The Bone & Joint Journal
Vol. 99-B, Issue 11 | Pages 1467 - 1476
1 Nov 2017
van Hamersveld KT Marang-van de Mheen PJ Tsonaka R Valstar† ER Toksvig-Larsen S

Aims

The optimal method of tibial component fixation remains uncertain in total knee arthroplasty (TKA). Hydroxyapatite coatings have been applied to improve bone ingrowth in uncemented designs, but may only coat the directly accessible surface. As peri-apatite (PA) is solution deposited, this may increase the coverage of the implant surface and thereby fixation. We assessed the tibial component fixation of uncemented PA-coated TKAs versus cemented TKAs.

Patients and Methods

Patients were randomised to PA-coated or cemented TKAs. In 60 patients (30 in each group), radiostereometric analysis of tibial component migration was evaluated as the primary outcome at baseline, three months post-operatively and at one, two and five years. A linear mixed-effects model was used to analyse the repeated measurements.


The Bone & Joint Journal
Vol. 99-B, Issue 7 | Pages 880 - 886
1 Jul 2017
Mohaddes M Shareghi B Kärrholm J

Aims

The aim of this study was to compare the incidence of aseptic loosening after the use of a cemented acetabular component and a Trabecular Metal (TM) acetabular component (Zimmer Inc., Warsaw, Indiana) at acetabular revision with bone impaction grafting.

Patients and Methods

A total of 42 patients were included in the study. Patients were randomised to receive an all- polyethylene cemented acetabular component (n = 19) or a TM component (n = 23). Radiostereometric analysis and conventional radiographic examinations were performed regularly up to two years post-operatively or until further revision.


Bone & Joint Research
Vol. 6, Issue 5 | Pages 270 - 276
1 May 2017
Gosiewski JD Holsgrove TP Gill HS

Objectives

Fractures of the proximal femur are a common clinical problem, and a number of orthopaedic devices are available for the treatment of such fractures. The objective of this study was to assess the rotational stability, a common failure predictor, of three different rotational control design philosophies: a screw, a helical blade and a deployable crucifix.

Methods

Devices were compared in terms of the mechanical work (W) required to rotate the implant by 6° in a bone substitute material. The substitute material used was Sawbones polyurethane foam of three different densities (0.08 g/cm3, 0.16 g/cm3 and 0.24 g/cm3). Each torsion test comprised a steady ramp of 1°/minute up to an angular displacement of 10°.


Bone & Joint 360
Vol. 5, Issue 6 | Pages 37 - 39
1 Dec 2016


The Bone & Joint Journal
Vol. 99-B, Issue 4 | Pages 458 - 464
1 Apr 2017
Abrahams JM Kim YS Callary SA De Ieso C Costi K Howie DW Solomon LB

Aims

This study aimed to determine the diagnostic performance of radiographic criteria to detect aseptic acetabular loosening after revision total hip arthroplasty (THA). Secondary aims were to determine the predictive values of different thresholds of migration and to determine the predictive values of radiolucency criteria.

Patients and Methods

Acetabular component migration to re-revision was measured retrospectively using Ein-Bild-Rontgen-Analyse (EBRA-Cup) and manual measurements (Sutherland method) in two groups: Group A, 52 components (48 patients) found not loose at re-revision and Group B, 42 components (36 patients) found loose at re-revision between 1980 and 2015. The presence and extent of radiolucent lines was also assessed.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 5 | Pages 665 - 677
1 May 2011
Sköldenberg OG Salemyr MO Bodén HS Lundberg A Ahl TE Adolphson PY

Our aim in this pilot study was to evaluate the fixation of, the bone remodelling around, and the clinical outcome after surgery of a new, uncemented, fully hydroxyapatite-coated, collared and tapered femoral component, designed specifically for elderly patients with a fracture of the femoral neck.

We enrolled 50 patients, of at least 70 years of age, with an acute displaced fracture of the femoral neck in this prospective single-series study. They received a total hip replacement using the new component and were followed up regularly for two years.

Fixation was evaluated by radiostereometric analysis and bone remodelling by dual-energy x-ray absorptiometry. Hip function and the health-related quality of life were assessed using the Harris hip score and the EuroQol-5D.

