The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 12 | Pages 1562 - 1569
1 Dec 2008
Mäkelä K Eskelinen A Pulkkinen P Paavolainen P Remes V

We have analysed from the Finnish Arthroplasty Register the long-term survivorship of the 12 most commonly-used cemented implants between 1980 and 2005 in patients aged 55 years or older with osteoarthritis.

Only two designs of femoral component, the Exeter Universal and the Müller Straight femoral component had a survivorship of over 95% at ten years with revision for aseptic loosening as the endpoint. At 15 years of the femoral and acetabular component combinations, only the Exeter Universal/Exeter All-poly implant had a survival rate of over 90% with revision for aseptic loosening as the endpoint. In the subgroup of patients aged between 55 and 64 years, survivorship overall was less than 90% at ten years.

The variation in the long-term rates of survival of different cemented hip implants was considerable in patients aged 55 years or older. In those aged between 55 and 64 years, none of the cemented prostheses studied yielded excellent long-term survival rates (≥ 90% at 15 years).


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 5 | Pages 635 - 639
1 May 2005
Ikeuchi M Kawakami T Kitaoka K Okanoue Y Tani T

We describe a new technique of reconstruction of the deficient acetabulum in cementless total hip arthroplasty. The outer iliac table just above the deficient acetabulum is osteotomised and slid downwards. We have termed this an iliac sliding graft. Between October 1997 and November 2001, cementless total hip arthroplasty with an iliac sliding graft was performed on 19 patients (19 hips) with acetabular dysplasia. The mean follow-up was 3.4 years (2 to 6).

The mean pre-operative Harris hip score was 45.1 which improved significantly to 85.3 at the time of the final follow-up. No patient had post-operative abductor dysfunction. Incorporation of the graft was seen after two to three months in all patients. Resorption of the graft and radiolucencies were infrequent. This technique is a useful alternative to femoral head autografting when the patient’s own femoral head cannot be used.


The Journal of Bone & Joint Surgery British Volume
Vol. 77-B, Issue 1 | Pages 3 - 5
1 Jan 1995
Murray D Carr A Bulstrode C


The Journal of Bone & Joint Surgery British Volume
Vol. 61-B, Issue 4 | Pages 419 - 421
1 Nov 1979
Hunter G Welsh R Cameron H Bailey W

The results of 140 total hip revision procedures for "non-septic" loosening, dislocation, and fracture of the femoral stem or shaft have been personally reviewed and rated by the Harris method. The minimum follow-up period was six months: thirty-three (24 per cent) showed excellent or good results, seventy-two (51 per cent) showed fair or poor results. Subsequent excision arthroplasty was performed in thirty-one patients. The infection rate for these revision procedures was very high, suggesting that many were already infected at the time of revision, and that every "loose" hip must be assumed to be infected until proved otherwise. The mortality rate of 3 per cent was surprisingly low after more than one major surgical procedure in these elderly patients.


The Bone & Joint Journal
Vol. 96-B, Issue 1 | Pages 43 - 47
1 Jan 2014
Craig P Bancroft G Burton A Collier S Shaylor P Sinha A

The issues surrounding raised levels of metal ions in the blood following large head metal-on-metal total hip replacement (THR), such as cobalt and chromium, have been well documented. Despite the national popularity of uncemented metal-on-polyethylene (MoP) THR using a large-diameter femoral head, few papers have reported the levels of metal ions in the blood following this combination. Following an isolated failure of a 44 mm Trident–Accolade uncemented THR associated with severe wear between the femoral head and the trunnion in the presence of markedly elevated levels of cobalt ions in the blood, we investigated the relationship between modular femoral head diameter and the levels of cobalt and chromium ions in the blood following this THR.

