header advert
Results 461 - 480 of 482
Results per page:
The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 12 | Pages 1596 - 1602
1 Dec 2006
Muller SD Deehan DJ Holland JP Outterside SE Kirk LMG Gregg PJ McCaskie AW

The role of modular tibial implants in total knee replacement is not fully defined. We performed a prospective randomised controlled clinical trial using radiostereophotogrammetric analysis to compare the performance of an all-polyethylene tibia with a metal-backed cruciate-retaining condylar design, PFC-∑ total knee replacement for up to 24 months. There were 51 patients who were randomised into two treatment groups. There were 10 subsequent withdrawals, leaving 21 all-polyethylene and 20 metal-backed tibial implants. No patient was lost to follow-up. There were no significant demographic differences between the groups. At two years one metal-backed implant showed migration > 1 mm, but no polyethylene implant reached this level. There was a significant increase in the SF-12 and Oxford knee scores after operation in both groups.

In an uncomplicated primary total knee replacement the all-polyethylene PFC-∑ tibial prosthesis showed no statistical difference in migration from that of the metal-backed counterpart. There was no difference in the clinical results as assessed by the SF-12, the Oxford knee score, alignment or range of movement at 24 months, although these assessment measures were not statistically powered in this study.


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 8 | Pages 1016 - 1021
1 Aug 2006
Delport HP Banks SA De Schepper J Bellemans J

Mobile-bearing posterior-stabilised knee replacements have been developed as an alternative to the standard fixed- and mobile-bearing designs. However, little is known about the in vivo kinematics of this new group of implants. We investigated 31 patients who had undergone a total knee replacement with a similar prosthetic design but with three different options: fixed-bearing posterior cruciate ligament-retaining, fixed-bearing posterior-stabilised and mobile-bearing posterior-stabilised. To do this we used a three-dimensional to two-dimensional model registration technique. Both the fixed- and mobile-bearing posterior-stabilised configurations used the same femoral component. We found that fixed-bearing posterior stabilised and mobile-bearing posterior-stabilised knee replacements demonstrated similar kinematic patterns, with consistent femoral roll-back during flexion. Mobile-bearing posterior-stabilised knee replacements demonstrated greater and more natural internal rotation of the tibia during flexion than fixed-bearing posterior-stabilised designs. Such rotation occurred at the interface between the insert and tibial tray for mobile-bearing posterior-stabilised designs. However, for fixed-bearing posterior-stabilised designs, rotation occurred at the proximal surface of the bearing. Posterior cruciate ligament-retaining knee replacements demonstrated paradoxical sliding forward of the femur.

We conclude that mobile-bearing posterior-stabilised knee replacements reproduce internal rotation of the tibia more closely during flexion than fixed-bearing posterior-stabilised designs. Furthermore, mobile-bearing posterior-stabilised knee replacements demonstrate a unidirectional movement which occurs at the upper and lower sides of the mobile insert. The femur moves in an anteroposterior direction on the upper surface of the insert, whereas the movement at the lower surface is pure rotation. Such unidirectional movement may lead to less wear when compared with the multidirectional movement seen in fixed-bearing posterior-stabilised knee replacements, and should be associated with more evenly applied cam-post stresses.


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 1 | Pages 35 - 39
1 Jan 2006
Beaulé PE Campbell PA Hoke R Dorey F

During hip resurfacing arthroplasty, excessive valgus positioning or surgical technique can result in notching of the femoral neck. Although mechanical weakening and subsequent fracture of the femoral neck are well described, the potential damage to the retinacular vessels leading to an ischaemic event is relatively unknown. Using laser Doppler flowmetry, we measured the blood flow in 14 osteoarthritic femoral heads during routine total hip replacement surgery, before and after notching of the femoral neck. In ten hips there was a reduction in blood flow of more than 50% from the baseline value after simulated notching of the femoral neck. Our results suggest that femoral head vascularity in the osteoarthritic state is similar to the non-arthritic state, where damage to the extraosseous vessels can predispose to avascular necrosis. Surgeons who perform resurfacing arthroplasty of the hip should pay careful attention to these vessels by avoiding excessive dissection around the femoral neck and/or notching.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 1 | Pages 34 - 38
1 Jan 2007
Epinette J Manley MT

