective femoral osteotomy was avoided, as were significant subsidence, major stress shielding and persistent thigh pain.

Because of these reliable results, cemented long collarless double-taper femoral components are recommended for routine revision THR in older patients.


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 8 | Pages 1110 - 1115
1 Aug 2006
Ong KL Kurtz SM Manley MT Rushton N Mohammed NA Field RE

The effects of the method of fixation and interface conditions on the biomechanics of the femoral component of the Birmingham hip resurfacing arthroplasty were examined using a highly detailed three-dimensional computer model of the hip. Stresses and strains in the proximal femur were compared for the natural femur and for the femur resurfaced with the Birmingham hip resurfacing. A comparison of cemented versus uncemented fixation showed no advantage of either with regard to bone loading. When the Birmingham hip resurfacing femoral component was fixed to bone, proximal femoral stresses and strains were non-physiological. Bone resorption was predicted in the inferomedial and superolateral bone within the Birmingham hip resurfacing shell. Resorption was limited to the superolateral region when the stem was not fixed. The increased bone strain observed adjacent to the distal stem should stimulate an increase in bone density at that location. The remodelling of bone seen during revision of failed Birmingham hip resurfacing implants appears to be consistent with the predictions of our finite element analysis.


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 3 | Pages 306 - 309
1 Mar 2005
Ito H Matsuno T Minami A

We present the medium-term results of hybrid total hip arthroplasties using pre-coated stems with a second-generation cementing technique. The 128 hips in 111 patients (18 men and 93 women) were followed up at a mean of 11 years after surgery. The mean age at the time of surgery was 61 years. Both components of one hip were removed at ten months after surgery for infection. None of the other 127 femoral components showed possible, probable, or definite loosening at the most recent follow-up. Five acetabular components were revised for aseptic loosening, recurrent dislocation, or displacement of the polyethylene liner from the metal shell. The mean Harris hip score at follow-up was 84 points. A pre-coated femoral component with a second-generation cementing technique provides good clinical function and survival in the medium term.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 6 | Pages 738 - 745
1 Jun 2011
Davda K Lali FV Sampson B Skinner JA Hart AJ

We retrospectively analysed concentrations of chromium and cobalt ions in samples of synovial fluid and whole blood taken from a group of 92 patients with failed current-generation metal-on-metal hip replacements. We applied acid oxidative digestion to our trace metal analysis protocol, which found significantly higher levels of metal ion concentrations in blood and synovial fluid than a non-digestive method. Patients were subcategorised by mode of failure as either ‘unexplained pain’ or ‘defined causes’. Using this classification, chromium and cobalt ion levels were present over a wider range in synovial fluid and not as strongly correlated with blood ion levels as previously reported. There was no significant difference between metal ion concentrations and manufacturer of the implant, nor femoral head size below or above 50 mm. There was a moderately positive correlation between metal ion levels and acetabular component inclination angle as measured on three-dimensional CT imaging.

Our results suggest that acid digestion of samples of synovial fluid samples is necessary to determine metal ion concentrations accurately so that meaningful comparisons can be made between studies.