Advertisement for orthosearch.org.uk
Results 1261 - 1280 of 1506
Results per page:
The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 4 | Pages 434 - 440
1 Apr 2007
Board TN Karva A Board RE Gambhir AK Porter ML

Heterotopic ossification following joint replacement in the lower limb occurs in 3% to 90% of cases. Higher grades of heterotopic ossification can result in significant limitation of function and can negate the benefits of joint replacement. The understanding of the pathophysiology of this condition has improved in recent years. It would appear to be related to a combination of systemic and local factors, including over-expression of bone morphogenetic protein-4. There is currently little evidence to support the routine use of prophylaxis for heterotopic ossification in arthroplasty patients, but prophylaxis is recommended by some for high-risk patients. Radiotherapy given as one dose of 7 Gy to 8 Gy, either pre-operatively (< four hours before) or post-operatively (within 72 hours of surgery), appears to be more effective than indometacin therapy (75 mg daily for six weeks). In cases of prophylaxis against recurrent heterotopic ossification following excision, recent work has suggested that a combination of radiotherapy and indometacin is effective. Advances in our understanding of this condition may permit the development of newer, safer treatment modalities.


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 3 | Pages 337 - 341
1 Mar 2010
Yamasaki T Yasunaga Y Ishikawa M Hamaki T Ochi M

We have investigated the effectiveness of the transplantation of bone-marrow-derived mononuclear cells (BMMNCs) with interconnected porous calcium hydroxyapatite (IP-CHA) on early bone repair for osteonecrosis of the femoral head. We studied 22 patients (30 hips) who had osteonecrosis with a minimum follow-up of one year after implantation of BMMNCs. The mean age at surgery was 41 years (18 to 64) and the mean period of follow-up was 29 months (19 to 48). In a control group, cell-free IP-CHA was implanted into a further eight patients (9 hips) with osteonecrosis of the femoral head and the outcomes were compared.

A reduction in the size of the osteonecrotic lesion was observed subsequent to hypertrophy of the bone in the transition zone in the BMMNC group. In three patients in the treatment group progression to extensive collapse was detected. In the control group subtle bone hypertrophy was observed, but severe collapse of the femoral head occurred in six of eight hips.

In this limited study the implantation of BMMNCs and IP-CHA appears to confer benefit in the repair of osteonecrosis and in the prevention of collapse.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 9 | Pages 1125 - 1126
1 Sep 2009
Oussedik S Haddad F

Recent publication of reports showing high revision rates for hip and knee replacements carried out in Independent Sector Treatment Centres (ISTCs) has raised doubts regarding their ability to provide high quality healthcare. The high revision rates also create a financial burden to the NHS. The poor quality of data collected at ISTCs makes their performance difficult to evaluate. Funds may be better spent improving existing NHS facilities rather than establishing parallel ISTCs.


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 11 | Pages 1501 - 1508
1 Nov 2010
Donell ST Darrah C Nolan JF Wimhurst J Toms A Barker THW Case CP Tucker JK

Metal-on-metal total hip replacement has been targeted at younger patients with anticipated long-term survival, but the effect of the production of metal ions is a concern because of their possible toxicity to cells. We have reviewed the results of the use of the Ultima hybrid metal-on-metal total hip replacement, with a cemented polished tapered femoral component with a 28 mm diameter and a cobalt-chrome (CoCr) modular head, articulating with a 28 mm CoCr acetabular bearing surface secured in a titanium alloy uncemented shell.

Between 1997 and 2004, 545 patients with 652 affected hips underwent replacement using this system. Up to 31 January 2008, 90 (13.8%) hips in 82 patients had been revised. Pain was the sole reason for revision in 44 hips (48.9%) of which 35 had normal plain radiographs. Peri-prosthetic fractures occurred in 17 hips (18.9%) with early dislocation in three (3.3%) and late dislocation in 16 (17.8%). Infection was found in nine hips (10.0%).