This study describes 146 primary total knee replacements, either fully or partially coated with hydroxyapatite of which 74 knees in 68 patients were available for clinical and radiological assessment at a mean of 11.2 years (10 to 15). The global failure rate was 1.37% and survival rate with mechanical failure as the end-point was 98.14%. Radiological assessment indicated intimate contact between bone and the hydroxyapatite coating. Over time the hydroxyapatite coating appears to encourage filling of interface gaps remaining after surgery. Our results compare favourably with those of series describing cemented or porous-coated knee replacements, and suggest that fixation with hydroxyapatite is a reliable option in primary total knee replacement.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 10 | Pages 1317 - 1323
1 Oct 2007
Kim Y Yoon S Kim J

We compared the results of 146 patients who received an anatomic modular knee fixed-bearing total knee replacement (TKR) in one knee and a low contact stress rotating platform mobile-bearing TKR in the other. There were 138 women and eight men with a mean age of 69.8 years (42 to 80). The mean follow-up was 13.2 years (11.0 to 14.5). The patients were assessed clinically and radiologically using the rating systems of the Hospital for Special Surgery and the Knee Society at three months, six months, one year, and annually thereafter.

The assessment scores of both rating systems pre-operatively and at the final review did not show any statistically significant differences between the two designs of implant. In the anatomic modular knee group, one knee was revised because of aseptic loosening of the tibial component and one because of infection. In addition, three knees were revised because of wear of the polyethylene tibial bearing. In the low contact stress group, two knees were revised because of instability requiring exchange of the polyethylene insert and one because of infection.

The radiological analysis found no statistical difference in the incidence of radiolucent lines at the final review (Student’s t-test, p = 0.08), most of which occurred at tibial zone 1. The Kaplan-Meier survivorship for aseptic loosening of the anatomic modular knee and the low contact stress implants at 14.5 years was 99% and 100%, respectively, with a 95% confidence interval of 94% to 100% for both designs.

We found no evidence of the superiority of one design over the other at long-term follow-up.


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 6 | Pages 796 - 799
1 Jun 2005
Lakdawala A Todo S Scott G

We investigated the changes in surface roughness of retrieved femoral components in 18 men and four women at revision knee surgery. The mean age at revision was 68.4 years and the mean period of implantation was for 55.6 months. Eighteen implants were retrieved for aseptic loosening and four for infection. The surface changes in the articulating areas were inspected visually and the roughness (Ra) analysed with a profilometer. Parallel scratching and burnishing were the two main forms of damage. The mean Ra measurements in the articulating areas showed no statistically significant difference when compared with those in a control area on either side of the patellar groove at the apex of the femoral flange. This suggests that it is not essential to revise a well-fixed and correctly aligned femoral component where the polished surface has become burnished or bears fine parallel scratches, if the revision is conducted solely for failure of the tibial component.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 4 | Pages 449 - 454
1 Apr 2007
Kim Y Kim J Yoon S

We studied prospectively the long-term results of the Charnley Elite-Plus femoral stem in 184 consecutive young patients (194 hips). There were 130 men and 54 women with a mean age of 49.1 years (21 to 60). The predominant diagnosis was osteonecrosis of the femoral head (63.6%, 117 patients). Clinical and radiological evaluation was undertaken at each follow-up. The mean follow-up was 11.2 years (10 to 12).

The mean pre-operative Harris hip score was 43.4 (12 to 49) which improved to 91 (59 to 100) at the final follow-up. The survival of the femoral stem at 12 years was 99% with revision as the end-point. The mean annual linear wear of the polyethylene liner was 0.17 mm (0.13 to 0.22). The prevalence of acetabular osteolysis was 10.8% (21 hips) and osteolysis of the calcar femorale 12.9% (25 hips).

