We have examined the outcome of 400 consecutive patients who underwent total knee replacement with the Low Contact Stress mobile-bearing system between 1993 and 1994 and were followed up for a minimum of ten years. All operations were performed by surgeons in Christchurch, New Zealand, who used no other knee prosthesis during the study time. At ten years after operation 238 patients (244 knees) remained for independent clinical and radiological assessment. There was a significant improvement (p <
0.001) in the postoperative knee scores at one, three, seven and ten years, although a slight deterioration in the score occurred between seven and ten years which did not reach statistical significance. The survival for polyethylene wear or loosening was 97% (95% CI 96 to 98) and survival using reoperation for any cause was 92% (95% CI 90 to 94) at 12 years. Polyethylene wear was more common in the meniscal-bearing component, with five knees requiring revision and a further eight demonstrating early wear. Osteolysis was not seen in the rotating platform component, but was present in three of the meniscal-bearing knees. There was no association between the radiological alignment at one year and those knees that subsequently required revision. However, there was an association between the overall limb alignment and the Western Ontario McMasters University score (p <
0.001). The Low Contact Stress mobile-bearing total knee replacement has proved to be a reliable implant at ten years when used in primary knee replacement irrespective of the deformity and diagnosis.
This study reviews the predisposing features, the clinical, and laboratory findings at the time of diagnosis and the results of single-stage revision of prosthetic replacement of the elbow for infection. Deep infection occurred in six of 305 (1.9%) primary total elbow replacements. The mean follow-up after revision was 6.8 years (6 months to 16 years) and the mean age at the time of revision was 62.7 years (56 to 74). All six cases with infection had rheumatoid arthritis and had received steroid therapy. The infective organism was
This prospective randomised controlled double-blind
trial compared two types of PFC Sigma total knee replacement (TKR),
differing in three design features aimed at improving flexion. The
outcome of a standard fixed-bearing posterior cruciate ligament-preserving
design (FB-S) was compared with that of a high-flexion rotating-platform
posterior-stabilised design (RP-F) at one year after TKR. The study group of 77 patients with osteoarthritis of the knee
comprised 37 men and 40 women, with a mean age of 69 years (44.9
to 84.9). The patients were randomly allocated either to the FB-S
or the RP-F group and assessed pre-operatively and at one year post-operatively.
The mean post-operative non-weight-bearing flexion was 107° (95%
confidence interval (CI) 104° to 110°)) for the FB-S group and 113°
(95% CI 109° to 117°) for the RP-F group, and this difference was
statistically significant (p = 0.032). However, weight-bearing range
of movement during both level walking and ascending a slope as measured
during flexible electrogoniometry was a mean of 4° lower in the RP-F
group than in the FB-S group, with 58° (95% CI 56° to 60°) Although the RP-F group achieved higher non-weight-bearing knee
flexion, patients in this group did not use this during activities
of daily living and reported more pain one year after surgery