Impaction bone grafting for the reconstitution
of bone stock in revision hip surgery has been used for nearly 30 years.
Between 1995 and 2001 we used this technique in acetabular reconstruction,
in combination with a cemented component, in 304 hips in 292 patients
revised for aseptic loosening. The only additional supports used
were stainless steel meshes placed against the medial wall or laterally
around the acetabular rim to contain the graft. All Paprosky grades
of defect were included. Clinical and radiographic outcomes were
collected in surviving patients at a minimum of ten years after
the index operation. Mean follow-up was 12.4 years ( Cite this article:
The custom triflange is a patient-specific implant
for the treatment of severe bone loss in revision total hip arthroplasty
(THA). Through a process of three-dimensional modelling and prototyping,
a hydroxyapatite-coated component is created for acetabular reconstruction.
There are seven level IV studies describing the clinical results
of triflange components. The most common complications include dislocation
and infection, although the rates of implant removal are low. Clinical
results are promising given the challenging problem. We describe
the design, manufacture and implantation process and review the
clinical results, contrasting them to other methods of acetabular
reconstruction in revision THA. Cite this article:
Patients from a randomised trial on resurfacing
hip arthroplasty (RHA) (n = 36, 19 males; median age 57 years, 24
to 65) comparing a conventional 28 mm metal-on-metal total hip arthroplasty
(MoM THA) (n = 28, 17 males; median age 59 years, 37 to 65) and
a matched control group of asymptomatic patients with a 32 mm ceramic-on-polyethylene
(CoP) THA (n = 33, 18 males; median age 63 years, 38 to 71) were
cross-sectionally screened with metal artefact reducing sequence-MRI
(MARS-MRI) for pseudotumour formation at a median of 55 months (23
to 72) post-operatively. MRIs were scored by consensus according
to three different classification systems for pseudotumour formation. Clinical scores were available for all patients and metal ion
levels for MoM bearing patients. Periprosthetic lesions with a median volume of 16 mL (1.5 to
35.9) were diagnosed in six patients in the RHA group (17%), one
in the MoM THA group (4%) and six in the CoP group (18%). The classification
systems revealed no clear differences between the groups. Solid
lesions (n = 3) were exclusively encountered in the RHA group. Two patients
in the RHA group and one in the MoM THA group underwent a revision
for pseudotumour formation. There was no statistically significant
relationship between clinical scoring, metal ion levels and periprosthetic
lesions in any of the groups. Periprosthetic fluid collections are seen on MARS-MRI after conventional
CoP THA and RHA and may reflect a soft-tissue collection or effusion. Currently available MRI classification systems seem to score
these collections as pseudotumours, causing an-overestimatation
of the incidence of pseudotumours. Cite this article:
Large-head metal-on-metal (MoM) total hip replacements
(THR) have given rise to concern. Comparative studies of small-head
MoM THRs over a longer follow-up period are lacking. Our objective
was to compare the incidence of complications such as infection,
dislocation, revision, adverse local tissue reactions, mortality
and radiological and clinical outcomes in small-head (28 mm) MoM
and ceramic-on-polyethylene (CoP) THRs up to 12 years post-operatively. A prospective cohort study included 3341 THRs in 2714 patients.
The mean age was 69.1 years (range 24 to 98) and 1848 (55.3%) were
performed in women, with a mean follow-up of 115 months (18 to 201).
