The purpose of this study was to undertake a
meta-analysis to determine whether there is lower polyethylene wear and
longer survival when using mobile-bearing implants in total knee
replacement when compared with fixed-bearing implants. Of 975 papers
identified, 34 trials were eligible for data extraction and meta-analysis
comprising 4754 patients (6861 knees). We found no statistically
significant differences between the two designs in terms of the incidence
of
Aims. We report the incidence of
Aims. We conducted a prospective study of a delta ceramic total hip
arthroplasty (THA) to determine the rate of ceramic fracture, to
characterise post-operative noise, and to evaluate the mid-term
results and survivorship. Patients and Methods. Between March 2009 and March 2011, 274 patients (310 hips) underwent
cementless THA using a delta ceramic femoral head and liner. At
each follow-up, clinical and radiological outcomes were recorded.
A Kaplan-Meier analysis was undertaken to estimate survival. Results. Four patients (four hips) died and 18 patients (20 hips) were
lost to follow-up within five years. The remaining 252 patients
(286 hips) were followed for a mean of 66.5 months (60 to 84). There
were 144 men (166 hips) and 108 women (120 hips) with a mean age
of 49.7 years (16 to 83) at surgery. The mean pre-operative Harris
Hip Score of 47.1 points improved to 93.8 points at final follow-up.
Six patients reported squeaking in seven hips; however, none were
audible.
Aims. This study aimed to determine the diagnostic performance of radiographic
criteria to detect aseptic acetabular loosening after revision total
hip arthroplasty (THA). Secondary aims were to determine the predictive
values of different thresholds of migration and to determine the
predictive values of radiolucency criteria. Patients and Methods. Acetabular component migration to re-revision was measured retrospectively
using Ein-Bild-Rontgen-Analyse (EBRA-Cup) and manual measurements
(Sutherland method) in two groups: Group A, 52 components (48 patients) found
not loose at re-revision and Group B, 42 components (36 patients)
found loose at re-revision between 1980 and 2015. The presence and
extent of
Aims. Our aim was to examine the clinical and radiographic outcomes
in 257 consecutive Oxford unicompartmental knee arthroplasties (OUKAs)
(238 patients), five years post-operatively. Patients and Methods. A retrospective evaluation was undertaken of patients treated
between April 2008 and October 2010 in a regional centre by two
non-designing surgeons with no previous experience of UKAs. The
Oxford Knee Scores (OKSs) were recorded and fluoroscopically aligned
radiographs were assessed post-operatively at one and five years. Results. The median age of the 238 patients was 65.0 years (interquartile
range (IQR) 59.0 to 73.0), the median body mas index was 30.0 (IQR
27.5 to 33.0) and 51.7% were male. There were no intra-operative
complications. There was a significant improvement in the median
OKS at six weeks (34, IQR 31.0 to 37.0), one year (38, IQR 29.0
to 43.0) and five years (37, IQR 27.0 to 42.0) when compared with
the pre-operative scores (16, IQR 13.0 to 19.0) (all p = <
0.01).
No patient had progressive
The clinical and radiological results of 50 consecutive acetabular reconstructions in 48 patients using impaction grafting have been retrospectively reviewed. A 1:1 mixture of frozen, ground irradiated bone graft and Apapore 60, a synthetic bone graft substitute, was used in all cases. There were 13 complex primary and 37 revision procedures with a mean follow-up of five years (3.4 to 7.6). The clinical survival rate was 100%, with improvements in the mean Harris Hip Scores for pain and function. Radiologically, 30 acetabular grafts showed evidence of incorporation, ten had
Aims. We report on the outcome of the Exeter Contemporary flanged cemented
all-polyethylene acetabular component with a mean follow-up of 12
years (10 to 13.9). This study reviewed 203 hips in 194 patients.
129 hips in 122 patients are still in situ; 66
hips in 64 patients were in patients who died before ten years,
and eight hips (eight patients) were revised. Clinical outcome scores
were available for 108 hips (104 patients) and radiographs for 103
hips (100 patients). Patients and Methods. A retrospective review was undertaken of a consecutive series
of 203 routine primary cemented total hip arthroplasties (THA) in
194 patients. Results. There were no acetabular component revisions for aseptic loosening.
Acetabular revision was undertaken in eight hips. In four hips revision
was necessitated by periprosthetic femoral fractures, in two hips
by recurrent dislocation, in one hip for infection and in one hip
for unexplained ongoing pain. . Oxford and Harris hip scores demonstrated significant clinical
improvement (all p <
0.001).
Ceramic-on-metal (CoM) is a relatively new bearing
combination for total hip arthroplasty (THA) with few reported outcomes.
A total of 287 CoM THAs were carried out in 271 patients (mean age
55.6 years (20 to 77), 150 THAs in female patients, 137 in male)
under the care of a single surgeon between October 2007 and October
2009. With the issues surrounding metal-on-metal bearings the decision
was taken to review these patients between March and November 2011,
at a mean follow-up of 34 months (23 to 45) and to record pain,
outcome scores, radiological analysis and blood ion levels. The
mean Oxford Hip Score was 19.2 (12 to 53), 254 patients with 268
hips (95%) had mild/very mild/no pain, the mean angle of inclination
of the acetabular component was 44.8. o. (28. o. to
63. o. ), 82 stems (29%) had evidence of radiolucent lines
of >
1 mm in at least one Gruen zone and the median levels of cobalt
and chromium ions in the blood were 0.83 μg/L (0.24 μg/L to 27.56 μg/L)
and 0.78 μg/L (0.21 μg/L to 8.84 μg/L), respectively. The five-year
survival rate is 96.9% (95% confidence interval 94.7% to 99%). Due to the presence of
The incidence of loosening of a cemented glenoid componentin total shoulder arthroplasty, detected by means of
We report the outcome of the flangeless, cemented all-polyethylene Exeter acetabular component at a mean of 14.6 years (10 to 17) after operation. Of the 263 hips in 243 patients, 122 prostheses are still in situ; 112 patients (119 hips) have died, 18 hips have been revised, and three patients (four hips) were lost to follow-up. Radiographs at the final review were available for 110 of the 122 surviving hips. There were acetabular
We report the long-term radiological results of 58 total hip arthroplasties (THA) using the Charnley offset-bore acetabular socket. Wear was measured at four sites and
We analysed at a mean follow-up of 7.25 years the clinical and radiological outcome of 117 patients (125 knees) who had undergone a primary, cemented, modular Freeman-Samuelson total knee replacement. While the tibial and femoral components were cemented, the patellar component was uncemented. A surface-cementing technique was used to secure the tibial components. A total of 82 knees was available for radiological assessment.
Between January 1988 and January 1991 we performed 100 consecutive cemented total hip replacements using a zirconia head, a titanium alloy stem and a polyethylene cup. We reviewed 78 of these hips in 61 patients in detail at a mean of 5.8 years (1 to 9). Aseptic loosening was seen in 11 hips (14%). Eight needed revision. In total, 37 cups (47.5%) showed
We studied 11 patients (14 elbows) with gross rheumatoid deformity of the elbow, treated by total arthroplasty using the Kudo type-5 unlinked prosthesis, and who were evaluated between five and 11 years after operation. Massive bone defects were augmented by autogenous bone grafts. There were no major complications such as infection, subluxation or loosening. In most elbows relief from pain and stability were achieved. The results, according to the Mayo Elbow Performance Score, were excellent in eight, good in five and fair in one. In most elbows there was minimal or no resorption of the grafted bone. There were no
Mechanical loosening which begins with early-onset migration of the prosthesis is the major reason for failure of the Souter-Strathclyde elbow replacement. In a prospective study of 18 Souter-Strathclyde replacements we evaluated the patterns of migration using roentgen stereophotogrammetric analysis. We had previously reported the short-term results after a follow-up of two years which we have now extended to a mean follow-up of 8.2 years (1 to 11.3). Migration was assessed along the co-ordinal axes and overall micromovement was expressed as the maximum total point movement. The alignment of the prosthesis and the presence of
Between 1993 and 1996, we undertook 35 Kudo 5 total elbow replacements in a consecutive series of 31 rheumatoid patients. A total of 25 patients (29 procedures) was evaluated at a mean follow-up of six years (5 to 7.5) using the Mayo Clinic performance index. In addition, all patients were assessed for loosening using standard anteroposterior and lateral radiographs. At review, 19 elbows (65%) had either no pain or mild pain, ten (35%) had moderate pain and none had severe pain. The mean arc of flexion/extension was 94° (35 to 130) and supination/pronation was 128° (30 to 165). A fracture of the medial epicondyle occurred during surgery in one patient. This was successfully treated with a single AO screw and a standard Kudo 5 implant was inserted. Postoperatively, there were no infections. One patient had a dislocation which was treated by closed reduction and five had neurapraxia of the ulnar nerve. Radiologically, there was no evidence of loosening of the humeral component, but two ulnar components had progressive
We report the radiological and clinical outcome of a press-fit (SLF) acetabular component at two to three years in two groups of patients having primary total hip replacement. In 69 the implant was coated with hydroxyapatite (HA) and in 40 it was uncoated. The stability of the cup was assessed by measurement of proximal migration and change in the angle of inclination. The clinical results in the two groups did not differ significantly, and the mean proximal linear wear was similar in both. Fewer
We report a consecutive series of 16 revision total knee arthroplasties using the Total Condylar III system in 14 patients with inflammatory arthritis which were performed between 1994 and 2000. There were 11 women and three men with a mean age of 59 years (36 to 78). The patients were followed up for 74 months (44 to 122). The mean pre-operative Knee Society score of 37 points (0 to 77) improved to 88 (61 to 100) at follow-up (t-test, p <
0.001) indicating very good overall results. The mean range of flexion improved from 62° (0° to 120°) to 98° (0° to 145°) (t-test, p <
0.05) allowing the patients to stand from a sitting position. The mean Knee Society pain score improved from 22 (10 to 45) to 44 (20 to 50) (t-test, p <
0.05). No knee had definite loosening, although five showed asymptomatic
We treated 34 patients with recurrent dislocation of the hip with a constrained acetabular component. Roentgen stereophotogrammetric analysis was performed to assess migration of the prosthesis. The mean clinical follow-up was 3.0 years (2.2 to 4.8) and the radiological follow-up was 2.7 years (2.0 to 4.8). At the latest review six patients had died and none was lost to follow-up. There were four acetabular revisions, three for aseptic loosening and one for deep infection. Another acetabular component was radiologically loose with progressive