Early failure associated with adverse reactions to metal debris is an emerging problem after hip resurfacing but the exact mechanism is unclear. We analysed our entire series of 660 metal-on-metal resurfacings (Articular Surface Replacement (ASR) and Birmingham Hip Resurfacing (BHR)) and large-bearing ASR total hip replacements, to establish associations with metal debris-related failures. Clinical and radiological outcomes, metal ion levels, explant studies and lymphocyte transformation tests were performed. A total of 17 patients (3.4%) were identified (all ASR bearings) with adverse reactions to metal debris, for which revision was required. This group had significantly smaller components, significantly higher acetabular component anteversion, and significantly higher whole concentrations of blood and joint chromium and cobalt ions than asymptomatic patients did (all p <
0.001). Post-revision lymphocyte transformation tests on this group showed no reactivity to chromium or cobalt ions. Explants from these revisions had greater surface wear than retrievals for uncomplicated fractures. The absence of adverse reactions to metal debris in patients with well-positioned implants usually implies high component wear. Surgeons must consider implant design, expected component size and acetabular component positioning in order to reduce early failures when performing large-bearing metal-on-metal hip resurfacing and replacement.
The December 2014 Knee Roundup360 looks at: national guidance on arthroplasty thromboprophylaxis is effective; unicompartmental knee replacement has the edge in terms of short-term complications; stiff knees, timing and manipulation; neuropathic pain and total knee replacement; synovial fluid α-defensin and CRP: a new gold standard in joint infection diagnosis?; how to assess anterior knee pain?; where is the evidence? Five new implants under the spotlight; and a fresh look at ACL reconstruction
The June 2014 Hip &
Pelvis Roundup360 looks at: Modular femoral necks: early signs are not good; is corrosion to blame for modular neck failures; metal-on-metal is not quite a closed book; no excess failures in fixation of displaced femoral neck fractures; noise no problem in hip replacement; heterotopic ossification after hip arthroscopy: are NSAIDs the answer?; thrombotic and bleeding events surprisingly low in total joint replacement; and the elephant in the room: complications and surgical volume.
We conducted a retrospective study to assess
the prevalence of adverse reactions to metal debris (ARMD) in patients
operated on at our institution with metal-on-metal (MoM) total hip
replacements with 36 mm heads using a Pinnacle acetabular shell.
A total of 326 patients (150 males, 175 hips; 176 females, 203 hips)
with a mean age of 62.7 years (28 to 85) and mean follow-up of 7.5
years (0.1 to 10.8) participating in our in-depth modern MoM follow-up
programme were included in the study, which involved recording whole
blood cobalt and chromium ion measurements, Oxford hip scores (OHS)
and plain radiographs of the hip and targeted cross-sectional imaging. Elevated
blood metal ion levels (>
5 parts per billion) were seen in 32 (16.1%)
of the 199 patients who underwent unilateral replacement. At 23
months after the start of our modern MoM follow-up programme, 29
new cases of ARMD had been revealed. Hence, the nine-year survival
of this cohort declined from 96% (95% CI 95 to 98) with the old
surveillance routine to 86% (95% CI 82 to 90) following the new
protocol. Although ARMD may not be as common in 36 mm MoM THRs as
in those with larger heads, these results support the Medicines
and Healthcare Products Regulatory Agency guidelines on regular
reviews and further investigations, and emphasise the need for specific
a follow-up programme for patients with MoM THRs. Cite this article:
The issues surrounding raised levels of metal
ions in the blood following large head metal-on-metal total hip replacement
(THR), such as cobalt and chromium, have been well documented. Despite
the national popularity of uncemented metal-on-polyethylene (MoP)
THR using a large-diameter femoral head, few papers have reported
the levels of metal ions in the blood following this combination.
Following an isolated failure of a 44 mm Trident–Accolade uncemented
THR associated with severe wear between the femoral head and the
trunnion in the presence of markedly elevated levels of cobalt ions
in the blood, we investigated the relationship between modular femoral head
diameter and the levels of cobalt and chromium ions in the blood
following this THR. A total of 69 patients received an uncemented Trident–Accolade
MoP THR in 2009. Of these, 43 patients (23 men and 20 women, mean
age 67.0 years) were recruited and had levels of cobalt and chromium
ions in the blood measured between May and June 2012. The patients
were then divided into three groups according to the diameter of
the femoral head used: 12 patients in the 28 mm group (controls),
18 patients in the 36 mm group and 13 patients in the 40 mm group.