A total of 69 patients received an uncemented Trident–Accolade MoP THR in 2009. Of these, 43 patients (23 men and 20 women, mean age 67.0 years) were recruited and had levels of cobalt and chromium ions in the blood measured between May and June 2012. The patients were then divided into three groups according to the diameter of the femoral head used: 12 patients in the 28 mm group (controls), 18 patients in the 36 mm group and 13 patients in the 40 mm group. A total of four patients had identical bilateral prostheses in situ at phlebotomy: one each in the 28 mm and 36 mm groups and two in the 40 mm group.

There was a significant increase in the mean levels of cobalt ions in the blood in those with a 36 mm diameter femoral head compared with those with a 28 mm diameter head (p = 0.013). The levels of cobalt ions in the blood were raised in those with a 40 mm diameter head but there was no statistically significant difference between this group and the control group (p = 0.152). The levels of chromium ions in the blood were normal in all patients.

The clinical significance of this finding is unclear, but we have stopped using femoral heads with a diameter of ≤ 36 mm, and await further larger studies to clarify whether, for instance, this issue particularly affects this combination of components.

Cite this article: Bone Joint J 2014;96-B:43–7.


The Journal of Bone & Joint Surgery British Volume
Vol. 86-B, Issue 8 | Pages 1214 - 1219
1 Nov 2004
Jafri AA Green SM Partington PF McCaskie AW Muller SD

Fatigue fractures which originate at stress-concentrating voids located at the implant-cement interface are a potential cause of septic loosening of cemented femoral components. Heating of the component to 44°C is known to reduce the porosity of the cement-prosthesis interface.

The temperature of the cement-bone interface was recorded intra-operatively as 32.3°C. A simulated femoral model was devised to study the effect of heating of the component on the implant-cement interface.

Heating of the implant and vacuum mixing have a synergistic effect on the porosity of the implant-cement interface, and heating also reverses the gradients of microhardness in the mantle.

Heating of the implant also reduces porosity at the interface depending on the temperature. A minimum difference in temperature between the implant and the bone of 3°C was required to produce this effect. The optimal difference was 7°C, representing a balance between maximal reduction of porosity and an increased risk of thermal injury. Using contemporary cementing techniques, heating the implant to 40°C is recommended to produce an optimum effect.


The Bone & Joint Journal
Vol. 95-B, Issue 9 | Pages 1178 - 1183
1 Sep 2013
Li H Mao Y Oni JK Dai K Zhu Z

In developmental dysplasia of the hip (DDH), a bone defect is often observed superior to the acetabulum after the reconstruction at the level of the true acetabulum during total hip replacement (THR). However, the essential amount of uncemented acetabular component coverage required for a satisfactory outcome remains controversial. The purpose of this study was to assess the stability and function of acetabular components with a lack of coverage > 30% (31% to 50%). A total of 760 DDH patients underwent THR with acetabular reconstruction at the level of the true floor. Lack of coverage above the acetabular component of > 30% occurred in 56 patients. Intra-operatively, autogenous morcellised bone grafts were used to fill the uncovered portion. Other than two screws inserted through the acetabular shell, no additional structural supports were used in these hips. In all, four patients were lost to follow-up. Therefore, 52 patients (52 hips, 41 women and 11 men) with a mean age of 60.1 years (42 to 78) were available for this study at a mean of 4.8 years (3 to 7). There were no instances of prosthesis revision or marked loosening during the follow-up. The Harris hip score improved from a mean of 40.7 points (sd 12.2) pre-operatively to 91.1 (sd 5.0) at the last follow-up. Radiological analysis with medical imaging software allowed us to calculate the extent of the uncoverage in terms of the uncovered arc of the implant as viewed on the anteroposterior pelvic radiograph. From this we propose that up to 17 mm of lateral undercoverage in the presence of a stable initial implantation in the presence of bone autografting, with an inclination angle of the acetabular component between 40° and 55°, is acceptable. This represents undercoverage of ≤ 50%.

Cite this article: Bone Joint J 2013;95-B:1178–83.


The Journal of Bone & Joint Surgery British Volume
Vol. 85-B, Issue 5 | Pages 775 - 776
1 Jul 2003
WATKINS ND


The Journal of Bone & Joint Surgery British Volume
Vol. 84-B, Issue 6 | Pages 832 - 838
1 Aug 2002
Espehaug B Furnes O Havelin LI Engesæter LB Vollset SE

Using data from the Norwegian Arthroplasty Register, we have the assessed survival of 17 323 primary Charnley hip prostheses in patients with osteoarthritis based upon the type of cement used for the fixation of the implant.

Overall, 9.2% had been revised after follow-up for ten years; 71% of the failures involved aseptic loosening of the femoral component. We observed significantly increased rates of failure for prostheses inserted with CMW1 and CMW3 cements. Using implants fixed with gentamicin-containing Palacos cement as the reference, the adjusted Cox regression failure rate ratios were 1.1 (95% CI 0.9 to 1.4) for implants cemented with plain Palacos, 1.1 (95% CI 0.7 to 1.6) for Simplex, 2.1 (95% 1.5 to 2.9) for gentamicin-containing CMW1, 2.0 (95% CI 1.6 to 2.4) for plain CMW1 and 3.0 (95% CI 2.3 to 3.9) for implants fixed with CMW3 cement. The adjusted failure rate at ten years varied from 5.9% for implants fixed with gentamicin-containing Palacos to 17% for those fixed with CMW3.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 8 | Pages 1036 - 1044
1 Aug 2012
Penny JO Brixen K Varmarken JE Ovesen O Overgaard S

It is accepted that resurfacing hip replacement preserves the bone mineral density (BMD) of the femur better than total hip replacement (THR). However, no studies have investigated any possible difference on the acetabular side.

Between April 2007 and March 2009, 39 patients were randomised into two groups to receive either a resurfacing or a THR and were followed for two years. One patient’s resurfacing subsequently failed, leaving 19 patients in each group.

Resurfaced replacements maintained proximal femoral BMD and, compared with THR, had an increased bone mineral density in Gruen zones 2, 3, 6, and particularly zone 7, with a gain of 7.5% (95% confidence interval (CI) 2.6 to 12.5) compared with a loss of 14.6% (95% CI 7.6 to 21.6). Resurfacing replacements maintained the BMD of the medial femoral neck and increased that in the lateral zones between 12.8% (95% CI 4.3 to 21.4) and 25.9% (95% CI 7.1 to 44.6).

On the acetabular side, BMD was similar in every zone at each point in time. The mean BMD of all acetabular regions in the resurfaced group was reduced to 96.2% (95% CI 93.7 to 98.6) and for the total hip replacement group to 97.6% (95% CI 93.7 to 101.5) (p = 0.4863). A mean total loss of 3.7% (95% CI 1.0 to 6.5) and 4.9% (95% CI 0.8 to 9.0) of BMD was found above the acetabular component in W1 and 10.2% (95% CI 0.9 to 19.4) and 9.1% (95% CI 3.8 to 14.4) medial to the implant in W2 for resurfaced replacements and THRs respectively. Resurfacing resulted in a mean loss of BMD of 6.7% (95% CI 0.7 to 12.7) in W3 but the BMD inferior to the acetabular component was maintained in both groups.

These results suggest that the ability of a resurfacing hip replacement to preserve BMD only applies to the femoral side.


The Journal of Bone & Joint Surgery British Volume
Vol. 83-B, Issue 4 | Pages 579 - 579
1 May 2001
Furnes O Lie SA Espehaug B Vollset SE Engesaeter LB Havelin LI

