A variety of radiological methods of measuring
version of the acetabular component after total hip replacement (THR)
have been described. The aim of this study was to evaluate the reliability
and validity of six methods (those of Lewinnek; Widmer; Hassan et
al; Ackland, Bourne and Uhthoff; Liaw et al; and Woo and Morrey)
that are currently in use. In 36 consecutive patients who underwent
THR, version of the acetabular component was measured by three independent
examiners on plain radiographs using these six methods and compared
with measurements using CT scans. The intra- and interobserver reliabilities
of each measurement were estimated. All measurements on both radiographs
and CT scans had excellent intra- and interobserver reliability
and the results from each of the six methods correlated well with
the CT measurements. However, measurements made using the methods
of Widmer and of Ackland, Bourne and Uhthoff were significantly
different from the CT measurements (both p <
0.001), whereas
measurements made using the remaining four methods were similar
to the CT measurements. With regard to reliability and convergent
validity, we recommend the use of the methods described by Lewinnek,
Hassan et al, Liaw et al and Woo and Morrey for measurement of version
of the acetabular component.
We report the use of porous metal acetabular
revision shells in the treatment of contained bone loss. The outcomes of
53 patients with
The purpose of this study was to assess N-acetyl aspartate changes
in the thalamus in patients with osteoarthritis of the hip using
proton magnetic resonance spectroscopy. Nine patients with osteoarthritis of the hip (symptomatic group,
nine women; mean age 61.4 years (48 to 78)) and nine healthy volunteers
(control group, six men, three women; mean age 30.0 years (26 to
38)) underwent proton magnetic resonance spectroscopy to assess
the changes of N-acetyl aspartate in the thalamus. Objectives
Methods
Peripheral nerve injury is an uncommon but serious
complication of hip surgery that can adversely affect the outcome.
Several studies have described the use of electromyography and intra-operative
sensory evoked potentials for early warning of nerve injury. We
assessed the results of multimodal intra-operative monitoring during
complex hip surgery. We retrospectively analysed data collected
between 2001 and 2010 from 69 patients who underwent complex hip
surgery by a single surgeon using multimodal intra-operative monitoring
from a total pool of 7894 patients who underwent hip surgery during
this period. In 24 (35%) procedures the surgeon was alerted to a
possible lesion to the sciatic and/or femoral nerve. Alerts were
observed most frequently during peri-acetabular osteotomy. The surgeon
adapted his approach based on interpretation of the neurophysiological changes.
From 69 monitored surgical procedures, there was only one true positive
case of post-operative nerve injury. There were no false positives
or false negatives, and the remaining 68 cases were all true negative.
The sensitivity for predicting post-operative nerve injury was 100%
and the specificity 100%. We conclude that it is possible and appropriate
to use this method during complex hip surgery and it is effective
for alerting the surgeon to the possibility of nerve injury.
We report the long-term results of revision total
hip replacement using femoral impaction allografting with both uncemented
and cemented Freeman femoral components. A standard design of component
was used in both groups, with additional proximal hydroxyapatite
coating in the uncemented group. A total of 33 hips in 30 patients received
an uncemented component and 31 hips in 30 patients a cemented component.
The mean follow-up was 9.8 years (2 to 17) in the uncemented group
and 6.2 years (1 to 11) in the cemented group. Revision procedures
(for all causes) were required in four patients (four hips) in the
uncemented group and in five patients (five hips) in the cemented
group. Harris hip scores improved significantly in both groups and
were maintained independently of the extent of any migration of
the femoral component within the graft or graft–cement mantle.
We obtained pre-operative and six-month post-operative
Oxford hip (OHS) and knee scores (OKS) for 1523 patients who underwent
total hip replacement and 1784 patients who underwent total knee
replacement. They all also completed a six-month satisfaction question. Scatter plots showed no relationship between pre-operative Oxford
scores and six-month satisfaction scores. Spearman’s rank correlation
coefficients were -0.04 (95% confidence interval (CI) -0.09 to 0.01)
between OHS and satisfaction and 0.04 (95% CI -0.01 to 0.08) between
OKS and satisfaction. A receiver operating characteristic (ROC) curve
analysis was used to identify a cut-off point for the pre-operative
OHS/OKS that identifies whether or not a patient is satisfied with
surgery. We obtained an area under the ROC curve of 0.51 (95% CI
0.45 to 0.56) for hip replacement and 0.56 (95% CI 0.51 to 0.60)
for knee replacement, indicating that pre-operative Oxford scores
have no predictive accuracy in distinguishing satisfied from dissatisfied
patients. In the NHS widespread attempts are being made to use patient-reported
outcome measures (PROMs) data for the purpose of prioritising patients
for surgery. Oxford hip and knee scores have no predictive accuracy
in relation to post-operative patient satisfaction. This evidence
does not support their current use in prioritising access to care.
A 70-year-old man with an uncemented metal-on-polyethylene
total hip prosthesis underwent revision arthroplasty 33 months later
because of pain, swelling and recurrent dislocation. There appeared
to be corrosion and metal release from the prosthetic head, resulting
in pseudotumour formation and severe local soft-tissue destruction.
The corrosion occurred at the junction between the titanium-molybdenum-zirconium-iron
taper and the cobalt-chrome-molybdenum head, but the mechanism was unproven.
We evaluated the incidence of heterotopic ossification
following total ankle replacement to determine whether the degree
of ossification was associated with the clinical outcome. We evaluated
90 ankles in 81 consecutive patients who underwent total ankle replacement,
and heterotopic ossification was assessed according to proportional involvement
of the ankle joint. Correlation analysis was used to investigate
the association between heterotopic ossification and outcome. No significant association was found between the formation of
heterotopic ossification and the clinical outcome. The degree of
heterotopic ossification in the posterior ankle joint was not significantly
correlated with posterior ankle pain (p = 0.929), the American Orthopaedic
Foot and Ankle Society score (p = 0.454) or range of movement (p
= 0.283). This study indicates that caution should be observed in attributing
symptoms and functional limitation to the presence of heterotopic
ossification in the posterior ankle joint when considering excision
of heterotopic bone after total ankle replacement.
The aim of this study was to investigate the
possible benefit of large-head metal-on-metal bearing on a stem
for primary hip replacement compared with a 28 mm diameter conventional
metal-on-polyethylene bearing in a prospective randomised controlled
trial. We investigated cemented stem behaviour between these two
different bearings using Einzel-Bild-Röntgen-Analyse, clinical and
patient reported measures (Harris hip score, Western Ontario and
McMaster Universities osteoarthritis index, Short Form-36 and satisfaction)
and whole blood metal ion levels at two years. A power study indicated
that 50 hips were needed in each group to detect subsidence of >
5 mm at two years with a
p-value of <
0.05. Significant improvement (p <
0.001) was found in the mean
clinical and patient reported outcomes at two years for both groups.
Comparison of outcomes between the groups at two years showed no
statistically significant difference for mean stem migration, clinical
and patient reported outcomes; except overall patient satisfaction which
was higher for metal-on-metal group (p = 0.05). Metal ion levels
were raised above the Medicines and Healthcare products Regulatory
Agency advised safety level (7 µg per litre) in 20% of the metal-on-metal
group and in one patient in metal-on-polyethylene group (who had
a metal-on-metal implant on the contralateral side). Two patients
in the metal-on-metal group were revised, one for pseudotumour and
one for peri-prosthetic fracture. Use of large modular heads is associated with a risk of raised
whole blood metal ion levels despite using a proven bearing from
resurfacing. The head-neck junction or excess stem micromotion are
possibly the weak links warranting further research.
Antibiotic prophylaxis is routinely administered during joint replacement surgery and may predispose patients to
Free vascularised fibular grafting has been reported
to be successful for adult patients with osteonecrosis of the femoral
head (ONFH). However, its benefit in teenage patients with post-traumatic
ONFH has not been determined. We evaluated the effectiveness of
free vascularised fibular grafting in the treatment of this condition
in children and adolescents. We retrospectively analysed 28 hips
in 28 patients in whom an osteonecrotic femoral head had been treated
with free vascularised fibular grafting between 2002 and 2008. Their
mean age was 16.3 years (13 to 19). The stage of the disease at
time of surgery, and results of treatment including pre- and post-operative
Harris hip scores, were studied. We defined clinical failure as
conversion to total hip replacement. All patients were followed
up for a mean of four years (2 to 7). The mean Harris hip score
improved from 60.4 (37 to 84) pre-operatively to 94.2 (87 to 100)
at final follow-up. At the latest follow-up we found improved or
unchanged radiographs in all four initially stage II hips and in
23 of 24 stage III or IV hips. Only one hip (stage V) deteriorated.
No patient underwent total hip replacement. Free vascularised fibular grafting is indicated for the treatment
of post-traumatic ONFH in teenage patients.
We report the results of 79 patients (81 hips)
who underwent impaction grafting at revision hip replacement using the
Exeter femoral stem. Their mean age was 64 years (31 to 83). According
to the Endoklinik classification, 20 hips had a type 2 bone defect,
40 had type 3, and 21 had type 4. The mean follow-up for unrevised
stems was 10.4 years (5 to 17). There were 12 re-operations due to intra- and post-operative
fractures, infection (one hip) and aseptic loosening (one hip).
All re-operations affected type 3 (6 hips) and 4 (6 hips) bone defects.
The survival rate for re-operation for any cause was 100% for type
2, 81.2% (95% confidence interval (CI) 67.1 to 95.3) for type 3,
and 70.8% (95% CI 51.1 to 90.5) for type 4 defects at 14 years.
The survival rate with further revision for aseptic loosening as
the end point was 98.6% (95% CI 95.8 to 100). The final clinical
score was higher for patients with type 2 bone defects than type
4 regarding pain, function and range of movement. Limp was most
frequent in the type 4 group (p <
0.001). The mean subsidence
of the stem was 2.3 mm ( The impacted bone grafting technique has good clinical results
in femoral revision. However, major bone defects affect clinical
outcome and also result in more operative complications.
Arthroscopy of the knee is one of the most commonly
performed orthopaedic procedures worldwide. Large-volume outcome
data have not previously been available for English NHS patients.
Prospectively collected admissions data, routinely collected on
every English NHS patient, were analysed to determine the rates
of complications within 30 days (including re-operation and re-admission),
90-day symptomatic venous thromboembolism and all-cause mortality.
There were 301 701 operations performed between 2005 and 2010 –
an annual incidence of 9.9 per 10 000 English population. Of these,
16 552 (6%) underwent ligament reconstruction and 106 793 (35%)
underwent meniscal surgery. The 30-day re-admission rate was 0.64%
(1662) and 30-day wound complication rate was 0.26% (677). The overall
30-day re-operation rate was 0.40% (1033) and the 90-day pulmonary embolism
rate was 0.08% (230), of which six patients died. 90-day mortality
was 0.02% (47). Age <
40 years, male gender and ligament reconstruction
were significantly associated with an increased rate of 30-day re-admission
and unplanned re-operation. In addition, a significant increase
in 30-day admission rates were seen with Charlson comorbidity scores
of 1 (p = 0.037) and ≥ 2 (p <
0.001) compared with scores of
0, and medium volume units compared with high volume units (p <
0.001). Complications following arthroscopy of the knee are rare. It
is a safe procedure, which in the majority of cases is performed
as day case surgery. These data can be used for quality benchmarking,
in terms of consent, consultant re-validation and individual unit
performance.
There is widespread concern regarding the incidence of adverse soft-tissue reactions after metal-on-metal (MoM) hip replacement. Recent National Joint Registry data have shown clear differences in the rates of failure of different designs of hip resurfacing. Our aim was to update the failure rates related to metal debris for the Articular Surface Replacement (ASR). A total of 505 of these were implanted. Kaplan-Meier analysis showed a failure rate of 25% at six years for the ASR resurfacing and of 48.8% for the ASR total hip replacement (THR). Of 257 patients with a minimum follow-up of two years, 67 (26.1%) had a serum cobalt concentration which was greater than 7 μg/l. Co-ordinate measuring machine analysis of revised components showed that all patients suffering adverse tissue reactions in the resurfacing group had abnormal wear of the bearing surfaces. Six THR patients had relatively low rates of articular wear, but were found to have considerable damage at the trunion-taper interface. Our results suggest that wear at the modular junction is an important factor in the development of adverse tissue reactions after implantation of a large-diameter MoM THR.
We prospectively followed 191 consecutive collarless
polished tapered (CPT) femoral stems, implanted in 175 patients
who had a mean age at operation of 64.5 years (21 to 85). At a mean
follow-up of 15.9 years (14 to 17.5), 86 patients (95 hips) were
still alive. The fate of all original stems is known. The 16-year
survivorship with re-operation for any reason was 80.7% (95% confidence
interval 72 to 89.4). There was no loss to follow-up, with clinical
data available on all 95 hips and radiological assessment performed
on 90 hips (95%). At latest follow-up, the mean Harris hip score
was 78 (28 to 100) and the mean Oxford hip score was 36 (15 to 48).
Stems subsided within the cement mantle, with a mean subsidence
of 2.1 mm (0.4 to 19.2). Among the original cohort, only one stem
(0.5%) has been revised due to aseptic loosening. In total seven
stems were revised for any cause, of which four revisions were required
for infection following revision of the acetabular component. A
total of 21 patients (11%) required some sort of revision procedure;
all except three of these resulted from failure of the acetabular
component. Cemented acetabular components had a significantly lower
revision burden (three hips, 2.7%) than Harris Galante uncemented
components (17 hips, 21.8%) (p <
0.001). The CPT stem continues to provide excellent radiological and
clinical outcomes at 15 years following implantation. Its results
are consistent with other polished tapered stem designs.
The objective of this study was to determine the effectiveness of screening and successful treatment of methicillin-resistant We screened 5933 elective orthopaedic in-patients for MRSA at pre-operative assessment. Of these, 108 (1.8%) were colonised with MRSA and 90 subsequently underwent surgery. Despite effective eradication therapy, six of these (6.7%) had an SSI within one year of surgery. Among these infections, deep sepsis occurred in four cases (4.4%) and superficial infection in two (2.2%). The responsible organism in four of the six cases was MRSA. Further analysis showed that patients undergoing surgery for joint replacement of the lower limb were at significantly increased risk of an SSI if previously colonised with MRSA. We conclude that previously MRSA-colonised patients undergoing elective surgery are at an increased risk of an SSI compared with other elective patients, and that this risk is significant for those undergoing joint replacement of the lower limb. Furthermore, when an infection occurs, it is likely to be due to MRSA.
We present the development and results of a nationwide, prospective, observational follow-up programme including patient-reported outcome measures (PROMs) for the Swedish Hip Arthroplasty Register. The programme started in 2002 and has gradually expanded to include all units performing total hip replacement in Sweden. The self-administered PROMs protocol comprises the EQ-5D instrument, the Charnley class categorisation and visual analogue scales for pain and satisfaction. These current analyses include 34 960 total hip replacements with complete pre- and one-year post-operative questionnaires. Patients eligible for total hip replacement generally report low health-related quality of life and suffer from pain. One year post-operatively the mean EQ-5D index increased to above the level of an age- and gender-matched population, with a considerable reduction of pain (p <
0.001). Females, younger patients and those with Charnley category C reported a lower EQ-5D index pre-operatively than males, older patients and Charnley category A or B, respectively (all p <
0.001). In a multivariable regression analysis Charnley category C, male gender and higher age were associated with less improvement in health-related quality of life (p <
0.001). Nationwide implementation of a PROMs programme requires a structured organisation and effective data capture. Patients’ response rates to the Registry are good. The continuous collection of PROMs permits local and national improvement work and allows for further health-economic evaluation.
Bilateral simultaneous total knee replacement (TKR) has been considered by some to be associated with increased morbidity and mortality. Our study analysed the outcome of 150 consecutive, but selected, bilateral simultaneous TKRs and compared them with that of 271 unilateral TKRs in a standardised fast-track setting. The procedures were performed between 2003 and 2009. Apart from staying longer in hospital (mean 4.7 days (2 to 16)
More than a million hip replacements are carried out each year worldwide, and the number of other artificial joints inserted is also rising, so that infections associated with arthroplasties have become more common. However, there is a paucity of literature on infections due to haematogenous seeding following dental procedures. We reviewed the published literature to establish the current knowledge on this problem and to determine the evidence for routine antibiotic prophylaxis prior to a dental procedure. We found that antimicrobial prophylaxis before dental interventions in patients with artificial joints lacks evidence-based information and thus cannot be universally recommended.
Diagnostic and operative codes are routinely collected for every patient admitted to hospital in the English NHS. Data on post-operative complications following foot and ankle surgery have not previously been available in large numbers. Data on symptomatic venous thromboembolism events and mortality within 90 days were extracted for patients undergoing fixation of an ankle fracture, first metatarsal osteotomy, hindfoot fusions and total ankle replacement over a period of 42 months. For ankle fracture surgery (45 949 patients), the rates of deep-vein thrombosis (DVT), pulmonary embolism and mortality were 0.12%, 0.17% and 0.37%, respectively. For first metatarsal osteotomy (33 626 patients), DVT, pulmonary embolism and mortality rates were 0.01%, 0.02% and 0.04%, and for hindfoot fusions (7033 patients) the rates were 0.03%, 0.11% and 0.11%, respectively. The rate of pulmonary embolism in 1633 total ankle replacement patients was 0.06%, and there were no recorded DVTs and no deaths. Statistical analysis could only identify risk factors for venous thromboembolic events of increasing age and multiple comorbidities following fracture surgery. Venous thromboembolism following foot and ankle surgery is extremely rare, but this subset of fracture patients is at a higher risk. However, there is no evidence that thromboprophylaxis reduces this risk, and these national data suggest that prophylaxis is not required in most of these patients.