We carried out a clinical and radiological review of 103 cementless primary hip arthroplasties with a tapered rectangular grit-blasted titanium press-fit femoral component and a threaded conical titanium acetabular component at a mean follow-up of 14.4 years (10.2 to 17.1). The mean Harris hip score at the last follow-up was 89.2 (32 to 100). No early loosening and no fracture of the implant were found. One patient needed revision surgery because of a late deep infection. In 11 hips (10.7%), the reason for revision was progressive wear of the polyethylene liner. Exchange of the acetabular component because of aseptic loosening without detectable liner wear was carried out in three hips (2.9%). After 15 years the survivorship with aseptic loosening as the definition for failure was 95.6% for the acetabular component and 100% for the femoral component.
The purpose of this prospective, randomised study
was to evaluate the clinical and radiological results comparing the
identical cemented or cementless NexGen total knee prostheses implanted
bilaterally in the same patient. Sequential simultaneous bilateral
total knee replacements were performed in 50 patients (100 knees).
There were 39 women and 11 men with a mean age of 58.4 years (51
to 67) who received a cemented prosthesis in one knee and a cementless
prosthesis in the other. The mean follow-up was 13.6 years (13 to
14). At final review, the mean Knee Society scores (96.2 (82 to
100)
Revision after failed femoral components may
be technically demanding due to loss of peri-prosthetic bone. This retrospective
study evaluated the long-term results of femoral revision using
the cementless Wagner Self-Locking stem. Between 1992 and 1998,
68 consecutive hips in 66 patients underwent femoral revision using
this implant. A total of 25 patients died from unrelated causes
without further revision; the remaining 41 hips in 41 patients (12 men
and 29 women) with a mean age of 61 years (29 to 80) were reviewed
at a mean follow-up of 13.9 years (10.4 to 15.8). A transfemoral
approach was used in 32 hips. A total of five stems required further
revision because of infection in two, progressive subsidence in
two and recurrent dislocation in one. Four hips had dislocated and
eight stems had subsided ≥ 10 mm. The mean Harris hip score improved
from 33 points pre-operatively to 75 points at final follow-up (p
<
0.001). In all, 33 stems (91.7%) showed radiological signs
of stable bone fixation. The cumulative survival rates at 15.8 years
with femoral revision for any reason and for stem failure as the
endpoints were 92.0% (95% confidence interval (CI) 86.0% to 98.4%)
and 96.6% (95% CI 92.2% to 100%), respectively. The survivorship
with revision and ≥ 10 mm migration of the stem as the endpoint
was 83.6% (95% CI 76.6% to 91.4%). This study shows quite good survival and moderate clinical outcome
when using a monoblock tapered titanium stem for supporting the
regeneration of bone in complex revision hip surgery.
There have been comparatively few studies of the incidence of osteolysis and the survival of hybrid and cementless total hip replacements (THRs) in patients younger than 50 years of age. We prospectively reviewed 78 patients (109 hips) with a hybrid THR having a mean age of 43.4 years (21 to 50) and 79 patients (110 hips) with a cementless THR with a mean age of 46.8 years (21 to 49). The patients were evaluated clinically using the Harris hip score, the Western Ontario and McMaster Universities (WOMAC) osteoarthritis score and the University of California, Los Angeles (UCLA) activity score. Radiographs and CT scans were assessed for loosening and osteolysis. The mean follow-up was for 18.4 years (16 to 19) in both groups. The mean post-operative Harris hip scores (91 points Although the long-term fixation of the acetabular metallic shell and the cemented and cementless femoral components was outstanding, wear and peri-acetabular osteolysis constitute the major challenges of THR in young patients.
Total shoulder replacement is a successful procedure for degenerative or some inflammatory diseases of the shoulder. However, fixation of the glenoid seems to be the main weakness with a high rate of loosening. The results using all-polyethylene components have been better than those using metal-backed components. We describe our experience with 35 consecutive total shoulder replacements using a new metal-backed glenoid component with a mean follow-up of 75.4 months (48 to 154). Our implant differs from others because of its mechanism of fixation. It has a convex metal-backed bone interface and the main stabilising factor is a large hollow central peg. The patients were evaulated with standard radiographs and with the Constant Score, the Simple Shoulder Test and a visual analogue scale. All the scores improved and there was no loosening, no polyethylene-glenoid disassembly and no other implant-related complications. We conclude that a metal-backed glenoid component is a good option in total shoulder replacement with no worse results than of those using a cemented all-polyethylene prosthesis.