We studied the rates of revision for 53 698 primary total hip replacements (THRs) in nine different groups of disease. Factors which have previously been shown to be associated with increased risk of revision, such as male gender, young age, or certain types of uncemented prosthesis, showed important differences between the diagnostic groups. Without adjustment for these factors we observed an increased risk of revision in patients with paediatric hip diseases and in a small heterogeneous ‘other’ group, compared with patients with primary osteoarthritis. Most differences were reduced or disappeared when an adjustment for the prognostic factors was made. After adjustment, an increased relative risk (RR) of revision compared with primary osteoarthritis was seen in hips with complications after fracture of the femoral neck (RR = 1.3, p = 0.0005), in hips with congenital dislocation (RR = 1.3, p = 0.03), and in the heterogenous ‘other’ group. The analyses were also undertaken in a more homogenous subgroup of 16 217 patients which had a Charnley prosthesis implanted with high-viscosity cement. The only difference in this group was an increased risk for revision in patients who had undergone THR for complications after fracture of the femoral neck (RR = 1.5, p = 0.0005).

THR for diagnoses seen mainly among young patients had a good prognosis, but they had more often received inferior uncemented implants. If a cemented Charnley prosthesis is used, the type of disease leading to THR seems in most cases to have only a minor influence on the survival of the prosthesis.


The Journal of Bone & Joint Surgery British Volume
Vol. 82-B, Issue 4 | Pages 469 - 472
1 May 2000
Thomas DP


The Journal of Bone & Joint Surgery British Volume
Vol. 81-B, Issue 2 | Pages 371 - 371
1 Mar 1999
DESHMUKH RG


The Journal of Bone & Joint Surgery British Volume
Vol. 80-B, Issue 5 | Pages 876 - 879
1 Sep 1998
Rodriguez JA Ranawat CS Maniar RN Umlas ME

We report 16 cases of erythematous eruption on the skin within the flaps of the surgical incision after primary total hip replacement over an eight-year period. The symptoms began within nine months of operation in 13 hips, and two to three years after in three. Four patients had recurrent episodes. All were treated with antibiotics (15 intravenous, one oral) with complete resolution of the eruption within one to six days. The mean follow-up after the last episode of cellulitis was 27 months (14 to 76). There were no cases of periprosthetic sepsis or other sequelae.


The Journal of Bone & Joint Surgery British Volume
Vol. 80-B, Issue 3 | Pages 404 - 410
1 May 1998
Matsui M Nakata K Masuhara K Ohzono K Sugano N Ochi T

We implanted 51 Metal-Cancellous Cementless Lübeck (MCCL) prostheses into 45 patients with dysplastic hips and followed 49 hips (96.1%) for five to nine years. One had needed revision for stem fracture and one for infection; the clinical outcome of the other 47 hips was assessed using the Merle d’Aubigné and Postel hip score.

All hips were either excellent (63%) or good (37%). Three patients (6%) had mild thigh pain at six months, but this had settled within two years. Serial radiographs showed stable fixation with bone ingrowth in all hips, with increased density of the cancellous bone in contact with the implant and some trabecular ingrowth. There was early varus shift of the stem in one hip, but this stabilised in three months. Osteolysis of the femoral cortex was seen in one hip at seven years after surgery, and mild bone resorption due to stress shielding in 31 (63%). Acetabular bone grafting with autogenous bone from the femoral head gave successful support to the socket in 13 hips.

The MCCL prosthesis gave satisfactory mid-term results in patients with osteoarthritis secondary to hip dysplasia.


The Journal of Bone & Joint Surgery British Volume
Vol. 80-B, Issue 2 | Pages 370 - 371
1 Mar 1998
WARWICK D


The Journal of Bone & Joint Surgery British Volume
Vol. 80-B, Issue 1 | Pages 180 - 180
1 Jan 1998
GAMBHIR AK SOCHART DH


The Journal of Bone & Joint Surgery British Volume
Vol. 79-B, Issue 6 | Pages 1037 - 1038
1 Nov 1997
AMSTUTZ HC DOREY F


The Journal of Bone & Joint Surgery British Volume
Vol. 79-B, Issue 5 | Pages 878 - 879
1 Sep 1997
MCNALLY MA


The Journal of Bone & Joint Surgery British Volume
Vol. 79-B, Issue 5 | Pages 880 - 880
1 Sep 1997
RANSTAM J SWIERSTRA BA