Up to six weeks post-operatively there was a mean subsidence of 0.2 mm (−2.1 to +0.5) and a retroversion of a mean of 1.2° (−8.2° to +1.5°). No component migrated after three months. The patients had a continuous loss of peri-prosthetic bone which amounted to a mean of 16% (−49% to +10%) at two years. The mean Harris hip score was 82 (51 to 100) after two years.

The two-year results from this pilot study indicate that this new, uncemented femoral component can be used for elderly patients with osteoporotic fractures of the femoral neck.


Bone & Joint 360
Vol. 5, Issue 4 | Pages 20 - 22
1 Aug 2016


The Bone & Joint Journal
Vol. 97-B, Issue 11 | Pages 1445 - 1446
1 Nov 2015
Haddad FS


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 9 | Pages 1143 - 1148
1 Sep 2006
Hallan G Aamodt A Furnes O Skredderstuen A Haugan K Havelin LI

We performed a randomised, radiostereometric study comparing two different bone cements, one of which has been sparsely clinically documented. Randomisation of 60 total hip replacements (57 patients) into two groups of 30 was undertaken. All the patients were operated on using a cemented Charnley total hip replacement, the only difference between groups being the bone cement used to secure the femoral component. The two cements used were Palamed G and Palacos R with gentamicin. The patients were followed up with repeated clinical and radiostereometric examinations for two years to assess the micromovement of the femoral component and the clinical outcome.

The mean subsidence was 0.18 mm and 0.21 mm, and the mean internal rotation was 1.7° and 2.0° at two years for the Palamed G and Palacos R with gentamicin bone cements, respectively. We found no statistically significant differences between the groups. Micromovement occurred between the femoral component and the cement, while the cement mantle was stable inside the bone. The Harris hip score improved from a mean of 38 points (14 to 54) and 36 (10 to 57) pre-operatively to a mean of 92 (77 to 100) and 91 (63 to 100) at two years in the Palamed G and Palacos R groups, respectively. No differences were found between the groups.

Both bone cements provided good initial fixation of the femoral component and good clinical results at two years.


The Bone & Joint Journal
Vol. 98-B, Issue 12 | Pages 1611 - 1619
1 Dec 2016
Wilson MJ Hook S Whitehouse SL Timperley AJ Gie GA

Aims

Femoral impaction bone grafting was first developed in 1987 using morselised cancellous bone graft impacted into the femoral canal in combination with a cemented, tapered, polished stem. We describe the evolution of this technique and instrumentation since that time.

Patients and Methods

Between 1987 and 2005, 705 revision total hip arthroplasties (56 bilateral) were performed with femoral impaction grafting using a cemented femoral stem. All surviving patients were prospectively followed for a mean of 14.7 years (9.8 to 28.3) with no loss to follow-up. By the time of the final review, 404 patients had died.


The Bone & Joint Journal
Vol. 98-B, Issue 5 | Pages 616 - 621
1 May 2016
Teeter MG Thoren J Yuan X McCalden RW MacDonald SJ Lanting BA Naudie DDR

Aims

The purpose of the present study was to examine the long-term fixation of a cemented fixed-bearing polished titanium tibial baseplate (Genesis ll).

Patients and Methods

Patients enrolled in a previous two-year prospective trial (n = 35) were recalled at ten years. Available patients (n = 15) underwent radiostereometric analysis (RSA) imaging in a supine position using a conventional RSA protocol. Migration of the tibial component in all planes was compared between initial and ten-year follow-up. Outcome scores including the Knee Society Score, Western Ontario and McMaster Universities Arthritis Index, 12-item Short Form Health Survey, Forgotten Joint Score, and University of California, Los Angeles Activity Score were recorded.


The Bone & Joint Journal
Vol. 98-B, Issue 7 | Pages 867 - 873
1 Jul 2016
Dalury DF

As the number of younger and more active patients treated with total knee arthroplasty (TKA) continues to increase, consideration of better fixation as a means of improving implant longevity is required. Cemented TKA remains the reference standard with the largest body of evidence and the longest follow-up to support its use. However, cementless TKA, may offer the opportunity of a more bone-sparing procedure with long lasting biological fixation to the bone. We undertook a review of the literature examining advances of cementless TKA and the reported results.

Cite this article: Bone Joint J 2016;98-B:867–73.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 6 | Pages 817 - 823
1 Jun 2011
Solomon LB Callary SA Stevenson AW McGee MA Chehade MJ Howie DW

We investigated the stability of seven Schatzker type II fractures of the lateral tibial plateau treated by subchondral screws and a buttress plate followed by immediate partial weight-bearing. In order to assess the stability of the fracture, weight-bearing inducible displacements of the fracture fragments and their migration over a one-year period were measured by differentially loaded radiostereometric analysis and standard radiostereometric analysis, respectively. The mean inducible craniocaudal fracture fragment displacements measured −0.30 mm (−0.73 to 0.02) at two weeks and 0.00 mm (−0.12 to 0.15) at 52 weeks. All inducible displacements were elastic in nature under all loads at each examination during follow-up. At one year, the mean craniocaudal migration of the fracture fragments was −0.34 mm (−1.64 to 1.51).

Using radiostereometric methods, this case series has shown that in the Schatzker type II fractures investigated, internal fixation with subchondral screws and a buttress plate provided adequate stability to allow immediate post-operative partial weight-bearing, without harmful consequences.


The Bone & Joint Journal
Vol. 96-B, Issue 11_Supple_A | Pages 118 - 121
1 Nov 2014
Lachiewicz PF Watters TS

Metaphyseal bone loss is common with revision total knee replacement (RTKR). Using the Anderson Orthopaedic Research Institute (AORI) classification, type 2-B and type 3 defects usually require large metal blocks, bulk structural allograft or highly porous metal cones. Tibial and femoral trabecular metal metaphyseal cones are a unique solution for large bone defects. These cones substitute for bone loss, improve metaphyseal fixation, help correct malalignment, restore the joint line and may permit use of a shorter stem. The technique for insertion involves sculpturing of the remaining bone with a high speed burr and rasp, followed by press-fit of the cone into the metaphysis. The fixation and osteoconductive properties of the porous cone outer surface allow ingrowth and encourage long-term biological fixation. The revision knee component is then cemented into the porous cone inner surface, which provides superior fixation compared with cementing into native but deficient metaphyseal bone. The advantages of the cone compared with allograft include: technical ease, biological fixation, no resorption, and possibly a lower risk of infection. The disadvantages include: difficult extraction and relatively short-term follow-up. Several studies using cones report promising short-term results for the reconstruction of large bone defects in RTKR.

Cite this article: Bone Joint J 2014;96-B(11 Suppl A):118–21.


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 11 | Pages 1502 - 1507
1 Nov 2006
Lauge-Pedersen H Hägglund G Johnsson R

Percutaneous physiodesis is an established technique for treating mild leg-length discrepancy and problems of expected extreme height. Angular deformities resulting from incomplete physeal arrest have been reported, and little is known about the time interval from percutaneous physiodesis to actual physeal arrest. This procedure was carried out in ten children, six with leg-length discrepancy and four with expected extreme height. Radiostereometric analysis was used to determine the three-dimensional dynamics of growth retardation. Errors of measurement of translation were less than 0.05 mm and of rotation less than 0.06°.

Physeal arrest was obtained in all but one child within 12 weeks after physiodesis and no clinically-relevant angular deformities occurred.

This is a suitable method for following up patients after percutaneous physiodesis. Incomplete physeal arrest can be detected at an early stage and the procedure repeated before corrective osteotomy is required.


The Bone & Joint Journal
Vol. 96-B, Issue 8 | Pages 1077 - 1081
1 Aug 2014
Nuttall D Birch A Haines JF Trail IA

Resurfacing of the humeral head is commonly used within the UK to treat osteoarthritis (OA) of the shoulder. We present the results of a small prospective randomised study of this procedure using the Global CAP prosthesis with two different coatings, Porocoat and DuoFix hydroxyapatite (HA). We followed two groups of ten patients with OA of the shoulder for two years after insertion of the prosthesis with tantalum marker beads, recording pain, Constant–Murley and American Shoulder and Elbow Surgeons (ASES) outcome scores, and using radiostereometric analysis to assess migration. The outcomes were similar to those of other series, with significant reductions in pain (p = 0.003) and an improvement in the Constant (p = 0.001) and ASES scores (p = 0.006). The mean migration of the prosthesis three months post-operatively was 0.78 mm (0.51 to 1.69) and 0.72 mm (0.33 to 1.45) for the Porocoat and DuoFix groups, respectively. Analysis of variance indicated that the rate of migration reached a plateau after three months post-operatively in both groups. At follow-up of two years the mean migration was 1 mm (sd 1 (0.25 to 3.32)); in the Porocoat group and 0.8 mm (sd 0.4 (0.27 to 1.45)) in the DuoFix HA group. Significant migration of the prosthesis was seen in one patient who had received an anterior humeral bone graft. This prosthesis was later revised after 2.7 years.

The addition of a coating of HA to the sintered surface does not improve fixation of this prosthesis.

Cite this article: Bone Joint J 2014;96-B:1077–81.


Bone & Joint 360
Vol. 2, Issue 3 | Pages 41 - 41
1 Jun 2013
Ollivere BJ


Bone & Joint 360
Vol. 3, Issue 4 | Pages 35 - 38
1 Aug 2014
Hammerberg EM


The Bone & Joint Journal
Vol. 97-B, Issue 1 | Pages 35 - 44
1 Jan 2015
Flivik G Kristiansson I Ryd L

We hypothesised that the removal of the subchondral bone plate (SCBP) for cemented acetabular component fixation in total hip arthroplasty (THA) offers advantages over retention by improving the cement-bone interface, without jeopardising implant stability. We have previously published two-year follow-up data of a randomised controlled trial (RCT), in which 50 patients with primary osteoarthritis were randomised to either retention or removal of the SCBP. The mean age of the retention group (n = 25, 13 males) was 70.0 years (sd 6.8). The mean age in the removal group (n = 25, 16 males) was 70.3 years (sd 7.9). Now we have followed up the patients at six (retention group, n = 21; removal group, n = 20) and ten years (retention group: n = 17, removal group: n = 18), administering clinical outcome questionnaires and radiostereometric analysis (RSA), and determining the presence of radiolucent lines (RLLs) on conventional radiographs. RSA demonstrated similar translation and rotation patterns up to six years. Between six and ten years, proximal acetabular component migration and changes of inclination were larger in the retention group, although the mean differences did not reach statistical significance. Differences in migration were driven by two patients in the SCBP retention group with extensive migration versus none in the SCBP removal group. The significant difference (p < 0.001) in the development of radiolucent lines in the retention group, previously observed at two years, increased even further during the course of follow-up (p < 0.001). While recognising SCBP removal is a more demanding technique, we conclude that, wherever possible, the SCBP should be removed to improve the cement–bone interface in order to maximise acetabular component stability and longevity.

Cite this article: Bone Joint J 2015;97-B:35–44.


The Bone & Joint Journal
Vol. 97-B, Issue 2 | Pages 185 - 191
1 Feb 2015
Kendrick BJL Kaptein BL Valstar ER Gill HS Jackson WFM Dodd CAF Price AJ Murray DW

The most common reasons for revision of unicompartmental knee arthroplasty (UKA) are loosening and pain. Cementless components may reduce the revision rate. The aim of this study was to compare the fixation and clinical outcome of cementless and cemented Oxford UKAs.

A total of 43 patients were randomised to receive either a cemented or a cementless Oxford UKA and were followed for two years with radiostereometric analysis (RSA), radiographs aligned with the bone–implant interfaces and clinical scores.

The femoral components migrated significantly during the first year (mean 0.2 mm) but not during the second. There was no significant difference in the extent of migration between cemented and cementless femoral components in either the first or the second year. In the first year the cementless tibial components subsided significantly more than the cemented components (mean 0.28 mm (sd 0.17) vs. 0.09 mm (sd 0.19 mm)). In the second year, although there was a small amount of subsidence (mean 0.05 mm) there was no significant difference (p = 0.92) between cemented and cementless tibial components. There were no femoral radiolucencies. Tibial radiolucencies were narrow (< 1 mm) and were significantly (p = 0.02) less common with cementless (6 of 21) than cemented (13 of 21) components at two years. There were no complete radiolucencies with cementless components, whereas five of 21 (24%) cemented components had complete radiolucencies. The clinical scores at two years were not significantly different (p = 0.20).

As second-year migration is predictive of subsequent loosening, and as radiolucency is suggestive of reduced implant–bone contact, these data suggest that fixation of the cementless components is at least as good as, if not better than, that of cemented devices.

Cite this article: Bone Joint J 2015; 97-B:185–91.