Recently the National Patient Safety Agency in the United Kingdom published a report entitled “Mitigating surgical risk in patients undergoing hip arthroplasty for fractures of the proximal femur”. A total of 26 deaths had been reported to them when cement was used at hemiarthroplasty between October 2003 and October 2008. This paper considers the evidence for using cement fixation of a hemiarthroplasty in the treatment of hip fractures.
The outcome of total hip replacement (THR) is potentially affected by the body mass index (BMI) of the patient. We studied the outcome of 2026 consecutive primary cementless THRs performed for osteoarthritis. The mean follow-up was 6.3 years (0 to 11.71) and no patient was lost to follow-up for survival analysis. The patients were divided into two groups according to their BMI as follows: non-obese (BMI <
30 kg/m2) and obese (BMI ≥ 30 kg/m2). The obese patient undergoing surgery was found to be significantly younger (p <
0.001). The log-rank test for equality of survival showed no difference in the mid-term survival (p = 0.552) with an estimated survival at 11 years of 95.2% (95% CI 92.5 to 98.0) in the non-obese and 96.7% (95% CI 94.9 to 98.5) in the obese groups. The clinical and radiological outcome was determined in a case-matched study performed on 134 obese individuals closely matched with 134 non-obese controls. The non-obese group was found to have a significantly higher post-operative Harris hip score (p <
0.001) and an increased range of movement, but overall satisfaction with surgery was comparable with that of the obese patients. Radiological analysis of the acetabular and femoral components showed no significant differences with regard to radiolucent lines, osteolysis, ingrowth of the femoral component, the acetabular inclination angle or alignment of the femoral component. Our results suggest that the survival of cementless THR is not adversely affected by obesity. Obese patients can therefore be counselled that despite a lower clinical score, they should expect to be satisfied with the result of their THR with a mid-term survival rate equivalent to that of non-obese patients.
Implantation of a large-diameter femoral head prosthesis with a metal-on-metal bearing surface reduces the risk of dislocation, increases the range of movement, minimises the risk of impingement and, in theory, results in little wear. Between February 2004 and March 2007 we implanted 100 consecutive total hip replacements with a metal-on-metal bearing and a large femoral head into 92 patients. There were 51 men and 41 women with a mean age of 50 years (18 to 70) at the time of surgery. Outcome was assessed using the Western Ontario McMaster University osteoarthritis index and the Harris hip score as well as the Devane activity score. These all improved significantly (p <
0.0001). At the last follow-up there were no cases of dislocation, no impingement, a good range of movement and no osteolysis, but seven revisions, two for infection and five for aseptic loosening. The probability of groin pain increased if the other acetabular component inclination exceeded 50° (p = 0.0007). At 4.8 years of follow-up, the projected survival of the Durom acetabular component, with revision for any reason, was 92.4% ( The design of the component made it difficult both to orientate and seat, which when combined with a poor porous coating, produced unpredictable fixation and a low survival at five years.
We report the use of an allograft prosthetic composite for reconstruction of the skeletal defect in complex revision total hip replacement for severe proximal femoral bone loss. Between 1986 and 1999, 72 patients (20 men, 52 women) with a mean age of 59.9 years (38 to 78) underwent reconstruction using this technique. At a mean follow-up of 12 years (8 to 20) 57 patients were alive, 14 had died and one was lost to follow-up. Further revision was performed in 19 hips at a mean of 44.5 months (11 to 153) post-operatively. Causes of failure were aseptic loosening in four, allograft resorption in three, allograft nonunion in two, allograft fracture in four, fracture of the stem in one, and deep infection in five. The survivorship of the allograft-prosthesis composite at ten years was 69.0% (95% confidence interval 67.7 to 70.3) with 26 patients remaining at risk. Survivorship was statistically significantly affected by the severity of the pre-operative bone loss (Paprosky type IV; p = 0.019), the number of previous hip revisions exceeding two (p = 0.047), and the length of the allograft used (p = 0.005).
We report the outcome at ten to 15 years of two-stage revision for hip infection in 99 patients using the Prostalac articulated hip spacer system. All the patients were contacted to determine their current functional and infection status using the Oxford-12, Short form-12, and Western Ontario and McMaster University Osteoarthritis Index questionnaires. A total of 11 of the 99 patients had a further infection, of whom seven responded to repeat surgery with no further sequelae. The mean interval between the stages was five months (1 to 36). We were able to review 48 living patients, with a mean age of 72 years (46 to 86), 34 (71%) of whom provided health-related quality-of-life outcome scores. The mean follow-up was 12 years (10 to 15). The long-term success rate was 89% and with additional surgery this rose to 96%. The mean global Western Ontario and McMaster University Osteoarthritis Index score was 80.6 ( Two-stage revision for hip infection using a Prostalac interim spacer offers a predictable and lasting solution for patients with this difficult problem.
Resurfacing arthroplasty of the hip is being performed more frequently in the United Kingdom. The majority of these patients are younger than 55 years of age, and in this group the key benefits include conservation of femoral bone stock and the potential reduction in the rate of dislocation afforded by the larger resurfacing head. Early aseptic loosening is well recognised in patients younger than 55 years of age, and proponents of resurfacing believe that the improved wear characteristics of the metal-on-metal bearing may improve the long-term survival of this implant. There has been some concern, however, that resurfacing may not be conservative of acetabular bone. We compared a series of 33 consecutive patients who had a hybrid total hip arthroplasty with an uncemented acetabular component and a cemented femoral implant, with 35 patients undergoing a Birmingham hip resurfacing arthroplasty. We compared the diameter of the implanted acetabulum in both groups and, because they were not directly comparable, we corrected for patient size by measuring the diameter of the contralateral femoral head. The data were analysed using unpaired There was a significantly larger acetabulum in the Birmingham arthroplasty group (mean diameter 56.6 mm
Avascular necrosis of the femoral head creates considerable morbidity in successful renal transplant recipients who are generally young and expect active lifestyles. Total hip replacement is considered the treatment of choice in these patients, but surgeons may be wary because of a supposed increase in the risk of infection and other complications. A review of the literature reveals that cemented hip arthroplasty provides good to excellent functional outcomes for renal transplant patients. Most authors have found that the risk of infection is not increased despite chronic immunosuppression, but the rates of general complications are and should be anticipated and treated. There is a high rate of early failure in these patients because of their young age and diffuse osteopenia as a result of secondary hyperparathyroidism related to the underlying renal disease and chronic steroid use. Recent studies have found that despite decreased bone stock in these patients, porous-coated prostheses are not contraindicated.
We describe the survival of 134 consecutive JRI Furlong hydroxyapatite-coated uncemented total hip replacements. The mean follow-up was for 14.2 years (13 to 15). Patients were assessed clinically, using the Merle d’Aubigné and Postel score. Radiographs were evaluated using Gruen zones for the stem and DeLee and Charnley zones for the cup. Signs of subsidence, radiolucent lines, endosteal bone formation (spot welds) and pedestal formation were used to assess fixation and stability of the stem according to Engh’s criteria. Cup angle, migration and radiolucency were used to assess loosening of the cup. The criteria for failure were revision, or impending revision because of pain or loosening. Survival analysis was performed using a life table and the Kaplan-Meier curve. The mean total Merle d’Aubigné and Postel score was 7.4 pre-operatively and 15.9 at follow-up. During the study period 22 patients died and six were lost to follow-up. None of the cups was revised. One stem was revised for a periprosthetic fracture following a fall but none was revised for loosening, giving a 99% survival at 13 years. Our findings suggest that the long-term results of these hydroxyapatite-coated prostheses are more than satisfactory.
This study describes 146 primary total knee replacements, either fully or partially coated with hydroxyapatite of which 74 knees in 68 patients were available for clinical and radiological assessment at a mean of 11.2 years (10 to 15). The global failure rate was 1.37% and survival rate with mechanical failure as the end-point was 98.14%. Radiological assessment indicated intimate contact between bone and the hydroxyapatite coating. Over time the hydroxyapatite coating appears to encourage filling of interface gaps remaining after surgery. Our results compare favourably with those of series describing cemented or porous-coated knee replacements, and suggest that fixation with hydroxyapatite is a reliable option in primary total knee replacement.
We have analysed from the Finnish Arthroplasty Register the long-term survivorship of the 12 most commonly-used cemented implants between 1980 and 2005 in patients aged 55 years or older with osteoarthritis. Only two designs of femoral component, the Exeter Universal and the Müller Straight femoral component had a survivorship of over 95% at ten years with revision for aseptic loosening as the endpoint. At 15 years of the femoral and acetabular component combinations, only the Exeter Universal/Exeter All-poly implant had a survival rate of over 90% with revision for aseptic loosening as the endpoint. In the subgroup of patients aged between 55 and 64 years, survivorship overall was less than 90% at ten years. The variation in the long-term rates of survival of different cemented hip implants was considerable in patients aged 55 years or older. In those aged between 55 and 64 years, none of the cemented prostheses studied yielded excellent long-term survival rates (≥ 90% at 15 years).
