To investigate whether chronic kidney disease (CKD) is associated with the risk of all-cause revision or revision due to a periprosthetic joint infection (PJI) after primary hip or knee arthroplasty. This retrospective cohort study comprised 18,979 consecutive hip and knee arthroplasties from a single high-volume academic hospital. At a median of 5.6 years (interquartile range (IQR) 3.5 to 8.1), all deaths and revisions were counted. To overcome the competing risk of death, competing risk analysis using the cumulative incidence function (CIF) was applied to analyze the association between different stages of CKD and revisions. Confounding factors such as diabetes and BMI were considered using either a stratified CIF or the Fine and Gray model.Aims
Methods
Currently, there is no single, comprehensive national guideline for analgesic strategies for total joint replacement. We compared inpatient and outpatient opioid requirements following total hip arthroplasty (THA) versus total knee arthroplasty (TKA) in order to determine risk factors for increased inpatient and outpatient opioid requirements following total hip or knee arthroplasty. Outcomes after 92 primary total knee (n = 49) and hip (n = 43) arthroplasties were analyzed. Patients with repeat surgery within 90 days were excluded. Opioid use was recorded while inpatient and 90 days postoperatively. Outcomes included total opioid use, refills, use beyond 90 days, and unplanned clinical encounters for uncontrolled pain. Multivariate modelling determined the effect of surgery, regional nerve block (RNB) or neuraxial anesthesia (NA), and non-opioid medications after adjusting for demographics, ength of stay, and baseline opioid use.Aims
Methods
Enhanced perioperative protocols have significantly improved patient recovery following primary total knee arthroplasty (TKA). Little has been investigated the effectiveness of these protocols for revision TKA (RTKA). We report on a matched group of aseptic revision and primary TKA patients treated with an identical pain and rehabilitation programmes. Overall, 40 aseptic full-component RTKA patients were matched (surgical date, age, sex, and body mass index (BMI)) to a group of primary cemented TKA patients. All RTKAs had new uncemented stemmed femoral and tibial components with metaphyseal sleeves. Both groups were treated with an identical postoperative pain protocol. Patients were followed for at least two years. Knee Society Scores (KSS) at six weeks and at final follow-up were recorded for both groups.Aims
Methods
A key to the success of
Instability is a common indication for early
revision after both primary and
The aims of this retrospective study were to
compare the mid-term outcomes following
A retrospective study was conducted to measure short-term Nonweightbearing supine RSA exams were performed postoperatively and at six, 12, and 24 months. Weightbearing standing RSA exams were performed on select patients at 12 and 24 months. Wear was measured both linearly (joint space) and volumetrically (digital model overlap) at each available follow-up. Precision of both methods was assessed by comparing double RSA exams. Patient age, sex, body mass index, and Oxford Knee Scores were analyzed for any association with PE wear.Aims
Patients and Methods
The aim of this study was to establish the results of isolated exchange of the tibial polyethylene insert in revision total knee arthroplasty (RTKA) in patients with well-fixed femoral or tibial components. We report on a series of RTKAs where only the polyethylene was replaced, and the patients were followed for a mean of 13.2 years (10.0 to 19.1). Our study group consisted of 64 non-infected, grossly stable TKA patients revised over an eight-year period (1998 to 2006). The mean age of the patients at time of revision was 72.2 years (48 to 88). There were 36 females (56%) and 28 males (44%) in the cohort. All patients had received the same cemented, cruciate-retaining patella resurfaced primary TKA. All subsequently underwent an isolated polyethylene insert exchange. The mean time from the primary TKA to RTKA was 9.1 years (2.2 to 16.1).Aims
Patients and Methods
The use of vitamin E-infused highly crosslinked polyethylene (HXLPE) in total knee prostheses is controversial. In this paper we have compared the clinical and radiological results between conventional polyethylene and vitamin E-infused HXLPE inserts in total knee arthroplasty (TKA). The study included 200 knees (175 patients) that underwent TKA using the same total knee prostheses. In all, 100 knees (77 patients) had a vitamin E-infused HXLPE insert (study group) and 100 knees (98 patients) had a conventional polyethylene insert (control group). There were no significant differences in age, sex, diagnosis, preoperative knee range of movement (ROM), and preoperative Knee Society Score (KSS) between the two groups. Clinical and radiological results were evaluated at two years postoperatively.Aims
