The Bone & Joint Journal
Vol. 96-B, Issue 11 | Pages 1503 - 1509
1 Nov 2014
Ritter MA Davis KE Small SR Merchun JG Farris A

The relationship between post-operative bone density and subsequent failure of total knee replacement (TKR) is not known. This retrospective study aimed to determine the relationship between bone density and failure, both overall and according to failure mechanism. All 54 aseptic failures occurring in 50 patients from 7760 consecutive primary cemented TKRs between 1983 and 2004 were matched with non-failing TKRs, and 47 failures in 44 patients involved tibial failures with the matching characteristics of age (65.1 for failed and 69.8 for non-failed), gender (70.2% female), diagnosis (93.6% OA), date of operation, bilaterality, pre-operative alignment (0.4 and 0.3 respectively), and body mass index (30.2 and 30.0 respectively). In each case, the density of bone beneath the tibial component was assessed at each follow-up interval using standardised, calibrated radiographs. Failing knees were compared with controls both overall and, as a subgroup analysis, by failure mechanism. Knees were compared with controls using univariable linear regression.

Significant and continuous elevation in tibial density was found in knees that eventually failed by medial collapse (p < 0.001) and progressive radiolucency (p < 0.001) compared with controls, particularly in the medial region of the tibia. Knees failing due to ligamentous instability demonstrated an initial decline in density (p = 0.0152) followed by a non-decreasing density over time (p = 0.034 for equivalence). Non-failing knees reported a decline in density similar to that reported previously using dual-energy x-ray absorptiometry (DEXA). Differences between failing and non-failing knees were observable as early as two months following surgery. This tool may be used to identify patients at risk of failure following TKR, but more validation work is needed.

Cite this article: Bone Joint J 2014;96-B:1503–9.


The Bone & Joint Journal
Vol. 95-B, Issue 3 | Pages 290 - 294
1 Mar 2013
MacLeod K Lingham A Chatha H Lewis J Parkes A Grange S Smitham PJ

Clinicians are often asked by patients, “When can I drive again?” after lower limb injury or surgery. This question is difficult to answer in the absence of any guidelines. This review aims to collate the currently available evidence and discuss the factors that influence the decision to allow a patient to return to driving. Medline, Web of Science, Scopus, and EMBASE were searched using the following terms: ‘brake reaction time’, ‘brake response time’, ‘braking force’, ‘brake pedal force’, ‘resume driving’, ‘rate of application of force’, ‘driving after injury’, ‘joint replacement and driving’, and ‘fracture and driving’. Of the relevant literature identified, most studies used the brake reaction time and total brake time as the outcome measures. Varying recovery periods were proposed based on the type and severity of injury or surgery. Surveys of the Driver and Vehicle Licensing Agency, the Police, insurance companies in the United Kingdom and Orthopaedic Surgeons offered a variety of opinions.

There is currently insufficient evidence for any authoritative body to determine fitness to drive. The lack of guidance could result in patients being withheld from driving for longer than is necessary, or returning to driving while still unsafe.

Cite this article: Bone Joint J 2013;95-B:290–4.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 11 | Pages 1479 - 1486
1 Nov 2011
Park J Kim Y

The purpose of this prospective, randomised study was to evaluate the clinical and radiological results comparing the identical cemented or cementless NexGen total knee prostheses implanted bilaterally in the same patient. Sequential simultaneous bilateral total knee replacements were performed in 50 patients (100 knees). There were 39 women and 11 men with a mean age of 58.4 years (51 to 67) who received a cemented prosthesis in one knee and a cementless prosthesis in the other. The mean follow-up was 13.6 years (13 to 14). At final review, the mean Knee Society scores (96.2 (82 to 100) versus 97.7 (90 to 100)), the mean Western Ontario and McMaster Universities osteoarthritis index (34.5 (4 to 59) versus 35.6 (5 to 51)), the mean ranges of knee movement (124° (100° to 140°) versus 128° (110° to 140°)), mean patient satisfaction (8.1 (sd 1.9) versus 8.3 (sd 1.7)), and radiological results were similar in both groups. The rate of survival of the femoral components was 100% in both groups at 14 years. The rate of survival of the cemented tibial component was 100% and 98% in the cementless tibial component. No osteolysis was identified in either group. Our data have shown no advantage of cementless over cemented components in total knee replacement.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 3 | Pages 399 - 403
1 Mar 2011
Griffiths D Gikas PD Jowett C Bayliss L Aston W Skinner J Cannon S Blunn G Briggs TWR Pollock R

Between 1997 and 2007, 68 consecutive patients underwent replacement of the proximal humerus for tumour using a fixed-fulcrum massive endoprosthesis. Their mean age was 46 years (7 to 87). Ten patients were lost to follow-up and 16 patients died. The 42 surviving patients were assessed using the Musculoskeletal Tumor Society (MSTS) Score and the Toronto Extremity Salvage Score (TESS) at a mean follow-up of five years and 11 months (one year to ten years and nine months). The mean MSTS score was 72.3% (53.3% to 100%) and the mean TESS was 77.2% (58.6% to 100%).

Four of 42 patients received a new constrained humeral liner to reduce the risk of dislocation. This subgroup had a mean MSTS score of 77.7% and a mean TESS of 80.0%. The dislocation rate for the original prosthesis was 25.9; none of the patients with the new liner had a dislocation at a mean of 14.5 months (12 to 18).

Endoprosthetic replacement for tumours of the proximal humerus using this prosthesis is a reliable operation yielding good results without the documented problems of unconstrained prostheses. The performance of this prosthesis is expected to improve further with a new constrained humeral liner, which reduces the risk of dislocation.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 6 | Pages 760 - 768
1 Jun 2011
ten Broeke RHM Alves A Baumann A Arts JJC Geesink RGT

Four uncemented Symax hip stems were extracted at three weeks and nine, 13 and 32 months, respectively, for reasons other than loosening. The reasons for implant removal were infection in two cases, recurrent dislocation in one and acetabular fracture in one. They were analysed to assess the effect and behaviour of an electrochemically deposited, completely resorbable biomimetic BONIT-hydroxyapatite (HA) coating (proximal part) and a DOTIZE surface treatment (distal part) using qualitative histology, quantitative histomorphometry and scanning electron microscopy (SEM). Early and direct bone-implant bonding with signs of active remodelling of bone and the HA coating were demonstrated by histology and SEM. No loose BONIT-HA particles or delamination of the coating were observed, and there was no inflammation or fibrous interposition at the interface.

Histomorphometry showed bone-implant contact varying between 26.5% at three weeks and 83.5% at 13 months at the HA-coated implant surface. The bone density in the area of investigation was between 24.6% at three weeks and 41.1% at 32 months. The DOTIZE surface treatment of the distal part of the stem completely prevented tissue and bone apposition in all cases, thereby optimising proximal stress transfer.

The overall features of this implant, in terms of geometry and surface texture, suggest a mechanically stable design with a highly active biomimetic coating, resulting in rapid and extensive osseo-integration, exclusively in the metaphyseal part of the stem. Early remodelling of the HA coating does not seem to have a detrimental effect on short-term bone-implant coupling. There were no adverse effects identified from either the BONIT-HA coating or the DOTIZE surface treatment.


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 8 | Pages 1079 - 1084
1 Aug 2010
Muirhead-Allwood S Sandiford N Skinner JA Hua J Kabir C Walker PS

We present the 10- to 17-year results of 112 computer-assisted design computer-assisted manufacture femoral components. The total hip replacements were performed between 1992 and 1998 in 111 patients, comprising 53 men and 58 women. Their mean age was 46.2 years (24.6 to 62.2) with a mean follow-up of 13 years (10 to 17). The mean Harris Hip Score improved from 42.4 (7 to 99) to 90.3 (38 to 100), the mean Oxford Hip Score from 43.1 (12 to 59) to 18.2 (12 to 51) and the mean Western Ontario MacMasters University Osteoarthritis Index score from 57.0 (7 to 96) to 11.9 (0 to 85). There was one revision due to failure of the acetabular component but no failures of the femoral component. There were no revisions for aseptic loosening. The worst-case survival in this cohort of custom femoral components at 13.2 years follow-up was 98.2% (95% confidence interval 95 to 99). Overall survival of this series of total hip replacements was 97.3% (95% confidence interval 95 to 99).