We present the clinical and radiographic outcome of 68 consecutive primary total hip replacements performed in 54 patients under the age of 55 years using a hydroxyapatite-coated femoral component and threaded cup with a modular ceramic head (JRI-Furlong). We reviewed 62 (91%) hips at a median follow-up of 8.8 years (5 to 13.8) after implantation; six (9%) were lost to follow-up. At review there had been four (6%) revisions but only one for aseptic loosening (acetabulum). Radiographic review of the remaining hips did not identify any evidence of femoral or acetabular loosening. The median Harris and Merle d’Aubigné and Postel hip scores were 95.9 (42.7 to 100) and 17 (3 to 18) respectively. The JRI-Furlong hip gives promising functional and radiographic results in young patients in the medium term.
Between January 1984 and January 1986, 131 porous-coated anatomic total hip replacements were performed in 119 consecutive patients. Of these, 100 patients (110 hips) who were alive at a minimum of 18 years after replacement were included in the study. The mean age of the patients at surgery was 48.4 years and that of surviving patients at the latest follow-up was 67.8 years. The mean duration of the clinical and radiological follow-up was 19.4 years (18 to 20). The mean Harris hip score initially improved from 55 points before to 95 points at two years after operation, but subsequently decreased to 91 points after six years, 87 points after 11.2 years, and 85 points after 19.4 years. At the final follow-up, 23 hips (18%) of the entire cohort and 21% of survivors had undergone revision because of loosening or osteolysis of the acetabular component and eight hips (6%) of the entire cohort and 7% of survivors for loosening or osteolysis of the femoral component. Only four femoral components (4%) were revised for isolated aseptic loosening without osteolysis and two (2%) for recurrent dislocation. On the basis of these long-term results, the porous-coated anatomic femoral component survived for a minimum of 18 years after operation while the acetabular component was less durable. The findings identify the principles of uncemented joint replacement which can be applied to current practice.
We performed 114 consecutive primary total hip arthroplasties with a cementless expansion acetabular component in 101 patients for advanced osteonecrosis of the femoral head. The mean age of the patients at surgery was 51 years (36 to 62) and the mean length of follow-up was 110 months (84 to 129). The mean pre-operative Harris hip score of 47 points improved to 93 points at final follow-up. The polyethylene liner was exchanged in two hips during this period and one broken acetabular component was revised. The mean linear wear rate of polyethylene was 0.07 mm/year and peri-acetabular osteolysis was seen in two hips (1.9%). Kaplan-Meier analysis indicated that the survival of the acetabular component without revision was 97.8% (95% confidence interval 0.956 to 1.000) at ten years. Our study has shown that the results of THA with a cementless expansion acetabular component and an alumina-polyethylene bearing surface are good.
This study explored the relationship between the initial stability of the femoral component and penetration of cement into the graft bed following impaction allografting. Impaction allografting was carried out in human cadaveric femurs. In one group the cement was pressurised conventionally but in the other it was not pressurised. Migration and micromotion of the implant were measured under simulated walking loads. The specimens were then cross-sectioned and penetration of the cement measured. Around the distal half of the implant we found approximately 70% and 40% of contact of the cement with the endosteum in the pressure and no-pressure groups, respectively. The distal migration/micromotion, and valgus/varus migration were significantly higher in the no-pressure group than in that subjected to pressure. These motion components correlated negatively with the mean area of cement and its contact with the endosteum. The presence of cement at the endosteum appears to play an important role in the initial stability of the implant following impaction allografting.
In an attempt to increase the life of cementless prostheses, an hydroxyapatite-coated implant which releases a bisphosphonate has been suggested as a drug-delivery system. Our Our findings demonstrated that zoledronate did not impair the proliferation of human osteoblasts when used at concentrations below 1 μ A concentration of 0.01% of titanium particles did not impair the proliferation of either cell line. Zoledronate affected the alkaline phosphatase activity of murine osteoblasts through a chelation phenomenon. The presence of titanium particles strongly decreased the alkaline phosphatase activity of murine osteoblasts. We did not detect any synergic effect of zoledronate and titanium particles on the behaviour of both human and murine osteoblasts.