Patient-specific instrumentation of total knee arthroplasty (TKA) is a technique permitting the targeting of individual kinematic alignment, but deviation from a neutral mechanical axis may have implications on implant fixation and therefore survivorship. The primary objective of this randomized controlled study was to compare the fixation of tibial components implanted with patient-specific instrumentation targeting kinematic alignment (KA+PSI) A total of 47 patients due to undergo TKA were randomized to KA+PSI (n = 24) or MA+CAS (n = 23). In the KA+PSI group, there were 16 female and eight male patients with a mean age of 64 years (Aims
Patients and Methods
The purpose of the present study was to compare patient-specific instrumentation (PSI) and conventional surgical instrumentation (CSI) for total knee arthroplasty (TKA) in terms of early implant migration, alignment, surgical resources, patient outcomes, and costs. The study was a prospective, randomized controlled trial of 50 patients undergoing TKA. There were 25 patients in each of the PSI and CSI groups. There were 12 male patients in the PSI group and seven male patients in the CSI group. The patients had a mean age of 69.0 years (Aims
Patients and Methods
The aim of this study was to evaluate the long-term inducible
displacement of cemented tibial components ten years after total
knee arthroplasty (TKA). A total of 15 patients from a previously reported prospective
trial of fixation using radiostereometric analysis (RSA) were examined
at a mean of 11 years (10 to 11) postoperatively. Longitudinal supine
RSA examinations were acquired at one week, one year, and two years
postoperatively and at final follow-up. Weight-bearing RSA examinations
were also undertaken with the operated lower limb in neutral and
in maximum internal rotation positions. Maximum total point motion
(MTPM) was calculated for the longitudinal and inducible displacement examinations
(supine Aims
Patients and Methods
An important measure for the diagnosis and monitoring of knee osteoarthritis is the minimum joint space width (mJSW). This requires accurate alignment of the x-ray beam with the tibial plateau, which may not be accomplished in practice. We investigate the feasibility of a new mJSW measurement method from stereo radiographs using 3D statistical shape models (SSM) and evaluate its sensitivity to changes in the mJSW and its robustness to variations in patient positioning and bone geometry. A validation study was performed using five cadaver specimens. The actual mJSW was varied and images were acquired with variation in the cadaver positioning. For comparison purposes, the mJSW was also assessed from plain radiographs. To study the influence of SSM model accuracy, the 3D mJSW measurement was repeated with models from the actual bones, obtained from CT scans.Objectives
Materials and Methods
This paper describes the methodology, validation and reliability
of a new computer-assisted method which uses models of the patient’s
bones and the components to measure their migration and polyethylene
wear from radiographs after total hip arthroplasty (THA). Models of the patient’s acetabular and femoral component obtained
from the manufacturer and models of the patient’s pelvis and femur
built from a single computed tomography (CT) scan, are used by a
computer program to measure the migration of the components and
the penetration of the femoral head from anteroposterior and lateral radiographs
taken at follow-up visits. The program simulates the radiographic
setup and matches the position and orientation of the models to
outlines of the pelvis, the acetabular and femoral component, and
femur on radiographs. Changes in position and orientation reflect
the migration of the components and the penetration of the femoral
head. Validation was performed using radiographs of phantoms simulating
known migration and penetration, and the clinical feasibility of
measuring migration was assessed in two patients.Aims
Materials and Methods
To compare the structural durability of hydroxyapatite-tricalcium
phosphate (HATCP) to autologous iliac crest bone graft in calcaneal
lengthening osteotomy (CLO) for pes planovalgus in childhood. We present the interim results of ten patients (HATCP, n = 6
and autograft, n = 5) with a mean age of 11.5 years (8.2 to 14.2)
from a randomised controlled non-inferiority trial with six months
follow-up. The primary outcome was the stability of the osteotomy
as measured by radiostereometric analysis. A non-inferiority margin
of ≤ 2 mm osteotomy compression was set.Aims
Patients and Methods
Lengthening osteotomies of the calcaneus in children are in general
grafted with bone from the iliac crest. Artificial bone grafts have
been introduced, however, their structural and clinical durability
has not been documented. Radiostereometric analysis (RSA) is a very
accurate and precise method for measurements of rigid body movements including
the evaluation of joint implant and fracture stability, however,
RSA has not previously been used in clinical studies of calcaneal
osteotomies. We assessed the precision of RSA as a measurement tool
in a lateral calcaneal lengthening osteotomy (LCLO). LCLO was performed in six fixed adult cadaver feet. Tantalum
markers were inserted on each side of the osteotomy and in the cuboideum.
Lengthening was done with a plexiglas wedge. A total of 24 radiological
double examinations were obtained. Two feet were excluded due to
loose and poorly dispersed markers. Precision was assessed as systematic
bias and 95% repeatability limits.Objectives
Methods
The aim of this study was to quantify the stability
of fracture-implant complex in fractures after fixation. A total
of 15 patients with an undisplaced fracture of the femoral neck,
treated with either a dynamic hip screw or three cannulated hip
screws, and 16 patients with an AO31-A2 trochanteric fracture treated
with a dynamic hip screw or a Gamma Nail, were included. Radiostereometric
analysis was used at six weeks, four months and 12 months post-operatively
to evaluate shortening and rotation. Migration could be assessed in ten patients with a fracture of
the femoral neck and seven with a trochanteric fracture. By four
months post-operatively, a mean shortening of 5.4 mm (-0.04 to 16.1)
had occurred in the fracture of the femoral neck group and 5.0 mm
(-0.13 to 12.9) in the trochanteric fracture group. A wide range
of rotation occurred in both types of fracture. Right-sided trochanteric
fractures seem more rotationally stable than left-sided fractures. This prospective study shows that migration at the fracture site
occurs continuously during the first four post-operative months,
after which stabilisation occurs. This information may allow the
early recognition of patients at risk of failure of fixation. Cite this article:
Wear of polyethylene inserts plays an important role in failure
of total knee replacement and can be monitored Before revision, the minimum joint space width values and their
locations on the insert were measured in 15 fully weight-bearing
radiographs. These measurements were compared with the actual minimum
thickness values and locations of the retrieved tibial inserts after
revision. Introduction
Method