Advertisement for orthosearch.org.uk
Results 81 - 100 of 349
Results per page:
The Bone & Joint Journal
Vol. 99-B, Issue 4_Supple_B | Pages 33 - 40
1 Apr 2017
Galea VP Laaksonen I Matuszak SJ Connelly JW Muratoglu O Malchau H

Aims. Our first aim was to determine whether there are significant changes in the level of metal ions in the blood at mid-term follow-up, in patients with an Articular Surface Replacement (ASR) arthroplasty. Secondly, we sought to identify risk factors for any increases. Patients and Methods. The study involved 435 patients who underwent unilateral, metal-on-metal (MoM) hip resurfacing (HRA) or total hip arthroplasty (THA). These patients all had one measurement of the level of metal ions in the blood before seven years had passed post-operatively (early evaluation) and one after seven years had passed post-operatively (mid-term evaluation). Changes in ion levels were tested using a Wilcoxon signed-rank test. We identified subgroups at the highest risk of increase using a multivariable linear logistic regression model. Results. There were significant increases in the levels of metal ions for patients who underwent both MoM HRA (Chromium (Cr): 0.5 parts per billion (ppb); Cobalt (Co): 1.1 ppb) and MoM THA (Cr: 0.5 ppb; Co: 0.7 ppb). In a multivariable model considering MoM HRAs, the change in the levels of metal ions was influenced by female gender (Co: Odds Ratio (OR) 1.42; p = 0.002 and Cr: OR 1.08; p = 0.006). The change was found to be irrespective of the initial level for the MoM HRAs, whereas there was a negative relationship between the initial level and the change in the level for those with a MoM THA (Co: OR -0.43; p <  0.001 and Cr: OR -0.14; p = 0.033). Conclusion. The levels of metal ions in the blood increase significantly over the period until mid-term follow-up in patients with both a MoM HRA and those with a MoM THA. We recommend that the levels of metal ions be measured most frequently for women with a MoM HRA. While those with a MoM THA appear to stabilise at a certain level, the accuracy of this trend is not yet clear. Vigilant follow-up is still recommended. . Cite this article: Bone Joint J 2017;99-B(4 Supple B):33–40


Bone & Joint 360
Vol. 11, Issue 4 | Pages 38 - 40
1 Aug 2022


The Journal of Bone & Joint Surgery British Volume
Vol. 63-B, Issue 3 | Pages 435 - 440
1 Aug 1981
Rae T

Some of the component metals of the alloys used for total joint prostheses are toxic and dissolve in the body fluids. It is important to establish how toxic these metals are and to assess the risk of localised tissue necrosis around the prostheses. This has been investigated by incubating primary monolayer cultures of human synovial fibroblasts with various preparations of metals for periods up to 18 days. Morphological changes were evident after exposure to cobalt chloride at a concentration of 50 nanomoles per millilitre and to nickel chloride at 200 nanomoles per millilitre. Chromic chloride, ammonium molybdate and ferric chloride produced no changes up to 500 nanomoles per millilitre. Cultures exposed to particulate pure metals were poisoned by cobalt and vanadium but were not affected under the same conditions by nickel, chromium, molybdenum, titanium or aluminium. Particulate cobalt and vanadium were probably toxic due to their relatively high solubility (four and one micromoles per millilitre respectively after seven days incubation). Particulate nickel also dissolved (three nanomoles per millilitre after seven days) but not in sufficient quantities to be toxic. It appears, therefore, that potentially the most harmful components are cobalt from cobalt-chromium alloy, nickel from stainless steel, and vanadium from titanium alloy. As far as can be estimated, the only combination of materials which is likely to give rise to toxic levels of metal under clinical conditions, is cobalt-chromium alloy articulating against itself to produce relatively high levels of cobalt


Bone & Joint 360
Vol. 13, Issue 3 | Pages 35 - 36
3 Jun 2024

The June 2024 Spine Roundup360 looks at: Intraoperative navigation increases the projected lifetime cancer risk in patients undergoing surgery for adolescent idiopathic scoliosis; Intrawound vancomycin powder reduces delayed deep surgical site infections following posterior spinal fusion for adolescent idiopathic scoliosis; Characterizing negative online reviews of spine surgeons; Proximal junctional failure after surgical instrumentation in adult spinal deformity: biomechanical assessment of proximal instrumentation stiffness; Nutritional supplementation and wound healing: a randomized controlled trial.


Bone & Joint 360
Vol. 13, Issue 1 | Pages 16 - 18
1 Feb 2024

The February 2024 Knee Roundup360 looks at: Do patients with hypoallergenic total knee arthroplasty implants for metal allergy do worse? An analysis of healthcare utilizations and patient-reported outcome measures; Defining a successful total knee arthroplasty; Incidence, microbiological studies, and factors associated with periprosthetic joint infection after total knee arthroplasty; A modified Delphi consensus statement on patellar instability; Cause for concern? Significant cement coverage in retrieved metaphyseal cones after revision total knee arthroplasty; Prevalence of post-traumatic osteoarthritis after anterior cruciate ligament injury remains high despite advances in surgical techniques; Cost-effectiveness of arthroscopic partial meniscectomy versus physical therapy for traumatic meniscal tears in patients aged under 45 years.


Bone & Joint 360
Vol. 12, Issue 6 | Pages 49 - 51
1 Dec 2023
Burden EG Whitehouse MR Evans JT


The Bone & Joint Journal
Vol. 106-B, Issue 1 | Pages 11 - 15
1 Jan 2024
Jain S Lamb JN Pandit H

Polished taper-slip (PTS) cemented stems have an excellent clinical track record and are the most common stem type used in primary total hip arthroplasty (THA) in the UK. Due to low rates of aseptic loosening, they have largely replaced more traditional composite beam (CB) cemented stems. However, there is now emerging evidence from multiple joint registries that PTS stems are associated with higher rates of postoperative periprosthetic femoral fracture (PFF) compared to their CB stem counterparts. The risk of both intraoperative and postoperative PFF remains greater with uncemented stems compared to either of these cemented stem subtypes. PFF continues to be a devastating complication following primary THA and is associated with high complication and mortality rates. Recent efforts have focused on identifying implant-related risk factors for PFF in order to guide preventative strategies, and therefore the purpose of this article is to present the current evidence on the effect of cemented femoral stem design on the risk of PFF.

Cite this article: Bone Joint J 2024;106-B(1):11–15.


Bone & Joint 360
Vol. 11, Issue 3 | Pages 17 - 20
1 Jun 2022


Bone & Joint Open
Vol. 5, Issue 4 | Pages 286 - 293
9 Apr 2024
Upadhyay PK Kumar V Mirza SB Shah N

Aims

This study reports the results of 38 total hip arthroplasties (THAs) in 33 patients aged less than 50 years, using the JRI Furlong hydroxyapatite ceramic (HAC)-coated femoral component.

Methods

We describe the survival, radiological, and functional outcomes of 33 patients (38 THAs) at a mean follow-up of 27 years (25 to 32) between 1988 and 2018.


Bone & Joint Open
Vol. 4, Issue 11 | Pages 839 - 845
6 Nov 2023
Callary SA Sharma DK D’Apollonio TM Campbell DG

Aims

Radiostereometric analysis (RSA) is the most accurate radiological method to measure in vivo wear of highly cross-linked polyethylene (XLPE) acetabular components. We have previously reported very low wear rates for a sequentially irradiated and annealed X3 XLPE liner (Stryker Orthopaedics, USA) when used in conjunction with a 32 mm femoral heads at ten-year follow-up. Only two studies have reported the long-term wear rate of X3 liners used in conjunction with larger heads using plain radiographs which have poor sensitivity. The aim of this study was to measure the ten-year wear of thin X3 XLPE liners against larger 36 or 40 mm articulations with RSA.

Methods

We prospectively reviewed 19 patients who underwent primary cementless THA with the XLPE acetabular liner (X3) and a 36 or 40 mm femoral head with a resultant liner thickness of at least 5.8 mm. RSA radiographs at one week, six months, and one, two, five, and ten years postoperatively and femoral head penetration within the acetabular component were measured with UmRSA software. Of the initial 19 patients, 12 were available at the ten-year time point.


Aims

This study investigated vancomycin-microbubbles (Vm-MBs) and meropenem (Mp)-MBs with ultrasound-targeted microbubble destruction (UTMD) to disrupt biofilms and improve bactericidal efficiency, providing a new and promising strategy for the treatment of device-related infections (DRIs).

Methods

A film hydration method was used to prepare Vm-MBs and Mp-MBs and examine their characterization. Biofilms of methicillin-resistant Staphylococcus aureus (MRSA) and Escherichia coli were treated with different groups. Biofilm biomass differences were determined by staining. Thickness and bacterial viability were observed with confocal laser scanning microscope (CLSM). Colony counts were determined by plate-counting. Scanning electron microscopy (SEM) observed bacterial morphology.


