There remains confusion in the literature with regard to the spinopelvic relationship, and its contribution to ideal acetabular component position. Critical assessment of the literature has been limited by use of conflicting terminology and definitions of new concepts that further confuse the topic. In 2017, the concept of a Hip-Spine Workgroup was created with the first meeting held at the American Academy of Orthopedic Surgeons Annual Meeting in 2018. The goal of this workgroup was to first help standardize terminology across the literature so that as a topic, multiple groups could produce literature that is immediately understandable and applicable. This consensus review from the Hip-Spine Workgroup aims to simplify the spinopelvic relationship, offer hip surgeons a concise summary of available literature, and select common terminology approved by both hip surgeons and spine surgeons for future research. Cite this article:
The purpose of this retrospective study was to evaluate the minimum
five-year outcome of revision total hip arthroplasty (THA) using
the Kerboull acetabular reinforcement device (KARD) in patients
with Paprosky type III acetabular defects and destruction of the
inferior margin of the acetabulum. We identified 36 patients (37 hips) who underwent revision THA
under these circumstances using the KARD, fresh frozen allograft
femoral heads, and reconstruction of the inferior margin of the
acetabulum. The Merle d’Aubigné system was used for clinical assessment.
Serial anteroposterior pelvic radiographs were used to assess migration
of the acetabular component.Aims
Patients and Methods
Between November 1997 and December 2000 we performed 27 total hip replacements in 22 patients with high congenital dislocation of the hip using porous tantalum monoblock
The aim of this study was to determine whether there is a difference
in the rate of wear between acetabular components positioned within
and outside the ‘safe zones’ of anteversion and inclination angle. We reviewed 100 hips in 94 patients who had undergone primary
total hip arthroplasty (THA) at least ten years previously. Patients
all had the same type of acetabular component with a bearing couple
which consisted of a 28 mm cobalt-chromium head on a highly crosslinked
polyethylene (HXLPE) liner. A supine radiostereometric analysis
(RSA) examination was carried out which acquired anteroposterior
(AP) and lateral paired images. Acetabular component anteversion
and inclination angles were measured as well as total femoral head
penetration, which was divided by the length of implantation to
determine the rate of polyethylene wear.Aims
Patients and Methods
The aims of this study were to determine the clinical and radiographic
outcomes, implant survivorship, and complications of patients with
a history of poliomyelitis undergoing total hip arthroplasty (THA)
in affected limbs and unaffected limbs of this same population. A retrospective review identified 51 patients (27 male and 24
female, 59 hips) with a mean age of 66 years (38 to 88) and with
the history of poliomyelitis who underwent THA for degenerative
arthritis between 1970 and 2012. Immigrant status, clinical outcomes,
radiographic results, implant survival, and complications were recorded.Aims
Patients and Methods
We have reviewed 70 Harris-Galante uncemented
In order to address acetabular defects, porous metal revision acetabular components and augments have been developed, which require fixation to each other. The fixation technique that results in the smallest relative movement between the components, as well as its influence on the primary stability with the host bone, have not previously been determined. A total of 18 composite hemipelvises with a Paprosky IIB defect were implanted using a porous titanium 56 mm multihole acetabular component and 1 cm augment. Each acetabular component and augment was affixed to the bone using two screws, while the method of fixation between the acetabular component and augment varied for the three groups of six hemipelvises: group S, screw fixation only; group SC, screw plus cement fixation; group C, cement fixation only. The implanted hemipelvises were cyclically loaded to three different loading maxima (0.5 kN, 0.9 kN, and 1.8 kN).Objectives
Methods
Reconstruction of the acetabulum after resection of a periacetabular
malignancy is technically challenging and many different techniques
have been used with varying success. Our aim was to prepare a systematic
review of the literature dealing with these techniques in order
to clarify the management, the rate of complications and the outcomes. A search of PubMed and MEDLINE was conducted for English language
articles published between January 1990 and February 2017 with combinations
of key search terms to identify studies dealing with periacetabular
resection with reconstruction in patients with a malignancy. Studies
in English that reported radiographic or clinical outcomes were
included. Data collected from each study included: the number and
type of reconstructions, the pathological diagnosis of the lesions,
the mean age and follow-up, gender distribution, implant survivorship, complications,
functional outcome, and mortality. The results from individual studies
were combined for the general analysis, and then grouped according
to the type of reconstruction. Aims
Patients and Methods
The aim of this study was to examine the results of the acetabular
distraction technique in achieving implantation of a stable construct,
obtaining biological fixation, and producing healing of chronic
pelvic discontinuity at revision total hip arthroplasty. We identified 32 patients treated between 2006 and 2013 who underwent
acetabular revision for a chronic pelvic discontinuity using acetabular
distraction, and who were radiographically evaluated at a mean of
62 months (25 to 160). Of these patients, 28 (87.5%) were female.
The mean age at the time of revision was 67 years (44 to 86). The patients
represented a continuous series drawn from two institutions that
adhered to an identical operative technique.Aims
Patients and Methods
This review summarises the technique of impaction
grafting with mesh augmentation for the treatment of uncontained
acetabular defects in revision hip arthroplasty. The ideal acetabular revision should restore bone stock, use
a small socket in the near-anatomic position, and provide durable
fixation. Impaction bone grafting, which has been in use for over
40 years, offers the ability to achieve these goals in uncontained
defects. The precepts of modern, revision impaction grafting are
that the segmental or cavitary defects must be supported with a
mesh; the contained cavity is filled with vigorously impacted morselised
fresh-frozen allograft; and finally, acrylic cement is used to stabilise
the graft and provide rigid, long-lasting fixation of the revised
acetabular component. Favourable results have been published with this technique. While
having its limitations, it is a viable option to address large acetabular
defects in revision arthroplasty. Cite this article:
A contact patch to rim (CPR) distance of <
10 mm has been
associated with edge-loading and excessive wear. However, not all
arthroplasties with a low CPR distance show problems with wear.
Therefore, CPR distance may not be the only variable affecting the
post-operative metal ion concentrations. We used multiple logistic regression to determine what variables
differed between the patients who had high and low cobalt (CoS)
and chromium (CrS) serum ion concentrations within a cohort of patients
with low (<
10 mm) CPR distances. A total of 56 patients treated
with unilateral hip resurfacing arthroplasty (HRA) had CoS and CrS
ion studies performed more than one year after surgery. The mean
age of the patients at the time of surgery was 51.7 years (29 to
70), with 38 women (68%) and 18 men (32%).Aims
Patients and Methods
Periprosthetic fracture (PF) after primary total hip arthroplasty
(THA) is an uncommon but potentially devastating complication. This
study aims to investigate the influence of cemented stem designs
on the risk of needing a revision for a PF. We analysed data on 257 202 primary THAs with cemented stems
and 390 linked first revisions for PF recorded in the National Joint
Registry (NJR) of England, Wales and Northern Ireland to determine
if a cemented femoral stem brand was associated with the risk of
having revision for a PF after primary THA. All cemented femoral
stem brands with more than 10 000 primary operations recorded in
the NJR were identified. The four most commonly used cemented femoral
stems were the Exeter V40 (n = 146 409), CPT (n = 24 300), C-Stem
(n = 15 113) and Charnley (n = 20 182). We compared the revision risk ratios due to PF amongst the stems
using a Poisson regression model adjusting for patient factors.
Compared with the Exeter V40, the age, gender and ASA grade adjusted
revision rate ratio was 3.89 for the cemented CPT stem (95% confidence
interval (CI) 3.07 to 4.93), 0.89 for the C-Stem (95% CI 0.57 to
1.41) and 0.41 for the Charnley stem (95% CI 0.24 to 0.70).Aims
Patients and Methods
We hypothesised that the removal of the subchondral
bone plate (SCBP) for cemented acetabular component fixation in
total hip arthroplasty (THA) offers advantages over retention by
improving the cement-bone interface, without jeopardising implant
stability. We have previously published two-year follow-up data
of a randomised controlled trial (RCT), in which 50 patients with
primary osteoarthritis were randomised to either retention or removal
of the SCBP. The mean age of the retention group (n = 25, 13 males)
was 70.0 years ( Cite this article:
This study reports the results of 38 total hip
arthroplasties (THAs) in 33 patients aged <
50 years, using the
JRI Furlong hydroxyapatite ceramic (HAC)-coated femoral component.
This represents an update of previous reports of the same cohort
at ten and 16 years, which were reported in 2004 and 2009, respectively.
We describe the survival, radiological and functional outcomes at
a mean follow-up of 21 years (17 to 25). Of the surviving 34 THAs,
one underwent femoral revision for peri-prosthetic fracture after
21 years, and one patient (one hip) was lost to follow-up. Using
aseptic loosening as the end-point, 12 hips (31.5%) needed acetabular
revision but none needed femoral revision, demonstrating 100% survival
(95% confidence interval 89 to 100). In young patients with high demands, the Furlong HAC–coated femoral
component gives excellent long-term results. Cite this article:
In this retrospective study we evaluated the
proficiency of shelf autograft in the restoration of bone stock
as part of primary total hip replacement (THR) for hip dysplasia,
and in the results of revision arthroplasty after failure of the primary
arthroplasty. Of 146 dysplastic hips treated by THR and a shelf
graft, 43 were revised at an average of 156 months, 34 of which
were suitable for this study (seven hips were excluded because of
insufficient bone-stock data and two hips were excluded because
allograft was used in the primary THR). The acetabular bone stock
of the hips was assessed during revision surgery. The mean implant–bone
contact was 58% (50% to 70%) at primary THR and 78% (40% to 100%)
at the time of the revision, which was a significant improvement
(p <
0.001). At primary THR all hips had had a segmental acetabular
defect >
30%, whereas only five (15%) had significant segmental
bone defects requiring structural support at the time of revision.
