The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 10 | Pages 1308 - 1313
1 Oct 2011
Hart AJ Sabah SA Bandi AS Maggiore P Tarassoli P Sampson B A. Skinner J

Blood metal ions have been widely used to investigate metal-on-metal hip replacements, but their ability to discriminate between well-functioning and failed hips is not known. The Medicines and Healthcare products Regulatory Agency (MHRA) has suggested a cut-off level of 7 parts per billion (ppb).

We performed a pair-matched, case-control study to investigate the sensitivity and specificity of blood metal ion levels for diagnosing failure in 176 patients with a unilateral metal-on-metal hip replacement. We recruited 88 cases with a pre-revision, unexplained failed hip and an equal number of matching controls with a well-functioning hip. We investigated the 7 ppb cut-off level for the maximum of cobalt or chromium and determined optimal mathematical cut-off levels from receiver-operating characteristic curves.

The 7 ppb cut-off level for the maximum of cobalt or chromium had a specificity of 89% and sensitivity 52% for detecting a pre-operative unexplained failed metal on metal hip replacement. The optimal cut-off level for the maximum of cobalt or chromium was 4.97 ppb and had sensitivity 63% and specificity 86%.

Blood metal ions had good discriminant ability to separate failed from well-functioning hip replacements. The MHRA cut-off level of 7 ppb provides a specific test but has poor sensitivity.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 10 | Pages 1289 - 1295
1 Oct 2011
Yoon RS Hwang JS Beebe KS

For over a decade, bisphosphonate administration has evolved and become the cornerstone of the prevention and treatment of fragility fractures. Millions of post-menopausal women have relied on, and continue to depend on, the long-acting, bone density-maintaining pharmaceutical drug to prevent low-energy fractures. In return, we have seen the number of fragility fractures decrease, along with associated costs and emotional benefits. However, with any drug, there are often concerns with side effects and complications, and this unique drug class is seeing one such complication in atypical subtrochanteric femoral fracture, counterproductive to that which it was designed to prevent. This has created concern over long-term bisphosphonate administration and its potential link to these atypical fractures. There is controversial evidence surrounding such a definitive link, and no protocol for managing these fractures.

This review offers the latest information regarding this rare but increasingly controversial adverse effect and its potential connection to one of the most successful forms of treatment that is available for the management of fragility fractures.


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 10 | Pages 1309 - 1315
1 Oct 2006
Shetty N Hamer AJ Stockley I Eastell R Willkinson JM

Bisphosphonates reduce peri-prosthetic bone loss in the short term after total hip replacement but the mid- and longer term effects are not known. The aims of this randomised trial were to examine the effect of a single dose of 90 mg of pamidronate on the clinical and radiological outcome and peri-prosthetic bone mineral density in 50 patients (56 hips) over a five-year period, following total hip replacement.

At five years, 37 patients (42 hips) returned for assessment. The Harris hip scores were similar in the pamidronate and placebo groups throughout the study. Also at five years, four patients, two from each group had osteolytic lesions on plain radiography. These were located around the acetabular component in three patients and in the femoral calcar in one. The femoral and acetabular peri-prosthetic bone mineral density in the pamidronate group and the control group was similar at five years.

Pamidronate given as a single post-operative dose does not appear to influence the clinical outcome or prevent the development of osteolytic lesions at five years after total hip replacement.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 12 | Pages 1577 - 1585
1 Dec 2011
Alshryda S Sarda P Sukeik M Nargol A Blenkinsopp J Mason JM

We conducted a systematic review and meta-analysis of randomised controlled trials evaluating the effect of tranexamic acid (TXA) upon blood loss and transfusion in primary total knee replacement. The review used the generic evaluation tool designed by the Cochrane Bone, Joint and Muscle Trauma Group. A total of 19 trials were eligible: 18 used intravenous administration, one also evaluated oral dosing and one trial evaluated topical use. TXA led to a significant reduction in the proportion of patients requiring blood transfusion (risk ratio (RR) 2.56, 95% confidence interval (CI) 2.1 to 3.1, p < 0.001; heterogeneity I2 = 75%; 14 trials, 824 patients). Using TXA also reduced total blood loss by a mean of 591 ml (95% CI 536 to 647, p < 0.001; I2 = 78%; nine trials, 763 patients). The clinical interpretation of these findings is limited by substantial heterogeneity. However, subgroup analysis of high-dose (> 4 g) TXA showed a plausible consistent reduction in blood transfusion requirements (RR 5.33; 95% CI 2.44 to 11.65, p < 0.001; I2 = 0%), a finding that should be confirmed by a further well-designed trial. The current evidence from trials does not support an increased risk of deep-vein thrombosis (13 trials, 801 patients) or pulmonary embolism (18 trials, 971 patients) due to TXA administration.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 4 | Pages 487 - 493
1 Apr 2009
Dayer R Assal M

We studied a cohort of 26 diabetic patients with chronic ulceration under the first metatarsal head treated by a modified Jones extensor hallucis longus and a flexor hallucis longus transfer. If the first metatarsal was still plantar flexed following these two transfers, a peroneus longus to the peroneus brevis tendon transfer was also performed. Finally, if ankle dorsiflexion was < 5° with the knee extended, a Strayer-type gastrocnemius recession was performed.

The mean duration of chronic ulceration despite a minimum of six months’ conservative care was 16.2 months (6 to 31). A total of 23 of the 26 patients were available for follow-up at a mean of 39.6 months (12 to 61) after surgery. All except one achieved complete ulcer healing at a mean of 4.4 weeks (2 to 8) after surgery, and there was no recurrence of ulceration under the first metatarsal.

We believe that tendon balancing using modified Jones extensor hallucis longus and flexor hallucis longus transfers, associated in selected cases with a peroneus longus to brevis transfer and/or Strayer procedure, can promote rapid and sustained healing of chronic diabetic ulcers under the first metatarsal head.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 11 | Pages 1537 - 1544
1 Nov 2011
McGovern PD Albrecht M G. Belani K Nachtsheim C Partington PF Carluke I Reed MR

We investigated the capacity of patient warming devices to disrupt the ultra-clean airflow system. We compared the effects of two patient warming technologies, forced-air and conductive fabric, on operating theatre ventilation during simulated hip replacement and lumbar spinal procedures using a mannequin as a patient. Infection data were reviewed to determine whether joint infection rates were associated with the type of patient warming device that was used.

Neutral-buoyancy detergent bubbles were released adjacent to the mannequin’s head and at floor level to assess the movement of non-sterile air into the clean airflow over the surgical site. During simulated hip replacement, bubble counts over the surgical site were greater for forced-air than for conductive fabric warming when the anaesthesia/surgery drape was laid down (p = 0.010) and at half-height (p < 0.001). For lumbar surgery, forced-air warming generated convection currents that mobilised floor air into the surgical site area. Conductive fabric warming had no such effect.

A significant increase in deep joint infection, as demonstrated by an elevated infection odds ratio (3.8, p = 0.024), was identified during a period when forced-air warming was used compared to a period when conductive fabric warming was used. Air-free warming is, therefore, recommended over forced-air warming for orthopaedic procedures.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 10 | Pages 1355 - 1361
1 Oct 2011
Bollars P Luyckx J Innocenti B Labey L Victor J Bellemans J

High-flexion total knee replacement (TKR) designs have been introduced to improve flexion after TKR. Although the early results of such designs were promising, recent literature has raised concerns about the incidence of early loosening of the femoral component. We compared the minimum force required to cause femoral component loosening for six high-flexion and six conventional TKR designs in a laboratory experiment.

Each TKR design was implanted in a femoral bone model and placed in a loading frame in 135° of flexion. Loosening of the femoral component was induced by moving the tibial component at a constant rate of displacement while maintaining the same angle of flexion. A stereophotogrammetric system registered the relative movement between the femoral component and the underlying bone until loosening occurred.

Compared with high-flexion designs, conventional TKR designs required a significantly higher force before loosening occurred (p < 0.001). High-flexion designs with closed box geometry required significantly higher loosening forces than high-flexion designs with open box geometry (p = 0.0478). The presence of pegs further contributed to the fixation strength of components.

We conclude that high-flexion designs have a greater risk for femoral component loosening than conventional TKR designs. We believe this is attributable to the absence of femoral load sharing between the prosthetic component and the condylar bone during flexion.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 9 | Pages 1217 - 1222
1 Sep 2011
Bonner TJ Eardley WGP Patterson P Gregg PJ

Correct positioning and alignment of components during primary total knee replacement (TKR) is widely accepted to be an important predictor of patient satisfaction and implant durability. This retrospective study reports the effect of the post-operative mechanical axis of the lower limb in the coronal plane on implant survival following primary TKR.

A total of 501 TKRs in 396 patients were divided into an aligned group with a neutral mechanical axis (± 3°) and a malaligned group where the mechanical axis deviated from neutral by > 3°. At 15 years’ follow-up, 33 of 458 (7.2%) TKRs were revised for aseptic loosening. Kaplan-Meier survival analysis showed a weak tendency towards improved survival with restoration of a neutral mechanical axis, but this did not reach statistical significance (p = 0.47).

We found that the relationship between survival of a primary TKR and mechanical axis alignment is weaker than that described in a number of previous reports.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 9 | Pages 1259 - 1264
1 Sep 2011
Wähnert D Windolf M Brianza S Rothstock S Radtke R Brighenti V Schwieger K

We investigated the static and cyclical strength of parallel and angulated locking plate screws using rigid polyurethane foam (0.32 g/cm3) and bovine cancellous bone blocks. Custom-made stainless steel plates with two conically threaded screw holes with different angulations (parallel, 10° and 20° divergent) and 5 mm self-tapping locking screws underwent pull-out and cyclical pull and bending tests. The bovine cancellous blocks were only subjected to static pull-out testing. We also performed finite element analysis for the static pull-out test of the parallel and 20° configurations. In both the foam model and the bovine cancellous bone we found the significantly highest pull-out force for the parallel constructs. In the finite element analysis there was a 47% more damage in the 20° divergent constructs than in the parallel configuration. Under cyclical loading, the mean number of cycles to failure was significantly higher for the parallel group, followed by the 10° and 20° divergent configurations.

In our laboratory setting we clearly showed the biomechanical disadvantage of a diverging locking screw angle under static and cyclical loading.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 8 | Pages 1001 - 1009
1 Aug 2007
Ahrens PM Boileau P

This paper describes the current views on the pathology of lesions of the tendon of the long head of biceps and their management. Their diagnosis is described and their surgical management classified, with details of the techniques employed.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 6 | Pages 722 - 725
1 Jun 2011
Buckley R Leighton R Trask K

The Canadian Orthopaedic Trauma Society was started in an endeavour to answer the difficult problem of obtaining enough patients to perform top-quality research into fractures. By maintaining a high standard, including randomised study design, inclusivity, open discussion among surgeons and excellent long-term follow-up, this group has become a leader in the orthopaedic research community. This annotation describes the short history, important components and spirit necessary to build a research community or team which will function well despite the difficult research environment facing individual surgeons.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 8 | Pages 1049 - 1053
1 Aug 2011
Putzer D Mayr E Haid C Reinthaler A Nogler M

In revision total hip replacement, bone loss can be managed by impacting porous bone chips. In order to guarantee sufficient mechanical strength, the bone chips have to be compacted. The aim of this study was to determine in an in vitro simulation whether the use of a pneumatic hammer leads to higher primary stability than manual impaction. Bone mass characteristics were measured by force and distance variation of a penetrating punch, which was lowered into a plastic cup filled with bone chips. From these measurements bulk density, contact stiffness, impaction hardness and penetration resistance were calculated for different durations of impaction.

We found that the pneumatic method reached higher values of impaction hardness, contact stiffness and bulk density suggesting an increase in stability of the implant. No significant differences were found between the two different methods concerning the penetration resistance. The pneumatic method might reduce the risk of fracture in vivo, as force peaks are smaller and applied for a shorter period. Results from manual impaction showed higher variability and depend much on the experience of the surgeon. The pneumatic hammer is a suitable tool to standardise the impaction process.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 8 | Pages 1054 - 1059
1 Aug 2011
van Jonbergen HPW Scholtes VAB van Kampen A Poolman RW

The efficacy of circumpatellar electrocautery in reducing the incidence of post-operative anterior knee pain is unknown. We conducted a single-centre, outcome-assessor and patient-blinded, parallel-group, randomised, controlled trial to compare circumpatellar electrocautery with no electrocautery in total knee replacement in the absence of patellar resurfacing. Patients requiring knee replacement for primary osteoarthritis were randomly assigned circumpatellar electrocautery (intervention group) or no electrocautery (control group). The primary outcome measure was the incidence of anterior knee pain. A secondary measure was the standardised clinical and patient-reported outcomes determined by the American Knee Society scores and the Western Ontario and McMaster Universities (WOMAC) osteoarthritis index. A total of 131 knees received circumpatellar electrocautery and 131 had no electrocautery.

The overall incidence of anterior knee pain at follow-up at one year was 26% (20% to 31%), with 19% (12% to 26%) in the intervention group and 32% (24% to 40%) in the control group (p = 0.02). The relative risk reduction from electrocautery was 40% (9% to 61%) and the number needed to treat was 7.7 (4.3 to 41.4). The intervention group had a better mean total WOMAC score at follow-up at one year compared with the control group (16.3 (0 to 77.7) versus 21.6 (0 to 76.7), p = 0.04). The mean post-operative American Knee Society knee scores and function scores were similar in the intervention and control groups (knee score: 92.4 (55 to 100) versus 90.4 (51 to 100), respectively (p = 0.14); function score: 86.5 (15 to 100) versus 84.5 (30 to 100), respectively (p = 0.49)).

Our study suggests that in the absence of patellar resurfacing electrocautery around the margin of the patella improves the outcome of total knee replacement.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 9 | Pages 1210 - 1216
1 Sep 2011
Mitsuyasu H Matsuda S Fukagawa S Okazaki K Tashiro Y Kawahara S Nakahara H Iwamoto Y

We investigated whether the extension gap in total knee replacement (TKR) would be changed when the femoral component was inserted. The extension gap was measured with and without the femoral component in place in 80 patients with varus osteoarthritis undergoing posterior-stabilised TKR. The effect of a post-operative increase in the size of the femoral posterior condyles was also evaluated. The results showed that placement of the femoral component significantly reduced the medial and lateral extension gaps by means of 1.0 mm and 0.9 mm, respectively (p < 0.0001). The extension gap was reduced when a larger femoral component was selected relative to the thickness of the resected posterior condyle. When the post-operative posterior lateral condyle was larger than that pre-operatively, 17 of 41 knees (41%) showed a decrease in the extension gap of > 2.0 mm. When a specially made femoral trial component with a posterior condyle enlarged by 4 mm was tested, the medial and lateral extension gaps decreased further by means of 2.1 mm and 2.8 mm, respectively.

If the thickness of the posterior condyle is expected to be larger than that pre-operatively, it should be recognised that the extension gap is likely to be altered. This should be taken into consideration when preparing the extension gap.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 4 | Pages 498 - 502
1 Apr 2011
Gwilym SE Oag HCL Tracey I Carr AJ

Impingement syndrome in the shoulder has generally been considered to be a clinical condition of mechanical origin. However, anomalies exist between the pathology in the subacromial space and the degree of pain experienced. These may be explained by variations in the processing of nociceptive inputs between different patients. We investigated the evidence for augmented pain transmission (central sensitisation) in patients with impingement, and the relationship between pre-operative central sensitisation and the outcomes following arthroscopic subacromial decompression.

We recruited 17 patients with unilateral impingement of the shoulder and 17 age- and gender-matched controls, all of whom underwent quantitative sensory testing to detect thresholds for mechanical stimuli, distinctions between sharp and blunt punctate stimuli, and heat pain. Additionally Oxford shoulder scores to assess pain and function, and PainDETECT questionnaires to identify ‘neuropathic’ and referred symptoms were completed. Patients completed these questionnaires pre-operatively and three months post-operatively.

A significant proportion of patients awaiting subacromial decompression had referred pain radiating down the arm and had significant hyperalgesia to punctate stimulus of the skin compared with controls (unpaired t-test, p < 0.0001). These are felt to represent peripheral manifestations of augmented central pain processing (central sensitisation).

The presence of either hyperalgesia or referred pain pre-operatively resulted in a significantly worse outcome from decompression three months after surgery (unpaired t-test, p = 0.04 and p = 0.005, respectively).

These observations confirm the presence of central sensitisation in a proportion of patients with shoulder pain associated with impingement. Also, if patients had relatively high levels of central sensitisation pre-operatively, as indicated by higher levels of punctate hyperalgesia and/or referred pain, the outcome three months after subacromial decompression was significantly worse.


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 8 | Pages 981 - 987
1 Aug 2008
Whittaker JP Dharmarajan R Toms AD

The management of bone loss in revision replacement of the knee remains a challenge despite an array of options available to the surgeon. Bone loss may occur as a result of the original disease, the design of the prosthesis, the mechanism of failure or technical error at initial surgery. The aim of revision surgery is to relieve pain and improve function while addressing the mechanism of failure in order to reconstruct a stable platform with transfer of load to the host bone. Methods of reconstruction include the use of cement, modular metal augmentation of prostheses, custom-made, tumour-type or hinged implants and bone grafting.

The published results of the surgical techniques are summarised and a guide for the management of bone defects in revision surgery of the knee is presented.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 4 | Pages 479 - 483
1 Apr 2011
Chang C Lai K Yang C Lan S

Between April 2004 and July 2007, we performed 241 primary total knee replacements in 204 patients using the e.motion posterior cruciate-retaining, multidirectional mobile-bearing prosthesis. Of these, 100 were carried out using an image-free navigation system, and the remaining 141 with the conventional technique. We conducted a retrospective study from the prospectively collected data of these patients to assess the early results of this new mobile-bearing design.

At a mean follow-up of 49 months (32 to 71), 18 knees (7.5%) had mechanical complications of which 13 required revision. Three of these had a peri-prosthetic fracture, and were removed from the study. The indication for revision in the remaining ten was loosening of the femoral component in two, tibiofemoral dislocation in three, disassociation of the polyethylene liner in four, and a broken polyethyene liner in one. There were eight further mechanically unstable knees which presented with recurrent disassociation of the polyethylene liner. There was no significant difference in the incidence of mechanical instability between the navigation-assisted procedures (8 of 99, 8.1%) and the conventionally implanted knees (10 of 139, 7.2%).

In our view, the relatively high rate of mechanical complications and revision within 30 months precludes the further use of new design of knee replacement.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 5 | Pages 695 - 699
1 May 2011
Spiro AS Babin K Lipovac S Stenger P Mladenov K Rupprecht M Rueger JM Stuecker R

The use of recombinant human bone morphogenetic protein-2 (rhBMP-2) for the treatment of congenital pseudarthrosis of the tibia has been investigated in only one previous study, with promising results. The aim of this study was to determine whether rhBMP-2 might improve the outcome of this disorder. We reviewed the medical records of five patients with a mean age of 7.4 years (2.3 to 21) with congenital pseudarthrosis of the tibia who had been treated with rhBMP-2 and intramedullary rodding. Ilizarov external fixation was also used in four of these patients. Radiological union of the pseudarthrosis was evident in all of them at a mean of 3.5 months (3.2 to 4) post-operatively. The Ilizarov device was removed after a mean of 4.2 months (3.0 to 5.3). These results indicate that treatment of congenital pseudarthrosis of the tibia using rhBMP-2 in combination with intramedullary stabilisation and Ilizarov external fixation may improve the initial rate of union and reduce the time to union.

Further studies with more patients and longer follow-up are necessary to determine whether this surgial procedure may significantly enhance the outcome of congenital pseudarthrosis of the tibia, considering the refracture rate (two of five patients) in this small case series.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 6 | Pages 713 - 719
1 Jun 2011
Duckworth AD Ring D McQueen MM

A suspected fracture of the scaphoid remains difficult to manage despite advances in knowledge and imaging methods. Immobilisation and restriction of activities in a young and active patient must be balanced against the risks of nonunion associated with an undiagnosed and undertreated fracture of the scaphoid.

The assessment of diagnostic tests for a suspected fracture of the scaphoid must take into account two important factors. First, the prevalence of true fractures among suspected fractures is low, which greatly reduces the probability that a positive test will correspond with a true fracture, as false positives are nearly as common as true positives. This situation is accounted for by Bayesian statistics. Secondly, there is no agreed reference standard for a true fracture, which necessitates the need for an alternative method of calculating diagnostic performance characteristics, based upon a statistical method which identifies clinical factors tending to associate (latent classes) in patients with a high probability of fracture.

The most successful diagnostic test to date is MRI, but in low-prevalence situations the positive predictive value of MRI is only 88%, and new data have documented the potential for false positive scans. The best strategy for improving the diagnosis of true fractures among suspected fractures of the scaphoid may well be to develop a clinical prediction rule incorporating a set of demographic and clinical factors which together increase the pre-test probability of a fracture of the scaphoid, in addition to developing increasingly sophisticated radiological tests.


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 12 | Pages 1631 - 1635
1 Dec 2005
von Schewelov T Sanzén L Önsten I Carlsson Å Besjakov J

We investigated the wear characteristics and clinical performance of four different total hip joint articulations in 114 patients. Wear and migration was measured by roentgenstereophotogrammetric analysis at five years or at the last follow-up. The mean annual wear was 0.11 mm for a stainless steel/Enduron articulation, 0.34 mm for stainless steel/Hylamer cup, 0.17 mm for zirconium oxide ceramic/Enduron and 0.40 mm for zirconium oxide ceramic/Hylamer. The difference between the groups was significant (p < 0.008) except for stainless steel/Hylamer vs zirconium oxide ceramic/Hylamer (p = 0.26).

At present, 12 patients have undergone a revision procedure, four at five years and eight thereafter. No patient who received a stainless steel/Enduron articulation at their primary replacement required revision.

Conflicting results have been reported about the performance of the zirconium oxide ceramic femoral head, but our findings suggest that it should not be used with a polymethylmethacrylate acetabular component. Hylamer has already been withdrawn from the market.