Patients and Methods
Anterior cruciate ligament (ACL) and multiligament knee (MLK) injuries increase the risk of development of knee osteoarthritis and eventual need for total knee arthroplasty (TKA). There are limited data regarding implant use and outcomes in these patients. The aim of this study was to compare the use of constrained implants and outcomes among patients undergoing TKA with a history of prior knee ligament reconstruction (PKLR) Patients with a history of ACL or MLK reconstruction who underwent TKA between 2007 and 2017 were identified in a single-institution registry. There were 223 patients who met inclusion criteria (188 ACL reconstruction patients, 35 MLK reconstruction patients). A matched cohort, also of 223 patients, was identified based on patient age, body mass index (BMI), sex, and year of surgery. There were 144 male patients and 79 female patients in both cohorts. Mean age at the time of TKA was 57.2 years (31 to 88). Mean BMI was 29.7 kg/m2 (19.5 to 55.7).Aims
Patients and Methods
Bone void fillers are increasingly being used for dead space management in arthroplasty revision surgery. The aim of this study was to investigate the influence of calcium sulphate bone void filler (CS-BVF) on the damage and wear of total knee arthroplasty using experimental wear simulation. A total of 18 fixed-bearing U2 total knee arthroplasty system implants (United Orthopedic Corp., Hsinchu, Taiwan) were used. Implants challenged with CS-BVF were compared with new implants (negative controls) and those intentionally scratched with a diamond stylus (positive controls) representative of severe surface damage (n = 6 for each experimental group). Three million cycles (MC) of experimental simulation were carried out to simulate a walking gait cycle. Wear of the ultra-high-molecular-weight polyethylene (UHMWPE) tibial inserts was measured gravimetrically, and damage to articulating surfaces was assessed using profilometry.Objectives
Methods
The aim of this study was to evaluate the prosthesis characteristics and associated conditions that may modify the survival of total femoral endoprosthetic replacements (TFEPR). In all, 81 patients treated with TFEPR from 1976 to 2017 were retrospectively evaluated and failures were categorized according to the Henderson classification. There were 38 female patients (47%) and 43 male patients (53%) with a mean age at diagnosis of 43 years (12 to 86). The mean follow-up time was 10.3 years (0 to 31.7). A survival analysis was performed followed by univariate and multivariate Cox regression to identify independent implant survival factors.Aims
Patients and Methods
The purpose of the present study was to compare patient-specific instrumentation (PSI) and conventional surgical instrumentation (CSI) for total knee arthroplasty (TKA) in terms of early implant migration, alignment, surgical resources, patient outcomes, and costs. The study was a prospective, randomized controlled trial of 50 patients undergoing TKA. There were 25 patients in each of the PSI and CSI groups. There were 12 male patients in the PSI group and seven male patients in the CSI group. The patients had a mean age of 69.0 years (Aims
Patients and Methods
Calcium sulphate has traditionally been used as a filler of dead space arising during surgery. Various complications have been described following the use of Stimulan bio-absorbable calcium sulphate beads. This study is a prospective observational study to assess the safety profile of these beads when used in revision arthroplasty, comparing the complication rates with those reported in the literature. A total of 755 patients who underwent 456 revision total knee arthroplasties (TKA) and 299 revision total hip arthroplasties (THA), with a mean follow-up of 35 months (0 to 78) were included in the study.Aims
Methods
The aim of this review was to evaluate the available literature
and to calculate the pooled sensitivity and specificity for the
different alpha-defensin test systems that may be used to diagnose
prosthetic joint infection (PJI). Studies using alpha-defensin or Synovasure (Zimmer Biomet, Warsaw,
Indiana) to diagnose PJI were identified from systematic searches
of electronic databases. The quality of the studies was evaluated
using the Quality Assessment of Studies of Diagnostic Accuracy (QUADAS)
tool. Meta-analysis was completed using a bivariate model.Aims
Materials and Methods