There were 883 MoM and 2458 CoP bearings. Crude incidence rates
(cases/1000 person-years) were: infection 1.3 In conclusion, we found similar results for small-head MoM and
CoP bearings up to ten years post-operatively, but after ten years
MoM THRs had a higher risk of all-cause revision. Furthermore, the
presence of an adverse response to metal debris seen in the small-head
MOM group at revision is a cause for concern. Cite this article:
Acetabular bone loss is a challenging problem
facing the revision total hip replacement surgeon. Reconstruction
of the acetabulum depends on the presence of anterosuperior and
posteroinferior pelvic column support for component fixation and
stability. The Paprosky classification is most commonly used when
determining the location and degree of acetabular bone loss. Augments
serve the function of either providing primary construct stability
or supplementary fixation. When a pelvic discontinuity is encountered we advocate the use
of an acetabular distraction technique with a jumbo cup and modular
porous metal acetabular augments for the treatment of severe acetabular
bone loss and associated chronic pelvic discontinuity. Cite this article:
The purpose of this study was to compare the long-term results
of primary total hip arthroplasty (THA) in young patients using
either a conventional (CPE) or a highly cross-linked (HXLPE) polyethylene
liner in terms of functional outcome, incidence of osteolysis, radiological
wear and rate of revision. We included all patients between the ages of 45 and 65 years
who, between January 2000 and December 2001, had undergone a primary
THA for osteoarthritis at our hospital using a CPE or HXLPE acetabular
liner and a 28 mm cobalt-chrome femoral head. From a total of 160 patients, 158 (177 hips) were available for
review (CPE 89; XLPE 88). The mean age, body mass index (BMI) and
follow-up in each group were: CPE: 56.8 years (46 to 65); 30.7 kg/m2 (19
to 58); 13.2 years (2.1 to 14.7) and HXLPE: 55.6 years (45 to 65);
BMI: 30 kg/m2 (18 to 51); 13.1 years (5.7 to 14.4).Aims
Methods
The National Joint Registry for England, Wales and Northern Ireland
(NJR) has extended its scope to report on hospital, surgeon and
implant performance. Data linkage of the NJR to the London Implant
Retrieval Centre (LIRC) has previously evaluated data quality for
hip primary procedures, but did not assess revision records. We analysed metal-on-metal hip revision procedures performed
between 2003 and 2013. A total of 69 929 revision procedures from
the NJR and 929 revised pairs of components from the LIRC were included.Aims
Methods
We retrospectively reviewed 44 consecutive patients
(50 hips) who underwent acetabular re-revision after a failed previous
revision that had been performed using structural or morcellised
allograft bone, with a cage or ring for uncontained defects. Of
the 50 previous revisions, 41 cages and nine rings were used with
allografts for 14 minor-column and 36 major-column defects. We routinely
assessed the size of the acetabular bone defect at the time of revision
and re-revision surgery. This allowed us to assess whether host
bone stock was restored. We also assessed the outcome of re-revision
surgery in these circumstances by means of radiological characteristics,
rates of failure and modes of failure. We subsequently investigated
the factors that may affect the potential for the restoration of bone
stock and the durability of the re-revision reconstruction using
multivariate analysis. At the time of re-revision, there were ten host acetabula with
no significant defects, 14 with contained defects, nine with minor-column,
seven with major-column defects and ten with pelvic discontinuity.
When bone defects at re-revision were compared with those at the
previous revision, there was restoration of bone stock in 31 hips, deterioration
of bone stock in nine and remained unchanged in ten. This was a
significant improvement (p <
0.001). Morselised allografting
at the index revision was not associated with the restoration of
bone stock. In 17 hips (34%), re-revision was possible using a simple acetabular
component without allograft, augments, rings or cages. There were
47 patients with a mean follow-up of 70 months (6 to 146) available
for survival analysis. Within this group, the successful cases had
a minimum follow-up of two years after re-revision. There were 22 clinical
or radiological failures (46.7%), 18 of which were due to aseptic
loosening. The five and ten year Kaplan–Meier survival rate was
75% (95% CI, 60 to 86) and 56% (95% CI, 40 to 70) respectively with
aseptic loosening as the endpoint. The rate of aseptic loosening
was higher for hips with pelvic discontinuity (p = 0.049) and less
when the allograft had been in place for longer periods (p = 0.040). The use of a cage or ring over structural allograft bone for
massive uncontained defects in acetabular revision can restore host
bone stock and facilitate subsequent re-revision surgery to a certain
extent. Cite this article:
The aim of this study was to establish what happens to patients
in the long term after endoprosthetic replacement for a primary
malignant tumour of bone. We conducted a retrospective analysis of a prospectively maintained
database to identify all patients who had undergone an endoprosthetic
replacement more than 25 years ago and who were still alive. Their
outcomes were investigated with reference to their complications and
need for further surgery. A total of 230 patients were identified.
