Advertisement for orthosearch.org.uk
Results 61 - 69 of 69
Results per page:
The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 3 | Pages 349 - 355
1 Mar 2010
Babis GC Sakellariou VI O’Connor MI Hanssen AD Sim FH

We report the use of an allograft prosthetic composite for reconstruction of the skeletal defect in complex revision total hip replacement for severe proximal femoral bone loss. Between 1986 and 1999, 72 patients (20 men, 52 women) with a mean age of 59.9 years (38 to 78) underwent reconstruction using this technique.

At a mean follow-up of 12 years (8 to 20) 57 patients were alive, 14 had died and one was lost to follow-up. Further revision was performed in 19 hips at a mean of 44.5 months (11 to 153) post-operatively. Causes of failure were aseptic loosening in four, allograft resorption in three, allograft nonunion in two, allograft fracture in four, fracture of the stem in one, and deep infection in five. The survivorship of the allograft-prosthesis composite at ten years was 69.0% (95% confidence interval 67.7 to 70.3) with 26 patients remaining at risk. Survivorship was statistically significantly affected by the severity of the pre-operative bone loss (Paprosky type IV; p = 0.019), the number of previous hip revisions exceeding two (p = 0.047), and the length of the allograft used (p = 0.005).


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 4 | Pages 441 - 446
1 Apr 2009
Ornstein E Linder L Ranstam J Lewold S Eisler T Torper M

We identified 1305 femoral impaction bone grafting revisions using the Exeter stem performed between 1989 and 2002 in 30 hospitals throughout Sweden. There were 1188 patients with a mean age of 71 years (29 to 94) followed up for between five and 18 years.

The participating departments reported 70 further revisions in total, of which 57 could also be identified on the Swedish National Arthroplasty Registry.

Kaplan-Meier survivorship for all causes of failure was 94.0% (95% confidence interval (CI) 92 to 96) for women and 94.7% (95% CI, 92 to 96) for men at 15 years. Survivorship at 15 years for aseptic loosening was 99.1% (95% CI 98.4 to 99.5), for infection 98.6% (95% CI 97.6 to 99.2), for subsidence 99.0% (95% CI 98.2 to 99.4) and for fracture 98.7% (95% CI 97.9 to 99.2)

Statistically significant predictors of failure were the year in which revision was conducted (p < 0.001). The number of previous revisions was slightly above the level of signifance (p = 0.056). Age, gender, the length of the stem and previous septic loosening were not predictors of failure (p = 0.213, p = 0.399, p = 0.337, p = 0.687, respectively). The difference in survivorship between high- and low-volume departments was only 3% at ten years.

We conclude that impaction bone grafting with the Exeter stem has an excellent long-term survivorship following revision arthroplasty. The technique of impaction grafting appears to be reliable, can be learned rapidly and produces a predictably low incidence of aseptic loosening.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 8 | Pages 1015 - 1018
1 Aug 2007
Wroblewski BM Siney PD Fleming PA

We studied survival to 38 years after Charnley low-friction arthroplasty of the hip. We used revision as an end-point, while adopting a policy of regular follow-up and early revision for radiological changes alone if indicated. Between November 1962 and June 2005, 22 066 primary low-friction arthroplasties (17 409 patients) had been performed at Wrightington Hospital by more than 330 surgeons. By June 2006, 1001 (4.5%) hips had been revised and 1490 patients (2662 hips, 12%) had died. At 31 years, where a minimum of 40 hips were still attending follow-up, survival with revision for infection as an endpoint was 95%, for dislocation 98%, for a fractured stem 88.6%, for a loose stem 72.5% and for a loose acetabular component 53.7%. Wear and loosening of the ultra-high-molecular-weight polyethylene acetabular component were the main long-term problems. We conclude that regular follow-up after hip replacement is essential and that all operative findings should be recorded at revision


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 8 | Pages 1022 - 1026
1 Aug 2006
Langlais F Belot N Ropars M Lambotte JC Thomazeau H

We evaluated the long-term fixation of 64 press-fit cemented stems of constrained total knee prostheses in 32 young patients with primary malignant bone tumours. Initial stable fixation, especially in rotation, was achieved by precise fit of the stem into the reamed endosteum, before cementation. Complementary fixation, especially in migration and rotation, was obtained by pressurised antibiotic-loaded cement. The mean age at operation was 33 years (13 to 61). No patient was lost to follow-up; 13 patients died and the 19 survivors were examined at a mean follow-up of 12.5 years (4 to 21). Standard revision press-fit cemented stems were used on the side of the joint which was not involved with tumour (26 tibial and six femoral), on this side there was no loosening or osteolysis and stem survival was 100%. On the reconstruction side, custom-made press-fit stems were used and the survival rate, with any cause for revision as an end point, was 88%, but 97% for loosening or osteolysis. This longevity is similar to that achieved at 20 years with the Charnley-Kerboull primary total hip replacement with press-fit cemented femoral components.

We recommend this type of fixation when extensive reconstruction of the knee is required. It may also be suitable for older patients requiring revision of a total knee replacement or in difficult situations such as severe deformity and complex articular fractures.


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 10 | Pages 1333 - 1336
1 Oct 2005
Busch CA Charles MN Haydon CM Bourne RB Rorabeck CH MacDonald SJ McCalden RW

We identified five (2.3%) fractures of the stem in a series of 219 revision procedures using a cementless, cylindrical, extensively porous-coated, distally-fixed femoral stem. Factors relating to the patients, the implant and the operations were compared with those with intact stems. Finite-element analysis was performed on two of the fractured implants.

Factors associated with fracture of the stem were poor proximal bone support (type III–type IV; p = 0.001), a body mass index > 30; (p = 0.014), a smaller diameter of stem (< 13.5 mm; p = 0.007) and the use of an extended trochanteric osteotomy (ETO 4/5: p = 0.028). Finite-element analysis showed that the highest stresses on the stem occurred adjacent to the site of the fracture. The use of a strut graft wired over an extended trochanteric osteotomy in patients lacking proximal femoral cortical support decreased the stresses on the stem by 48%.

We recommend the use of a strut allograft in conjunction with an extended trochanteric osteotomy in patients with poor proximal femoral bone stock.


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 7 | Pages 865 - 869
1 Jul 2006
Comba F Buttaro M Pusso R Piccaluga F

We reviewed the clinical and radiological results of 131 patients who underwent acetabular revision for aseptic loosening with impacted bone allograft and a cemented acetabular component. The mean follow-up was 51.7 months (24 to 156).

The mean post-operative Merle D’Aubigné and Postel scores were 5.7 points (4 to 6) for pain, 5.2 (3 to 6) for gait and 4.5 (2 to 6) for mobility. Radiological evaluation revealed migration greater than 5 mm in four acetabular components. Radiological failure matched clinical failure. Asymptomatic radiolucent lines were observed in 31 of 426 areas assessed (7%). Further revision was required in six patients (4.5%), this was due to infection in three and mechanical failure in three. The survival rate for the reconstruction was 95.8% (95% confidence interval 92.3 to 99.1) overall, and 98%, excluding revision due to sepsis.

Our study, from an independent centre, has reproduced the results of the originators of the method.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 6 | Pages 730 - 737
1 Jun 2009
Carrington NC Sierra RJ Gie GA Hubble MJW Timperley AJ Howell JR

The first 325 Exeter Universal stems (309 patients) implanted at the originating centre were inserted between March 1988 and February 1990 by a group of surgeons with differing experience. In this report we describe the clinical and radiological results at a mean of 15.7 years (14.7 to 17.3) after operation with no loss to follow-up. There were 97 patients (108 hips) with replacements still in situ and 31 (31 hips) who had undergone a further procedure. With an endpoint of revision for aseptic loosening, the survivorship at 17 years was 100% and 90.4% for the femoral and acetabular component, respectively. The mean Merle D’Aubigné and Postel scores at review were 5.4 (sd 0.97) for pain and 4.5 (sd 1.72) for function. The mean Oxford score was 38.4 (sd 9.8) (0 to 48 worst-to-best scale) and the mean combined Harris pain and function score was 73.2 (sd 16.9). Radiological review showed excellent preservation of bone stock in the proximal femur and no failures of the femoral component.


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 12 | Pages 1617 - 1621
1 Dec 2005
O’Shea K Quinlan JF Kutty S Mulcahy D Brady OH

We assessed the outcome of patients with Vancouver type B2 and B3 periprosthetic fractures treated with femoral revision using an uncemented extensively porous-coated implant. A retrospective clinical and radiographic assessment of 22 patients with a mean follow-up of 33.7 months was performed. The mean time from the index procedure to fracture was 10.8 years. There were 17 patients with a satisfactory result. Complications in four patients included subsidence in two, deep sepsis in one, and delayed union in one. Concomitant acetabular revision was required in 19 patients. Uncemented extensively porous-coated femoral stems incorporate distally allowing stable fixation. We found good early survival rates and a low incidence of nonunion using this implant.


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 3 | Pages 315 - 320
1 Mar 2006
Field RE Singh PJ Latif AMH Cronin MD Matthews DJ

We describe the results at five years of a prospective study of a new tri-tapered polished, cannulated, cemented femoral stem implanted in 51 patients (54 hips) with osteoarthritis. The mean age and body mass index of the patients was 74 years and 27.9, respectively. Using the anterolateral approach, half of the stems were implanted by a consultant orthopaedic surgeon and half by six different registrars. There were three withdrawals from the study because of psychiatric illness, a deep infection and a recurrent dislocation. Five deaths occurred prior to five-year follow-up and one patient withdrew from clinical review.

In the remaining 51 hips the mean pre-operative Oxford hip score was 47 points which decreased to 19 points at five years (45 hips). Of the stems 49 (98%) were implanted within 1° of neutral in the femoral canal. The mean migration of the stem at five years was 1.9 mm and the survivorship for aseptic loosening was 100%. There was no significant difference in outcome between the consultant and registrar groups. At five years, the results were comparable with those of other polished, tapered, cemented stems. Long-term surveillance continues.