A third-generation cementing technique, accurate alignment of the stem and the use of a 22 mm zirconia head were important factors in the prevention of aseptic loosening of the Elite Plus femoral stem in these high-risk young patients.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 7 | Pages 904 - 910
1 Jul 2007
Kim Y Kim D Kim J

We conducted a randomised prospective study to evaluate the clinical and radiological results of a mobile- and fixed-bearing total knee replacement of similar design in 174 patients who had bilateral simultaneous knee replacement. The mean follow-up was for 5.6 years (5.2 to 6.1).

The total knee score, pain score, functional score and range of movement were not statistically different (p > 0.05) between the two groups. Osteolysis was not seen in any knee in either group. Two knees (1%) in the mobile-bearing group required revision because of infection; none in the fixed-bearing group needed revision. Excellent results can be achieved with both mobile- and fixed-bearing prostheses of similar design at mid-term follow-up. We could demonstrate no significant clinical advantage for a mobile bearing.


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 12 | Pages 1585 - 1593
1 Dec 2008
Henricson A Linder L Nilsson KG

We compared the performance of uncemented trabecular metal tibial components in total knee replacement with that of cemented tibial components in patients younger than 60 years over two years using radiostereophotogrammetric analysis (RSA). A total of 22 consecutive patients (mean age 53 years, 33 to 59, 26 knees) received an uncemented NexGen trabecular metal cruciate-retaining monobloc tibial component and 19 (mean 53 years, 44 to 59, 21 knees) a cemented NexGen Option cruciate-retaining modular tibial component.

All the trabecular metal components migrated during the initial three months and then stabilised. The exception was external rotation, which did not stabilise until 12 months. Unlike conventional metal-backed implants which displayed a tilting migration comprising subsidence and lift-off from the tibial tray, most of the trabecular metal components showed subsidence only, probably due to the elasticity of the implant.

This pattern of subsidence is regarded as being beneficial for uncemented fixation.


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 3 | Pages 320 - 323
1 Mar 2005
Little CP Ruiz AL Harding IJ McLardy-Smith P Gundle R Murray DW Athanasou NA

We present the histological findings of bone retrieved from beneath the femoral components of failed metal-on-metal hip resurfacing arthroplasties. Of a total of 377 patients who underwent resurfacing arthroplasty, 13 required revision; for fracture of the femoral neck in eight, loosening of a component in three and for other reasons in two. None of these cases had shown histological evidence of osteonecrosis in the femoral bone at the time of the initial implantation.

Bone from the remnant of the femoral head showed changes of osteonecrosis in all but one case at revision. In two cases of fracture which occurred within a week of implantation, the changes were compatible with early necrosis of the edge of the fracture. In the remaining six fractures, there were changes of established osteonecrosis. In all but one of the non-fracture cases, patchy osteonecrosis was seen.

We conclude that histological evidence of osteonecrosis is a common finding in failed resurfaced hips. Given that osteonecrosis is extensive in resurfaced femoral heads which fail by fracture, it is likely to play a role in the causation of these fractures.


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 11 | Pages 1441 - 1447
1 Nov 2006
Cui W Won Y Baek M Kim K Cho J

The appearance of the ‘grand-piano sign’ on the anterior resected surface of the femur has been considered to be a marker for correct femoral rotational alignment during total knee replacement. Our study was undertaken to assess quantitatively the morphological patterns on the resected surface after anterior femoral resection with various angles of external rotation, using a computer-simulation technique. A total of 50 right distal femora with varus osteoarthritis in 50 Korean patients were scanned using computerised tomography. Computer image software was used to simulate the anterior femoral cut, which was applied at an external rotation of 0°, 3° and 6° relative to the posterior condylar axis, and parallel to the surgical and clinical epicondylar axes in each case. The morphological patterns on the resected surface were quantified and classified as the ‘grand-piano sign’, ‘the boot sign’ and the ‘butterfly sign’. The surgeon can use the analogy of these quantified sign patterns to ensure that a correct rotational alignment has been obtained intra-operatively.


