The aims of this retrospective study were to report the feasibility
of using 3D-printing technology for patients with a pelvic tumour
who underwent reconstruction. A total of 35 patients underwent resection of a pelvic tumour
and reconstruction using 3D-printed endoprostheses between September
2013 and December 2015. According to Enneking’s classification of
bone defects, there were three Type I lesions, 12 Type II+III lesions,
five Type I+II lesions, two Type I+II+III lesions, ten type I+II+IV
lesions and three type I+II+III+IV lesions. A total of three patients
underwent reconstruction using an iliac prosthesis, 12 using a standard
hemipelvic prosthesis and 20 using a screw-rod connected hemipelvic
prosthesis.Aims
Patients and Methods
Instability remains a challenging problem in both primary and
revision total hip arthroplasty (THA). Dual mobility components
confer increased stability, but there are concerns about the unique
complications associated with these designs, as well as the long-term
survivorship. We performed a systematic review of all English language articles
dealing with dual mobility THAs published between 2007 and 2016
in the MEDLINE and Embase electronic databases. A total of 54 articles
met inclusion criteria for the final analysis of primary and revision
dual mobility THAs and dual mobility THAs used in the treatment
of fractures of the femoral neck. We analysed the survivorship and
rates of aseptic loosening and of intraprosthetic and extra-articular
dislocation.Aims
Materials and Methods
Several studies have reported the safety and efficacy of subcapital
re-alignment for patients with slipped capital femoral epiphysis
(SCFE) using surgical dislocation of the hip and an extended retinacular
flap. Instability of the hip and dislocation as a consequence of
this surgery has only recently gained attention. We discuss this
problem with some illustrative cases. We explored the literature on the possible pathophysiological
causes and surgical steps associated with the risk of post-operative
instability and articular damage. In addition, we describe supplementary
steps that could be used to avoid these problems.Aims
Materials and Methods
The aims of this study were to determine the clinical and radiographic
outcomes, implant survivorship, and complications of patients with
a history of poliomyelitis undergoing total hip arthroplasty (THA)
in affected limbs and unaffected limbs of this same population. A retrospective review identified 51 patients (27 male and 24
female, 59 hips) with a mean age of 66 years (38 to 88) and with
the history of poliomyelitis who underwent THA for degenerative
arthritis between 1970 and 2012. Immigrant status, clinical outcomes,
radiographic results, implant survival, and complications were recorded.Aims
Patients and Methods
Posterior tilt of the pelvis with sitting provides biological
acetabular opening. Our goal was to study the post-operative interaction
of skeletal mobility and sagittal acetabular component position. This was a radiographic study of 160 hips (151 patients) who
prospectively had lateral spinopelvic hip radiographs for skeletal
and implant measurements. Intra-operative acetabular component position
was determined according to the pre-operative spinal mobility. Sagittal
implant measurements of ante-inclination and sacral acetabular angle were
used as surrogate measurements for the risk of impingement, and
intra-operative acetabular component angles were compared with these.Aims
Materials and Methods
The stability of cementless acetabular components is an important
factor for surgical planning in the treatment of patients with pelvic
osteolysis after total hip arthroplasty (THA). However, the methods
for determining the stability of the acetabular component from pre-operative
radiographs remain controversial. Our aim was to develop a scoring
system to help in the assessment of the stability of the acetabular
component under these circumstances. The new scoring system is based on the mechanism of failure of
these components and the location of the osteolytic lesion, according
to the DeLee and Charnley classification. Each zone is evaluated
and scored separately. The sum of the individual scores from the
three zones is reported as a total score with a maximum of 10 points.
The study involved 96 revision procedures which were undertaken
for wear or osteolysis in 91 patients between July 2002 and December
2012. Pre-operative anteroposterior pelvic radiographs and Judet
views were reviewed. The stability of the acetabular component was
confirmed intra-operatively.Aims
Patients and Methods
The primary aim of this study was to define and quantify three
new measurements to indicate the position of the greater trochanter.
Secondary aims were to define ‘functional antetorsion’ as it relates
to abductor function in populations both with and without torsional
abnormality. Three new measurements, functional antetorsion, posterior tilt,
and posterior translation of the greater trochanter, were assessed
from 61 CT scans of cadaveric femurs, and their reliability determined.
These measurements and their relationships were also evaluated in
three groups of patients: a control group (n = 22), a ‘high-antetorsion’ group
(n = 22) and a ‘low-antetorsion’ group (n = 10).Aims
Patients and Methods
To validate the precision of digitally reconstructed radiograph (DRR) radiostereometric analysis (RSA) and the model-based method (MBM) RSA with respect to benchmark marker-based (MM) RSA for evaluation of kinematics in the native hip joint. Seven human cadaveric hemipelves were CT scanned and bone models were segmented. Tantalum beads were placed in the pelvis and proximal femoral bone. RSA recordings of the hips were performed during flexion, adduction and internal rotation. Stereoradiographic recordings were all analyzed with DRR, MBM and MM. Migration results for the MBM and DRR with respect to MM were compared. Precision was assessed as systematic bias (mean difference) and random variation (Pitman’s test for equal variance).Objectives
Methods
This review summarises the technique of impaction
grafting with mesh augmentation for the treatment of uncontained
acetabular defects in revision hip arthroplasty. The ideal acetabular revision should restore bone stock, use
a small socket in the near-anatomic position, and provide durable
fixation. Impaction bone grafting, which has been in use for over
40 years, offers the ability to achieve these goals in uncontained
defects. The precepts of modern, revision impaction grafting are
that the segmental or cavitary defects must be supported with a
mesh; the contained cavity is filled with vigorously impacted morselised
fresh-frozen allograft; and finally, acrylic cement is used to stabilise
the graft and provide rigid, long-lasting fixation of the revised
acetabular component. Favourable results have been published with this technique. While
having its limitations, it is a viable option to address large acetabular
defects in revision arthroplasty. Cite this article:
Periacetabular osteotomy is an effective way of treating symptomatic
hip dysplasia. We describe a new minimally invasive technique using
a modification of the Smith-Peterson approach. We performed a prospective, longitudinal cohort study to assess
for any compromise in acetabular correction when using this approach,
and to see if the procedure would have a higher complication rate
than that quoted in the literature for other approaches. We also
assessed for any improvement in functional outcome. From 168 consecutive patients (189 hips) who underwent acetabular
correction between March 2010 and March 2013 we excluded those who
had undergone previous pelvic surgery for DDH and those being treated
for acetabular retroversion. The remaining 151 patients (15 men,
136 women) (166 hips) had a mean age of 32 years (15 to 56) and the
mean duration of follow-up was 2.8 years (1.2 to 4.5). In all 90%
of cases were Tönnis grade 0 or 1. Functional outcomes were assessed
using the Non Arthritic Hip Score (NAHS), University of California,
Los Angeles (UCLA) and Tegner activity scores.Aims
Patients and Methods
Modular or custom-made femoral components have been preferred
for total hip arthroplasty (THA) in patients with a history of Perthes’
disease because of the distortion in the anatomy of the proximal
femur. However, it has not been established whether a monobloc cementless
stem will fit the distorted proximal femur or whether the results
of the procedure are satisfactory in this group of patients. We reviewed 68 consecutive patients who had undergone THA for
childhood Perthes’ disease between June 2003 and December 2008.
There were 35 men and 33 women with a mean age of 48 years (16 to
73) at the time of index arthroplasty. Their mean body mass index
was 24.4 (18.3 to 32.9). Of the 68 hips, 32 were classified as Stulberg
class III and 36 as class IV. The mean pre-operative shortening
of the affected leg was 17.2 mm (5 to 34). The minimum follow-up
was five years (mean 8.5 years; 5.2 to 10).Aims
Patients and Methods
Our aim was to evaluate the radiographic characteristics of patients
undergoing total hip arthroplasty (THA) for the potential of posterior
bony impingement using CT simulations. Virtual CT data from 112 patients who underwent THA were analysed.
There were 40 men and 72 women. Their mean age was 59.1 years (41
to 76). Associations between radiographic characteristics and posterior
bony impingement and the range of external rotation of the hip were
evaluated. In addition, we investigated the effects of pelvic tilt
and the neck/shaft angle and femoral offset on posterior bony impingement.Aims
Patients and Methods
Few reconstructive techniques are available for patients requiring
complex acetabular revisions such as those involving Paprosky type
2C, 3A and 3B deficiencies and pelvic discontinuity. Our aim was
to describe the development of the patient specific Triflange acetabular
component for use in these patients, the surgical technique and
mid-term results. We include a description of the pre-operative
CT scanning, the construction of a model, operative planning, and
surgical technique. All implants were coated with porous plasma
spray and hydroxyapatite if desired. A multicentre, retrospective review of 95 complex acetabular
reconstructions in 94 patients was performed. A total of 61 (64.2%)
were female. The mean age of the patients was 66 (38 to 85). The
mean body mass index was 29 kg/m2 (18 to 51). Outcome
was reported using the Harris Hip Score (HHS), complications, failures
and survival.Aims
Patients and Methods
Transtrochanteric rotational osteotomy (TRO) is performed for
young patients with non-traumatic osteonecrosis of the femoral head
(ONFH) to preserve the hip. We aimed to investigate the long-term
outcomes and the risk factors for failure 15 years after this procedure. This study included 95 patients (111 hips) with a mean age of
40 years (21 to 64) who underwent TRO for ONFH. The mean follow-up
was 18.2 years (3 to 26). Kaplan–Meier survivorship analyses were
performed with conversion to total hip arthroplasty (THA) and radiological
failure due to secondary collapse of the femoral head or osteoarthritic changes
as the endpoint. Multivariate analyses were performed to assess
risk factors for each outcome.Aims
Patients and Methods
Ideal placement of the acetabular component remains
elusive both in terms of defining and achieving a target. Our aim
is to help restore original anatomy by using the transverse acetabular
ligament (TAL) to control the height, depth and version of the component.
In the normal hip the TAL and labrum extend beyond the equator of
the femoral head and therefore, if the definitive acetabular component
is positioned such that it is cradled by and just deep to the plane
of the TAL and labrum and is no more than 4mm larger than the original
femoral head, the centre of the hip should be restored. If the face
of the component is positioned parallel to the TAL and psoas groove
the patient specific version should be restored. We still use the
TAL for controlling version in the dysplastic hip because we believe
that the TAL and labrum compensate for any underlying bony abnormality. The TAL should not be used as an aid to inclination. Worldwide,
>
75% of surgeons operate with the patient in the lateral decubitus
position and we have shown that errors in post-operative radiographic
inclination (RI) of >
50° are generally caused by errors in patient positioning.
Consequently, great care needs to be taken when positioning the
patient. We also recommend 35° of apparent operative inclination
(AOI) during surgery, as opposed to the traditional 45°. Cite this article:
The ‘jumbo’ acetabular component is now commonly
used in acetabular revision surgery where there is extensive bone
loss. It offers high surface contact, permits weight bearing over
a large area of the pelvis, the need for bone grafting is reduced
and it is usually possible to restore centre of rotation of the
hip. Disadvantages of its use include a technique in which bone
structure may not be restored, a risk of excessive posterior bone
loss during reaming, an obligation to employ screw fixation, limited
bone ingrowth with late failure and high hip centre, leading to increased
risk of dislocation. Contraindications include unaddressed pelvic
dissociation, inability to implant the component with a rim fit,
and an inability to achieve screw fixation. Use in acetabulae with
<
50% bone stock has also been questioned. Published results
have been encouraging in the first decade, with late failures predominantly because
of polyethylene wear and aseptic loosening. Dislocation is the most
common complication of jumbo acetabular revisions, with an incidence
of approximately 10%, and often mandates revision. Based on published results,
a hemispherical component with an enhanced porous coating, highly
cross-linked polyethylene, and a large femoral head appears to represent
the optimum tribology for jumbo acetabular revisions. Cite this article: