Advertisement for orthosearch.org.uk
Results 101 - 120 of 268
Results per page:
The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 8 | Pages 1072 - 1078
1 Aug 2010
Grammatopoulos G Pandit H Glyn-Jones S McLardy-Smith P Gundle R Whitwell D Gill HS Murray DW

Pseudotumours are a rare complication of hip resurfacing. They are thought to be a response to metal debris which may be caused by edge loading due to poor orientation of the acetabular component. Our aim was to determine the optimal acetabular orientation to minimise the risk of pseudotumour formation.

We matched 31 hip resurfacings revised for pseudotumour formation with 58 controls who had a satisfactory outcome from this procedure. The radiographic inclination and anteversion angles of the acetabular component were measured on anteroposterior radiographs of the pelvis using Einzel-Bild-Roentgen-Analyse software. The mean inclination angle (47°, 10° to 81°) and anteversion angle (14°, 4° to 34°) of the pseudotumour cases were the same (p = 0.8, p = 0.2) as the controls, 46° (29° to 60°) and 16° (4° to 30°) respectively, but the variation was greater. Assuming an accuracy of implantation of ± 10° about a target position, the optimal radiographic position was found to be approximately 45° of inclination and 20° of anteversion. The incidence of pseudotumours inside the zone was four times lower (p = 0.007) than outside the zone.

In order to minimise the risk of pseudotumour formation we recommend that surgeons implant the acetabular component at an inclination of 45° (± 10) and anteversion of 20° (± 10) on post-operative radiographs. Because of differences between the radiographic and the operative angles, this may be best achieved by aiming for an inclination of 40° and an anteversion of 25°.


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 7 | Pages 989 - 993
1 Jul 2010
Johnston AT Barnsdale L Smith R Duncan K Hutchison JD

We investigated the excess mortality risk associated with fractures of the hip. Data related to 29 134 patients who underwent surgery following a fracture of the hip were obtained from the Scottish Hip Fracture Audit database. Fractures due to primary or metastatic malignancy were excluded. An independent database (General Register Office (Scotland)) was used to validate dates of death. The observed deaths per 100 000 of the population were then calculated for each group (gender, age and fracture type) at various time intervals up to eight years. A second database (Interim Life Tables for Scotland, Scottish Government) was then used to create standardised mortality ratios. Analysis showed that mortality in patients aged > 85 years with a fracture of the hip tended to return to the level of the background population between two and five years after the fracture. In those patients aged < 85 years excess mortality associated with hip fracture persisted beyond eight years. Extracapsular hip fractures and male gender also conferred increased risk.


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 6 | Pages 835 - 841
1 Jun 2010
Holt G Smith R Duncan K McKeown DW

We have investigated how medical postponement, the time to surgery and the correction of medical abnormalities, according to McLaughlin criteria, before operation affected peri-operative mortality after fracture of the hip. From February to December 2007, in addition to core data, the Scottish Hip Fracture Audit collected information relating to surgical delay. Data were available for 4284 patients which allowed 30-day survival analysis to be performed. Multivariable logistic regression models were used to control for differences in case-mix.

Patients with major clinical abnormalities were more likely to have a postponement and had a lower unadjusted 30-day survival. The time to operation and postponement were not associated with higher mortality after adjustment for case-mix. Correction of major clinical abnormalities before surgery improved the adjusted survival, but this improvement was not significant (p = 0.10). Postponement without correction of a medical abnormality before surgery was associated with a significantly lower (p = 0.006) 30-day adjusted survival. The possible benefits of postponement need to be weighed against prolonged discomfort for the patient and the possibility of the development of other complications.


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 3 | Pages 356 - 361
1 Mar 2010
Kwon Y- Glyn-Jones S Simpson DJ Kamali A McLardy-Smith P Gill HS Murray DW

The presence of pseudotumours, which are soft-tissue masses relating to the hip, after metal-on-metal hip resurfacing arthroplasty has been associated with elevated levels of metal ions in serum, suggesting that pseudotumours occur when there is increased wear. We aimed to quantify the wear in vivo of implants revised for pseudotumours (eight) and of a control group of implants (22) revised for other reasons of failure.

