The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 12 | Pages 1561 - 1565
1 Dec 2009
Blakey CM Eswaramoorthy VK Hamilton LC Biant LC Field RE

We report the minimum five-year follow-up of 352 primary total hip replacements using the uncemented hydroxyapatite-coated ANCA-Fit femoral component with a modular neck and head. The series comprised 319 patients (212 men, 107 women) with a mean age at operation of 64.4 years (28 to 97). The principal diagnosis was osteoarthritis. A total of 18 patients (21 hips) died before their follow-up at five years, nine patients (11 hips) were lost to follow-up, and four (four hips) declined further follow-up. Patient-reported outcomes have been recorded for 288 patients (316 hips).

Their mean Oxford Hip Score improved significantly from 41 points (16 to 57) pre-operatively to 20 points (12 to 44) at five-year follow-up. Radiological assessment showed good bony stability in 98% of implants. There were two cases of aseptic loosening of the femoral component. There were no clinical or radiological complications related to modularity. In our series we did not see the high rate of intra-operative fracture previously reported for this implant.

This medium-term follow-up study demonstrates that the clinical outcome of the ANCA-Fit femoral component is, to date, comparable with that of other metaphyseal loading femoral components.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 2 | Pages 158 - 163
1 Feb 2011
Baker RP Pollard TCB Eastaugh-Waring SJ Bannister GC

We compared the medium-term clinical and radiological results of hybrid total hip replacement (THR) with metal-on-metal Birmingham hip resurfacing (BHR) in two groups of 54 young patients matched for age, gender, body mass index and pre-operative levels of activity.

The clinical outcome was assessed by the University of California, Los Angeles (UCLA) activity score, the Oxford Hip Score (OHS) and the EuroQol scores. Radiologically, all hips were assessed for migration and osteolysis, the hybrid THRs for polyethylene wear and the BHRs for a pedestal sign. The mean follow-up of the patients with a hybrid THR was ten years and for those with a BHR, nine years. Four patients with a hybrid THR and one with a BHR had died. In each group five were lost to follow-up. The revision rate of the hybrid THRs was 16.7% (9 of 54) and of the BHRs 9.3% (5 of 54) (p = 0.195). Radiographs of a further eight hybrid THRs demonstrated wear and osteolysis, and they await revision (p = 0.008). Of the unrevised BHRs 90% had radiological changes, of which approximately 50% had progressed over the previous four years. All hybrid THRs demonstrated linear polyethylene wear with a mean of 1.24 mm (0.06 to 3.03). The BHRs recorded superior OHS (p = 0.013), UCLA (p = 0.008), and EuroQol visual analogue scores (p = 0.009).

After nine years, patients with BHRs remained more active and had a lower rate of revision than those with hybrid THRs. Both groups demonstrated progressive radiological changes at medium-term follow-up.


The Bone & Joint Journal
Vol. 95-B, Issue 11_Supple_A | Pages 46 - 50
1 Nov 2013
Issa K Pivec R Kapadia BH Banerjee S Mont MA

Symptomatic hip osteonecrosis is a disabling condition with a poorly understood aetiology and pathogenesis. Numerous treatment options for hip osteonecrosis are described, which include non-operative management and joint preserving procedures, as well as total hip replacement (THR). Non-operative or joint preserving treatment may improve outcomes when an early diagnosis is made before the lesion has become too large or there is radiographic evidence of femoral head collapse. The presence of a crescent sign, femoral head flattening, and acetabular involvement indicate a more advanced-stage disease in which joint preserving options are less effective than THR. Since many patients present after disease progression, primary THR is often the only reliable treatment option available. Prior to the 1990s, outcomes of THR for osteonecrosis were poor. However, according to recent reports and systemic reviews, it is encouraging that with the introduction of newer ceramic and/or highly cross-linked polyethylene bearings as well as highly-porous fixation interfaces, THR appears to be a reliable option in the management of end-stage arthritis following hip osteonecrosis in this historically difficult to treat patient population.

Cite this article: Bone Joint J 2013;95-B, Supple A:46–50.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 11 | Pages 1454 - 1458
1 Nov 2009
Eswaramoorthy VK Biant LC Field RE

We report the outcome of total hip replacement in 29 failed metal-on-metal resurfacing hip replacements in which the primary surgery was performed between August 1995 and February 2005. The mean length of follow-up was five years (1.7 to 11.7). Of the 29 hip resurfacings, 19 acetabular components and all the femoral components were revised (28 uncemented stems and one cemented stem). There were no deaths and none of the patients was lost to follow-up. None of the hips underwent any further revision. The results of the revision resurfacing group were compared with those of a control group of age-matched patients. In the latter group there were 236 primary total hip replacements and 523 resurfacings performed during the same period by the same surgeons.

The outcome of the revision resurfacing group was comparable with that of the stemmed primary hip replacement group but was less good than that of the primary hip resurfacing group. Long-term follow-up is advocated to monitor the outcome of these cases.


The Bone & Joint Journal
Vol. 95-B, Issue 11_Supple_A | Pages 88 - 91
1 Nov 2013
Su EP Su SL

Surface hip replacement (SHR) is generally used in younger, active patients as an alternative conventional total hip replacement in part because of the ability to preserve femoral bone. This major benefit of surface replacement will only hold true if revision procedures of SHRs are found to provide good clinical results.

A retrospective review of SHR revisions between 2007 and 2012 was presented, and the type of revision and aetiologies were recorded. There were 55 SHR revisions, of which 27 were in women. At a mean follow-up of 2.3 years (0.72 to 6.4), the mean post-operative Harris hip score (HHS) was 94.8 (66 to 100). Overall 23 were revised for mechanical reasons, nine for impingement, 13 for metallosis, nine for unexplained pain and one for sepsis. Of the type of revision surgery performed, 14 were femoral-only revisions; four were acetabular-only revisions, and 37 were complete revisions.

We did not find that clinical scores were significantly different between gender or different types of revisions. However, the mean post-operative HHS was significantly lower in patients revised for unexplained pain compared with patients revised for mechanical reasons (86.9 (66 to 100) versus 99 (96 to 100); p = 0.029). There were two re-revisions for infection in the entire cohort.

Based on the overall clinical results, we believe that revision of SHR can have good or excellent results and warrants a continued use of the procedure in selected patients. Close monitoring of these patients facilitates early intervention, as we believe that tissue damage may be related to the duration of an ongoing problem. There should be a low threshold to revise a surface replacement if there is component malposition, rising metal ion levels, or evidence of soft-tissue abnormalities.

Cite this article: Bone Joint J 2013;95-B, Supple A:88–91.


The Bone & Joint Journal
Vol. 95-B, Issue 11 | Pages 1458 - 1463
1 Nov 2013
Won S Lee Y Ha Y Suh Y Koo K

Pre-operative planning for total hip replacement (THR) is challenging in hips with severe acetabular deformities, including those with a hypoplastic acetabulum or severe defects and in the presence of arthrodesis or ankylosis. We evaluated whether a Rapid Prototype (RP) model, which is a life-sized reproduction based on three-dimensional CT scans, can determine the feasibility of THR and provide information about the size and position of the acetabular component in severe acetabular deformities. THR was planned using an RP model in 21 complex hips in five men (five hips) and 16 women (16 hips) with a mean age of 47.7 years (24 to 70) at operation. An acetabular component was implanted successfully and THR completed in all hips. The acetabular component used was within 2 mm of the predicted size in 17 hips (80.9%). All of the acetabular components and femoral stems had radiological evidence of bone ingrowth and stability at the final follow-up, without any detectable wear or peri-prosthetic osteolysis. The RP model allowed a simulated procedure pre-operatively and was helpful in determining the feasibility of THR pre-operatively, and to decide on implant type, size and position in complex THRs.

Cite this article: Bone Joint J 2013;95-B:1458–63.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 8 | Pages 1019 - 1024
1 Aug 2009
Grammatopolous G Pandit H Kwon Y Gundle R McLardy-Smith P Beard DJ Murray DW Gill HS

Inflammatory pseudotumours occasionally occur after metal-on-metal hip resurfacing and often lead to revision. Our aim was to determine the severity of this complication by assessing the outcome of revision in these circumstances and by comparing this with the outcome of other metal-on-metal hip resurfacing revisions as well as that of matched primary total hip replacements.

We identified 53 hips which had undergone metal-on-metal hip resurfacing and required revision at a mean of 1.59 years (0.01 to 6.69) after operation. Of these, 16 were revised for pseudotumours, 21 for fracture and 16 for other reasons. These were matched by age, gender and diagnosis with 103 patients undergoing primary total hip replacement with the Exeter implant.

