We report a 12- to 15-year implant survival assessment
of a prospective single-surgeon series of Birmingham Hip Resurfacings
(BHRs). The earliest 1000 consecutive BHRs including 288 women (335
hips) and 598 men (665 hips) of all ages and diagnoses with no exclusions
were prospectively followed-up with postal questionnaires, of whom
the first 402 BHRs (350 patients) also had clinical and radiological
review. Mean follow-up was 13.7 years (12.3 to 15.3). In total, 59 patients
(68 hips) died 0.7 to 12.6 years following surgery from unrelated
causes. There were 38 revisions, 0.1 to 13.9 years (median 8.7)
following operation, including 17 femoral failures (1.7%) and seven
each of infections, soft-tissue reactions and other causes. With
revision for any reason as the end-point Kaplan–Meier survival analysis
showed 97.4% (95% confidence interval (CI) 96.9 to 97.9) and 95.8%
(95% CI 95.1 to 96.5) survival at ten and 15 years, respectively.
Radiological assessment showed 11 (3.5%) femoral and 13 (4.1%) acetabular
radiolucencies which were not deemed failures and one radiological
femoral failure (0.3%). Our study shows that the performance of the BHR continues to
be good at 12- to 15-year follow-up. Men have better implant survival
(98.0%; 95% CI 97.4 to 98.6) at 15 years than women (91.5%; 95%
CI 89.8 to 93.2), and women <
60 years (90.5%; 95% CI 88.3 to
92.7) fare worse than others. Hip dysplasia and osteonecrosis are
risk factors for failure. Patients under 50 years with osteoarthritis
fare best (99.4%; 95% CI 98.8 to 100 survival at 15 years), with
no failures in men in this group. Cite this article:
Ceramic-on-metal (CoM) is a relatively new bearing
combination for total hip arthroplasty (THA) with few reported outcomes.
A total of 287 CoM THAs were carried out in 271 patients (mean age
55.6 years (20 to 77), 150 THAs in female patients, 137 in male)
under the care of a single surgeon between October 2007 and October
2009. With the issues surrounding metal-on-metal bearings the decision
was taken to review these patients between March and November 2011,
at a mean follow-up of 34 months (23 to 45) and to record pain,
outcome scores, radiological analysis and blood ion levels. The
mean Oxford Hip Score was 19.2 (12 to 53), 254 patients with 268
hips (95%) had mild/very mild/no pain, the mean angle of inclination
of the acetabular component was 44.8o (28o to
63o), 82 stems (29%) had evidence of radiolucent lines
of >
1 mm in at least one Gruen zone and the median levels of cobalt
and chromium ions in the blood were 0.83 μg/L (0.24 μg/L to 27.56 μg/L)
and 0.78 μg/L (0.21 μg/L to 8.84 μg/L), respectively. The five-year
survival rate is 96.9% (95% confidence interval 94.7% to 99%). Due to the presence of radiolucent lines and the higher than
expected levels of metal ions in the blood, we would not recommend
the use of CoM THA without further long-term follow-up. We plan
to monitor all these patients regularly. Cite this article:
PCL retaining fixed-bearing TKA is a highly successful
operation with the need for more surgery occurring at the rate of
approximately 0.4% per year over the first 27 years. The most common
cause for revision surgery is related to polyethylene insert failure
and accounts for approximately 50% of re-operations. Late metastatic
infection is the next most frequent cause followed by patellar problems,
late instability and component loosening in decreasing frequency.
A myriad of rare miscellaneous problems can also occur.
We carried out a prospective investigation into
the radiological outcomes of uncemented Oxford medial compartment
unicondylar replacement in 220 consecutive patients (231 knees)
performed in a single centre with a minimum two-year follow-up.
The functional outcomes using the mean Oxford knee score and the
mean high-activity arthroplasty score were significantly improved
over the pre-operative scores (p <
0.001). There were 196 patients
with a two-year radiological examination performed under fluoroscopic
guidance, aiming to provide images acceptable for analysis of the
bone–implant interface. Of the six tibial zones examined on each
knee on the anteroposterior radiograph, only three had a partial
radiolucent line. All were in the medial aspect of the tibial base plate
(zone 1) and all measured <
1 mm. All of these patients were
asymptomatic. There were no radiolucent lines seen around the femoral
component or on the lateral view. There was one revision for loosening
at one year due to initial inadequate seating of the tibial component.
