The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 8 | Pages 997 - 1002
1 Aug 2006
Vendittoli P Lavigne M Girard J Roy AG

We have undertaken a prospective, randomised study to compare conservation of acetabular bone after total hip replacement and resurfacing arthroplasty of the hip. We randomly assigned 210 hips to one of the two treatment groups. Uncemented, press-fit acetabular components were used for both.

No significant difference was found in the mean diameter of acetabular implant inserted in the groups (54.74 mm for total hip replacement and 54.90 mm for resurfacing arthroplasty). In seven resurfacing procedures (6.8%), the surgeon used a larger size of component in order to match the corresponding diameter of the femoral component.

With resurfacing arthroplasty, conservation of bone is clearly advantageous on the femoral side. Our study has shown that, with a specific design of acetabular implant and by following a careful surgical technique, removal of bone on the acetabular side is comparable with that of total hip replacement.


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 8 | Pages 1079 - 1084
1 Aug 2010
Muirhead-Allwood S Sandiford N Skinner JA Hua J Kabir C Walker PS

We present the 10- to 17-year results of 112 computer-assisted design computer-assisted manufacture femoral components. The total hip replacements were performed between 1992 and 1998 in 111 patients, comprising 53 men and 58 women. Their mean age was 46.2 years (24.6 to 62.2) with a mean follow-up of 13 years (10 to 17). The mean Harris Hip Score improved from 42.4 (7 to 99) to 90.3 (38 to 100), the mean Oxford Hip Score from 43.1 (12 to 59) to 18.2 (12 to 51) and the mean Western Ontario MacMasters University Osteoarthritis Index score from 57.0 (7 to 96) to 11.9 (0 to 85). There was one revision due to failure of the acetabular component but no failures of the femoral component. There were no revisions for aseptic loosening. The worst-case survival in this cohort of custom femoral components at 13.2 years follow-up was 98.2% (95% confidence interval 95 to 99). Overall survival of this series of total hip replacements was 97.3% (95% confidence interval 95 to 99).

These results are comparable with the best medium- to long-term results for femoral components used in primary total hip replacement with any means of fixation.


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 11 | Pages 1502 - 1506
1 Nov 2005
Arora J Ogden AC

We analysed at a mean follow-up of 7.25 years the clinical and radiological outcome of 117 patients (125 knees) who had undergone a primary, cemented, modular Freeman-Samuelson total knee replacement. While the tibial and femoral components were cemented, the patellar component was uncemented. A surface-cementing technique was used to secure the tibial components. A total of 82 knees was available for radiological assessment. Radiolucent lines were seen in 41 knees (50%) and osteolytic lesions were seen in 13 knees (16%). Asymptomatic, rotational loosening of the patellar implant was seen in four patients and osteolysis was more common in patients with a patellar resurfacing. Functional outcome scores were available for 41 patients (41 knees, 35%) and the mean Western Ontario McMasters Universities score was 77.5 (sd 19.5) and the cumulative survival was 93.4% at ten years with revision for aseptic loosening as an endpoint. Increased polyethylene wear from modular components, a rotationally-loose patella, and the surface-cementing technique may have contributed to the high rate of osteolysis seen in our study.


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 6 | Pages 894 - 899
1 Jun 2010
Khattak MJ Ahmad T Rehman R Umer M Hasan SH Ahmed M

The nervous system is known to be involved in inflammation and repair. We aimed to determine the effect of physical activity on the healing of a muscle injury and to examine the pattern of innervation. Using a drop-ball technique, a contusion was produced in the gastrocnemius in 20 rats. In ten the limb was immobilised in a plaster cast and the remaining ten had mobilisation on a running wheel. The muscle and the corresponding dorsal-root ganglia were studied by histological and immunohistochemical methods.

In the mobilisation group, there was a significant reduction in lymphocytes (p = 0.016), macrophages (p = 0.008) and myotubules (p = 0.008) between three and 21 days. The formation of myotubules and the density of nerve fibres was significantly higher (both p = 0.016) compared with those in the immobilisation group at three days, while the density of CGRP-positive fibres was significantly lower (p = 0.016) after 21 days.