The Bone & Joint Journal
Vol. 99-B, Issue 10 | Pages 1298 - 1303
1 Oct 2017
Schouten R Malone AA Frampton CM Tiffen C Hooper G

Aims . The primary aim of this independent prospective randomised trial was to compare serum metal ion levels for ceramic-on-metal (CoM) and metal-on-metal (MoM) bearing surfaces in total hip arthroplasty (THA). Our one-year results demonstrated elevation in metal ion levels above baseline with no significant difference between the CoM and MoM groups. This paper reviews the five-year data. Patients and Methods. The implants used in each patient differed only in respect to the type of femoral head (ceramic or metal). At five-year follow-up of the 83 enrolled patients, data from 67 (36 CoM, 31 MoM) was available for comparison. Results. The mean serum cobalt (Co) and chromium (Cr) ion levels remained above baseline in both groups (CoM: Co 1.16 μg/l (0.41 to 14.67), Cr 1.05 μg/l (0.16 to 12.58); MoM: Co 2.93 μg/l (0.35 to 30.29), Cr 1.85 μg/l (0.36 to 17.00)) but the increase was significantly less in the CoM cohort (Co difference p = 0.001, Cr difference p = 0.002). These medium-term results, coupled with lower revision rates from national joint registries, suggest that the performance of CoM THA may be superior to that of MoM. . Conclusion. While both bearing combinations have since been withdrawn these results provide useful information for planning clinical surveillance of CoM THAs and warrants continued monitoring. Cite this article: Bone Joint J 2017;99-B:1298–1303


The Bone & Joint Journal
Vol. 104-B, Issue 3 | Pages 309 - 310
1 Mar 2022
Haddad FS


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 10 | Pages 1291 - 1297
1 Oct 2008
De Haan R Pattyn C Gill HS Murray DW Campbell PA De Smet K

We examined the relationships between the serum levels of chromium and cobalt ions and the inclination angle of the acetabular component and the level of activity in 214 patients implanted with a metal-on-metal resurfacing hip replacement. Each patient had a single resurfacing and no other metal in their body. All serum measurements were performed at a minimum of one year after operation. The inclination of the acetabular component was considered to be steep if the abduction angle was greater than 55°. There were significantly higher levels of metal ions in patients with steeply-inclined components (p = 0.002 for chromium, p = 0.003 for cobalt), but no correlation was found between the level of activity and the concentration of metal ions. A highly significant (p < 0.001) correlation with the arc of cover was found. Arcs of cover of less than 10 mm were correlated with a greater risk of high concentrations of serum metal ions. The arc of coverage was also related to the design of the component and to size as well as to the abduction angle of the acetabular component. Steeply-inclined acetabular components, with abduction angles greater than 55°, combined with a small size of component are likely to give rise to higher serum levels of cobalt and chromium ions. This is probably due to a greater risk of edge-loading


The Bone & Joint Journal
Vol. 106-B, Issue 6 | Pages 555 - 564
1 Jun 2024
Leal J Holland CT Cochrane NH Seyler TM Jiranek WA Wellman SS Bolognesi MP Ryan SP

Aims

This study aims to assess the relationship between history of pseudotumour formation secondary to metal-on-metal (MoM) implants and periprosthetic joint infection (PJI) rate, as well as establish ESR and CRP thresholds that are suggestive of infection in these patients. We hypothesized that patients with a pseudotumour were at increased risk of infection.

Methods

A total of 1,171 total hip arthroplasty (THA) patients with MoM articulations from August 2000 to March 2014 were retrospectively identified. Of those, 328 patients underwent metal artefact reduction sequence MRI and had minimum two years’ clinical follow-up, and met our inclusion criteria. Data collected included demographic details, surgical indication, laterality, implants used, history of pseudotumour, and their corresponding preoperative ESR (mm/hr) and CRP (mg/dl) levels. Multivariate logistic regression modelling was used to evaluate PJI and history of pseudotumour, and receiver operating characteristic curves were created to assess the diagnostic capabilities of ESR and CRP to determine the presence of infection in patients undergoing revision surgery.


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 1 | Pages 38 - 46
1 Jan 2010
Langton DJ Jameson SS Joyce TJ Hallab NJ Natu S Nargol AVF

Early failure associated with adverse reactions to metal debris is an emerging problem after hip resurfacing but the exact mechanism is unclear. We analysed our entire series of 660 metal-on-metal resurfacings (Articular Surface Replacement (ASR) and Birmingham Hip Resurfacing (BHR)) and large-bearing ASR total hip replacements, to establish associations with metal debris-related failures. Clinical and radiological outcomes, metal ion levels, explant studies and lymphocyte transformation tests were performed. A total of 17 patients (3.4%) were identified (all ASR bearings) with adverse reactions to metal debris, for which revision was required. This group had significantly smaller components, significantly higher acetabular component anteversion, and significantly higher whole concentrations of blood and joint chromium and cobalt ions than asymptomatic patients did (all p < 0.001). Post-revision lymphocyte transformation tests on this group showed no reactivity to chromium or cobalt ions. Explants from these revisions had greater surface wear than retrievals for uncomplicated fractures. The absence of adverse reactions to metal debris in patients with well-positioned implants usually implies high component wear. Surgeons must consider implant design, expected component size and acetabular component positioning in order to reduce early failures when performing large-bearing metal-on-metal hip resurfacing and replacement


Bone & Joint 360
Vol. 11, Issue 3 | Pages 14 - 17
1 Jun 2022


The Bone & Joint Journal
Vol. 104-B, Issue 3 | Pages 359 - 367
1 Mar 2022
Deere K Matharu GS Ben-Shlomo Y Wilkinson JM Blom AW Sayers A Whitehouse MR

Aims

A recent report from France suggested an association between the use of cobalt-chrome (CoCr) femoral heads in total hip arthroplasties (THAs) and an increased risk of dilated cardiomyopathy and heart failure. CoCr is a commonly used material in orthopaedic implants. If the reported association is causal, the consequences would be significant given the millions of joint arthroplasties and other orthopaedic procedures in which CoCr is used annually. We examined whether CoCr-containing THAs were associated with an increased risk of all-cause mortality, heart outcomes, cancer, and neurodegenerative disorders in a large national database.

Methods

Data from the National Joint Registry was linked to NHS English hospital inpatient episodes for 374,359 primary THAs with up to 14.5 years' follow-up. We excluded any patients with bilateral THAs, knee arthroplasties, indications other than osteoarthritis, aged under 55 years, and diagnosis of one or more outcome of interest before THA. Implants were grouped as either containing CoCr or not containing CoCr. The association between implant construct and the risk of all-cause mortality and incident heart failure, cancer, and neurodegenerative disorders was examined.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 7 | Pages 895 - 900
1 Jul 2012
Gill IPS Webb J Sloan K Beaver RJ

We present a series of 35 patients (19 men and 16 women) with a mean age of 64 years (36.7 to 75.9), who underwent total hip replacement using the ESKA dual-modular short stem with metal on-polyethylene bearing surfaces. This implant has a modular neck section in addition to the modular head. Of these patients, three presented with increasing post-operative pain due to pseudotumour formation that resulted from corrosion at the modular neck-stem junction. These patients underwent further surgery and aseptic lymphocytic vaculitis associated lesions were demonstrated on histological analysis. Retrieval analysis of two modular necks showed corrosion at the neck-stem taper. Blood cobalt and chromium levels were measured at a mean of nine months (3 to 28) following surgery. These were compared with the levels in seven control patients (three men and four women) with a mean age of 53.4 years (32.1 to 64.1), who had an identical prosthesis and articulation but with a prosthesis that had no modularity at neck-stem junction. The mean blood levels of cobalt in the study group were raised at 50.75 nmol/l (5 to 145) compared with 5.6 nmol/l (2 to 13) in control patients. Corrosion at neck-stem tapers has been identified as an important source of metal ion release and pseudotumour formation requiring revision surgery. Finite element modelling of the dual modular stem demonstrated high stresses at the modular stem-neck junction. Dual modular cobalt-chrome hip prostheses should be used with caution due to these concerns


The Bone & Joint Journal
Vol. 104-B, Issue 5 | Pages 598 - 603
1 May 2022
Siljander MP Gausden EB Wooster BM Karczewski D Sierra RJ Trousdale RT Abdel MP

Aims

The aim of this study was to evaluate the incidence of liner malseating in two commonly used dual-mobility (DM) designs. Secondary aims included determining the risk of dislocation, survival, and clinical outcomes.

Methods

We retrospectively identified 256 primary total hip arthroplasties (THAs) that included a DM component (144 Stryker MDM and 112 Zimmer-Biomet G7) in 233 patients, performed between January 2012 and December 2019. Postoperative radiographs were reviewed independently for malseating of the liner by five reviewers. The mean age of the patients at the time of THA was 66 years (18 to 93), 166 (65%) were female, and the mean BMI was 30 kg/m2 (17 to 57). The mean follow-up was 3.5 years (2.0 to 9.2).