At operation, a range of changes was noted including cavities containing cloudy fluid under pressure, necrotic soft tissues with avulsed tendons and denuded osteonecrotic upper femora. Corrosion was frequently observed on the retrieved cemented part of the femoral component. Typically, the peri-operative findings confirmed those found on pre-operative metal artefact reduction sequence MRI and histological examination showed severe necrosis.

Metal artefact reduction sequence MRI proved to be useful when investigating these patients with pain in the absence of adverse plain radiological features.


Bone & Joint 360
Vol. 1, Issue 2 | Pages 14 - 16
1 Apr 2012

The April 2012 Hip & Pelvis Roundup360 looks at osteoporotic hip fractures, retrotrochanteric pain, fibrin adhesive and reattachment of articular cartilage, autologous bone marrow mononuclear cells and avascular necrosis, bearing surfaces, stability after THR, digital templating, pelvic tilt after THR, custom-made sockets for DDH, and dogs and THR


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 12 | Pages 1561 - 1565
1 Dec 2009
Blakey CM Eswaramoorthy VK Hamilton LC Biant LC Field RE

We report the minimum five-year follow-up of 352 primary total hip replacements using the uncemented hydroxyapatite-coated ANCA-Fit femoral component with a modular neck and head. The series comprised 319 patients (212 men, 107 women) with a mean age at operation of 64.4 years (28 to 97). The principal diagnosis was osteoarthritis. A total of 18 patients (21 hips) died before their follow-up at five years, nine patients (11 hips) were lost to follow-up, and four (four hips) declined further follow-up. Patient-reported outcomes have been recorded for 288 patients (316 hips).

Their mean Oxford Hip Score improved significantly from 41 points (16 to 57) pre-operatively to 20 points (12 to 44) at five-year follow-up. Radiological assessment showed good bony stability in 98% of implants. There were two cases of aseptic loosening of the femoral component. There were no clinical or radiological complications related to modularity. In our series we did not see the high rate of intra-operative fracture previously reported for this implant.

This medium-term follow-up study demonstrates that the clinical outcome of the ANCA-Fit femoral component is, to date, comparable with that of other metaphyseal loading femoral components.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 6 | Pages 768 - 773
1 Jun 2012
Wang Q Zhang XL Chen YS Shen H Shao JJ

In this prospective study a total of 80 consecutive Chinese patients with Crowe type I or II developmental dysplasia of the hip were randomly assigned for hip resurfacing arthroplasty (HRA) or total hip replacement (THR).

Three patients assigned to HRA were converted to THR, and three HRA patients and two THR patients were lost to follow-up. This left a total of 34 patients (37 hips) who underwent HRA and 38 (39 hips) who underwent THR. The mean follow-up was 59.4 months (52 to 70) in the HRA group and 60.6 months (50 to 72) in the THR group. There was no failure of the prosthesis in either group. Flexion of the hip was significantly better after HRA, but there was no difference in the mean post-operative Harris hip scores between the groups. The mean size of the acetabular component in the HRA group was significantly larger than in the THR group (49.5 mm vs 46.1 mm, p = 0.001). There was no difference in the mean abduction angle of the acetabular component between the two groups.

Although the patients in this series had risk factors for failure after HRA, such as low body weight, small femoral heads and dysplasia, the clinical results of resurfacing in those with Crowe type I or II hip dysplasia were satisfactory. Patients in the HRA group had a better range of movement, although neck-cup impingement was observed. However, more acetabular bone was sacrificed in HRA patients, and it is unclear whether this will have an adverse effect in the long term.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 8 | Pages 1051 - 1057
1 Aug 2012
Nutton RW Wade FA Coutts FJ van der Linden ML

This prospective randomised controlled double-blind trial compared two types of PFC Sigma total knee replacement (TKR), differing in three design features aimed at improving flexion. The outcome of a standard fixed-bearing posterior cruciate ligament-preserving design (FB-S) was compared with that of a high-flexion rotating-platform posterior-stabilised design (RP-F) at one year after TKR.

The study group of 77 patients with osteoarthritis of the knee comprised 37 men and 40 women, with a mean age of 69 years (44.9 to 84.9). The patients were randomly allocated either to the FB-S or the RP-F group and assessed pre-operatively and at one year post-operatively. The mean post-operative non-weight-bearing flexion was 107° (95% confidence interval (CI) 104° to 110°)) for the FB-S group and 113° (95% CI 109° to 117°) for the RP-F group, and this difference was statistically significant (p = 0.032). However, weight-bearing range of movement during both level walking and ascending a slope as measured during flexible electrogoniometry was a mean of 4° lower in the RP-F group than in the FB-S group, with 58° (95% CI 56° to 60°) versus 54° (95% CI 51° to 57°) for level walking (p = 0.019) and 56° (95% CI 54° to 58°) versus 52° (95% CI 48° to 56°) for ascending a slope (p = 0.044). Further, the mean post-operative pain score of the Western Ontario and McMaster Universities Osteoarthritis Index was significantly higher in the RP-F group (2.5 (95% CI 1.5 to 3.5) versus 4.2 (95% CI 2.9 to 5.5), p = 0.043).

Although the RP-F group achieved higher non-weight-bearing knee flexion, patients in this group did not use this during activities of daily living and reported more pain one year after surgery.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 10 | Pages 1330 - 1338
1 Oct 2012
Namba RS Inacio MCS Paxton EW

We examined patient and surgical factors associated with deep surgical site infection (SSI) following total hip replacement (THR) in a large integrated healthcare system. A retrospective review of a cohort of primary THRs performed between 2001 and 2009 was conducted. Patient characteristics, surgical details, surgeon and hospital volumes, and SSIs were identified using the Kaiser Permanente Total Joint Replacement Registry (TJRR). Proportional-hazard regression models were used to assess risk factors for SSI. The study cohort consisted of 30 491 THRs, of which 17 474 (57%) were performed on women. The mean age of the patients in the whole series was 65.5 years (13 to 97; sd 11.8) and the mean body mass index was 29.3 kg/m2 (15 to 67; sd 5.9). The incidence of SSI was 0.51% (155 of 30 491). Patient factors associated with SSI included female gender, obesity, and American Society of Anesthesiologists (ASA) score 3. Age, diagnosis, diabetes and race were not associated with SSI. The only surgical factor associated with SSI was a bilateral procedure. Surgeon and hospital volumes, use of antibiotic-laden cement, fixation method, laminar flow, body exhaust suits, surgical approach and fellowship training were not associated with risk of SSI.

A comprehensive infection surveillance system, combined with a TJRR, identified patient and surgical factors associated with SSI. Obesity and chronic medical conditions should be addressed prior to THR. The finding of increased SSI risk with bilateral THR requires further investigation.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 7 | Pages 901 - 907
1 Jul 2012
Esposito CI Walter WL Roques A Tuke MA Zicat BA Walsh WR Walter WK

We analysed 54 alumina ceramic-on-ceramic bearings from total hip replacements retrieved at one centre after a mean duration of 3.5 years (0.2 to 10.6) in situ. These implants were obtained from 54 patients (16 men and 38 women) with a mean age of 67 years (33 to 88) who underwent revision for a variety of reasons. Posterior edge loading was found in the majority of these retrievals (32 out of 54). Anterosuperior edge loading occurred less often but produced a higher rate of wear. Stripe wear on the femoral heads had a median volumetric wear rate of 0.2 mm3/year (0 to 7.2). The wear volume on the femoral heads corresponded to the width of edge wear on the matching liner. Anteversion of the acetabular component was found to be a more important determinant than inclination for wear in ceramic bearings. Posterior edge loading may be considered to be a normal occurrence in ceramic-on-ceramic bearings, with minimal clinical consequences. Edge loading should be defined as either anterosuperior or posterior, as each edge loading mechanism may result in different clinical implications.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 9 | Pages 1225 - 1228
1 Sep 2007
van de Pol GJ Sturm PDJ van Loon CJ Verhagen C Schreurs BW

Allografts of bone from the femoral head are often used in orthopaedic procedures. Although the donated heads are thoroughly tested microscopically before release by the bone bank, some surgeons take additional cultures in the operating theatre before implantation. There is no consensus about the need to take these cultures. We retrospectively assessed the clinical significance of the implantation of positive-cultured bone allografts. The contamination rate at retrieval of the allografts was 6.4% in our bone bank. Intra-operative cultures were taken from 426 femoral head allografts before implantation; 48 (11.3%) had a positive culture. The most frequently encountered micro-organism was coagulase-negative staphylococcus. Deep infection occurred in two of the 48 patients (4.2%). In only one was it likely that the same micro-organism caused the contamination and the subsequent infection.

In our study, the rate of infection in patients receiving positive-cultured allografts at implantation was not higher than the overall rate of infection in allograft surgery suggesting that the positive cultures at implantation probably represent contamination and that the taking of additional cultures is not useful.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 12 | Pages 1545 - 1549
1 Dec 2009
Migliore A Perrini MR Romanini E Fella D Cavallo A Cerbo M Jefferson T

This study evaluated the feasibility of using published data from more than one register to define the performance of different hip implants. In order to obtain estimates of performance for specific types of hip system from different register, we analysed data from the annual reports of five national and one Italian regional register. We extracted the number of implants and rates of implant survival at different periods of follow-up. Our aim was to assess whether estimates of cumulative survival rate were comparable with data from registers from different countries, and our conclusion was that such a comparison could only be performed incompletely.


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 6 | Pages 838 - 838
1 Jun 2006
SPENCER RF


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 2 | Pages 202 - 208
1 Feb 2010
Berton C Girard J Krantz N Migaud H

Implantation of a large-diameter femoral head prosthesis with a metal-on-metal bearing surface reduces the risk of dislocation, increases the range of movement, minimises the risk of impingement and, in theory, results in little wear.

Between February 2004 and March 2007 we implanted 100 consecutive total hip replacements with a metal-on-metal bearing and a large femoral head into 92 patients. There were 51 men and 41 women with a mean age of 50 years (18 to 70) at the time of surgery.

Outcome was assessed using the Western Ontario McMaster University osteoarthritis index and the Harris hip score as well as the Devane activity score. These all improved significantly (p < 0.0001). At the last follow-up there were no cases of dislocation, no impingement, a good range of movement and no osteolysis, but seven revisions, two for infection and five for aseptic loosening. The probability of groin pain increased if the other acetabular component inclination exceeded 50° (p = 0.0007). At 4.8 years of follow-up, the projected survival of the Durom acetabular component, with revision for any reason, was 92.4% (sd 2.8) (95% confidence interval 89.6 to 95.2).

The design of the component made it difficult both to orientate and seat, which when combined with a poor porous coating, produced unpredictable fixation and a low survival at five years.


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 11 | Pages 1545 - 1546
1 Nov 2006
DANIEL J ZIAEE H PRADHAN C McMINN DJW


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 7 | Pages 971 - 976
1 Jul 2007
Kampa RJ Prasthofer A Lawrence-Watt DJ Pattison RM

In order to determine the potential for an internervous safe zone, 20 hips from human cadavers were dissected to map out the precise pattern of innervation of the hip capsule. The results were illustrated in the form of a clock face. The reference point for measurement was the inferior acetabular notch, representing six o’clock. Capsular branches from between five and seven nerves contributed to each hip joint, and were found to innervate the capsule in a relatively constant pattern. An internervous safe zone was identified anterosuperiorly in an arc of 45° between the positions of one o’clock and half past two.

Our study shows that there is an internervous zone that could be safely used in a capsule-retaining anterior, anterolateral or lateral approach to the hip, or during portal placement in hip arthroscopy.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 4 | Pages 483 - 487
1 Apr 2012
Prosser GH Shears E O’Hara JN

The painful subluxed or dislocated hip in adults with cerebral palsy presents a challenging problem. Prosthetic dislocation and heterotopic ossification are particular concerns. We present the first reported series of 19 such patients (20 hips) treated with hip resurfacing and proximal femoral osteotomy. The pre-operative Gross Motor Function Classification System (GMFCS) was level V in 13 (68%) patients, level IV in three (16%), level III in one (5%) and level II in two (11%). The mean age at operation was 37 years (13 to 57).

The mean follow-up was 8.0 years (2.7 to 11.6), and 16 of the 18 (89%) contactable patients or their carers felt that the surgery had been worthwhile. Pain was relieved in 16 of the 18 surviving hips (89%) at the last follow-up, and the GMFCS level had improved in seven (37%) patients. There were two (10%) early dislocations; three hips (15%) required revision of femoral fixation, and two hips (10%) required revision, for late traumatic fracture of the femoral neck and extra-articular impingement, respectively. Hence there were significant surgical complications in a total of seven hips (35%). No hips required revision for instability, and there were no cases of heterotopic ossification.

We recommend hip resurfacing with proximal femoral osteotomy for the treatment of the painful subluxed or dislocated hip in patients with cerebral palsy.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 9 | Pages 1285 - 1286
1 Sep 2011
Baker JF Vioreanu MH Khan HA

A variety of materials were used for early hip prostheses. The introduction of Vitallium by Smith-Petersen represented a further advance in this surgical field. We present the longest known follow-up of a Smith-Petersen Vitallium mould arthroplasty.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 2 | Pages 249 - 253
1 Feb 2012
Sorlí L Puig L Torres-Claramunt R González A Alier A Knobel H Salvadó M Horcajada JP

Patients with infected arthroplasties are normally treated with a two-stage exchange procedure using polymethylmethacrylate bone cement spacers impregnated with antibiotics. However, spacers may act as a foreign body to which micro-organisms may adhere and grow. In this study it was hypothesised that subclinical infection may be diagnosed with sonication of the surface biofilm of the spacer. The aims were to assess the presence of subclinical infection through sonication of the spacer at the time of a second-stage procedure, and to determine the relationship between subclinical infection and the clinical outcome. Of 55 patients studied, 11 (20%) were diagnosed with subclinical infection. At a mean follow-up of 12 months (interquartile range 6 to 18), clinical failure was found in 18 (32.7%) patients. Of the patients previously diagnosed with subclinical infection, 63% (7 of 11) had failed compared with 25% (11 of 44) of those without subclinical infection (odds ratio 5.25, 95% confidence interval 1.29 to 21.4, p = 0.021). Sonication of the biofilm of the surface of the spacer is useful in order to exclude subclinical infection and therefore contributes to improving the outcome after two-stage procedures.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 1 | Pages 37 - 42
1 Jan 2012
Affatato S Traina F De Fine M Carmignato S Toni A

Alumina–alumina bearings are among the most resistant to wear in total hip replacement. Examination of their surfaces is one way of comparing damage caused by wear of hip joints simulated in vitro to that seen in explanted bearings. The aim of this study was to determine whether second-generation ceramic bearings exhibited a better pattern of wear than those reported in the literature for first-generation bearings. We considered both macro- and microscopic findings.

We found that long-term alumina wear in association with a loose acetabular component could be categorised into three groups. Of 20 specimens, four had ‘low wear’, eight ‘crescent wear’ and eight ‘severe wear’, which was characterised by a change in the physical shape of the bearing and a loss of volume. This suggests that the wear in alumina–alumina bearings in association with a loose acetabular component may be variable in pattern, and may explain, in part, why the wear of a ceramic head in vivo may be greater than that seen after in vitro testing.