Mobilisation after contusional injury to the muscle resulted in early and increased formation of myotubules, early nerve regeneration and progressive reduction in inflammation, suggesting that it promoted a better healing response.


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 7 | Pages 922 - 928
1 Jul 2010
Bose VC Baruah BD

We performed 96 Birmingham resurfacing arthroplasties of the hip in 71 consecutive patients with avascular necrosis of the femoral head. A modified neck-capsule-preserving approach was used which is described in detail. The University of California, Los Angeles outcome score, the radiological parameters and survival rates were assessed. The mean follow-up was for 5.4 years (4.0 to 8.1). All the patients remained active with a mean University of California, Los Angeles activity score of 6.86 (6 to 9). Three hips failed, giving a cumulative survival rate of 95.4%. With failure of the femoral component as the endpoint, the cumulative survival rate was 98.0%. We also describe the combined abduction-valgus angle of the bearing couple, which is the sum of the inclination angle of the acetabular component and the stem-shaft angle, as an index of the optimum positioning of the components in the coronal plane.

Using a modified surgical technique, it is possible to preserve the femoral head in avascular necrosis by performing hip resurfacing in patients with good results.


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 4 | Pages 508 - 512
1 Apr 2010
Pearse AJ Hooper GJ Rothwell A Frampton C

We reviewed the rate of revision of unicompartmental knee replacements (UKR) from the New Zealand Joint Registry between 1999 and 2008. There were 4284 UKRs, of which 236 required revision, 205 to a total knee replacement (U2T) and 31 to a further unicompartmental knee replacement (U2U). We used these data to establish whether the survival and functional outcome for revised UKRs were comparable with those of primary total knee replacement (TKR). The rate of revision for the U2T cohort was four times higher than that for a primary TKR (1.97 vs 0.48; p < 0.05). The mean Oxford Knee Score was also significantly worse in the U2T group than that of the primary TKR group (30.02 vs 37.16; p < 0.01). The rate of revision for conversion of a failed UKR to a further UKR (U2U cohort) was 13 times higher than that for a primary TKR.

The poor outcome of a UKR converted to a primary TKR compared with a primary TKR should contra-indicate the use of a UKR as a more conservative procedure in the younger patient.


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 9 | Pages 1238 - 1244
1 Sep 2010
Nicoll D Rowley DI

This study used CT analysis to determine the rotational alignment of 39 painful and 26 painless fixed-bearing total knee replacements (TKRs) from a cohort of 740 NexGen Legacy posterior-stabilised and cruciate-retaining prostheses implanted between May 1996 and August 2003.

The mean rotation of the tibial component was 4.3° of internal rotation (25.4° internal to 13.9° external rotation) in the painful group and 2.2° of external rotation (8.5° internal to 18.2° external rotation) in the painfree group (p = 0.024). In the painful group 17 tibial components were internally rotated more than 9° compared with none in the painfree group (p < 0.001). Additionally, six femoral components in the painful group were internally rotated more than 6° compared with none in the painfree group (p = 0.017). External rotational errors were not found to be associated with pain.

Overall, 22 (56.4%) of the painful TKRs had internal rotational errors involving the femoral, the tibial or both components. It is estimated that at least 4.6% of all our TKRs have been implanted with significant internal rotational errors.


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 9 | Pages 1245 - 1252
1 Sep 2010
Song EK Seon JK Park SJ Jeong MS

We compared the incidence and severity of complications during and after closing- and opening-wedge high tibial osteotomy used for the treatment of varus arthritis of the knee, and identified the risk factors associated with the development of complications. In total, 104 patients underwent laterally based closing-wedge and 90 medial opening-wedge high tibial osteotomy between January 1993 and December 2006. The characteristics of each group were similar. All the patients were followed up for more than 12 months. We assessed the outcome using the Hospital for Special Surgery knee score, and recorded the complications. Age, gender, obesity (body mass index > 27.5 kg/m2), the type of osteotomy (closing versus opening) and the pre-operative mechanical axis were subjected to risk-factor analysis.