At a mean follow-up of three years (0.8 to 7.2) the outcome of metal-on-metal hip resurfacing revision for pseudotumour was poor with a mean Oxford hip score of 20.9 (sd 9.3) and was significantly worse (p < 0.001) than the outcome for fracture with a mean Oxford hip score of 40.2 (sd 9.2) or that for other causes with a mean Oxford hip score of 37.8 (sd 9.4). The clinical outcome of revision for pseudotumour was also significantly worse (p < 0.001) than the outcome of matched primary total hip replacements. By contrast, the outcome for fracture and other causes was not significantly different from that of matched primary total hip replacements (p = 0.065). After revision for pseudotumour there were three cases of recurrent dislocation, three of palsy of the femoral nerve, one of stenosis of the femoral artery and two of loosening of the component. Five hips required further revision. In three of these there was evidence of recurrent pseudotumour, and one is currently awaiting further revision. The incidence of major complications after revision for pseudotumour (50%) was significantly higher (p = 0.018) than that after revision for other causes (14%).

The outcome of revision for pseudotumour is poor and consideration should be given to early revision to limit the extent of the soft-tissue destruction. The outcome of resurfacing revision for other causes is good.


Bone & Joint 360
Vol. 2, Issue 5 | Pages 16 - 18
1 Oct 2013

The October 2013 Hip & Pelvis Roundup360 looks at: Young and impinging; Clothes, weather and femoral heads?; Go long, go cemented; Surgical repair of the abductors?; Aspirin for DVT prophylaxis?; Ceramic-on-polyethylene: a low-wear solution?; ALVAL and ASR™: the story continues….; Salvaging Legg-Calve-Perthes’ disease


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 4 | Pages 535 - 537
1 Apr 2007
Evangelista GT Fulkerson E Kummer F Di Cesare PE

During open reduction of an irreducible anterior dislocation of a total hip replacement with an Oxinium femoral head, it was observed that the head had been significantly damaged. Gross and scanning electron microscopic examination revealed cracking, gouging, and delamination of the surface. Because of the risk which this poses for damaging the polyethylene acetabular liner, it is strongly recommended that patients with this type of prosthetic head be carefully monitored after a dislocation.


The Bone & Joint Journal
Vol. 95-B, Issue 7 | Pages 881 - 886
1 Jul 2013
Bjorgul K Novicoff WN Andersen ST Ahlund OR Bunes A Wiig M Brevig K

A total of 397 hips were randomised to receive Metasul metal-on-metal (MoM), metal-on-conventional polyethylene (MoP) or ceramic-on-polyethylene (CoP) bearings using a cemented triple-tapered polished femoral component (MS-30). There were 129 MoM hips in 123 patients (39 male and 84 female, mean age 63.3 years (40.7 to 72.9)), 137 MoP hips in 127 patients (39 male and 88 female, mean age 62.8 years (24.5 to 72.7)) and 131 CoP hips in 124 patients (51 male and 73 female, mean age 63.9 years (30.6 to 73.8)). All acetabular components were cemented Weber polyethylene components with the appropriate inlay for the MoM articulation. Clinical evaluation was undertaken using the Harris hip score (HHS) and radiological assessments were made at two, five and seven years. The HHS and radiological analysis were available for 341 hips after seven years. The MoM group had the lowest mean HHS (p = 0.124), a higher rate of revision (p < 0.001) and a higher incidence of radiolucent lines in unrevised hips (p < 0.001). In all, 12 revisions had been performed in 12 patients: eight in the MoM group (four for infection, four for aseptic loosening, three in the MoP group (one each of infection, dislocation and pain) and one in the CoP group (infection).

Our findings reveal no advantage to the MoM bearing and identified a higher revision rate and a greater incidence of radiolucent lines than with the other articulations. We recommend that patients with a 28 mm Metasul MoM bearing be followed carefully.

Cite this article: Bone Joint J 2013;95-B:881–6.


The Bone & Joint Journal
Vol. 95-B, Issue 8 | Pages 1040 - 1044
1 Aug 2013
Kazi HA Perera JR Gillott E Carroll FA Briggs TWR

We prospectively assessed the efficacy of a ceramic-on-metal (CoM) hip bearing with uncemented acetabular and femoral components in which cobalt­–chrome acetabular liners and alumina ceramic heads were used.

