Compression and absolute stability are important in the management of intra-articular fractures. We compared tension band wiring with plate fixation for the treatment of fractures of the olecranon by measuring compression within the fracture. Identical transverse fractures were created in models of the ulna. Tension band wires were applied to ten fractures and ten were fixed with Acumed plates. Compression was measured using a Tekscan force transducer within the fracture gap. Dynamic testing was carried out by reproducing cyclical contraction of the triceps of 20 N and of the brachialis of 10 N. Both methods were tested on each sample. Paired t-tests compared overall compression and compression at the articular side of the fracture.

The mean compression for plating was 819 N (sd 602, 95% confidence interval (CI)) and for tension band wiring was 77 N (sd 19, 95% CI) (p = 0.039). The mean compression on the articular side of the fracture for plating was 343 N (sd 276, 95% CI) and for tension band wiring was 1 N (sd 2, 95% CI) (p = 0.038).

During simulated movements, the mean compression was reduced in both groups, with tension band wiring at −14 N (sd 7) and for plating −173 N (sd 32). No increase in compression on the articular side was detected in the tension band wiring group.

Pre-contoured plates provide significantly greater compression than tension bands in the treatment of transverse fractures of the olecranon, both over the whole fracture and specifically at the articular side of the fracture. In tension band wiring the overall compression was reduced and articular compression remained negligible during simulated contraction of the triceps, challenging the tension band principle.


Bone & Joint 360
Vol. 2, Issue 2 | Pages 30 - 32
1 Apr 2013

The April 2013 Children’s orthopaedics Roundup360 looks at: improving stress distribution in dysplastic hips; the dangers of fashion; the natural history of supracondylar fractures; ankles that perform well as knees; intra-articular hip pathology at osteotomy; the safe removal of flexible nails; supracondylar fracture fixation; and talipes.


The Bone & Joint Journal
Vol. 95-B, Issue 5 | Pages 643 - 648
1 May 2013
Wang J Hsu C Huang C Lin P Chen W

Structural allografts may be used to manage uncontained bone defects in revision total knee replacement (TKR). However, the availability of cadaver grafts is limited in some areas of Asia. The aim of this study was to evaluate the mid-term outcome of the use of femoral head allografts for the reconstruction of uncontained defects in revision TKR, focusing on complications related to the graft.

We retrospectively reviewed 28 patients (30 TKRs) with Anderson Orthopaedic Research Institute (AORI) type 3 bone defects, who underwent revision using femoral head allografts and stemmed components. The mean number of femoral heads used was 1.7 (1 to 3). The allograft–host junctions were packed with cancellous autograft.

At a mean follow-up of 76 months (38 to 136) the mean American Knee Society knee score improved from 37.2 (17 to 60) pre-operatively to 90 (83 to 100) (p < 0.001). The mean function score improved from 26.5 (0 to 50) pre-operatively to 81 (60 to 100) (p < 0.001). All the grafts healed to the host bone. The mean time to healing of the graft was 6.6 months (4 to 16). There have been no complications of collapse of the graft, nonunion, infection or implant loosening. No revision surgery was required.

The use of femoral head allografts in conjunction with a stemmed component and autogenous bone graft in revision TKR in patients with uncontained bone defects resulted in a high rate of healing of the graft with minimal complications and a satisfactory outcome. Longer follow-up is needed to observe the evolution of the graft.

Cite this article: Bone Joint J 2013;95-B:643–8.


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 10 | Pages 1416 - 1421
1 Oct 2010
Qureshi F Draviaraj KP Stanley D

Between September 1993 and September 1996, we performed 34 Kudo 5 total elbow replacements in 31 rheumatoid patients. All 22 surviving patients were reviewed at a mean of 11.9 years (10 to 14). Their mean age was 56 years (37 to 78) at the time of operation. All had Larsen grade IV or V rheumatoid changes on X-ray. Nine (three bilateral replacements and six unilateral) had died from unrelated causes. One who had died before ten years underwent revision for dislocation.

Of the 22 total elbow replacements reviewed six had required revision, four for aseptic loosening (one humeral and three ulnar) and two for infection. Post-operatively, one patient had neuropraxia of the ulnar nerve and one of the radial nerve. Two patients had valgus tilting of the ulnar component.

With revision as the endpoint, the mean survival time for the prosthesis was 11.3 years (95% confidence interval (10 to 13) and the estimated survival of the prosthesis at 12 years according to Kaplan-Meier survival analysis was 74% (95% confidence interval 0.53 to 0.91).

Of the 16 surviving implants, ten were free from pain, four had mild pain and two moderate. The mean arc of flexion/extension of the elbow was 106° (65° to 130°) with pronation/supination of 90° (30° to 150°) with the joint at 90° of flexion. The mean Mayo elbow performance score was 82 (60 to 100) with five excellent, ten good and one fair result.

Good long-term results can be expected using the Kudo 5 total elbow replacement in patients with rheumatoid disease, with a low incidence of loosening of the components.


The Bone & Joint Journal
Vol. 95-B, Issue 2 | Pages 177 - 180
1 Feb 2013
Arthur CHC Wood AM Keenan ACM Clayton RAE Walmsley P Brenkel I

We report ten-year clinical and radiological follow-up data for the Sigma Press Fit Condylar total knee replacement system (Sigma PFC TKR). Between October 1998 and October 1999 a total of 235 consecutive PFC Sigma TKRs were carried out in 203 patients. Patients were seen at a specialist nurse-led clinic seven to ten days before admission and at six and 18 months, three, five and eight to ten years after surgery. Data were recorded prospectively at each clinic visit. Radiographs were obtained at the five- and eight- to ten-year follow-up appointments. Of the 203 patients, 147 (171 knees) were alive at ten years and 12 were lost to follow-up. A total of eight knees (3.4%) were revised, five for infection and three to change the polyethylene insert. The survival at ten years with an endpoint of revision for any reason was 95.9%, and with an endpoint of revision for aseptic failure was 98.7%. The mean American Knee Society Score (AKSS) was 79 (10 to 99) at eight to ten years, compared with 31 (2 to 62) pre-operatively. Of 109 knee with radiographs reviewed, 47 knees had radiolucent lines but none showed evidence of loosening.

Cite this article: Bone Joint J 2013;95-B:177–80.


The Bone & Joint Journal
Vol. 95-B, Issue 7 | Pages 966 - 971
1 Jul 2013
Pumberger M Froemel D Aichmair A Hughes AP Sama AA Cammisa FP Girardi FP

The purpose of this study was to investigate the clinical predictors of surgical outcome in patients with cervical spondylotic myelopathy (CSM). We reviewed a consecutive series of 248 patients (71 women and 177 men) with CSM who had undergone surgery at our institution between January 2000 and October 2010. Their mean age was 59.0 years (16 to 86). Medical records, office notes, and operative reports were reviewed for data collection. Special attention was focused on pre-operative duration and severity as well as post-operative persistence of myelopathic symptoms. Disease severity was graded according to the Nurick classification.

Our multivariate logistic regression model indicated that Nurick grade 2 CSM patients have the highest chance of complete symptom resolution (p < 0.001) and improvement to normal gait (p = 0.004) following surgery. Patients who did not improve after surgery had longer duration of myelopathic symptoms than those who did improve post-operatively (17.85 months (1 to 101) vs 11.21 months (1 to 69); p = 0.002). More advanced Nurick grades were not associated with a longer duration of symptoms (p = 0.906).

Our data suggest that patients with Nurick grade 2 CSM are most likely to improve from surgery. The duration of myelopathic symptoms does not have an association with disease severity but is an independent prognostic indicator of surgical outcome.

Cite this article: Bone Joint J 2013;95-B:966–71.


Bone & Joint 360
Vol. 2, Issue 2 | Pages 2 - 7
1 Apr 2013
Colton C

Ancient Egypt was a highly developed agrarian society with a massive civil engineering capability. Trauma and skeletal disease were common and vestiges of the evidence for that survive, largely in the form of hieratic images and papyri dedicated to the practice of medicine. The earliest treatise on trauma is the Edwin Smith papyrus, possibly the work of Imhotep. This study details some remarkable examples of musculoskeletal pathology including fatal open fractures, foot deformity of Tutankhamun, and the earliest recorded instances of child abuse.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 11_Supple_A | Pages 137 - 140
1 Nov 2012
Jones RE Russell RD Huo MH

Most problems encountered in complex revision total knee arthroplasty can be managed with the wide range of implant systems currently available. Modular metaphyseal sleeves, metallic augments and cones provide stability even with significant bone loss. Hinged designs substitute for significant ligamentous deficiencies. Catastrophic failure that precludes successful reconstruction can be encountered. The alternatives to arthroplasty in such drastic situations include knee arthrodesis, resection arthroplasty and amputation. The relative indications for the selection of these alternatives are recurrent deep infection, immunocompromised host, and extensive non-reconstructible bone or soft-tissue defects.


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 11 | Pages 1574 - 1579
1 Nov 2010
Hariri A Mascard E Atlan F Germain MA Heming N Dubousset JF Wicart P

We describe a retrospective review of 38 cases of reconstruction following resection of the metaphysiodiaphysis of the lower limb for malignant bone tumours using free vascularised fibular grafts. The mean follow-up was for 7.6 years (0.4 to 18.4). The mean Musculoskeletal Tumor Society score was 27.2 (20 to 30). The score was significantly higher when the graft was carried out in a one-stage procedure after resection of the tumour rather than in two stages. Bony union was achieved in 89% of the cases. The overall mean time to union was 1.7 years (0.2 to 10.3).

Free vascularised fibular transfer is a major operation with frequent, but preventable, complications which allows salvage of the limb with satisfactory functional results.


Bone & Joint Research
Vol. 2, Issue 8 | Pages 162 - 168
1 Aug 2013
Chia PH Gualano L Seevanayagam S Weinberg L

Objectives

To determine the morbidity and mortality outcomes of patients presenting with a fractured neck of femur in an Australian context. Peri-operative variables related to unfavourable outcomes were identified to allow planning of intervention strategies for improving peri-operative care.

Methods

We performed a retrospective observational study of 185 consecutive adult patients admitted to an Australian metropolitan teaching hospital with fractured neck of femur between 2009 and 2010. The main outcome measures were 30-day and one-year mortality rates, major complications and factors influencing mortality.


The Bone & Joint Journal
Vol. 95-B, Issue 3 | Pages 371 - 377
1 Mar 2013
Kugan R Aslam N Bose D McNally MA

Achieving arthrodesis of the ankle can be difficult in the presence of infection, deformity, poor soft tissues and bone loss. We present a series of 48 patients with complex ankle pathology, treated with the Ilizarov technique. Infection was present in 30 patients and 30 had significant deformity before surgery. Outcome was assessed clinically and with patient-reported outcome measures (Modified American Orthopaedic Foot and Ankle Society (MAOFAS) scale and the Short-Form (SF-36)).

Arthrodesis was achieved in 40 patients with the Ilizarov technique alone and in six further patients with additional surgery. Infection was eradicated in all patients at a mean follow-up of 46.6 months (13 to 162). Successful arthrodesis was less likely in those with comorbidities and in tibiocalcaneal fusion compared with tibiotalar fusion.

These patients had poor general health scores compared with the normal population before surgery. The mean MAOFAS score improved significantly from 24.3 (0 to 90) pre-operatively to 56.2 (30 to 90) post-operatively, but there was only a modest improvement in general health; the mean SF-36 improved from 44.8 (19 to 66) to 50.1 (21 to 76). There was a major benefit in terms of pain relief.

Arthrodesis using the Ilizarov technique is an effective treatment for complex ankle pathology, with good clinical outcomes and eradication of infection. However, even after successful arthrodesis general health scores remain limited.

Cite this article: Bone Joint J 2013;95-B:371–7.


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 4 | Pages 430 - 435
1 Apr 2008
Biant LC Bruce WJM Assini JB Walker PM Walsh WR

This prospective study presents the ten-year (5 to 16) clinical and radiological results of 55 primary total hip replacements (THR) using a cementless modular femoral component (S-ROM). All patients had a significant anatomical abnormality which rendered the primary THR difficult.

The mean Harris hip score was 36 (12 to 72) pre-operatively, 83 (44 to 100) at five years, and 85 (45 to 99) at ten years. The Western Ontario and McMaster Universities osteoarthritis index (WOMAC) and short-form (SF)-12 scores were recorded from the year 2000. The mean SF-12 score at five years after surgery was 45.24 (22.74 to 56.58) for the physical component and 54.14 (29.20 to 66.61) for the mental component. By ten years the SF-12 scores were 42.86 (21.59 to 58.95) and 51.03 (33.78 to 61.40), respectively. The mean WOMAC score at five years post-operatively was 25 (0 to 59), and at ten years was 27 (2 to 70).

No femoral components were radiologically loose, although five had osteolysis in Gruen zone 1, three had osteolysis in zone 7, and two showed osteolysis in both zones 1 and 7. No osteolysis was observed around or distal to the prosthetic sleeve. No femoral components were revised, although three hips underwent an acetabular revision and two required a liner exchange. At a mean of ten years’ follow-up the S-ROM femoral component implanted for an anatomically difficult primary THR has excellent clinical and radiological results.


The Bone & Joint Journal
Vol. 95-B, Issue 1 | Pages 70 - 74
1 Jan 2013
Dattani R Smith CD Patel VR

We investigated the incidence of and risk factors for venous thromboembolism (VTE) following surgery of the shoulder and elbow and assessed the role of thromboprophylaxis in upper limb surgery. All papers describing VTE after shoulder and elbow surgery published in the English language literature before 31 March 2012 were reviewed. A total of 14 papers were available for analysis, most of which were retrospective studies and case series. The incidence of VTE was 0.038% from 92 440 shoulder arthroscopic procedures, 0.52% from 42 261 shoulder replacements, and 0.64% from 4833 procedures for fractures of the proximal humerus (open reduction and internal fixation or hemiarthroplasty). The incidence following replacement of the elbow was 0.26% from 2701 procedures. Diabetes mellitus, rheumatoid arthritis and ischaemic heart disease were identified as the major risk factors.

The evidence that exists on thromboprophylaxis is based on level III and IV studies, and we therefore cannot make any recommendations on prophylaxis based on the current evidence. It seems reasonable to adopt a multimodal approach that involves all patients receiving mechanical prophylaxis, with chemical prophylaxis reserved for those who are at high risk for VTE.

Cite this article: Bone Joint J 2013;95-B:70–4.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 12 | Pages 1591 - 1594
1 Dec 2012
Cousins GR Obolensky L McAllen C Acharya V Beebeejaun A

We report the results of six trauma and orthopaedic projects to Kenya in the last three years. The aims are to deliver both a trauma service and teaching within two hospitals; one a district hospital near Mount Kenya in Nanyuki, the other the largest public hospital in Kenya in Mombasa. The Kenya Orthopaedic Project team consists of a wide range of multidisciplinary professionals that allows the experience to be shared across those specialties. A follow-up clinic is held three months after each mission to review the patients. To our knowledge there are no reported outcomes in the literature for similar projects.

A total of 211 operations have been performed and 400 patients seen during the projects. Most cases were fractures of the lower limb; we have been able to follow up 163 patients (77%) who underwent surgical treatment. We reflect on the results so far and discuss potential improvements for future missions.


Bone & Joint 360
Vol. 1, Issue 6 | Pages 30 - 32
1 Dec 2012

The December 2012 Research Roundup360 looks at: whether the rheumatoid factor is just a ‘quick test’; osteonecrosis in smokers; pasteurisation effect on bone reconstruction; venous thromboembolism risk in rheumatoids; whether stem cells reverse age-related osteopenia; the effect of running on rat knees; rapid fracture healing in rats with ultrasound; magnetic stem cells; and the safety of surgery.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 12 | Pages 1586 - 1591
1 Dec 2007
Flecher X Parratte S Aubaniac J Argenson J

A clinical and radiological study was conducted on 97 total hip replacements performed for congenital hip dislocation in 79 patients between 1989 and 1998 using a three-dimensional custom-made cementless stem. The mean age at operation was 48 years (17 to 72) and the mean follow-up was for 123 months (83 to 182).

According to the Crowe classification, there were 37 class I, 28 class II, 13 class III and 19 class IV hips. The mean leg lengthening was 25 mm (5 to 58), the mean pre-operative femoral anteversion was 38.6° (2° to 86°) and the mean correction in the prosthetic neck was −23.6° (−71° to 13°). The mean Harris hip score improved from 58 (15 to 84) to 93 (40 to 100) points. A revision was required in six hips (6.2%). The overall survival rate was 89.5% (95% confidence interval 89.2 to 89.8) at 13 years when two hips were at risk.

This custom-made cementless femoral component, which can be accommodated in the abnormal proximal femur and will correct the anteversion and frontal offset, provided good results without recourse to proximal femoral corrective osteotomy.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 10 | Pages 1409 - 1413
1 Oct 2012
Roberts VI Komarasamy B Pandey R

Controversy surrounds the management of displaced three- and four-part fractures of the proximal humerus. The percutaneous Resch technique of stabilisation involves minimal soft-tissue dissection and a reduced risk of stiffness and avascular necrosis. However, it requires a second operation to remove Kirschner wires and the humeral block. We describe a modification of this technique that dispenses with the need for this second operation and relies on a sequential pattern of screw placement. We report the outcome of 32 three- or four-part fractures of the proximal humerus treated in this way at a mean follow-up of 3.8 years (2 to 8)). There were 14 men and 18 women with a mean age of 56 years (28 to 83). At final follow-up the mean Oxford shoulder scores were 38 (31 to 44) and 39 (31 to 42), and the mean Constant scores were 79 (65 to 92) and 72 (70 to 80) for three- and four-part fractures, respectively. We further analysed the results in patients aged < 60 years with high-energy fractures and those aged ≥ 60 years with osteoporotic fractures. There were no cases of nonunion or avascular necrosis.

The results were good and comparable to those previously reported for the Resch technique and other means of fixation for proximal humeral fractures. We would recommend this modification of the technique for the treatment of displaced three-part and four-part fractures in patients both younger and older than 60 years of age.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 9 | Pages 1249 - 1251
1 Sep 2009
Huang K Yang R Hsieh C

Breast cancer is generally managed surgically with adjuvant agents which include hormone therapy, chemotherapy, radiotherapy and bisphosphonate therapy. However, some of these adjuvant therapies may cause adverse events, including wound infection, neutropenia, bone marrow suppression and fever. The simultaneous presentation of osteonecrosis and osteomyelitis has not previously been described in patients with breast cancer undergoing hormone therapy and chemotherapy.

We report a patient with breast cancer who developed bone infarcts in both legs as well as osteomyelitis in the right distal tibia after treatment which included a modified radical mastectomy, hormone therapy and chemotherapy. Simultaneous osteonecrosis and osteomyelitis should be considered in patients with breast cancer who are receiving chemotherapy and hormone therapy who present with severe bone pain, especially if there have been infective episodes during treatment.


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 1 | Pages 142 - 145
1 Jan 2010
Spencer SJ Holt G Clarke JV Mohammed A Leach WJ Roberts JLB

The humerus is a common site for skeletal metastases in the adult. Surgical stabilisation of such lesions is often necessary to relieve pain and restore function. These procedures are essentially palliative and should therefore provide effective relief from pain for the remainder of the patient’s life without the need for further surgical intervention.

We report a retrospective analysis of 35 patients (37 nails) with symptomatic metastases in the shaft of the humerus which were treated by locked, antegrade nailing. There were 27 true fractures (73.0%) and ten painful deposits (27.0%). Relief from pain was excellent in four (11.4%), good in 29 (82.9%) and fair in two (5.7%) on discharge. Function was improved in all but one patient. One case of palsy of the radial nerve was noted. The mean postoperative survival was 7.1 months (0.2 to 45.5) which emphasises the poor prognosis in this group of patients. There were no failures of fixation and no case in which further surgery was required.

Antegrade intramedullary nailing is an effective means of stabilising the humerus for the palliative treatment of metastases. It relieves pain and restores function to the upper limb with low attendant morbidity.