The mean Hospital for Special Surgery score in the closing and opening groups improved from 73.4 (54 to 86) to 91.8 (81 to 100) and from 73.8 (56 to 88) to 93 (84 to 100), respectively. The incidence of complications overall and of major complications in both groups was not significantly different (p = 0.20 overall complication, p = 0.29 major complication). Logistic regression analysis adjusting for obesity and the pre-operative mechanical axis showed that obesity remained a significant independent risk factor (odds ratio = 3.23) of a major complication after high tibial osteotomy.

Our results suggest that the opening-wedge high tibial osteotomy can be an alternative treatment option for young patients with medial compartment osteoarthritis and varus deformity.


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 11 | Pages 1527 - 1534
1 Nov 2010
Grammatopoulos G Pandit H Murray DW Gill HS

Pseudotumour is a rare but important complication of metal-on-metal hip resurfacing that occurs much more commonly in women than in men. We examined the relationship between head-neck ratio (HNR) and pseudotumour formation in 18 resurfaced hips (18 patients) revised for pseudotumour and 42 asymptomatic control resurfaced hips (42 patients).

Patients in whom pseudotumour formation had occurred had higher pre-operative HNR than the control patients (mean 1.37 (sd 0.10) vs mean 1.30 (sd 0.08) p = 0.001). At operation the patients with pseudotumours had a greater reduction in the size of their femoral heads (p = 0.035) and subsequently had greater neck narrowing (mean 10.1% (sd 7.2) vs mean 3.8% (sd 3.2) p < 0.001). No female patient with a pre-operative HNR ≤ 1.3 developed a pseudotumour.

We suggest that reducing the size of the femoral head, made possible by a high pre-operative HNR, increases the risk of impingement and edge loading, and may contribute to high wear and pseudotumour formation. As the incidence of pseudotumour is low in men, it appears safe to perform resurfacing in men. However, this study suggests that it is also reasonable to resurface in women with a pre-operative HNR ≤ 1.3.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 12 | Pages 1579 - 1582
1 Dec 2009
Starks I Roberts S White SH

We present a prospective review of the two-year functional outcome of 37 Avon patellofemoral joint replacements carried out in 29 patients with a mean age of 66 years (30 to 82) between October 2002 and March 2007. No patients were lost to follow-up. This is the first independent assessment of this prosthesis using both subjective and objective analysis of outcome. At two years the median Oxford knee score was 39 (interquartile range 32 to 44), the median American Knee Society objective score was 95 (interquartile range 90 to 100), the median American Knee Society functional score was 85 (interquartile range 60 to 100), and the median Melbourne Knee score was 28 (interquartile range 21 to 30). Two patients underwent further surgery. Only one patient reported an unsatisfactory outcome.

We conclude that the promising early results observed by the designing centre are reproducible and provide further support for the role of patellofemoral joint replacement.


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 3 | Pages 291 - 295
1 Mar 2005
Farmer SE Woollam PJ Patrick JH Roberts AP Bromwich W


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 5 | Pages 592 - 600
1 May 2006
Pollard TCB Baker RP Eastaugh-Waring SJ Bannister GC

We compared the five- to seven-year clinical and radiological results of the metal-on-metal Birmingham hip resurfacing with a hybrid total hip arthroplasty in two groups of 54 hips, matched for gender, age, body mass index and activity level.

