Advertisement for orthosearch.org.uk
Results 61 - 67 of 67
Results per page:
The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 11_Supple_A | Pages 123 - 125
1 Nov 2012
Munro JT Garbuz DS Masri BA Duncan CP

Antibiotic impregnated articulating spacers are used in two-stage revision total knee arthroplasty to deliver local antibiotic therapy while preserving function. We have observed infection control in greater than 95% of cases with functional outcomes approaching those seen in revision for aseptic loosening. Higher failure has been observed with methicillin resistant organisms.


The Bone & Joint Journal
Vol. 95-B, Issue 8 | Pages 1075 - 1082
1 Aug 2013
Choi GW Kim HJ Yeo ED Song SY

In a retrospective study we compared 32 HINTEGRA total ankle replacements (TARs) and 35 Mobility TARs performed between July 2005 and May 2010, with a minimum follow-up of two years. The mean follow-up for the HINTEGRA group was 53 months (24 to 76) and for the Mobility group was 34 months (24 to 45). All procedures were performed by a single surgeon.

There was no significant difference between the two groups with regard to the mean AOFAS score, visual analogue score for pain or range of movement of the ankle at the latest follow-up. Most radiological measurements did not differ significantly between the two groups. However, the most common grade of heterotopic ossification (HO) was grade 3 in the HINTEGRA group (10 of 13 TARs, 76.9%) and grade 2 in the Mobility group (four of seven TARs, 57.1%) (p = 0.025). Although HO was more frequent in the HINTEGRA group (40.6%) than in the Mobility group (20.0%), this was not statistically significant (p = 0.065).The difference in peri-operative complications between the two groups was not significant, but intra-operative medial malleolar fractures occurred in four (11.4%) in the Mobility group; four (12.5%) in the HINTEGRA group and one TAR (2.9%) in the Mobility group failed (p = 0.185).

Cite this article: Bone Joint J 2013;95-B:1075–82.


The Bone & Joint Journal
Vol. 95-B, Issue 1 | Pages 31 - 37
1 Jan 2013
Zywiel MG Brandt J Overgaard CB Cheung AC Turgeon TR Syed KA

Symptomatic cobalt toxicity from a failed total hip replacement is a rare but devastating complication. It has been reported following revision of fractured ceramic components, as well as in patients with failed metal-on-metal articulations. Potential clinical findings include fatigue, weakness, hypothyroidism, cardiomyopathy, polycythaemia, visual and hearing impairment, cognitive dysfunction, and neuropathy. We report a case of an otherwise healthy 46-year-old patient, who developed progressively worsening symptoms of cobalt toxicity beginning approximately six months following synovectomy and revision of a fractured ceramic-on-ceramic total hip replacement to a metal-on-polyethylene bearing. The whole blood cobalt levels peaked at 6521 µg/l. The patient died from cobalt-induced cardiomyopathy. Implant retrieval analysis confirmed a loss of 28.3 g mass of the cobalt–chromium femoral head as a result of severe abrasive wear by ceramic particles embedded in the revision polyethylene liner. Autopsy findings were consistent with heavy metal-induced cardiomyopathy.

We recommend using new ceramics at revision to minimise the risk of wear-related cobalt toxicity following breakage of ceramic components.

Cite this article: Bone Joint J 2013;95-B:31–7.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 10 | Pages 1366 - 1371
1 Oct 2012
Pijls BG Valstar ER Kaptein BL Nelissen RGHH

This single-blinded randomised controlled trial investigated whether one design of mobile-bearing (MB) total knee replacement (TKR) has any advantage over a fixed-bearing (FB) design on long-term fixation as measured by radiostereometry. The amount of wear underneath the mobile bearing was also evaluated. A series of 42 knees was randomised to MB or FB tibial components with appropriate polyethylene inserts and followed for between ten and 12 years, or until the death of the patient. The polyethylene in the MB group was superior in that it was gamma-irradiated in inert gas and was calcium-stearate free; the polyethylene in the FB group was gamma-irradiated in air and contained calcium stearate. In theory this should be advantageous to the wear rate of the MB group. At final follow-up the overall mean migration was 0.75 mm (sd 0.76) in the MB group and 0.66 mm (sd 0.4) in the FB group, with the FB group demonstrating more posterior tilt and the MB group more internal rotation. In the FB group there was one revision for aseptic loosening, but none in the MB group. There were no significant differences in clinical or radiological scores.

For the MB group, the mean linear wear rate on the under-surface was 0.026 mm/year (sd 0.014). This was significantly smaller than the wear rate of 0.11 mm/year (sd 0.06) in the MB between femur and polyethylene (p < 0.001). Nevertheless, even in a best-case setting the mobile bearings of this TKR design had no apparent advantage in terms of fixation over the FB knee prosthesis at ten to 12 years. The wear underneath the mobile bearing was small and is unlikely to be clinically relevant.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 10 | Pages 1321 - 1329
1 Oct 2012
Sarmah SS Patel S Hossain FS Haddad FS

Radiological assessment of total and unicompartmental knee replacement remains an essential part of routine care and follow-up. Appreciation of the various measurements that can be identified radiologically is important. It is likely that routine plain radiographs will continue to be used, although there has been a trend towards using newer technologies such as CT, especially in a failing knee, where it provides more detailed information, albeit with a higher radiation exposure.

The purpose of this paper is to outline the radiological parameters used to evaluate knee replacements, describe how these are measured or classified, and review the current literature to determine their efficacy where possible.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 5 | Pages 671 - 677
1 May 2012
Raiss P Sowa B Bruckner T Eck S Woerz S Rohr K Rickert M Kasten P

The aim of this study was to compare a third-generation cementing procedure for glenoid components with a new technique for cement pressurisation. In 20 pairs of scapulae, 20 keeled and 20 pegged glenoid components were implanted using either a third-generation cementing technique (group 1) or a new pressuriser (group 2). Cement penetration was measured by three-dimensional (3D) analysis of micro-CT scans. The mean 3D depth of penetration of the cement was significantly greater in group 2 (p < 0.001). The mean thickness of the cement mantle for keeled glenoids was 2.50 mm (2.0 to 3.3) in group 1 and 5.18 mm (4.4 to 6.1) in group 2, and for pegged glenoids it was 1.72 mm (0.9 to 2.3) in group 1 and 5.63 mm (3.6 to 6.4) in group 2. A cement mantle < 2 mm was detected less frequently in group 2 (p < 0.001). Using the cement pressuriser the proportion of cement mantles < 2 mm was significantly reduced compared with the third-generation cementing technique.


Bone & Joint Research
Vol. 1, Issue 8 | Pages 180 - 191
1 Aug 2012
Stilling M Kold S de Raedt S Andersen NT Rahbek O Søballe K

Objectives

The accuracy and precision of two new methods of model-based radiostereometric analysis (RSA) were hypothesised to be superior to a plain radiograph method in the assessment of polyethylene (PE) wear.

Methods

A phantom device was constructed to simulate three-dimensional (3D) PE wear. Images were obtained consecutively for each simulated wear position for each modality. Three commercially available packages were evaluated: model-based RSA using laser-scanned cup models (MB-RSA), model-based RSA using computer-generated elementary geometrical shape models (EGS-RSA), and PolyWare. Precision (95% repeatability limits) and accuracy (Root Mean Square Errors) for two-dimensional (2D) and 3D wear measurements were assessed.