We found that the implant group with pseudotumours had a significantly higher rate of median linear wear of the femoral component at 8.1 μm/year (2.75 to 25.4) than the 1.79 μm/year (0.82 to 4.15; p = 0.002) of the non-pseudotumour group. For the acetabular component a significantly higher rate of median linear wear of 7.36 μm/year (1.61 to 24.9) was observed in the pseudotumour group compared with 1.28 μm/year (0.81 to 3.33, p = 0.001) in the other group. Wear of the acetabular component in the pseudotumour group always involved the edge of the implant, indicating that edge-loading had occurred.

Our findings are the first direct evidence that pseudotumour is associated with increased wear at the metal-on-metal articulation. Furthermore, edge-loading with the loss of fluid-film lubrication may be an important mechanism of generation of wear in patients with a pseudotumour.


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 1 | Pages 83 - 91
1 Jan 2010
Moosmayer S Lund G Seljom U Svege I Hennig T Tariq R Smith H

In order to compare the outcome from surgical repair and physiotherapy, 103 patients with symptomatic small and medium-sized tears of the rotator cuff were randomly allocated to one of the two approaches. The primary outcome measure was the Constant score, and secondary outcome measures included the self-report section of the American Shoulder and Elbow Surgeons score, the Short Form 36 Health Survey and subscores for shoulder movement, pain, strength and patient satisfaction. Scores were taken at baseline and after six and 12 months by a blinded assessor. Nine patients (18%) with insufficient benefit from physiotherapy after at least 15 treatment sessions underwent secondary surgical treatment.

Analysis of between-group differences showed better results for the surgery group on the Constant scale (difference 13.0 points, p − 0.002), on the American Shoulder and Elbow surgeons scale (difference 16.1 points, p < 0.0005), for pain-free abduction (difference 28.8°, p = 0.003) and for reduction in pain (difference on a visual analogue scale −1.7 cm, p < 0.0005).


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 8 | Pages 1019 - 1024
1 Aug 2009
Grammatopolous G Pandit H Kwon Y Gundle R McLardy-Smith P Beard DJ Murray DW Gill HS

Inflammatory pseudotumours occasionally occur after metal-on-metal hip resurfacing and often lead to revision. Our aim was to determine the severity of this complication by assessing the outcome of revision in these circumstances and by comparing this with the outcome of other metal-on-metal hip resurfacing revisions as well as that of matched primary total hip replacements.

We identified 53 hips which had undergone metal-on-metal hip resurfacing and required revision at a mean of 1.59 years (0.01 to 6.69) after operation. Of these, 16 were revised for pseudotumours, 21 for fracture and 16 for other reasons. These were matched by age, gender and diagnosis with 103 patients undergoing primary total hip replacement with the Exeter implant.

At a mean follow-up of three years (0.8 to 7.2) the outcome of metal-on-metal hip resurfacing revision for pseudotumour was poor with a mean Oxford hip score of 20.9 (sd 9.3) and was significantly worse (p < 0.001) than the outcome for fracture with a mean Oxford hip score of 40.2 (sd 9.2) or that for other causes with a mean Oxford hip score of 37.8 (sd 9.4). The clinical outcome of revision for pseudotumour was also significantly worse (p < 0.001) than the outcome of matched primary total hip replacements. By contrast, the outcome for fracture and other causes was not significantly different from that of matched primary total hip replacements (p = 0.065). After revision for pseudotumour there were three cases of recurrent dislocation, three of palsy of the femoral nerve, one of stenosis of the femoral artery and two of loosening of the component. Five hips required further revision. In three of these there was evidence of recurrent pseudotumour, and one is currently awaiting further revision. The incidence of major complications after revision for pseudotumour (50%) was significantly higher (p = 0.018) than that after revision for other causes (14%).