The cohort comprised 94 total hip replacements (THRs) in 83 patients (38 women and 45 men) with a mean age of 58 years (42 to 70). Minimum follow-up was two years. All patients had pre- and post-operative assessment using the Western Ontario and McMaster Universities osteoarthritis index (WOMAC), Oxford hip score and Short-Form 12 scores. All showed a statistically significant improvement from three months post-operatively onwards (all p < 0.001).

After two years whole blood metal ion levels were measured and chromosomal analysis was performed. The levels of all metal ions were elevated except vanadium. Levels of chromium, cobalt, molybdenum and titanium were significantly higher in patients who underwent bilateral THR compared with those undergoing unilateral THR (p < 0.001). Chromosomal analysis demonstrated both structural and aneuploidy mutations. There were significantly more breaks and losses than in the normal population (p < 0.001). There was no significant difference in chromosomal aberration between those undergoing unilateral and bilateral procedures (all analyses p ≥ 0.62).

The use of a CoM THR is effective clinically in the short-term, with no concerns, but the significance of high metal ion levels and chromosomal aberrations in the long-term remains unclear.

Cite this article: Bone Joint J 2013;95-B:1040–44.


Bone & Joint 360
Vol. 2, Issue 2 | Pages 10 - 12
1 Apr 2013

The April 2013 Hip & Pelvis Roundup360 looks at: hip cartilage and magnets; labral repair or resection; who benefits from injection; rotational osteotomy for osteonecrosis; whether ceramic implants risk fracture; dual articulation; and hydroxyapatite.


The Bone & Joint Journal
Vol. 96-B, Issue 2 | Pages 147 - 156
1 Feb 2014
Rajpura A Kendoff D Board TN

We reviewed the literature on the currently available choices of bearing surface in total hip replacement (THR). We present a detailed description of the properties of articulating surfaces review the understanding of the advantages and disadvantages of existing bearing couples. Recent technological developments in the field of polyethylene and ceramics have altered the risk of fracture and the rate of wear, although the use of metal-on-metal bearings has largely fallen out of favour, owing to concerns about reactions to metal debris. As expected, all bearing surface combinations have advantages and disadvantages. A patient-based approach is recommended, balancing the risks of different options against an individual’s functional demands.

Cite this article: Bone Joint J 2014;96-B:147–56.


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 3 | Pages 299 - 305
1 Mar 2008
Kim Y

This study reviewed the results of a cementless anatomical femoral component to give immediate post-operative stability, and with a narrow distal section in order not to contact the femoral cortex in the diaphysis, ensuring exclusively metaphyseal loading.

A total of 471 patients (601 hips) who had a total hip replacement between March 1995 and February 2002 were included in the study. There were 297 men and 174 women. The mean age at the time of operation was 52.7 years (28 to 63). Clinical and radiological evaluation were performed at each follow-up. Bone densitometry was carried out on all patients two weeks after operation and at the final follow-up examination. The mean follow-up was 8.8 years (5 to 12).

The mean pre-operative Harris hip score was 41 points (16 to 54), which improved to a mean of 96 (68 to 100) at the final follow-up. No patient complained of thigh pain at any stage. No acetabular or femoral osteolysis was observed and no hip required revision for aseptic loosening of either component. Deep infection occurred in two hips (0.3%) which required revision. One hip (0.2%) required revision of the acetabular component for recurrent dislocation. Bone mineral densitometry revealed a minimal bone loss in the proximal femur.

This cementless anatomical femoral component with metaphyseal loading but without distal fixation produced satisfactory fixation and encourages proximal femoral loading.


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 10 | Pages 1279 - 1285
1 Oct 2006
Parvizi J Campfield A Clohisy JC Rothman RH Mont MA

Arthritis of the hip in the young adult can be a disabling condition. Recent years have witnessed extensive research related to the management of this condition. This article reviews the current status with regard to aetiology, diagnosis and treatment of arthritis of the hip in the young adult.


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 11 | Pages 1475 - 1479
1 Nov 2005
Goosen JHM Verheyen CCPM Tulp NJA

We investigated the rate of polyethylene wear of a cementless acetabular component at different periods of follow-up in order to test the hypothesis than an irrecoverable deformation process (creep) was followed by an initially low, but gradually increasing wear rate. We studied prospectively 93 uncemented total hip arthroplasties in 83 patients (mean age 50 years (22 to 63)) with a mean follow-up of 8.2 years (3 to 12). We measured the penetration of the femoral head from radiographs taken immediately after surgery at three, six and nine years, or at the latest follow-up.