Their mean age at diagnosis was 20.7 years (five to 62). The most
common diagnosis was osteosarcoma (132). The most common site was
the distal femur (102). Aims
Patients and Methods
We reviewed the literature on the currently available
choices of bearing surface in total hip replacement (THR). We present
a detailed description of the properties of articulating surfaces
review the understanding of the advantages and disadvantages of
existing bearing couples. Recent technological developments in the
field of polyethylene and ceramics have altered the risk of fracture
and the rate of wear, although the use of metal-on-metal bearings has
largely fallen out of favour, owing to concerns about reactions
to metal debris. As expected, all bearing surface combinations have
advantages and disadvantages. A patient-based approach is recommended,
balancing the risks of different options against an individual’s
functional demands. Cite this article:
Dysplasia of the hip, hypotonia, osteopenia,
ligamentous laxity, and mental retardation increase the complexity
of performing and managing patients with Down syndrome who require
total hip replacement (THR). We identified 14 patients (six males,
eight females, 21 hips) with Down syndrome and degenerative disease
of the hip who underwent THR, with a minimum follow-up of two years
from 1969 to 2009. In seven patients, bilateral THRs were performed
while the rest had unilateral THRs. The mean clinical follow-up
was 5.8 years (standard deviation ( Cite this article:
Revision of a cemented hemiarthroplasty of the
hip may be a hazardous procedure with high rates of intra-operative complications.
Removing well-fixed cement is time consuming and risks damaging
already weak bone or perforating the femoral shaft. The cement-in-cement
method avoids removal of intact cement and has shown good results
when used for revision total hip arthroplasty (THA). The use of
this technique for the revision of a hemiarthroplasty to THA has
not been previously reported. A total of 28 consecutive hemiarthroplasties (in 28 patients)
were revised to a THA using an Exeter stem and the cement-in-cement
technique. There were four men and 24 women; their mean age was
80 years (35 to 93). Clinical and radiographic data, as well as
operative notes, were collected prospectively and no patient was
lost to follow-up. Four patients died within two years of surgery. The mean follow
up of the remainder was 70 months (25 to 124). Intra-operatively
there was one proximal perforation, one crack of the
femoral calcar and one acetabular fracture. No femoral components
have required subsequent revision for aseptic loosening or are radiologically loose. Four patients with late complications (14%) have since undergone
surgery (two for a peri-prosthetic fracture, and one each for deep
infection and recurrent dislocation) resulting in an overall major
rate of complication of 35.7%. The cement-in-cement technique provides reliable femoral fixation
in this elderly population and may reduce operating time and rates
of complication. Cite this article:
In 2005, we demonstrated that the polished triple-tapered
C-stem at two years had migrated distally and rotated internally.
From that series, 33 patients have now been followed radiologically,
clinically and by radiostereometric analysis (RSA) for up to ten
years. The distal migration within the cement mantle had continued
and reached a mean of 2 mm (0.5 to 4.0) at ten years. Internal rotation,
also within the cement mantle, was a mean 3.8° (external 1.6° to
internal 6.6°) The cement mantle did not show any sign of migration
or loosening in relation to the femoral bone. There were no clinical
or radiological signs indicating that the migration or rotation
within the cement mantle had had any adverse effects for the patients. Cite this article:
We report a 12- to 15-year implant survival assessment
of a prospective single-surgeon series of Birmingham Hip Resurfacings
(BHRs). The earliest 1000 consecutive BHRs including 288 women (335
hips) and 598 men (665 hips) of all ages and diagnoses with no exclusions
were prospectively followed-up with postal questionnaires, of whom
the first 402 BHRs (350 patients) also had clinical and radiological
review. Mean follow-up was 13.7 years (12.3 to 15.3). In total, 59 patients
(68 hips) died 0.7 to 12.6 years following surgery from unrelated
causes. There were 38 revisions, 0.1 to 13.9 years (median 8.7)
following operation, including 17 femoral failures (1.7%) and seven
each of infections, soft-tissue reactions and other causes. With
revision for any reason as the end-point Kaplan–Meier survival analysis
showed 97.4% (95% confidence interval (CI) 96.9 to 97.9) and 95.8%
(95% CI 95.1 to 96.5) survival at ten and 15 years, respectively.
Radiological assessment showed 11 (3.5%) femoral and 13 (4.1%) acetabular
radiolucencies which were not deemed failures and one radiological
femoral failure (0.3%). Our study shows that the performance of the BHR continues to
be good at 12- to 15-year follow-up. Men have better implant survival
(98.0%; 95% CI 97.4 to 98.6) at 15 years than women (91.5%; 95%
CI 89.8 to 93.2), and women <
60 years (90.5%; 95% CI 88.3 to
92.7) fare worse than others. Hip dysplasia and osteonecrosis are
risk factors for failure. Patients under 50 years with osteoarthritis
fare best (99.4%; 95% CI 98.8 to 100 survival at 15 years), with
no failures in men in this group. Cite this article:
We report the clinical and radiological outcomes
of a series of contemporary cementless ceramic-on-ceramic total hip
replacements (THRs) at ten years in patients aged ≤ 55 years of
age. Pre- and post-operative activity levels are described. A total
of 120 consecutive ceramic cementless THRs were performed at a single
centre in 110 patients from 1997 to 1999. The mean age of the patients
at operation was 45 years (20 to 55). At ten years, four patients
had died and six were lost to follow-up, comprising ten hips. The
mean post-operative Harris hip score was 94.7 (55 to 100). Radiological
analysis was undertaken in 90 available THRs of the surviving 106
hips at final review: all had evidence of stable bony ingrowth,
with no cases of osteolysis. Wear was undetectable. There were four
revisions. The survival for both components with revision for any
cause as an endpoint was 96.5% (95% confidence interval 94.5 to
98.7). The mean modified University of California, Los Angeles activity
level rose from a mean of 6.4 (4 to 10) pre-operatively to 9.0 (6
to 10) at the ten-year post-operative period. Alumina ceramic-on-ceramic bearings in cementless primary THR
in this series have resulted in good clinical and radiological outcomes
with undetectable rates of wear and excellent function in the demanding
younger patient group at ten years. Cite this article:
The use of ilioischial cage reconstruction for
pelvic discontinuity has been replaced by the Trabecular Metal (Zimmer,
Warsaw, Indiana) cup-cage technique in our institution, due to the
unsatisfactory outcome of using a cage alone in this situation.
We report the outcome of 26 pelvic discontinuities in 24 patients
(20 women and four men, mean age 65 years (44 to 84)) treated by
the cup-cage technique at a mean follow-up of 82 months (12 to 113)
and compared them with a series of 19 pelvic discontinuities in
19 patients (18 women and one man, mean age 70 years (42 to 86))
treated with a cage at a mean follow-up of 69 months (1 to 170).
The clinical and radiological outcomes as well as the survivorship
of the groups were compared. In all, four of the cup-cage group
(15%) and 13 (68%) of the cage group failed due to septic or aseptic
loosening. The seven-year survivorship was 87.2% (95% confidence interval
(CI) 71 to 103) for the cup-cage group and 49.9% (95% CI 15 to 84)
for the cage-alone group (p = 0.009). There were four major complications
in the cup-cage group and nine in the cage group. Radiological union
of the discontinuity was found in all successful cases in the cup-cage
group and three of the successful cage cases. Three hips in the
cup-cage group developed early radiological migration of the components,
which stabilised with a successful outcome. Cup-cage reconstruction is a reliable technique for treating
pelvic discontinuity in mid-term follow-up and is preferred to ilioischial
cage reconstruction. If the continuity of the bone graft at the
discontinuity site is not disrupted, early migration of the components
does not necessarily result in failure. Cite this article:
There are few reports describing the technique
of managing acetabular protrusio in primary total hip replacement. Most
are small series with different methods of addressing the challenges
of significant medial and proximal migration of the joint centre,
deficient medial bone and reduced peripheral bony support to the
acetabular component. We describe our technique and the clinical
and radiological outcome of using impacted morsellised autograft
with a porous-coated cementless cup in 30 primary THRs with mild
(n = 8), moderate (n = 10) and severe (n = 12) grades of acetabular
protrusio. The mean Harris hip score had improved from 52 pre-operatively
to 85 at a mean follow-up of 4.2 years (2 to 10). At final follow-up,
27 hips (90%) had a good or excellent result, two (7%) had a fair
result and one (3%) had a poor result. All bone grafts had united
by the sixth post-operative month and none of the hips showed any
radiological evidence of recurrence of protrusio, osteolysis or
loosening. By using impacted morsellised autograft and cementless
acetabular components it was possible to achieve restoration of
hip mechanics, provide a biological solution to bone deficiency
and ensure long-term fixation without recurrence in arthritic hips
with protrusio undergoing THR. Cite this article:
The aim of this study was to assess the effect
of frictional torque and bending moment on fretting corrosion at
the taper interface of a modular femoral component and to investigate
whether different combinations of material also had an effect. The
combinations we examined were 1) cobalt–chromium (CoCr) heads on
CoCr stems 2) CoCr heads on titanium alloy (Ti) stems and 3) ceramic
heads on CoCr stems. In test 1 increasing torque was imposed by offsetting the stem
in the anteroposterior plane in increments of 0 mm, 4 mm, 6 mm and
8 mm when the torque generated was equivalent to 0 Nm, 9 Nm, 14
Nm and 18 Nm. In test 2 we investigated the effect of increasing the bending
moment by offsetting the application of axial load from the midline
in the mediolateral plane. Increments of offset equivalent to head
+ 0 mm, head + 7 mm and head + 14 mm were used. Significantly higher currents and amplitudes were seen with increasing
torque for all combinations of material. However, Ti stems showed
the highest corrosion currents. Increased bending moments associated
with using larger offset heads produced more corrosion: Ti stems
generally performed worse than CoCr stems. Using ceramic heads did
not prevent corrosion, but reduced it significantly in all loading
configurations. Cite this article:
Squeaking arising from a ceramic-on-ceramic (CoC)
total hip replacement (THR) may cause patient concern and in some
cases causes patients to seek revision surgery. We performed a meta-analysis
to determine the incidence of squeaking and the incidence of revision
surgery for squeaking. A total of 43 studies including 16 828 CoC
THR that reported squeaking, or revision for squeaking, were entered
into the analysis. The incidence of squeaking was 4.2% and the incidence
of revision for squeaking was 0.2%. The incidence of squeaking in
patients receiving the Accolade femoral stem was 8.3%, and the incidence
of revision for squeaking in these patients was 1.3%. Cite this article:
Femoral stem version has a major influence on
impingement and early post-operative stability after total hip arthroplasty
(THA). The main objective of this study was to evaluate the validity
of a novel radiological method for measuring stem version. Anteroposterior
(AP) radiographs and three-dimensional CT scans were obtained for
115 patients (female/male 63/72, mean age 62.5 years (50 to 75))
who had undergone minimally invasive, cementless THA. Stem version was
calculated from the AP hip radiograph by rotation-based change in
the projected prosthetic neck–shaft (NSA*) angle using the mathematical
formula ST = arcos [tan (NSA*) / tan (135)]. We used two independent
observers who repeated the analysis after a six-week interval. Radiological
measurements were compared with 3D-CT measurements by an independent,
blinded external institute. We found a mean difference of 1.2° ( We found that femoral tilt was associated with the mean radiological
measurement error (r = 0.22, p = 0.02). The projected neck–shaft angle is a reliable method for measuring
stem version on AP radiographs of the hip after a THA. However,
a highly standardised radiological technique is required for its
precise measurement. Cite this article: