We describe three cases of fracture of the titanium JRI-Furlong hydroxyapatite-ceramic (HAC)-coated femoral component. We have examined previous case reports of failure of this stem and conclude that fracture may occur in two places, namely at the neck-shoulder junction and at the conical-distal cylindrical junction. These breakages are the result of fatigue in a metallurgically-proven normal femoral component. All the cases of failure of the femoral component have occurred in patients with a body mass index of more than 25 in whom a small component, either size 9 or 10, had been used.

In patients with a body mass index above normal size 9 components should be avoided and the femoral canal should be reamed sufficiently to accept a large femoral component to ensure that there is adequate metaphyseal fixation.


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 3 | Pages 364 - 370
1 Mar 2008
Kubo T Sugita T Shimose S Matsuo T Arihiro K Ochi M

We investigated the use of hypoxia-inducible factor (HIF) proteins as prognostic markers in chondrosarcoma and the relationship of HIF to the biological characteristics of cartilage tumours. The expression of HIF-1α, HIF-2α, proliferating cell nuclear antigen (PCNA) and microvessel density (MVD) were measured immunohistochemically in 29 specimens of cartilage tumour.

There was no HIF-1α and HIF-2α staining in any of the nine benign cartilage tumours. In 20 specimens of chondrosarcoma, the rate of HIF-1α and HIF-2α expression was 40% and 25%, respectively. The tumour size (≥ 8 cm), histological grade (grade 2 and grade 3) surgical margin (marginal and intralesional) and HIF-1α expression (positive) correlated significantly with a shorter disease-free survival. There was a significant association between HIF-1α and the MVD and a strong trend towards a correlation between HIF-1α and the PCNA index or histological grade.

Our findings suggest that HIF-1α protein may be a useful objective marker in the assessment of the prognosis in chondrosarcoma, since it plays an important role in tumour angiogenesis and cell proliferation.


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 2 | Pages 209 - 214
1 Feb 2008
Pradhan A Cheung YC Grimer RJ Peake D Al-Muderis OA Thomas JM Smith M

We have investigated the oncological outcome of 63 patients with soft-tissue sarcomas of the hand managed at three major centres in the United Kingdom. There were 44 males and 19 females with a mean age of 45 years (11 to 92). The three most common diagnoses were synovial sarcoma, clear cell sarcoma and epithelioid sarcoma. Local excision was carried out in 45 patients (71%) and amputation in 18 (29%). All those treated by amputation had a wide margin of excision but this was only achieved in 58% of those treated by local excision. The risk of local recurrence was 6% in those treated by amputation compared with 42% for those who underwent attempted limb salvage. An inadequate margin of excision resulted in a 12 times greater risk of local recurrence when compared with those in whom a wide margin of excision had been achieved. We were unable to demonstrate any role for radiotherapy in decreasing the risk of local recurrence when there was an inadequate margin of excision. Patients with an inadequate margin of excision had a much higher risk of both local recurrence and metastasis than those with wide margins. The overall survival rate at five years was 87% and was related to the grade and size of the tumour and to the surgical margin.

We have shown that a clear margin of excision is essential to achieve local control of a soft-tissue sarcoma in the hand and that failure to achieve this results in a high risk of both local recurrence and metastastic disease.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 11 | Pages 1457 - 1461
1 Nov 2007
Han HS Kang S Yoon KS

We have examined the results obtained with 72 NexGen legacy posterior stabilised-flex fixed total knee replacements in 47 patients implanted by a single surgeon between March 2003 and September 2004.

Aseptic loosening of the femoral component was found in 27 (38%) of the replacements at a mean follow-up of 32 months (30 to 48) and 15 knees (21%) required revision at a mean of 23 months (11 to 45). We compared the radiologically-loose and revised knees with those which had remained well-fixed to identify the factors which had contributed to this high rate of aseptic loosening.

Post-operatively, the mean maximum flexion was 136° (110° to 140°) in the loosened group and 125° (95° to 140°) in the well-fixed group (independent t-test, p = 0.022). Squatting, kneeling, or sitting cross-legged could be achieved by 23 (85%) of the loosened knees, but only 22 (49%) of the well-fixed knees (chi-squared test, p = 0.001). The loosened femoral components were found to migrate into a more flexed position, but no migration was detected in the well-fixed group.

These implants allowed a high degree of flexion, but showed a marked rate of early loosening of the femoral component, which was associated with weight-bearing in maximum flexion.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 9 | Pages 1178 - 1183
1 Sep 2007
Culpan P Le Strat V Piriou P Judet T

We present a series of 16 patients treated between 1993 and 2006 who had a failed total ankle replacement converted to an arthrodesis using bone grafting with internal fixation. We used tricortical autograft from the iliac crest to preserve the height of the ankle, the malleoli and the subtalar joint. A successful arthrodesis was achieved at a mean of three months (1.5 to 4.5) in all patients except one, with rheumatoid arthritis and severe bone loss, who developed a nonunion and required further fixation with an intramedullary nail at one year after surgery, before obtaining satisfactory fusion. The post-operative American Orthopaedic Foot and Ankle Society score improved to a mean of 70 (41 to 87) with good patient satisfaction. From this series and an extensive review of the literature we have found that rates of fusion after failed total ankle replacement in patients with degenerative arthritis are high. We recommend our method of arthrodesis in this group of patients. A higher rate of nonunion is associated with rheumatoid arthritis which should be treated differently.


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 6 | Pages 818 - 821
1 Jun 2006
Girard J Vendittoli P Lavigne M Roy AG

A 34-year-old woman with a benign form of osteopetrosis developed osteoarthritis of the hip. In order to avoid the difficulties associated with inserting the femoral component of a conventional total hip arthroplasty, a hybrid metal-on-metal resurfacing was performed. There were several technical challenges associated with the procedure, including the sizing of the component, press-fit fixation of the acetabular component and femoral head preparation, as well as trying to avoid a fracture. No surgical complication occurred. After more than a year following surgery, the patient showed excellent clinical function and remained satisfied with the outcome. We conclude that the hybrid metal-on-metal resurfacing arthroplasty represents a valuable option for the treatment of patients with osteopetrosis and secondary hip osteoarthritis.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 5 | Pages 586 - 590
1 May 2007
Sauvé P Mountney J Khan T De Beer J Higgins B Grover M

Metal-on-metal bearings for total hip replacement (THR) are becoming increasingly popular. Improved wear characteristics mean that these articulations are being inserted into younger patients in the form of THR and resurfacing procedures. This has led to concerns regarding potential carcinogenicity because of the increased exposure to metal ions that the procedure brings.

We have studied the serum cobalt and chromium concentrations in patients who had primary, well-fixed Ring metal-on-metal THRs for more than 30 years. The levels of cobalt and chromium were elevated by five and three times, respectively compared with those in our reference groups. Metal-on-metal articulations appear to be the source of metal ions throughout the life of the prosthesis. In three patients who had undergone revision of a previous metal-on-metal THR to a metal-on-polyethylene replacement the levels of metal ions were within the normal range. The elevations of cobalt and chromium ions seen in our study were comparable with those in patients with modern metal-on-metal THRs.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 3 | Pages 330 - 334
1 Mar 2007
Cesar M Roussanne Y Bonnel F Canovas F

Between 1993 and 2002, 58 GSB III total elbow replacements were implanted in 45 patients with rheumatoid arthritis by the same surgeon. At the most recent follow-up, five patients had died (five elbows) and six (nine elbows) had been lost to follow-up, leaving 44 total elbow replacements in 34 patients available for clinical and radiological review at a mean follow-up of 74 months (25 to 143). There were 26 women and eight men with a mean age at operation of 55.7 years (24 to 77).

At the latest follow-up, 31 excellent (70%), six good (14%), three fair (7%) and four poor (9%) results were noted according to the Mayo elbow performance score. Five humeral (11%) and one ulnar (2%) component were loose according to radiological criteria (type III or type IV). Of the 44 prostheses, two (5%) had been revised, one for type-IV humeral loosening after follow-up for ten years and one for fracture of the ulnar component. Seven elbows had post-operative dysfunction of the ulnar nerve, which was transient in five and permanent in two.

Despite an increased incidence of loosening with time, the GSB III prosthesis has given favourable mid-term results in patients with rheumatoid arthritis.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 12 | Pages 1561 - 1566
1 Dec 2007
Nowicki P Chaudhary H

Avascular necrosis of the femoral head creates considerable morbidity in successful renal transplant recipients who are generally young and expect active lifestyles. Total hip replacement is considered the treatment of choice in these patients, but surgeons may be wary because of a supposed increase in the risk of infection and other complications.

A review of the literature reveals that cemented hip arthroplasty provides good to excellent functional outcomes for renal transplant patients. Most authors have found that the risk of infection is not increased despite chronic immunosuppression, but the rates of general complications are and should be anticipated and treated. There is a high rate of early failure in these patients because of their young age and diffuse osteopenia as a result of secondary hyperparathyroidism related to the underlying renal disease and chronic steroid use. Recent studies have found that despite decreased bone stock in these patients, porous-coated prostheses are not contraindicated.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 4 | Pages 521 - 526
1 Apr 2007
Myers GJC Abudu AT Carter SR Tillman RM Grimer RJ

We investigated whether improvements in design have altered the outcome for patients undergoing endoprosthetic replacement of the distal femur after resection of a tumour. Survival of the implant and ‘servicing’ procedures have been documented using a prospective database, review of the design of the implant and case records. In total, 335 patients underwent a distal femoral replacement, 162 having a fixed-hinge design and 173 a rotating-hinge. The median age of the patients was 24 years (interquartile range 17 to 48).

A total of 192 patients remained alive with a mean follow-up of 12 years (5 to 30). The risk of revision for any reason was 17% at five years, 33% at ten years and 58% at 20 years. Aseptic loosening was the main reason for revision of the fixed-hinge knees while infection and fracture of the stem were the most common for the rotating-hinge implant. The risk of revision for aseptic loosening was 35% at ten years with the fixed-hinge knee, which has, however, been replaced by the rotating-hinge knee with a hydroxyapatite collar. The overall risk of revision for any reason fell by 52% when the rotating-hinge implant was used.

Improvements in the design of distal femoral endoprostheses have significantly decreased the need for revision operations, but infection remains a serious problem. We believe that a cemented, rotating-hinge prosthesis with a hydroxyapatite collar offers the best chance of long-term survival of the prosthesis.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 12 | Pages 1608 - 1614
1 Dec 2007
Baker PN Khaw FM Kirk LMG Esler CNA Gregg PJ

We report the long-term survival of a prospective randomised consecutive series of 501 primary knee replacements using the press-fit condylar posterior cruciate ligament-retaining prosthesis. Patients received either cemented (219 patients, 277 implants) or cementless (177 patients, 224 implants) fixation. Altogether, 44 of 501 knees (8.8%) underwent revision surgery (24 cemented vs 20 cementless). For cemented knees the 15-year survival rate was 80.7% (95% confidence interval (CI) 71.5 to 87.4) and for cementless knees it was 75.3% (95% CI 63.5 to 84.3). There was no significant difference between the two groups (cemented vs cementless; hazard ratio (HR) 0.83, 95% CI 0.45 to 1.52, p = 0.55). When comparing the covariates there was no significant difference in the rates of survival between the side of operation (HR 0.58, p = 0.07), age (HR 0.97, p = 0.10) and diagnosis (HR 1.25 p = 0.72). However, there was a significant gender difference, with males having a higher failure rate with cemented fixation (HR 2.48, p = 0.004). Females had a similar failure rate in both groups.

This single-surgeon series, with no loss to follow-up, provides reliable data of the revision rates of one of the most commonly-used total knee replacements. The survival of the press-fit condylar total knee replacement remained good at 15 years, irrespective of the method of fixation.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 3 | Pages 366 - 371
1 Mar 2007
Krieg AH Davidson AW Stalley PD

Between 1996 and 2003, 16 patients (nine female, seven male) were treated for a primary bone sarcoma of the femur by wide local excision of the tumour, extracorporeal irradiation and re-implantation. An additional vascularised fibular graft was used in 13 patients (81%). All patients were free from disease when reviewed at a minimum of two years postoperatively (mean 49.7 months (24 to 96).

There were no cases of infection. Primary union was achieved after a median of nine months (interquartile range 7 to 11). Five host-donor junctions (16%) united only after a second procedure. Primary union recurred faster at metaphyseal junctions (94% (15) at a median of 7.5 months (interquartile range 4 to 12)) than at diaphyseal junctions (75% (12) at a median of 11.1 months (interquartile range 5 to 18)).

Post-operatively, the median Musculoskeletal Tumour Society score was 85% (interquartile range 75 to 96) and the median Toronto Extremity Salvage score 94% (interquartile range 82 to 99). The Mankin score gave a good or excellent result in 14 patients (88%).

The range of movement of the knee was significantly worse when the extracorporeally irradiated autografts were fixed by plates rather than by nails (p = 0.035).

A total of 16 (62%) of the junctions of the vascularised fibular grafts underwent hypertrophy, indicating union and loading.

Extracorporeal irradiation autografting with supplementary vascularised fibular grafting is a promising biological alternative for intercalary reconstruction after wide resection of malignant bone tumours of the femur.


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 10 | Pages 1279 - 1285
1 Oct 2006
Parvizi J Campfield A Clohisy JC Rothman RH Mont MA

Arthritis of the hip in the young adult can be a disabling condition. Recent years have witnessed extensive research related to the management of this condition. This article reviews the current status with regard to aetiology, diagnosis and treatment of arthritis of the hip in the young adult.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 1 | Pages 34 - 38
1 Jan 2007
Epinette J Manley MT

This study describes 146 primary total knee replacements, either fully or partially coated with hydroxyapatite of which 74 knees in 68 patients were available for clinical and radiological assessment at a mean of 11.2 years (10 to 15). The global failure rate was 1.37% and survival rate with mechanical failure as the end-point was 98.14%. Radiological assessment indicated intimate contact between bone and the hydroxyapatite coating. Over time the hydroxyapatite coating appears to encourage filling of interface gaps remaining after surgery. Our results compare favourably with those of series describing cemented or porous-coated knee replacements, and suggest that fixation with hydroxyapatite is a reliable option in primary total knee replacement.


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 6 | Pages 786 - 789
1 Jun 2005
Grobler GP Learmonth ID Bernstein BP Dower BJ

We retrospectively reviewed, ten years after surgery, 100 consecutive total hip replacements in which the Duraloc 300 cup had been used. Post-operative radiographs were analysed for placement of the cup and interface gaps and follow-up radiographs for lucent lines, osteolysis, wear and migration.

All the components were found to be stable with no evidence of loosening. The mean rate of wear was 0.12 mm/year. Three hips developed acetabular osteolysis at the level of the apex hole. Two have successfully undergone bone grafting without removal of the implants and one patient is awaiting surgery. The Duraloc 300 cup has a survival of 100% at ten years with no aseptic loosening and a low incidence of pelvic osteolysis.


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 10 | Pages 1309 - 1315
1 Oct 2006
Shetty N Hamer AJ Stockley I Eastell R Willkinson JM

Bisphosphonates reduce peri-prosthetic bone loss in the short term after total hip replacement but the mid- and longer term effects are not known. The aims of this randomised trial were to examine the effect of a single dose of 90 mg of pamidronate on the clinical and radiological outcome and peri-prosthetic bone mineral density in 50 patients (56 hips) over a five-year period, following total hip replacement.

At five years, 37 patients (42 hips) returned for assessment. The Harris hip scores were similar in the pamidronate and placebo groups throughout the study. Also at five years, four patients, two from each group had osteolytic lesions on plain radiography. These were located around the acetabular component in three patients and in the femoral calcar in one. The femoral and acetabular peri-prosthetic bone mineral density in the pamidronate group and the control group was similar at five years.

Pamidronate given as a single post-operative dose does not appear to influence the clinical outcome or prevent the development of osteolytic lesions at five years after total hip replacement.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 2 | Pages 174 - 179
1 Feb 2007
Kim Y Yoon S Kim J

Our aim in this prospective study was to compare the bone mineral density (BMD) around cementless acetabular and femoral components which were identical in geometry and had the same alumina modular femoral head, but differed in regard to the material of the acetabular liners (alumina ceramic or polyethylene) in 50 patients (100 hips) who had undergone bilateral simultaneous primary total hip replacement. Dual energy X-ray absorptiometry scans of the pelvis and proximal femur were obtained at one week, at one year, and annually thereafter during the five-year period of the study.

At the final follow-up, the mean BMD had increased significantly in each group in acetabular zone I of DeLee and Charnley (20% (15% to 26%), p = 0.003), but had decreased in acetabular zone II (24% (18% to 36%) in the alumina group and 25% (17% to 31%) in the polyethylene group, p = 0.001). There was an increase in the mean BMD in zone III of 2% (0.8% to 3.2%) in the alumina group and 1% (0.6% to 2.2%) in the polyethylene group (p = 0.315). There was a decrease in the mean BMD in the calcar region (femoral zone 7) of 15% (8% to 24%) in the alumina group and 14% (6% to 23%) in the polyethylene group (p < 0.001). The mean bone loss in femoral zone 1 of Gruen et al was 2% (1.1% to 3.1%) in the alumina group and 3% (1.3% to 4.3%) in the polyethylene group (p = 0.03), and in femoral zone 6, the mean bone loss was 15% (9% to 27%) in the alumina group and 14% (11% to 29%) in the polyethylene group compared with baseline values. There was an increase in the mean BMD on the final scans in femoral zones 2 (p = 0.04), 3 (p = 0.04), 4 (p = 0.12) and 5 (p = 0.049) in both groups.

There was thus no significant difference in the bone remodelling of the acetabulum and femur five years after total hip replacement in those two groups where the only difference was in the acetabular liner.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 5 | Pages 567 - 573
1 May 2007
Keegan GM Learmonth ID Case CP

The long-term effects of metal-on-metal arthroplasty are currently under scrutiny because of the potential biological effects of metal wear debris. This review summarises data describing the release, dissemination, uptake, biological activity, and potential toxicity of metal wear debris released from alloys currently used in modern orthopaedics. The introduction of risk assessment for the evaluation of metal alloys and their use in arthroplasty patients is discussed and this should include potential harmful effects on immunity, reproduction, the kidney, developmental toxicity, the nervous system and carcinogenesis.