In 15 hips (44%) no bone graft or metal augments were used during
revision. A total of 30 hips were eligible for the survival study. At a
mean follow-up of 103 months (27 to 228), two aseptic and two septic
failures had occurred. Kaplan-Meier survival analysis of the revision
procedures demonstrated a ten-year survival rate of 93.3% (95% confidence
interval (CI) 78 to 107) with clinical or radiological failure as
the endpoint. The mean Oxford hip score was 38.7 (26 to 46) for
non-revised cases at final follow-up. Our results indicate that the use of shelf autografts during
THR for dysplastic hips restores bone stock, contributing to the
favourable survival of the revision arthroplasty should the primary
procedure fail. Cite this article:
Acetabular bone loss is a challenging problem
facing the revision total hip replacement surgeon. Reconstruction
of the acetabulum depends on the presence of anterosuperior and
posteroinferior pelvic column support for component fixation and
stability. The Paprosky classification is most commonly used when
determining the location and degree of acetabular bone loss. Augments
serve the function of either providing primary construct stability
or supplementary fixation. When a pelvic discontinuity is encountered we advocate the use
of an acetabular distraction technique with a jumbo cup and modular
porous metal acetabular augments for the treatment of severe acetabular
bone loss and associated chronic pelvic discontinuity. Cite this article:
In 2005, we demonstrated that the polished triple-tapered
C-stem at two years had migrated distally and rotated internally.
From that series, 33 patients have now been followed radiologically,
clinically and by radiostereometric analysis (RSA) for up to ten
years. The distal migration within the cement mantle had continued
and reached a mean of 2 mm (0.5 to 4.0) at ten years. Internal rotation,
also within the cement mantle, was a mean 3.8° (external 1.6° to
internal 6.6°) The cement mantle did not show any sign of migration
or loosening in relation to the femoral bone. There were no clinical
or radiological signs indicating that the migration or rotation
within the cement mantle had had any adverse effects for the patients. Cite this article:
Large-head metal-on-metal (MoM) total hip replacements
(THR) have given rise to concern. Comparative studies of small-head
MoM THRs over a longer follow-up period are lacking. Our objective
was to compare the incidence of complications such as infection,
dislocation, revision, adverse local tissue reactions, mortality
and radiological and clinical outcomes in small-head (28 mm) MoM
and ceramic-on-polyethylene (CoP) THRs up to 12 years post-operatively. A prospective cohort study included 3341 THRs in 2714 patients.
The mean age was 69.1 years (range 24 to 98) and 1848 (55.3%) were
performed in women, with a mean follow-up of 115 months (18 to 201).
There were 883 MoM and 2458 CoP bearings. Crude incidence rates
(cases/1000 person-years) were: infection 1.3 In conclusion, we found similar results for small-head MoM and
CoP bearings up to ten years post-operatively, but after ten years
MoM THRs had a higher risk of all-cause revision. Furthermore, the
presence of an adverse response to metal debris seen in the small-head
MOM group at revision is a cause for concern. Cite this article:
The use of ilioischial cage reconstruction for
pelvic discontinuity has been replaced by the Trabecular Metal (Zimmer,
Warsaw, Indiana) cup-cage technique in our institution, due to the
unsatisfactory outcome of using a cage alone in this situation.
We report the outcome of 26 pelvic discontinuities in 24 patients
(20 women and four men, mean age 65 years (44 to 84)) treated by
the cup-cage technique at a mean follow-up of 82 months (12 to 113)
and compared them with a series of 19 pelvic discontinuities in
19 patients (18 women and one man, mean age 70 years (42 to 86))
treated with a cage at a mean follow-up of 69 months (1 to 170).
The clinical and radiological outcomes as well as the survivorship
of the groups were compared. In all, four of the cup-cage group
(15%) and 13 (68%) of the cage group failed due to septic or aseptic
loosening. The seven-year survivorship was 87.2% (95% confidence interval
(CI) 71 to 103) for the cup-cage group and 49.9% (95% CI 15 to 84)
for the cage-alone group (p = 0.009). There were four major complications
in the cup-cage group and nine in the cage group. Radiological union
of the discontinuity was found in all successful cases in the cup-cage
group and three of the successful cage cases. Three hips in the
cup-cage group developed early radiological migration of the components,
which stabilised with a successful outcome. Cup-cage reconstruction is a reliable technique for treating
pelvic discontinuity in mid-term follow-up and is preferred to ilioischial
cage reconstruction. If the continuity of the bone graft at the
discontinuity site is not disrupted, early migration of the components
does not necessarily result in failure. Cite this article: