Displaced fractures of the forearm in children are often treated conservatively, but there is a relatively high incidence of redisplacement, malunion and consequent limitation of function. We have performed percutaneous Kirschner (K) wire fixation in 72 such children under the age of 14 years, of which 57 were reviewed for our study. Both the radius and ulna were fractured in 45 (79%), the radius only in eight and the ulna only in four. The mean initial angulation was 19° in the lateral plane and 9° in the anteroposterior plane for the radius and 15° and 9°, respectively, for the ulna. In 42 patients (74%) we performed closed reduction. In the remaining 15 (26%) closed reduction failed and an open reduction, through a minimal approach, was required before K wiring. At a mean follow-up of 20 months all patients had good functional results with an excellent range of movement. Only five had angulation of from 10° to 15° and none had nonunion, premature epiphyseal closure or deep infection. Percutaneous
We report a prospective study of the use of
We report the use of a new method of locked
The Sheffield Expanding
This is a prospective randomised controlled trial comparing the
functional outcomes of plate fixation and elastic stable intramedullary
nailing (ESIN) of completely displaced mid-shaft fractures of the
clavicle in the active adult population. We prospectively recruited 123 patients and randomised them to
either plate fixation or ESIN. Patients completed the Quick Disabilities
of the Arm, Shoulder and Hand (DASH) score at one to six weeks post-operatively.
They were followed up at six weeks, three and six months and one
year with radiographs, and their clinical outcome was assessed using
both the DASH and the Constant Score.Aims
Patients and Methods
End caps are intended to prevent nail migration
(push-out) in elastic stable intramedullary nailing. The aim of
this study was to investigate the force at failure with and without
end caps, and whether different insertion angles of nails and end caps
would alter that force at failure. Simulated oblique fractures of the diaphysis were created in
15 artificial paediatric femurs. Titanium Elastic Nails with end
caps were inserted at angles of 45°, 55° and 65° in five specimens
for each angle to create three study groups. Biomechanical testing
was performed with axial compression until failure. An identical
fracture was created in four small adult cadaveric femurs harvested
from two donors (both female, aged 81 and 85 years, height 149 cm and
156 cm, respectively). All femurs were tested without and subsequently
with end caps inserted at 45°. In the artificial femurs, maximum force was not significantly
different between the three groups (p = 0.613). Push-out force was
significantly higher in the cadaveric specimens with the use of
end caps by an up to sixfold load increase (830 N, standard deviation
(SD) 280 These results indicate that the nail and end cap insertion angle
can be varied within 20° without altering construct stability and
that the risk of elastic stable intramedullary nailing push–out
can be effectively reduced by the use of end caps. Cite this article:
We present two cases of metastatic lung cancer which occurred at the site of a previously united tibial fracture. Both patients were treated with a locked intramedullary nail. The patients presented with metastases at the site of their initial fracture approximately 16 and 13 months after injury respectively. We discuss this unusual presentation and review the relevant literature. We are unaware of any previous reports of a metastatic tumour occurring at the site of an orthopaedic implant used to stabilise a non-pathological fracture. These cases demonstrate the similar clinical presentation of infection and malignancy: a diagnosis which should always be considered in such patients.
We report the outcomes of 20 patients (12 men,
8 women, 21 feet) with Charcot neuro-arthropathy who underwent correction
of deformities of the ankle and hindfoot using retrograde intramedullary
nail arthrodesis. The mean age of the patients was 62.6 years (46
to 83); their mean BMI was 32.7 (15 to 47) and their median American
Society of Anaesthetists score was 3 (2 to 4). All presented with
severe deformities and 15 had chronic ulceration. All were treated
with reconstructive surgery and seven underwent simultaneous midfoot
fusion using a bolt, locking plate or a combination of both. At
a mean follow-up of 26 months (8 to 54), limb salvage was achieved
in all patients and 12 patients (80%) with ulceration achieved healing
and all but one patient regained independent mobilisation. There was
failure of fixation with a broken nail requiring revision surgery
in one patient. Migration of distal locking screws occurred only
when standard screws had been used but not with hydroxyapatite-coated
screws. The mean American Academy of Orthopaedic Surgeons Foot and
Ankle (AAOS-FAO) score improved from 50.7 (17 to 88) to 65.2 (22
to 88), (p = 0.015). The mean Short Form (SF)-36 Health Survey Physical
Component Score improved from 25.2 (16.4 to 42.8) to 29.8 (17.7
to 44.2), (p = 0.003) and the mean Euroqol EQ‑5D‑5L score improved
from 0.63 (0.51 to 0.78) to 0.67 (0.57 to 0.84), (p = 0.012). Single-stage correction of deformity using an intramedullary
hindfoot arthrodesis nail is a good form of treatment for patients
with severe Charcot hindfoot deformity, ulceration and instability
provided a multidisciplinary care plan is delivered. Cite this article:
We have analysedin the results of 24 femoral lengthenings in 23 patients operated on between 1993 and 2000, using a gradual elongation intramedullary nail (Albizzia). Of the 23 patients, 22 had femoral deficiency and one was of short stature. Their mean age was 16.9 years. Technical difficulties encountered during the procedure were mild or moderate in 18 femora and severe in six femora. Distraction was obtained by 15 ratchetings per day (1 mm/day). There were 18 excellent results although in two patients this was achieved after the development of a pseudarthrosis which required further surgery. There were four good and two fair results in which the lengthening obtained was at least 3 cm less than had been projected. The consolidation index was 35.2 days/cm. No patient had associated long-term stiffness of the knee. Femoral lengthening using an elongation nail gives good results and is a comfortable procedure.
We performed transoesophageal echocardiography in 111 operations (110 patients) which included medullary reaming for fresh fractures of the femur and tibia, pathological lesions of the femur, and hemiarthroplasty of the hip. Embolic events of varying intensity were seen in 97 procedures and measured pulmonary responses correlated with the severity of embolic phenomena. Twenty-four out of the 25 severe embolic responses occurred while reaming pathological lesions or during cemented hemiarthroplasty of the hip and, overall, pathological lesions produced the most severe responses. Paradoxical embolisation occurred in four patients, all with pathological lesions of the femur (21%); two died. In 12 patients large coagulative masses became trapped in the heart. Extensive pulmonary thromboembolism with reamed bone and immature clot was found at post-mortem in two patients; there was severe systemic embolisation of fat and marrow in one who had a patent foramen ovale and widespread mild systemic fat embolisation in the other without associated foraminal defect. Sequential analysis of blood from the right atrium in five patients showed considerable activation of clotting cascades during reaming.
We evaluated the biomechanical properties of two different methods of fixation for unstable fractures of the proximal humerus. Biomechanical testing of the two groups, locking plate alone (LP), and locking plate with a fibular strut graft (LPSG), was performed using seven pairs of human cadaveric humeri. Cyclical loads between 10 N and 80 N at 5 Hz were applied for 1 000 000 cycles. Immediately after cycling, an increasing axial load was applied at a rate of displacement of 5 mm/min. The displacement of the construct, maximum failure load, stiffness and mode of failure were compared. The displacement was significantly less in the LPSG group than in the LP group (p = 0.031). All maximum failure loads and measures of stiffness in the LPSG group were significantly higher than those in the LP group (p = 0.024 and p = 0.035, respectively). In the LP group, varus collapse and plate bending were seen. In the LPSG group, the humeral head cut out and the fibular strut grafts fractured. No broken plates or screws were seen in either group. We conclude that strut graft augmentation significantly increases both the maximum failure load and the initial stiffness of this construct compared with a locking plate alone.
This was a retrospective analysis of the medium-
to long-term results of 46 TC3 Sigma revision total knee replacements
using long uncemented stems in press-fit mode. Clinical and radiological analysis took place pre-operatively,
at two years post-operatively, and at a mean follow-up of 8.5 years
(4 to 12). The mean pre-operative International Knee Society (IKS)
clinical score was 42 points (0 to 74), improving to 83.7 (52 to
100) by the final follow-up. The mean IKS score for function improved
from 34.3 points (0 to 80) to 64.2 (15 to 100) at the final follow-up.
At the final follow-up 30 knees (65.2%) had an excellent result, seven
(15.2%) a good result, one (2.2%) a medium and eight (17.4%) a poor
result. There were two failures, one with anteroposterior instability
and one with aseptic loosening. The TC3 revision knee system, when used with press-fit for long
intramedullary stems and cemented femoral and tibial components,
in both septic and aseptic revisions, results in a satisfactory
clinical and radiological outcome, and has a good medium- to long-term
survival rate.