These results confirm that the early uncemented Oxford medial unicompartmental
compartmental knee replacements were reliable and the incidence
of radiolucent lines was significantly decreased compared with the
reported results of cemented versions of this implant. These independent
results confirm those of the designing centre.
The primary purpose of this study of metal-on-metal
(MoM) hip resurfacing was to compare the effect of using a cementless
or cemented femoral component on the subsequent bone mineral density
(BMD) of the femoral neck. This was a single-centre, prospective, double-blinded control
trial which randomised 120 patients (105 men and 15 women) with
a mean age of 49.4 years (21 to 68) to receive either a cemented
or cementless femoral component. Follow-up was to two years. Outcome
measures included total and six-point region-of-interest BMD of
the femoral neck, radiological measurements of acetabular inclination,
neck-shaft and stem-shaft angles, and functional outcome scores
including the Harris hip score, the Western Ontario and McMaster
Universities Osteoarthritis Index and the University of California
at Los Angeles activity scale. In total, 17 patients were lost to follow-up leaving 103 patients
at two years. There were no revisions in the cementless group and
three revisions (5%) in the cemented group (two because of hip pain
and one for pseudotumour). The total BMD was significantly higher in the cementless group
at six months (p <
0.001) and one year (p = 0.01) than in the
cemented group, although there was a loss of statistical significance
in the difference at two years (p = 0.155). All patient outcomes improved significantly: there were no significant
differences between the two groups. The results show better preservation of femoral neck BMD with
a cementless femoral component after two years of follow-up. Further
investigation is needed to establish whether this translates into
improved survivorship. Cite this article:
This prospective study reports the 15-year survival and ten-year
functional outcome of a consecutive series of 1000 minimally invasive
Phase 3 Oxford medial UKAs (818 patients, 393 men, 48%, 425 women,
52%, mean age 66 years; 32 to 88). These were implanted by two surgeons
involved with the design of the prosthesis to treat anteromedial
osteoarthritis and spontaneous osteonecrosis of the knee, which
are recommended indications. Patients were prospectively identified
and followed up independently for a mean of 10.3 years (5.3 to 16.6). At ten years, the mean Oxford Knee Score was 40 (standard deviation
( This is the only large series of minimally invasive UKAs with
15-year survival data. The results support the continued use of
minimally invasive UKA for the recommended indications. Cite this article:
The Oxford unicompartmental knee replacement gives good results in patients with symptomatic osteoarthritis of the medial compartment. Previous studies have suggested that the presence of radiolucent lines (RLLs) does not reflect a poor outcome in such patients. However, the reliability and validity of this assessment have not been determined. Our aim was to assess the intra- and interobserver reliability and the sensitivity and specificity of the assessment of RLLs around both tibial and femoral components using standard radiographs. Two reviewers assessed the radiographs of 45 patients who had loosening of the tibial or femoral component confirmed at revision surgery and compared them with those of a series of 45 asymptomatic patients matched for age and gender. The results suggested that, using standard radiographs, tibial RLLs were 63.6% sensitive and 94.4% specific and femoral RLLs 63.9% sensitive and 72.7% specific for loosening. Overall intra- and interobserver reliability was highly variable, but zonal analysis showed that lucency at the tip of the femoral peg was significantly associated with loosening of the femoral component. Fluoroscopically guided radiographs may improve the zonal reliability of the assessment of RLLs, but further independent and comparative studies are required. In the meantime, the innocence of the physiological RLLs detected by standard radiographs should be viewed with caution.
Total hip replacement (THR) still is a rare intervention
in many African countries. In Burkina Faso it is not performed on
a regular basis. A visiting programme for THR was started in a district
hospital with no previous relevant experience. In this paper we
present an analysis of the surgical technical problems and peri-operative
complications of 152 THRs in 136 patients and three bipolar hemiarthroplasties
in three patients undertaken in this new programme with limited
orthopaedic equipment. There were 86 male and 53 female patients
with a mean age of 49 years (21 to 78). We identified 77 intra-operative
technical problems in 51 operations. There were 24 peri-operative complications
in 21 patients, 17 of which were bony in nature. So far, ten revision
THRs have been performed in nine patients. Regular analysis of the technical problems and complications
was used to improve quality, and we identified patient selection
adapted to the local circumstances as important to avoid complications.
Our reflections on the problems encountered in initiating such a
programme may be of help to other teams planning similar projects. Cite this article:
Persistent groin pain after seemingly successful
total hip replacement (THR) appears to have become more common.
Recent studies have indicated a high incidence after metal-on-polyethylene
and metal-on-metal conventional THR and it has been documented in
up to 18% of patients after metal-on-metal resurfacing. There are many
causes, including acetabular loosening, stress fracture, and iliopsoas
tendonitis and impingement. The evaluation of this problem requires
a careful history and examination, plain radiographs and an algorithmic approach
to special diagnostic imaging and tests. Non-operative treatment
is not usually successful. Specific operative treatment depending
on the cause of the pain usually involves revision of the acetabular
component, iliopsoas tenotomy or other procedures, and is usually
successful. Here, an appropriate algorithm is described.
The strain on clinic and surgeon resources resulting
from a rise in demand for total knee replacement (TKR) requires reconsideration
of when and how often patients need to be seen for follow-up. Surgeons
will otherwise require increased paramedical staff or need to limit
the number of TKRs they undertake. We reviewed the outcome data
of 16 414 primary TKRs undertaken at our centre to determine the
time to re-operation for any reason and for specific failure mechanisms.
Peak risk years for failure were determined by comparing the conditional
probability of failure, the number of failures divided by the total
number of TKRs cases, for each year. The median times to failure
for the most common failure mechanisms were 4.9 years (interquartile
range (IQR) 1.7 to 10.7) for femoral and tibial loosening, 1.9 years
(IQR 0.8 to 3.9) for infection, 3.1 years (IQR 1.6 to 5.5) for tibial
collapse and 5.6 years (IQR 3.4 to 9.3) for instability. The median
time to failure for all revisions was 3.3 years (IQR 1.2 to 8.5),
with an overall revision rate of 1.7% (n = 282). Results from our
patient population suggest that patients be seen for follow-up at
six months, one year, three years, eight years, 12 years, and every
five years thereafter. Patients with higher pain in the early post-operative
period or high body mass index (≥ 41 kg/m2) should be
monitored more closely. Cite this article:
We evaluated the clinical results and complications
after extra-articular resection of the distal femur and/or proximal
tibia and reconstruction with a tumour endoprosthesis (MUTARS) in
59 patients (mean age 33 years (11 to 74)) with malignant bone or
soft-tissue tumours. According to a Kaplan–Meier analysis, limb
survival was 76% (95% confidence interval (CI) 64.1 to 88.5) after
a mean follow-up of 4.7 years (one month to 17 years). Peri-prosthetic infection
was the most common indication for subsequent amputation (eight
patients). Survival of the prosthesis without revision was 48% (95%
CI 34.8 to 62.0) at two years and 25% (95% CI 11.1 to 39.9) at five years
post-operatively. Failure of the prosthesis was due to deep infection
in 22 patients (37%), aseptic loosening in ten patients (17%), and
peri-prosthetic fracture in six patients (10%). Wear of the bearings
made a minor revision necessary in 12 patients (20%). The mean Musculoskeletal
Tumor Society score was 23 (10 to 29). An extensor lag >
10° was
noted in ten patients (17%). These results suggest that limb salvage after extra-articular
resection with a tumour prosthesis can achieve good functional results
in most patients, although the rates of complications and subsequent
amputation are higher than in patients treated with intra-articular
resection. Cite this article:
The term developmental dysplasia of the hip (DDH)
describes a spectrum of disorders that results in abnormal development
of the hip joint. If not treated successfully in childhood, these
patients may go on to develop hip symptoms and/or secondary osteoarthritis
in adulthood. In this review we describe the altered anatomy encountered
in adults with DDH along with the management options, and the challenges
associated with hip arthroscopy, osteotomies and arthroplasty for
the treatment of DDH in young adults. Cite this article:
The purpose of this study was twofold: first,
to determine whether the five-year results of hip resurfacing arthroplasty
(HRA) in Canada justified the continued use of HRA; and second,
to identify whether greater refinement of patient selection was
warranted. This was a retrospective cohort study that involved a review
of 2773 HRAs performed between January 2001 and December 2008 at
11 Canadian centres. Cox’s proportional hazards models were used
to analyse the predictors of failure of HRA. Kaplan–Meier survival
analysis was performed to predict the cumulative survival rate at
five years. The factors analysed included age, gender, body mass
index, pre-operative hip pathology, surgeon’s experience, surgical
approach, implant sizes and implant types. The most common modes
of failure were also analysed. The 2773 HRAs were undertaken in 2450 patients: 2127 in men and
646 in women. The mean age at operation was 50.5 years ( The failure rates of HRA at five years justify the ongoing use
of this technique in men. Female gender is an independent predictor
of failure, and a higher failure rate at five years in women leads
the authors to recommend this technique only in exceptional circumstances
for women. Cite this article:
A total of 219 hips in 192 patients aged between
18 and 65 years were randomised to 28-mm metal-on-metal uncemented
total hip replacements (THRs, 107 hips) or hybrid hip resurfacing
(HR, 112 hips). At a mean follow-up of eight years (6.6 to 9.3)
there was no significant difference between the THR and HR groups
regarding rate of revision (4.0% (4 of 99) Cite this article:
Despite the increasing interest and subsequent published literature on hip resurfacing arthroplasty, little is known about the prevalence of its complications and in particular the less common modes of failure. The aim of this study was to identify the prevalence of failure of hip resurfacing arthroplasty and to analyse the reasons for it. From a multi-surgeon series (141 surgeons) of 5000 Birmingham hip resurfacings we have analysed the modes, prevalence, gender differences and times to failure of any hip requiring revision. To date 182 hips have been revised (3.6%). The most common cause for revision was a fracture of the neck of the femur (54 hips, prevalence 1.1%), followed by loosening of the acetabular component (32 hips, 0.6%), collapse of the femoral head/avascular necrosis (30 hips, 0.6%), loosening of the femoral component (19 hips, 0.4%), infection (17 hips, 0.3%), pain with aseptic lymphocytic vascular and associated lesions (ALVAL)/metallosis (15 hips, 0.3%), loosening of both components (five hips, 0.1%), dislocation (five hips, 0.1%) and malposition of the acetabular component (three hips, 0.1%). In two cases the cause of failure was unknown. Comparing men with women, we found the prevalence of revision to be significantly higher in women (women = 5.7%; men = 2.6%, p <
0.001). When analysing the individual modes of failure women had significantly more revisions for loosening of the acetabular component, dislocation, infection and pain/ALVAL/metallosis (p <
0.001, p = 0.004, p = 0.008, p = 0.01 respectively). The mean time to failure was 2.9 years (0.003 to 11.0) for all causes, with revision for fracture of the neck of the femur occurring earlier than other causes (mean 1.5 years, 0.02 to 11.0). There was a significantly shorter time to failure in men (mean 2.1 years, 0.4 to 8.7) compared with women (mean 3.6 years, 0.003 to 11.0) (p <
0.001).
We have reviewed the rate of revision of fully cemented, hybrid and uncemented primary total hip replacements (THRs) registered in the New Zealand Joint Registry between 1999 and December 2006 to determine whether there was any statistically significant difference in the early survival and reason for revision in these different types of fixation. The percentage rate of revision was calculated per 100 component years and compared with the reason for revision, the type of fixation and the age of the patients. Of the 42 665 primary THRs registered, 920 (2.16%) underwent revision requiring change of at least one component. Fully-cemented THRs had a lower rate of revision when considering all causes for failure (p <
0.001), but below the age of 65 years uncemented THRs had a lower rate (p <
0.01). The rate of revision of the acetabular component for aseptic loosening was less in the uncemented and hybrid groups compared with that in the fully cemented group (p <
0.001), and the rate of revision of cemented and uncemented femoral components was similar, except in patients over 75 years of age in whom revision of cemented femoral components was significantly less frequent (p <
0.02). Revision for infection was more common in patients aged below 65 years and in cemented and hybrid THRs compared with cementless THRs (p <
0.001). Dislocation was the most common cause of revision for all types of fixation and was more frequent in both uncemented acetabular groups (p <
0.001). The experience of the surgeon did not affect the findings. Although cemented THR had the lowest rate of revision for all causes in the short term (90 days), uncemented THR had the lowest rate of aseptic loosening in patients under 65 years of age and had rates comparable with international rates of aseptic loosening in those over 65 years.
There are no recent studies comparing cable with
wire for the fixation of osteotomies or fractures in total hip replacement
(THR). Our objective was to evaluate the five-year clinical and
radiological outcomes and complication rates of the two techniques.
We undertook a review including all primary and revision THRs performed
in one hospital between 1996 and 2005 using cable or wire fixation.
Clinical and radiological evaluation was performed five years post-operatively.
Cables were used in 51 THRs and wires in 126, and of these, 36 THRs
with cable (71%) and 101 with wire (80%) were evaluated at follow-up.
The five-year radiographs available for 33 cable and 91 wire THRs
revealed rates of breakage of fixation of 12 of 33 (36%) and 42
of 91 (46%), respectively. With cable there was a significantly
higher risk of metal debris (68% In conclusion, we found a higher incidence of complications and
a trend towards increased infection and foreign-body reaction with
the use of cables.
We carried out a prospective study to assess the clinical outcome, complications and survival of the NexGen Legacy posterior-stabilised-Flex total knee replacement (TKR) in a consecutive series of 278 knees between May 2003 and February 2005. Mean follow-up for 259 TKRs (98.2%) was 3.8 years (3.0 to 4.8). Annual follow-up showed improvement in the Knee Society scores (paired This relatively large study indicates that the legacy posterior stabilised-Flex design provides excellent short-term outcome but warrants ongoing evaluation to confirm the long-term durability and functioning of the implant.
Down’s syndrome is associated with a number of
musculoskeletal abnormalities, some of which predispose patients
to early symptomatic arthritis of the hip. The purpose of the present
study was to review the general and hip-specific factors potentially
compromising total hip replacement (THR) in patients with Down’s
syndrome, as well as to summarise both the surgical techniques that
may anticipate the potential adverse impact of these factors and
the clinical results reported to date. A search of the literature
was performed, and the findings further informed by the authors’
clinical experience, as well as that of the hip replacement in Down
Syndrome study group. The general factors identified include a high
incidence of ligamentous laxity, as well as associated muscle hypotonia
and gait abnormalities. Hip-specific factors include: a high incidence
of hip dysplasia, as well as a number of other acetabular, femoral
and combined femoroacetabular anatomical variations. Four studies
encompassing 42 hips, which reported the clinical outcomes of THR
in patients with Down’s syndrome, were identified. All patients
were successfully treated with standard acetabular and femoral components.
The use of supplementary acetabular screw fixation to enhance component
stability was frequently reported. The use of constrained liners
to treat intra-operative instability occurred in eight hips. Survival
rates of between 81% and 100% at a mean follow-up of 105 months
(6 to 292) are encouraging. Overall, while THR in patients with
Down’s syndrome does present some unique challenges, the overall
clinical results are good, providing these patients with reliable
pain relief and good function. Cite this article:
When performing total hip replacement (THR) in high dislocated hips, the presence of soft-tissue contractures means that most surgeons prefer to use a femoral shortening osteotomy in order to avoid the risk of neurovascular damage. However, this technique will sacrifice femoral length and reduce the extent of any leg-length equalisation. We report our experience of 74 THRs performed between 2000 and 2008 in 65 patients with a high dislocated hip without a femoral shortening osteotomy. The mean age of the patients was 55 years (46 to 72) and the mean follow-up was 42 months (12 to 78). All implants were cementless except for one resurfacing hip implant. We attempted to place the acetabular component in the anatomical position in each hip. The mean Harris hip score improved from 53 points (34 to 74) pre-operatively to 86 points (78 to 95) at final follow-up. The mean radiologically determined leg lengthening was 42 mm (30 to 66), and the mean leg-length discrepancy decreased from 36 mm (5 to 56) pre-operatively to 8.5 mm (0 to 18) postoperatively. Although there were four (5%) post-operative femoral nerve palsies, three had fully resolved by six months after the operation. No loosening of the implant was observed, and no dislocations or infections were encountered. Total hip replacement without a femoral shortening osteotomy proved to be a safe and effective surgical treatment for high dislocated hips.