The outcome of revision for pseudotumour is poor and consideration should be given to early revision to limit the extent of the soft-tissue destruction. The outcome of resurfacing revision for other causes is good.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 4 | Pages 522 - 529
1 Apr 2009
Ryzewicz M Morgan SJ Linford E Thwing JI de Resende GVP Smith WR

Nonunion of the tibia associated with bone loss, previous infection, obliteration of the intramedullary canal or located in the distal metaphysis poses a challenge to the surgeon and significant morbidity to patients. We retrospectively reviewed the records of 24 patients who were treated by central bone grafting and compared them to those of 20 who were treated with a traditional posterolateral graft. Central bone grafting entails a lateral approach, anterior to the fibula and interosseous membrane which is used to create a central space filled with cancellous iliac crest autograft. Upon consolidation, a tibiofibular synostosis is formed that is strong enough for weight-bearing. This procedure has advantages over other methods of treatment for selected nonunions.

Of the 24 patients with central bone grafting, 23 went on to radiographic and clinical union without further intervention. All healed within a mean of 20 weeks (10 to 48). No further bone grafts were required, and few complications were encountered. These results were comparable to those of the 20 patients who underwent posterolateral bone grafting who united at a mean of 31.3 weeks (16 to 60) but one of whom required below-knee amputation for intractable sepsis.

Central bone grafting is a safe and effective treatment for difficult nonunions of the tibia.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 2 | Pages 196 - 200
1 Feb 2009
Moosmayer S Smith H Tariq R Larmo A

We undertook clinical and ultrasonographic examination of the shoulders of 420 asymptomatic volunteers aged between 50 and 79 years. MRI was performed in selected cases. Full-thickness tears of the rotator cuff were detected in 32 subjects (7.6%). The prevalence increased with age as follows: 50 to 59 years, 2.1%; 60 to 69 years, 5.7%; and 70 to 79 years, 15%. The mean size of the tear was less than 3 cm and tear localisation was limited to the supraspinatus tendon in most cases (78%). The strength of flexion was reduced significantly in the group with tears (p = 0.01).

Asymptomatic tears of the rotator cuff should be regarded as part of the normal ageing process in the elderly but may be less common than hitherto believed.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 1 | Pages 37 - 43
1 Jan 2009
Hassaballa M Mehendale S Poniatowski S Kalantzis G Smith E Learmonth ID

Loss of bone stock is a major problem in revision surgery of the hip. Impaction bone grafting of the femur is frequently used when dealing with deficient bone stock. In this retrospective study a consecutive series of 68 patients (69 hips) who had revision of a hip replacement with femoral impaction grafting were reviewed. Irradiated bone allograft was used in all hips. Radiological measurement of subsidence of the stem, incorporation of the graft and remodelling was carried out and showed incorporation of the graft in 26 of 69 hips (38%). However, there was no evidence of trabecular remodelling. Moderate subsidence of 5 mm to 10 mm occurred in ten hips (14.5%), and massive subsidence of > 10 mm in five (7.2%).

The results of this study are less favourable than those of others describing studies of revision of the femoral stem using impaction bone grafting. The absence of the characteristic changes of graft remodelling noted in other series raises the question as to whether irradiated bone graft may be a significant factor influencing the post-operative outcome.


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 11 | Pages 1435 - 1440
1 Nov 2008
Smith IDM Elton R Ballantyne JA Brenkel IJ

In Scotland, the number of primary total knee replacements performed annually has been increasing steadily. The price of the implant is fixed but the length of hospital stay is variable.

We prospectively investigated all patients who underwent primary unilateral total knee replacement in the Scottish region of Fife, between December 1994 and February 2007 and assessed their recorded pre-operative details. The data were analysed using univariate and multiple linear regression statistical analysis.

Data on the length of stay were available from a total of 2106 unilateral total knee replacements. The median length of hospital stay was eight days. The significant pre-operative risk factors for an increased length of stay were the year of admission, details of the consultant looking after the patient, the stair score, the walking-aid score and age.

Awareness of the pre-operative factors which increase the length of hospital stay may provide the opportunity to influence them favourably and to reduce the time in hospital and the associated costs of unilateral total knee replacement.


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 10 | Pages 1357 - 1363
1 Oct 2008
Holt G Smith R Duncan K Finlayson DF Gregori A

We investigated the relationship between a number of patient and management variables and mortality after surgery for fracture of the hip. Data relating to 18 817 patients were obtained from the Scottish Hip Fracture Audit database. We divided variables into two categories, depending on whether they were case-mix (age; gender; fracture type; pre-fracture residence; pre-fracture mobility and ASA scores) or management variables (time from fracture to surgery; time from admission to surgery; grade of surgical and anaesthetic staff undertaking the procedure and anaesthetic technique).

Multivariate logistic regression analysis showed that all case-mix variables were strongly associated with post-operative mortality, even when controlling for the effects of the remaining variables. Inclusion of the management variables into the case-mix base regression model provided no significant improvement to the model. Patient case-mix variables have the most significant effect on post-operative mortality and unfortunately such variables cannot be modified by pre-operative medical interventions.


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 10 | Pages 1372 - 1379
1 Oct 2008
Robin J Graham HK Selber P Dobson F Smith K Baker R

There is much debate about the nature and extent of deformities in the proximal femur in children with cerebral palsy. Most authorities accept that increased femoral anteversion is common, but its incidence, severity and clinical significance are less clear. Coxa valga is more controversial and many authorities state that it is a radiological artefact rather than a true deformity.

We measured femoral anteversion clinically and the neck-shaft angle radiologically in 292 children with cerebral palsy. This represented 78% of a large, population-based cohort of children with cerebral palsy which included all motor types, topographical distributions and functional levels as determined by the gross motor function classification system.

The mean femoral neck anteversion was 36.5° (11° to 67.5°) and the mean neck-shaft angle 147.5° (130° to 178°). These were both increased compared with values in normally developing children. The mean femoral neck anteversion was 30.4° (11° to 50°) at gross motor function classification system level I, 35.5° (8° to 65°) at level II and then plateaued at approximately 40.0° (25° to 67.5°) at levels III, IV and V. The mean neck-shaft angle increased in a step-wise manner from 135.9° (130° to 145°) at gross motor function classification system level I to 163.0° (151° to 178°) at level V. The migration percentage increased in a similar pattern and was closely related to femoral deformity.

Based on these findings we believe that displacement of the hip in patients with cerebral palsy can be explained mainly by the abnormal shape of the proximal femur, as a result of delayed walking, limited walking or inability to walk. This has clinical implications for the management of hip displacement in children with cerebral palsy.


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 7 | Pages 847 - 851
1 Jul 2008
Pandit H Glyn-Jones S McLardy-Smith P Gundle R Whitwell D Gibbons CLM Ostlere S Athanasou N Gill HS Murray DW

We report 17 patients (20 hips) in whom metal-on-metal resurfacing had been performed and who presented with various symptoms and a soft-tissue mass which we termed a pseudotumour. Each patient underwent plain radiography and in some, CT, MRI and ultrasonography were also performed. In addition, histological examination of available samples was undertaken.

All the patients were women and their presentation was variable. The most common symptom was discomfort in the region of the hip. Other symptoms included spontaneous dislocation, nerve palsy, a noticeable mass or a rash. The common histological features were extensive necrosis and lymphocytic infiltration. To date, 13 of the 20 hips have required revision to a conventional hip replacement. Two are awaiting revision.