Function was excellent in both groups, as measured by the Oxford hip score, but the Birmingham hip resurfacings had higher University of California at Los Angeles activity scores and better EuroQol quality of life scores. The total hip arthroplasties had a revision or intention-to-revise rate of 8%, and the Birmingham hip resurfacings of 6%. Both groups demonstrated impending failure on surrogate end-points. Of the total hip arthroplasties, 12% had polyethylene wear and osteolysis under observation, and 8% of Birmingham hip resurfacings showed migration of the femoral component. Polyethylene wear was present in 48% of the hybrid hips without osteolysis. Of the femoral components in the Birmingham hip resurfacing group which had not migrated, 66% had radiological changes of unknown significance.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 6 | Pages 812 - 814
1 Jun 2009
Ramaswamy R Kosashvili Y Cameron H

Osteogenesis imperfecta is a rare inherited disorder of connective tissue which may present with recurrent fractures which are prone to nonunion and malunion resulting in deformity. Some patients develop osteoarthritis of the hip. Formation of hyperplastic callus after recurrent fractures may deform the shape of the femur and preclude the use of standard implants at joint replacement. Replacement can thus be technically demanding.

We present a case of bilateral hip replacement in a patient with osteogenesis imperfecta and hyperplastic callus which was treated by the use of long femoral allografts and cemented femoral stems.


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 1 | Pages 88 - 94
1 Jan 2005
Hasler CC Von Laer L Hell AK

We reviewed 15 patients, nine girls and six boys, with chronic anterior dislocation of the radial head which was treated by ulnar osteotomy, external fixation and open reconstruction of the elbow joint but without repair of the annular ligament. Their mean age was 9.5 years (5 to 15) and the mean interval between the injury and reconstruction was 22 months (2 months to 7 years).

All radial heads remained reduced at a mean follow-up of 20 months (6 months to 5 years). Normal ranges of movement for flexion, extension, pronation and supination were unchanged in 96.1% (49/51) and worse in 3.9% (2/51). Limited ranges of movement were improved in 77.8% (7/9), unchanged in 11% (1/9) and further decreased in 11% (1/9).There were two superficial pin-track infections and two cases of delayed union but with no serious complications. Reconstruction of the radiocapitellar joint is easier using external fixation since accurate correction of the ulna can be determined empirically and active functional exercises started immediately. Only patients with a radial head of normal shape were selected for treatment by this method.


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 4 | Pages 602 - 608
1 Apr 2010
Drobnič M Radosavljevič D Cör A Brittberg M Stražar K

We compared the quality of debridement of chondral lesions performed by four arthroscopic (SH, shaver; CU, curette; SHCU, shaver and curette; BP, bipolar electrodes) and one open technique (OPEN, scalpel and curette) which are used prior to autologous chondrocyte implantation (ACI). The ex vivo simulation of all five techniques was carried out on six juvenile equine stifle joints. The OPEN, SH and SHCU techniques were tested on knees harvested from six adult human cadavers.

The most vertical walls with the least adjacent damage to cartilage were obtained with the OPEN technique. The CU and SHCU methods gave inferior, but still acceptable results whereas the SH technique alone resulted in a crater-like defect and the BP method undermined the cartilage wall. The subchondral bone was severely violated in all the equine samples which might have been peculiar to this model. The predominant depth of the debridement in the adult human samples was at the level of the calcified cartilage. Some minor penetrations of the subchondral end-plate were induced regardless of the instrumentation used.

Our study suggests that not all routine arthroscopic instruments are suitable for the preparation of a defect for ACI. We have shown that the preferred debridement technique is either open or arthroscopically-assisted manual curettage. The use of juvenile equine stifles was not appropriate for the study of the cartilage-subchondral bone interface.


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 9 | Pages 1171 - 1177
1 Sep 2005
Trieb K


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 12 | Pages 1623 - 1627
1 Dec 2009
Bubbar V Heras FL Amato D Pritzker KPH Gross AE

Total hip replacement in patients with Gaucher’s disease with symptomatic osteonecrosis of the femoral head is controversial because of the high early failure rates. We describe four patients who had an uncemented total hip replacement following enzyme replacement therapy for a median of two years and one month (1 to 9.8 years) prior to surgery, and who remained on treatment. At operation, the bone had a normal appearance and consistency. Histopathological examination showed that, compared with previous biopsies of untreated Gaucher’s disease, the Gaucher cell infiltrate had decreased progressively with therapy, being replaced by normal adipose tissue. The surfaces of viable bone beyond the osteonecrotic areas showed osteoblasts, indicating remodelling. In one case acetabular revision was carried out after 11 years and eight months. The three remaining patients had a mean follow-up of six years and four months (3.3 to 12 years). We recommend initiating enzyme replacement therapy at least one to two years prior to total hip replacement to facilitate bone remodelling and to allow implantation of uncemented components in these young patients.


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 1 | Pages 19 - 25
1 Jan 2006
Scheerlinck T de Mey J Deklerck R Noble PC

Using a modern cementing technique, we implanted 22 stereolithographic polymeric replicas of the Charnley-Kerboul stem in 11 pairs of human cadaver femora. On one side, the replicas were cemented line-to-line with the largest broach. On the other, one-size undersized replicas were used (radial difference, 0.89 mm sd 0.13).

CT analysis showed that the line-to-line stems without distal centralisers were at least as well aligned and centered as undersized stems with a centraliser, but were surrounded by less cement and presented more areas of thin (< 2 mm) or deficient (< 1 mm) cement. These areas were located predominantly at the corners and in the middle and distal thirds of the stem. Nevertheless, in line-to-line stems, penetration of cement into cancellous bone resulted in a mean thickness of cement of 3.1 mm (sd 0.6) and only 6.2% of deficient and 26.4% of thin cement. In over 90% of these areas, the cement was directly supported by cortical bone or cortical bone with less than 1 mm of cancellous bone interposed.

When Charnley-Kerboul stems are cemented line-to-line, good clinical results are observed because cement-deficient areas are limited and are frequently supported by cortical bone.


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 9 | Pages 1292 - 1297
1 Sep 2005
Lietman SA Inoue N Rafiee B Deitz LW Chao EYS

We used a canine intercalary bone defect model to determine the effects of recombinant human osteogenic protein 1 (rhOP-1) on allograft incorporation. The allograft was treated with an implant made up of rhOP-1 and type I collagen or with type I collagen alone.

Radiographic analysis showed an increased volume of periosteal callus in both test groups compared with the control group at weeks 4, 6, 8 and 10. Mechanical testing after 12 weeks revealed increased maximal torque and stiffness in the rhOP-1 treated groups compared with the control group.

These results indicate a benefit from the use of an rhOP-1 implant in the healing of bone allografts. The effect was independent of the position of the implant. There may be a beneficial clinical application for this treatment.


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 4 | Pages 527 - 534
1 Apr 2010
Streubel PN Gardner MJ Morshed S Collinge CA Gallagher B Ricci WM

It is unclear whether there is a limit to the amount of distal bone required to support fixation of supracondylar periprosthetic femoral fractures. This retrospective multicentre study evaluated lateral locked plating of periprosthetic supracondylar femoral fractures and compared the results according to extension of the fracture distal with the proximal border of the femoral prosthetic component.

Between 1999 and 2008, 89 patients underwent lateral locked plating of a supracondylar periprosthetic femoral fracture, of whom 61 patients with a mean age of 72 years (42 to 96) comprising 53 women, were available after a minimum follow-up of six months or until fracture healing. Patients were grouped into those with fractures located proximally (28) and those with fractures that extended distal to the proximal border of the femoral component (33).

Delayed healing and nonunion occurred respectively in five (18%) and three (11%) of more proximal fractures, and in two (6%) and five (15%) of the fractures with distal extension (p = 0.23 for delayed healing; p = 0.72 for nonunion, Fisher’s exact test). Four construct failures (14%) occurred in more proximal fractures, and three (9%) in fractures with distal extension (p = 0.51). Of the two deep infections that occurred in each group, one resolved after surgical debridement and antibiotics, and one progressed to a nonunion.

Extreme distal periprosthetic supracondylar fractures of the femur are not a contra-indication to lateral locked plating. These fractures can be managed with internal fixation, with predictable results, similar to those seen in more proximal fractures.