The median wear rate was 0.17 mm per year in the first three years, a finding which we considered to be caused by creep. Thereafter, the rate of wear declined to 0.07 mm per year (four- to six-year period) and then increased to 0.17 mm per year (seven to nine years) and 0.27 mm per year (more than nine years), which we considered to be a reflection of genuine polyethylene wear. After the nine-year follow-up the wear rates were higher in patients with marked osteolysis. We found no relationship between the inclination angle of the acetabular component or femoral head orientation and the rate of wear. No acetabular component required revision.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 2 | Pages 163 - 166
1 Feb 2012
Zahos K Mehendale S Ward AJ Smith EJ Nichols M

We report the use of a 15° face-changing cementless acetabular component in patients undergoing total hip replacement for osteoarthritis secondary to developmental dysplasia of the hip. The rationale behind its design and the surgical technique used for its implantation are described. It is distinctly different from a standard cementless hemispherical component as it is designed to position the bearing surface at the optimal angle of inclination, that is, < 45°, while maximising the cover of the component by host bone.


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 5 | Pages 592 - 600
1 May 2006
Pollard TCB Baker RP Eastaugh-Waring SJ Bannister GC

We compared the five- to seven-year clinical and radiological results of the metal-on-metal Birmingham hip resurfacing with a hybrid total hip arthroplasty in two groups of 54 hips, matched for gender, age, body mass index and activity level.

Function was excellent in both groups, as measured by the Oxford hip score, but the Birmingham hip resurfacings had higher University of California at Los Angeles activity scores and better EuroQol quality of life scores. The total hip arthroplasties had a revision or intention-to-revise rate of 8%, and the Birmingham hip resurfacings of 6%. Both groups demonstrated impending failure on surrogate end-points. Of the total hip arthroplasties, 12% had polyethylene wear and osteolysis under observation, and 8% of Birmingham hip resurfacings showed migration of the femoral component. Polyethylene wear was present in 48% of the hybrid hips without osteolysis. Of the femoral components in the Birmingham hip resurfacing group which had not migrated, 66% had radiological changes of unknown significance.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 11 | Pages 1462 - 1467
1 Nov 2012
Schouten R Malone AA Tiffen C Frampton CM Hooper G

In a double-blinded randomised controlled trial, 83 patients with primary osteoarthritis of the hip received either a ceramic-on-metal (CoM) or metal-on-metal (MoM) total hip replacement (THR). The implants differed only in the bearing surfaces used. The serum levels of cobalt and chromium and functional outcome scores were compared pre-operatively and at six and 12 months post-operatively.

Data were available for 41 CoM and 36 MoM THRs (four patients were lost to follow-up, two received incorrect implants). The baseline characteristics of both cohorts were similar. Femoral head size measured 36 mm in all but two patients who had 28 mm heads. The mean serum cobalt and chromium levels increased in both groups, with no difference noted between groups at six months (cobalt p = 0.67, chromium p = 0.87) and 12 months (cobalt p = 0.76, chromium p = 0.76) post-operatively. Similarly, the mean Oxford hip scores, Western Ontario and McMaster Universities Osteoarthritis index and University of California, Los Angeles activity scores showed comparable improvement at 12 months.

Our findings indicate that CoM and MoM couplings are associated with an equivalent increase in serum cobalt and chromium levels, and comparable functional outcome scores at six and 12-months follow-up.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 1 | Pages 37 - 42
1 Jan 2012
Affatato S Traina F De Fine M Carmignato S Toni A

Alumina–alumina bearings are among the most resistant to wear in total hip replacement. Examination of their surfaces is one way of comparing damage caused by wear of hip joints simulated in vitro to that seen in explanted bearings. The aim of this study was to determine whether second-generation ceramic bearings exhibited a better pattern of wear than those reported in the literature for first-generation bearings. We considered both macro- and microscopic findings.

We found that long-term alumina wear in association with a loose acetabular component could be categorised into three groups. Of 20 specimens, four had ‘low wear’, eight ‘crescent wear’ and eight ‘severe wear’, which was characterised by a change in the physical shape of the bearing and a loss of volume. This suggests that the wear in alumina–alumina bearings in association with a loose acetabular component may be variable in pattern, and may explain, in part, why the wear of a ceramic head in vivo may be greater than that seen after in vitro testing.