These results are comparable with the best medium- to long-term results for femoral components used in primary total hip replacement with any means of fixation.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 11 | Pages 1439 - 1445
1 Nov 2007
Triclot P Grosjean G El Masri F Courpied JP Hamadouche M

We carried out a prospective randomised study designed to compare the penetration rate of acetabular polyethylene inserts of identical design but different levels of cross-linking at a minimum of four years follow-up. A total of 102 patients (102 hips) were randomised to receive either highly cross-linked Durasul, or contemporary Sulene polyethylene inserts at total hip replacement. A single blinded observer used the Martell system to assess penetration of the femoral head. At a mean follow-up of 4.9 years (4.2 to 6.1) the mean femoral head penetration rate was 0.025 mm/year (sd 0.128) in the Durasul group compared with 0.106 mm/year (sd 0.109) in the Sulene group (Mann-Whitney test, p = 0.0027). The mean volumetric penetration rate was 29.24 mm3/year (sd 44.08) in the Durasul group compared with 53.32 mm3/year (sd 48.68) in the Sulene group. The yearly volumetric penetration rate was 55% lower in the Durasul group (Mann-Whitney test, p = 0.0058). Longer term results are needed to investigate whether less osteolysis will occur.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 9 | Pages 1134 - 1141
1 Sep 2009
Isaac GH Brockett C Breckon A van der Jagt D Williams S Hardaker C Fisher J Schepers A

This study reports on ceramic-on-metal (CoM) bearings in total hip replacement. Whole blood metal ion levels were measured. The median increase in chromium and cobalt at 12 months was 0.08 μg/1 and 0.22 μg/1, respectively, in CoM bearings. Comparable values for metal-on-metal (MoM) were 0.48 μg/1 and 0.32 μg/1. The chromium levels were significantly lower in CoM than in MoM bearings (p = 0.02). The cobalt levels were lower, but the difference was not significant. Examination of two explanted ceramic heads revealed areas of thin metal transfer. CoM bearings (one explanted head and acetabular component, one explanted head and new acetabular component, and three new heads and acetabular components) were tested in a hip joint simulator. The explanted head and acetabular component had higher bedding-in. However, after one million cycles all the wear rates were the same and an order of magnitude less than that reported for MoM bearings. There were four outliers in each clinical group, primarily related to component malposition.


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 12 | Pages 1567 - 1573
1 Dec 2006
Kovac S Trebse R Milosev I Pavlovcic V Pisot V

We present a retrospective series of 170 cemented titanium straight-stem femoral components combined with two types of femoral head: cobalt-chromium (CoCr) alloy (114 heads) and alumina ceramic (50 heads). Of the study group, 55 patients (55 stems) had died and six (six stems) were lost to follow-up. At a mean of 13.1 years (3 to 15.3) 26 stems had been revised for aseptic loosening. The mean follow-up time for stable stems was 15.1 years (12.1 to 16.6).

Survival of the stem at 15 years was 75.4% (95% confidence interval (CI) 67.3 to 83.5) with aseptic failure (including radiological failure) as the end-point, irrespective of the nature of the head and the quality of the cement mantle. Survival of the stem at 15 years was 79.1% (95% CI 69.8 to 88.4) and 67.1% (95% CI 51.3 to 82.9) with the CoCr alloy and ceramic heads, respectively. The quality of the cement mantle was graded as a function of stem coverage: stems with complete tip coverage (type 1) had an 84.9% (95% CI 77.6 to 92.2) survival at 15 years, compared with those with a poor tip coverage (type 2) which had a survival of only 22.4% (95% CI 2.4 to 42.4). The poor quality of the cement mantle and the implantation of an alumina head substantially lowered the survival of the stem.

In our opinion, further use of the cemented titanium alloy straight-stem femoral components used in our series is undesirable.


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 10 | Pages 1410 - 1415
1 Oct 2010
Castagna A Randelli M Garofalo R Maradei L Giardella A Borroni M

Total shoulder replacement is a successful procedure for degenerative or some inflammatory diseases of the shoulder. However, fixation of the glenoid seems to be the main weakness with a high rate of loosening. The results using all-polyethylene components have been better than those using metal-backed components. We describe our experience with 35 consecutive total shoulder replacements using a new metal-backed glenoid component with a mean follow-up of 75.4 months (48 to 154).

Our implant differs from others because of its mechanism of fixation. It has a convex metal-backed bone interface and the main stabilising factor is a large hollow central peg. The patients were evaulated with standard radiographs and with the Constant Score, the Simple Shoulder Test and a visual analogue scale. All the scores improved and there was no loosening, no polyethylene-glenoid disassembly and no other implant-related complications.

We conclude that a metal-backed glenoid component is a good option in total shoulder replacement with no worse results than of those using a cemented all-polyethylene prosthesis.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 3 | Pages 294 - 303
1 Mar 2009
Lindner T Kanakaris NK Marx B Cockbain A Kontakis G Giannoudis PV

Failure of fixation is a common problem in the treatment of osteoporotic fractures around the hip. The reinforcement of bone stock or of fixation of the implant may be a solution. Our study assesses the existing evidence for the use of bone substitutes in the management of these fractures in osteoporotic patients. Relevant publications were retrieved through Medline research and further scrutinised. Of 411 studies identified, 22 met the inclusion criteria, comprising 12 experimental and ten clinical reports. The clinical studies were evaluated with regard to their level of evidence. Only four were prospective and randomised.

Polymethylmethacrylate and calcium-phosphate cements increased the primary stability of the implant-bone construct in all experimental and clinical studies, although there was considerable variation in the design of the studies. In randomised, controlled studies, augmentation of intracapsular fractures of the neck of the femur with calcium-phosphate cement was associated with poor long-term results. There was a lack of data on the long-term outcome for trochanteric fractures. Because there were only a few, randomised, controlled studies, there is currently poor evidence for the use of bone cement in the treatment of fractures of the hip.


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 8 | Pages 1061 - 1065
1 Aug 2010
Cho W Cho SK Wu C

There are three basic concepts that are important to the biomechanics of pedicle screw-based instrumentation. First, the outer diameter of the screw determines pullout strength, while the inner diameter determines fatigue strength. Secondly, when inserting a pedicle screw, the dorsal cortex of the spine should not be violated and the screws on each side should converge and be of good length. Thirdly, fixation can be augmented in cases of severe osteoporosis or revision.

A trajectory parallel or caudal to the superior endplate can minimise breakage of the screw from repeated axial loading. Straight insertion of the pedicle screw in the mid-sagittal plane provides the strongest stability.

Rotational stability can be improved by adding transverse connectors. The indications for their use include anterior column instability, and the correction of rotational deformity.


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 7 | Pages 973 - 979
1 Jul 2008
Savadkoohi DG Sadeghipour P Attarian H Sardari S Eslamifar A Shokrgozar MA

Curettage and packing with polymethylmethacrylate cement is a routine treatment for giant-cell tumour (GCT) of bone. We performed an in vitro evaluation of the cytotoxic effect of a combination of cement and methotrexate, doxorubicin and cisplatin on primary cell cultures of stromal GCT cells obtained from five patients. Cement cylinders containing four different concentrations of each drug were prepared, and the effect of the eluted drugs was examined at three different time intervals.

We found that the cytotoxic effect of eluted drugs depended on their concentration and the time interval, with even the lowest dose of each drug demonstrating an acceptable rate of cytotoxicity. Even in low doses, cytotoxic drugs mixed with polymethylmethacrylate cement could therefore be considered as effective local adjuvant treatment for GCTs.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 10 | Pages 1303 - 1308
1 Oct 2007
Nixon M Taylor G Sheldon P Iqbal SJ Harper W

We matched 78 patients with a loose cemented Charnley Elite Plus total hip replacement (THR) by age, gender, race, prosthesis and time from surgery with 49 patients with a well-fixed stable hip replacement, to determine if poor bone quality predisposes to loosening. Clinical, radiological, biomechanical and bone mineral density indicators of bone quality were assessed.

Patients with loose replacements had more pain, were more likely to have presented with atrophic arthritis and to have a history of fragility fracture, narrower femoral cortices and lower peri-prosthetic or lumbar spine bone mineral density (all t-test, p < 0.01). They also tended to be smokers (chi-squared test, p = 0.08). Vitamin-D deficiency was common, but not significantly different between the two groups (t-test, p = 0.31)

In this series of cemented hip replacements performed between 1994 and 1998, aseptic loosening was associated with poor bone quality. Patients with a THR should be screened for osteoporosis and have regular radiological surveillance.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 1 | Pages 39 - 44
1 Jan 2007
Uvehammer J Kärrholm J Carlsson L

We have carried out a radiostereometric study of 50 patients (54 knees) with osteoarthritis of the knee who were randomly allocated to receive a cemented or a hydroxyapatite-coated femoral component for total knee replacement. The patients were also stratified to receive one of three types of articulating surface (standard, rotating platform, Freeman-Samuelson (FS)1000) all based on the Freeman-Samuelson design. The tibial components were cemented in all cases. Radiostereometry was performed post-operatively and at 3, 12 and 24 months. The analysis was restricted to rotation of the femoral component over time.

After two years, rotation of the femoral components in the transverse, longitudinal and sagittal planes did not differ between the cemented and the hydroxyapatite-coated implants (p = 0.2 to 0.9).

In total knee replacements with a rotating platform, the femoral component tended to tilt more posteriorly than in the other two designs, regardless of the choice of fixation (cemented or hydroxyapatite-coated, p = 0.04). The standard version of the femoral component, whether cemented or hydroxyapatite-coated, rotated more into valgus than was observed with the rotating-platform and FS1000 designs (p = 0.005). The increased constraint provided by the FS1000 component did not appear to have any adverse effect on fixation of the femoral component.


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 1 | Pages 20 - 27
1 Jan 2010
Daniel J Ziaee H Kamali A Pradhan C Band T McMinn DJW

Second-generation metal-on-metal bearings were introduced as a response to the considerable incidence of wear-induced failures associated with conventional replacements, especially in young patients. We present the results at ten years of a consecutive series of patients treated using a metal-on-metal hip resurfacing. A distinct feature of the bearings used in our series was that they had been subjected to double-heat treatments during the post-casting phase of their manufacture. In the past these bearings had not been subjected to thermal treatments, making this a unique metal-on-metal bearing which had not been used before in clinical practice.

We report the outcome of 184 consecutive hips (160 patients) treated using a hybrid-fixed metal-on-metal hip resurfacing during 1996. Patients were invited for a clinicoradiological follow-up at a minimum of ten years. The Oxford hip score and anteroposterior and lateral radiographs were obtained. The mean age at operation was 54 years (21 to 75). A series of 107 consecutive hips (99 patients) who received the same prosthesis, but subjected to a single thermal treatment after being cast, between March 1994 and December 1995, were used as a control group for comparison.

In the 1994 to 1995 group seven patients (seven hips) died from unrelated causes and there were four revisions (4%) for osteolysis and aseptic loosening. In the 1996 group nine patients died at a mean of 6.9 years after operation because of unrelated causes. There were 30 revisions (16%) at a mean of 7.3 years (1.2 to 10.9), one for infection at 1.2 years and 29 for osteolysis and aseptic loosening. Furthermore, in the latter group there were radiological signs of failure in 27 (24%) of the 111 surviving hips.

The magnitude of the problem of osteolysis and aseptic loosening in the 1996 cohort did not become obvious until five years after the operation. Our results indicate that double-heat treatments of metal-on-metal bearings can lead to an increased incidence of wear-induced osteolysis.


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 8 | Pages 1016 - 1021
1 Aug 2006
Delport HP Banks SA De Schepper J Bellemans J

Mobile-bearing posterior-stabilised knee replacements have been developed as an alternative to the standard fixed- and mobile-bearing designs. However, little is known about the in vivo kinematics of this new group of implants. We investigated 31 patients who had undergone a total knee replacement with a similar prosthetic design but with three different options: fixed-bearing posterior cruciate ligament-retaining, fixed-bearing posterior-stabilised and mobile-bearing posterior-stabilised. To do this we used a three-dimensional to two-dimensional model registration technique. Both the fixed- and mobile-bearing posterior-stabilised configurations used the same femoral component. We found that fixed-bearing posterior stabilised and mobile-bearing posterior-stabilised knee replacements demonstrated similar kinematic patterns, with consistent femoral roll-back during flexion. Mobile-bearing posterior-stabilised knee replacements demonstrated greater and more natural internal rotation of the tibia during flexion than fixed-bearing posterior-stabilised designs. Such rotation occurred at the interface between the insert and tibial tray for mobile-bearing posterior-stabilised designs. However, for fixed-bearing posterior-stabilised designs, rotation occurred at the proximal surface of the bearing. Posterior cruciate ligament-retaining knee replacements demonstrated paradoxical sliding forward of the femur.

We conclude that mobile-bearing posterior-stabilised knee replacements reproduce internal rotation of the tibia more closely during flexion than fixed-bearing posterior-stabilised designs. Furthermore, mobile-bearing posterior-stabilised knee replacements demonstrate a unidirectional movement which occurs at the upper and lower sides of the mobile insert. The femur moves in an anteroposterior direction on the upper surface of the insert, whereas the movement at the lower surface is pure rotation. Such unidirectional movement may lead to less wear when compared with the multidirectional movement seen in fixed-bearing posterior-stabilised knee replacements, and should be associated with more evenly applied cam-post stresses.


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 8 | Pages 997 - 1002
1 Aug 2006
Vendittoli P Lavigne M Girard J Roy AG

We have undertaken a prospective, randomised study to compare conservation of acetabular bone after total hip replacement and resurfacing arthroplasty of the hip. We randomly assigned 210 hips to one of the two treatment groups. Uncemented, press-fit acetabular components were used for both.

No significant difference was found in the mean diameter of acetabular implant inserted in the groups (54.74 mm for total hip replacement and 54.90 mm for resurfacing arthroplasty). In seven resurfacing procedures (6.8%), the surgeon used a larger size of component in order to match the corresponding diameter of the femoral component.

With resurfacing arthroplasty, conservation of bone is clearly advantageous on the femoral side. Our study has shown that, with a specific design of acetabular implant and by following a careful surgical technique, removal of bone on the acetabular side is comparable with that of total hip replacement.


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 11 | Pages 1475 - 1479
1 Nov 2005
Goosen JHM Verheyen CCPM Tulp NJA

We investigated the rate of polyethylene wear of a cementless acetabular component at different periods of follow-up in order to test the hypothesis than an irrecoverable deformation process (creep) was followed by an initially low, but gradually increasing wear rate. We studied prospectively 93 uncemented total hip arthroplasties in 83 patients (mean age 50 years (22 to 63)) with a mean follow-up of 8.2 years (3 to 12). We measured the penetration of the femoral head from radiographs taken immediately after surgery at three, six and nine years, or at the latest follow-up.

The median wear rate was 0.17 mm per year in the first three years, a finding which we considered to be caused by creep. Thereafter, the rate of wear declined to 0.07 mm per year (four- to six-year period) and then increased to 0.17 mm per year (seven to nine years) and 0.27 mm per year (more than nine years), which we considered to be a reflection of genuine polyethylene wear. After the nine-year follow-up the wear rates were higher in patients with marked osteolysis. We found no relationship between the inclination angle of the acetabular component or femoral head orientation and the rate of wear. No acetabular component required revision.


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 11 | Pages 1409 - 1418
1 Nov 2006
Scheerlinck T Casteleyn P

We undertook a review of the literature relating to the two basic stem designs in use in cemented hip replacement, namely loaded tapers or force-closed femoral stems, and the composite beam or shape-closed designs. The associated stem fixation theory as understood from in vitro studies and finite element modelling were examined with reference to the survivorship results for each of the concepts of fixation.

It is clear that both design principles are capable of producing successful long-term results, providing that their specific requirements of stem metallurgy, shape and surface finish, preparation of the bone and handling of the cement are observed.