A total of four patients had identical bilateral prostheses in situ
at phlebotomy: one each in the 28 mm and 36 mm groups and two in
the 40 mm group. There was a significant increase in the mean levels of cobalt
ions in the blood in those with a 36 mm diameter femoral head compared
with those with a 28 mm diameter head (p = 0.013). The levels of
cobalt ions in the blood were raised in those with a 40 mm diameter
head but there was no statistically significant difference between
this group and the control group (p = 0.152). The levels of chromium
ions in the blood were normal in all patients. The clinical significance of this finding is unclear, but we
have stopped using femoral heads with a diameter of ≤ 36 mm, and
await further larger studies to clarify whether, for instance, this
issue particularly affects this combination of components. Cite this article:
The October 2013 Hip &
Pelvis Roundup360 looks at: Young and impinging; Clothes, weather and femoral heads?; Go long, go cemented; Surgical repair of the abductors?; Aspirin for DVT prophylaxis?; Ceramic-on-polyethylene: a low-wear solution?; ALVAL and ASR™: the story continues….; Salvaging Legg-Calve-Perthes’ disease
We aimed to determine the effect of surgical approach on the
histology of the femoral head following resurfacing of the hip. We performed a histological assessment of the bone under the
femoral component taken from retrieval specimens of patients having
revision surgery following resurfacing of the hip. We compared the
number of empty lacunae in specimens from patients who had originally
had a posterior surgical approach with the number in patients having alternative
surgical approaches.Objectives
Methods
A total of 397 hips were randomised to receive
Metasul metal-on-metal (MoM), metal-on-conventional polyethylene (MoP)
or ceramic-on-polyethylene (CoP) bearings using a cemented triple-tapered
polished femoral component (MS-30). There were 129 MoM hips in 123
patients (39 male and 84 female, mean age 63.3 years (40.7 to 72.9)),
137 MoP hips in 127 patients (39 male and 88 female, mean age 62.8
years (24.5 to 72.7)) and 131 CoP hips in 124 patients (51 male
and 73 female, mean age 63.9 years (30.6 to 73.8)). All acetabular
components were cemented Weber polyethylene components with the
appropriate inlay for the MoM articulation. Clinical evaluation
was undertaken using the Harris hip score (HHS) and radiological
assessments were made at two, five and seven years. The HHS and radiological
analysis were available for 341 hips after seven years. The MoM
group had the lowest mean HHS (p = 0.124), a higher rate of revision
(p <
0.001) and a higher incidence of radiolucent lines in unrevised
hips (p <
0.001). In all, 12 revisions had been performed in
12 patients: eight in the MoM group (four for infection, four for
aseptic loosening, three in the MoP group (one each of infection,
dislocation and pain) and one in the CoP group (infection). Our findings reveal no advantage to the MoM bearing and identified
a higher revision rate and a greater incidence of radiolucent lines
than with the other articulations. We recommend that patients with
a 28 mm Metasul MoM bearing be followed carefully. Cite this article:
We have previously developed a radiographic technique, the oblique posterior condylar view, for assessment of the posterior aspect of the femoral condyles after total knee arthroplasty. The purpose of this study was to confirm the validity of this radiographic view based upon intra-operative findings at revision total knee arthroplasty. Lateral and oblique posterior condylar views were performed for 11 knees prior to revision total knee arthroplasty, and radiolucent lines or osteolysis of the posterior aspect of the femoral condyles were identified. These findings were compared with the intra-operative appearance of the posterior aspects of the femoral condyles. Statistical analysis showed that sensitivity and efficacy were significantly better for the oblique posterior condylar than the lateral view. This method can, therefore, be considered as suitable for routine follow-up radiographs of the femoral component and in the pre-operative planning of revision surgery.
Metal-on-metal hip arthroplasty gained significant
favor in the first decade of the millennium. However, the past several
years have seen increasing reports of failure, pseudotumor and other
adverse reactions. This study presents the results of a single center’s
15-year experience with metal-on-metal total hip arthroplasty as
strong evidence that metal-on-metal is going, going, gone.
We report on the long-term results of 163 bicruciate-retaining
Hermes 2C total knee replacements in 130 patients at a mean follow-up
of 22.4 years (20.3 to 23.5). Even when the anterior cruciate ligament
had a partially degenerative appearance it was preserved as long
as the knee had a normal anterior drawer and Lachman’s test pre-operatively. The
description and surgical technique of this minimally constrained
prosthesis were published in 1983 and the ten-year clinical results
in 1999. A total of 12% of the knees (20 of 163) in this study were revised
because of wear of the polyethylene tibial insert. Excellent stability
was achieved and the incidence of aseptic component loosening was
4.3% (seven of 163). The survival rate using revision for any reason
as the endpoint was 82% (95% confidence interval 76.2 to 88.0). Although this series included a relatively small number of replacements,
it demonstrated that the anterior cruciate ligament, even when partially
degenerated at the time of TKR, remained functional and provided
adequate stability at a long-term follow-up. Cite this article:
A moderator and panel of five experts led an
interactive session in discussing five challenging and interesting patient
case presentations involving surgery of the hip. The hip pathologies
reviewed included failed open reduction internal fixation of subcapital
femoral neck fracture, bilateral hip disease, evaluation of pain
after metal-on-metal hip arthroplasty, avascular necrosis, aseptic
loosening secondary to osteolysis and polyethylene wear, and management
of ceramic femoral head fracture.
Fracture of a ceramic component in total hip
replacement is a rare but potentially catastrophic complication.
The incidence is likely to increase as the use of ceramics becomes
more widespread. We describe such a case, which illustrates how
inadequate initial management will lead to further morbidity and
require additional surgery. We present the case as a warning that
fracture of a ceramic component should be revised to another ceramic-on-ceramic
articulation in order to minimise the risk of further catastrophic
wear.
Following the recall of modular neck hip stems
in July 2012, research into femoral modularity will intensify over
the next few years. This review aims to provide surgeons with an
up-to-date summary of the clinically relevant evidence. The development
of femoral modularity, and a classification system, is described.
The theoretical rationale for modularity is summarised and the clinical
outcomes are explored. The review also examines the clinically relevant problems
reported following the use of femoral stems with a modular neck. Joint replacement registries in the United Kingdom and Australia
have provided data on the failure rates of modular devices but cannot
identify the mechanism of failure. This information is needed to
determine whether modular neck femoral stems will be used in the
future, and how we should monitor patients who already have them implanted. Cite this article:
Stems improve the mechanical stability of tibial
components in total knee replacement (TKR), but come at a cost of stress
shielding along their length. Their advantages include resistance
to shear, reduced tibial lift-off and increased stability by reducing
micromotion. Longer stems may have disadvantages including stress
shielding along the length of the stem with associated reduction
in bone density and a theoretical risk of subsidence and loosening, peri-prosthetic
fracture and end-of-stem pain. These features make long stems unattractive
in the primary TKR setting, but often desirable in revision surgery
with bone loss and instability. In the revision scenario, stems
are beneficial in order to convey structural stability to the construct
and protect the reconstruction of bony defects. Cemented and uncemented
long stemmed implants have different roles depending on the nature
of the bone loss involved. This review discusses the biomechanics of the design of tibial
components and stems to inform the selection of the component and
the technique of implantation.
We report a case of a male patient presenting
with bilateral painful but apparently well-positioned and -fixed
large-diameter metal-on-metal hip replacements four years post-operatively.
Multiple imaging modes revealed a thick-walled, cystic expansile
mass in communication with the hip joint (a pseudotumour). Implant
retrieval analysis and tissue culture eliminated high bearing wear
or infection as causes for the soft-tissue reaction, but noted marked corrosion
of the modular neck taper adaptor and corrosion products in the
tissues. Therefore, we believe corrosion products from the taper
caused by mismatch of the implant components led to pseudotumour
formation requiring revision.
The painful subluxed or dislocated hip in adults
with cerebral palsy presents a challenging problem. Prosthetic dislocation
and heterotopic ossification are particular concerns. We present
the first reported series of 19 such patients (20 hips) treated
with hip resurfacing and proximal femoral osteotomy. The pre-operative
Gross Motor Function Classification System (GMFCS) was level V in
13 (68%) patients, level IV in three (16%), level III in one (5%) and
level II in two (11%). The mean age at operation was 37 years (13
to 57). The mean follow-up was 8.0 years (2.7 to 11.6), and 16 of the
18 (89%) contactable patients or their carers felt that the surgery
had been worthwhile. Pain was relieved in 16 of the 18 surviving
hips (89%) at the last follow-up, and the GMFCS level had improved
in seven (37%) patients. There were two (10%) early dislocations;
three hips (15%) required revision of femoral fixation, and two
hips (10%) required revision, for late traumatic fracture of the
femoral neck and extra-articular impingement, respectively. Hence
there were significant surgical complications in a total of seven
hips (35%). No hips required revision for instability, and there
were no cases of heterotopic ossification. We recommend hip resurfacing with proximal femoral osteotomy
for the treatment of the painful subluxed or dislocated hip in patients
with cerebral palsy.
Peri-articular soft-tissue masses or ‘pseudotumours’
can occur after large-diameter metal-on-metal (MoM) resurfacing
of the hip and conventional total hip replacement (THR). Our aim
was to assess the incidence of pseudotumour formation and to identify
risk factors for their formation in a prospective cohort study. A total of 119 patients who underwent 120 MoM THRs with large-diameter
femoral heads between January 2005 and November 2007 were included
in the study. Outcome scores, serum metal ion levels, radiographs
and CT scans were obtained. Patients with symptoms or an identified
pseudotumour were offered MRI and an ultrasound-guided biopsy. There were 108 patients (109 hips) eligible for evaluation by
CT scan at a mean follow-up of 3.6 years (2.5 to 4.5); 42 patients
(39%) were diagnosed with a pseudotumour. The hips of 13 patients
(12%) were revised to a polyethylene acetabular component with small-diameter
metal head. Patients with elevated serum metal ion levels had a
four times increased risk of developing a pseudotumour. This study shows a substantially higher incidence of pseudotumour
formation and subsequent revisions in patients with MoM THRs than
previously reported. Because most revision cases were identified
only after an intensive screening protocol, we recommend close monitoring
of patients with MoM THR.
Persistent groin pain after seemingly successful
total hip replacement (THR) appears to have become more common.
Recent studies have indicated a high incidence after metal-on-polyethylene
and metal-on-metal conventional THR and it has been documented in
up to 18% of patients after metal-on-metal resurfacing. There are many
causes, including acetabular loosening, stress fracture, and iliopsoas
tendonitis and impingement. The evaluation of this problem requires
a careful history and examination, plain radiographs and an algorithmic approach
to special diagnostic imaging and tests. Non-operative treatment
is not usually successful. Specific operative treatment depending
on the cause of the pain usually involves revision of the acetabular
component, iliopsoas tenotomy or other procedures, and is usually
successful. Here, an appropriate algorithm is described.
The aims of this piece of work were to: 1) record the background
concentrations of blood chromium (Cr) and cobalt (Co) concentrations
in a large group of subjects; 2) to compare blood/serum Cr and Co
concentrations with retrieved metal-on-metal (MoM) hip resurfacings;
3) to examine the distribution of Co and Cr in the serum and whole
blood of patients with MoM hip arthroplasties; and 4) to further
understand the partitioning of metal ions between the serum and
whole blood fractions. A total of 3042 blood samples donated to the local transfusion
centre were analysed to record Co and Cr concentrations. Also, 91
hip resurfacing devices from patients who had given pre-revision
blood/serum samples for metal ion analysis underwent volumetric
wear assessment using a coordinate measuring machine. Linear regression analysis
was carried out and receiver operating characteristic curves were
constructed to assess the reliability of metal ions to identify
abnormally wearing implants. The relationship between serum and
whole blood concentrations of Cr and Co in 1048 patients was analysed
using Bland-Altman charts. This relationship was further investigated
in an Objectives
Methods