We estimate that approximately 1% of patients who have a metal-on-metal resurfacing develop a pseudotumour within five years. The cause is unknown and is probably multifactorial. There may be a toxic reaction to an excess of particulate metal wear debris or a hypersensitivity reaction to a normal amount of metal debris. We are concerned that with time the incidence of these pseudotumours may increase. Further investigation is required to define their cause.


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 5 | Pages 556 - 561
1 May 2008
Glyn-Jones S McLardy-Smith P Gill HS Murray DW

The creep and wear behaviour of highly cross-linked polyethylene and standard polyethylene liners were examined in a prospective, double-blind randomised, controlled trial using radiostereometric analysis.

We randomised 54 patients to receive hip replacements with either highly cross-linked polyethylene or standard liners and determined the three-dimensional penetration of the liners over three years.

After three years the mean total penetration was 0.35 mm (SD 0.14) for the highly cross-linked polyethylene group and 0.45 mm (SD 0.19) for the standard group. The difference was statistically significant (p = 0.0184). From the pattern of penetration it was possible to discriminate creep from wear. Most (95%) of the creep occurred within six months of implantation and nearly all within the first year. There was no difference in the mean degree of creep between the two types of polyethylene (highly cross-linked polyethylene 0.26 mm, SD 0.17; standard 0.27 mm, SD 0.2; p = 0.83). There was, however, a significant difference (p = 0.012) in the mean wear rate (highly cross-linked polyethylene 0.03 mm/yr, SD 0.06; standard 0.07 mm/yr, SD 0.05). Creep and wear occurred in significantly different directions (p = 0.01); creep was predominantly proximal whereas wear was anterior, proximal and medial.

We conclude that penetration in the first six months is creep-dominated, but after one year virtually all penetration is due to wear. Highly cross-linked polyethylene has a 60% lower rate of wear than standard polyethylene and therefore will probably perform better in the long term.


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 4 | Pages 436 - 441
1 Apr 2008
Steffen RT Pandit HP Palan J Beard DJ Gundle R McLardy-Smith P Murray DW Gill HS

Few independent studies have reported the outcome of resurfacing arthroplasty of the hip. The aim of this study was to report the five-year clinical outcome and seven-year survival of an independent series.

A total of 610 Birmingham Hip Resurfacing arthroplasties were performed in 532 patients with a mean age of 51.8 years (16.5 to 81.6). They were followed for between two and eight years; 107 patients (120 hips) had been followed up for more than five years. Two patients were lost to follow-up. At a minimum of five years’ follow-up, 79 of 85 hips (93%) had an excellent or good outcome according to the Harris hip score. The mean Oxford hip score was 16.1 points (sd 7.7) and the mean University of California Los Angeles activity score was 6.6 points (sd 1.9). There were no patients with definite radiological evidence of loosening or of narrowing of the femoral neck exceeding 10% of its width. There were 23 revisions (3.8%), giving an overall survival of 95% (95% confidence interval 85.3 to 99.2) at seven years. Fractured neck of femur in 12 hips was the most common indication for revision, followed by aseptic loosening in four. In three hips (three patients) (0.5%), failure was possibly related to metal debris.

Considering that these patients are young and active these results are good, and support the use of resurfacing. Further study is needed to address the early failures, particularly those related to fracture and metal debris.


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 4 | Pages 480 - 483
1 Apr 2008
Holt G Smith R Duncan K Hutchison JD Gregori A

We report gender differences in the epidemiology and outcome after hip fracture from the Scottish Hip Fracture Audit, with data on admission and at 120 days follow-up from 22 orthopaedic units across the country between 1998 and 2005. Outcome measures included early mortality, length of hospital stay, 120-day residence and mobility. A multivariate logistic regression model compared outcomes between genders. The study comprised 25 649 patients of whom 5674 (22%) were men and 19 975 (78%) were women. The men were in poorer pre-operative health, despite being younger at presentation (mean 77 years (60 to 101) vs 81 years (50 to 106)). Pre-fracture residence and mobility were similar between genders.

Multivariate analysis indicated that the men were less likely to return to their home or mobilise independently at the 120-day follow-up. Mortality at 30 and 120 days was higher for men, even after differences in case-mix variables between genders were considered.


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 3 | Pages 393 - 399
1 Mar 2008
Morley JR Smith RM Pape HC MacDonald DA Trejdosiewitz LK Giannoudis PV

We have undertaken a prospective study in patients with a fracture of the femoral shaft requiring intramedullary nailing to test the hypothesis that the femoral canal could be a potential source of the second hit phenomenon. We determined the local femoral intramedullary and peripheral release of interleukin-6 (IL-6) after fracture and subsequent intramedullary reaming.

In all patients, the fracture caused a significant increase in the local femoral concentrations of IL-6 compared to a femoral control group. The concentration of IL-6 in the local femoral environment was significantly higher than in the patients own matched blood samples from their peripheral circulation. The magnitude of the local femoral release of IL-6 after femoral fracture was independent of the injury severity score and whether the fracture was closed or open.

In patients who underwent intramedullary reaming of the femoral canal a further significant local release of IL-6 was demonstrated, providing evidence that intramedullary reaming can cause a significant local inflammatory reaction.


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 2 | Pages 125 - 126
1 Feb 2008
Smith R

The application and interpretation of the impact factor of a journal is fraught with problems, some of which are explored in this editorial.


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 2 | Pages 209 - 214
1 Feb 2008
Pradhan A Cheung YC Grimer RJ Peake D Al-Muderis OA Thomas JM Smith M

We have investigated the oncological outcome of 63 patients with soft-tissue sarcomas of the hand managed at three major centres in the United Kingdom. There were 44 males and 19 females with a mean age of 45 years (11 to 92). The three most common diagnoses were synovial sarcoma, clear cell sarcoma and epithelioid sarcoma. Local excision was carried out in 45 patients (71%) and amputation in 18 (29%). All those treated by amputation had a wide margin of excision but this was only achieved in 58% of those treated by local excision. The risk of local recurrence was 6% in those treated by amputation compared with 42% for those who underwent attempted limb salvage. An inadequate margin of excision resulted in a 12 times greater risk of local recurrence when compared with those in whom a wide margin of excision had been achieved. We were unable to demonstrate any role for radiotherapy in decreasing the risk of local recurrence when there was an inadequate margin of excision. Patients with an inadequate margin of excision had a much higher risk of both local recurrence and metastasis than those with wide margins. The overall survival rate at five years was 87% and was related to the grade and size of the tumour and to the surgical margin.

We have shown that a clear margin of excision is essential to achieve local control of a soft-tissue sarcoma in the hand and that failure to achieve this results in a high risk of both local recurrence and metastastic disease.


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 2 | Pages 246 - 253
1 Feb 2008
Coathup M Smith N Kingsley C Buckland T Dattani R Ascroft GP Blunn G

An experimental sheep model was used for impaction allografting of 12 hemiarthroplasty femoral components placed into two equal-sized groups. In group 1, a 50:50 mixture of ApaPore hydroxyapatite bone-graft substitute and allograft was used. In group 2, ApaPore and allograft were mixed in a 90:10 ratio. Both groups were killed at six months. Ground reaction force results demonstrated no significant differences (p > 0.05) between the two groups at 8, 16 and 24 weeks post-operatively, and all animals remained active. The mean bone turnover rates were significantly greater in group 1, at 0.00206 mm/day, compared to group 2 at 0.0013 mm/day (p < 0.05). The results for the area of new bone formation demonstrated no significant differences (p > 0.05) between the two groups. No significant differences were found between the two groups in thickness of the cement mantle (p > 0.05) and percentage ApaPore-bone contact (p > 0.05).

The results of this animal study demonstrated that a mixture of ApaPore allograft in a 90:10 ratio was comparable to using a 50:50 mixture.