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 10 | Pages 1286 - 1292
1 Oct 2006
McLaughlin JR Lee KR

We studied a consecutive series of 285 uncemented total hip replacements in 260 patients using the Taperloc femoral component and the T-Tap acetabular component. The outcome of every hip was determined in both living and deceased patients. A complete clinical and radiological follow-up was obtained for 209 hips in 188 living patients, followed for a mean of 14.5 years (10 to 18.9). They were divided into two groups, obese and non-obese, as determined by their body mass index. There were 100 total hip replacements in 89 patients in the obese cohort (body mass index ≥ 30 kg/m2), and 109 in 99 non-obese (body mass index < 30 kg/m2) patients. A subgroup analysis of 31 patients of normal weight (body mass index 20 kg/m2 to 25 kg/m2) (33 hips) and 26 morbidly obese patients (body mass index ≥ 35 kg/m2) (30 hips) was also carried out.

In the obese group five femoral components (5%) were revised and one (1%) was loose by radiological criteria. Femoral cortical osteolysis was seen in eight hips (8%). The acetabular component was revised in 57 hips (57%) and a further 17 (17%) were loose. The mean Harris hip score improved from 52 (30 to 66) pre-operatively to 89 (49 to 100) at final follow-up. Peri-operative complications occurred in seven patients (7%).

In the non-obese group six (6%) femoral components were revised and one (1%) was loose. Femoral cortical osteolysis occurred in six hips (6%). The acetabular component was revised in 72 hips (66%) and a further 18 (17%) were loose. The mean Harris hip score increased from 53 (25 to 73) prior to surgery to 89 (53 to 100) at the time of each patient’s final follow-up radiograph.

No statistically significant difference was identified between the obese and non-obese patients with regards to clinical and radiological outcome or complications. The subgroup analysis of patients of normal weight and those who were morbidly obese showed no statistically significant difference in the rate of revision of either component.

Our findings suggest there is no evidence to support withholding total hip replacement from obese patients with arthritic hips on the grounds that their outcome will be less satisfactory than those who are not obese.


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 6 | Pages 755 - 758
1 Jun 2005
Nizard R Sedel L Hannouche D Hamadouche M Bizot P


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 1 | Pages 12 - 15
1 Jan 2005
Robertson A Lavalette D Morgan S Angus PD

We present the clinical and radiographic outcome of 68 consecutive primary total hip replacements performed in 54 patients under the age of 55 years using a hydroxyapatite-coated femoral component and threaded cup with a modular ceramic head (JRI-Furlong). We reviewed 62 (91%) hips at a median follow-up of 8.8 years (5 to 13.8) after implantation; six (9%) were lost to follow-up. At review there had been four (6%) revisions but only one for aseptic loosening (acetabulum). Radiographic review of the remaining hips did not identify any evidence of femoral or acetabular loosening. The median Harris and Merle d’Aubigné and Postel hip scores were 95.9 (42.7 to 100) and 17 (3 to 18) respectively. The JRI-Furlong hip gives promising functional and radiographic results in young patients in the medium term.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 9 | Pages 1158 - 1164
1 Sep 2007
Lusty PJ Watson A Tuke MA Walter WL Walter WK Zicat B

We studied 33 third generation, alumina ceramic-on-ceramic bearings retrieved from cementless total hip replacements after more than six months in situ. Wear volume was measured with a Roundtest machine, and acetabular orientation from the anteroposterior pelvic radiograph. The overall median early wear rate was 0.1 mm3/yr for the femoral heads, and 0.04 mm3/yr for the acetabular liners. We then excluded hips where the components had migrated. In this stable subgroup of 22 bearings, those with an acetabular anteversion of < 15° (seven femoral heads) had a median femoral head wear rate of 1.2 mm3/yr, compared with 0 mm3/yr for those with an anteversion of ≥15° (15 femoral heads, p < 0.001). Even under edge loading, wear volumes with ceramic-on-ceramic bearings are small in comparison to other bearing materials. Low acetabular anteversion is associated with greater wear.


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 4 | Pages 509 - 514
1 Apr 2006
Aldridge JM Lightdale NR Mallon WJ Coonrad RW

There have been few reports in the literature of total elbow arthroplasty extending beyond 10 to 15 years. We reviewed 40 patients (41 elbows) with a mean age of 56 years (19 to 83) who had undergone a Coonrad/Coonrad-Morrey elbow arthroplasty by one surgeon for various diagnoses between 1974 and 1994. Surgical selection excluded patients with previous elbow infection or who refused to accept a sedentary level of elbow activity postoperatively. Objective data were collected from charts, radiographs, clinical photographs and supplemented by the referring orthopaedic surgeons’ records and radiographs if health or distance prevented a patient from returning for final review. Subjective outcome was defined by patient satisfaction.

Of the 41 elbows, 21 were functional between 10 and 14 years after operation, ten between 15 and 19 years and ten between 20 and 31 years. There were 14 complications and 13 revisions, but no cases of acute infection, or permanent removal of any implant.


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 5 | Pages 623 - 627
1 May 2005
Kim Y

Between January 1984 and January 1986, 131 porous-coated anatomic total hip replacements were performed in 119 consecutive patients. Of these, 100 patients (110 hips) who were alive at a minimum of 18 years after replacement were included in the study. The mean age of the patients at surgery was 48.4 years and that of surviving patients at the latest follow-up was 67.8 years. The mean duration of the clinical and radiological follow-up was 19.4 years (18 to 20).

The mean Harris hip score initially improved from 55 points before to 95 points at two years after operation, but subsequently decreased to 91 points after six years, 87 points after 11.2 years, and 85 points after 19.4 years. At the final follow-up, 23 hips (18%) of the entire cohort and 21% of survivors had undergone revision because of loosening or osteolysis of the acetabular component and eight hips (6%) of the entire cohort and 7% of survivors for loosening or osteolysis of the femoral component. Only four femoral components (4%) were revised for isolated aseptic loosening without osteolysis and two (2%) for recurrent dislocation. On the basis of these long-term results, the porous-coated anatomic femoral component survived for a minimum of 18 years after operation while the acetabular component was less durable. The findings identify the principles of uncemented joint replacement which can be applied to current practice.


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 6 | Pages 776 - 780
1 Jun 2005
Kim Y Kim S Kim S Park B Kim P Ihn J

We performed 114 consecutive primary total hip arthroplasties with a cementless expansion acetabular component in 101 patients for advanced osteonecrosis of the femoral head. The mean age of the patients at surgery was 51 years (36 to 62) and the mean length of follow-up was 110 months (84 to 129).

The mean pre-operative Harris hip score of 47 points improved to 93 points at final follow-up. The polyethylene liner was exchanged in two hips during this period and one broken acetabular component was revised. The mean linear wear rate of polyethylene was 0.07 mm/year and peri-acetabular osteolysis was seen in two hips (1.9%). Kaplan-Meier analysis indicated that the survival of the acetabular component without revision was 97.8% (95% confidence interval 0.956 to 1.000) at ten years.

Our study has shown that the results of THA with a cementless expansion acetabular component and an alumina-polyethylene bearing surface are good.


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 5 | Pages 628 - 631
1 May 2005
Dunstan E Sanghrajka AP Tilley S Unwin P Blunn G Cannon SR Briggs TWR

Metal-on-metal hip bearings are being implanted into younger patients. The consequence of elevated levels of potentially carcinogenic metal ions is therefore a cause for concern. We have determined the levels of cobalt (Co), chromium (Cr), titanium (Ti) and vanadium (Va) in the urine and whole blood of patients who had had metal-on-metal and metal-on-polyethylene articulations in situ for more than 30 years. We compared these with each other and with the levels for a control group of subjects.

We found significantly elevated levels of whole blood Ti, Va and urinary Cr in all arthroplasty groups. The whole blood and urine levels of Co were grossly elevated, by a factor of 50 and 300 times respectively in patients with loose metal-on-metal articulations when compared with the control group. Stable metal-on-metal articulations showed much lower levels. Elevated levels of whole blood or urinary Co may be useful in identifying metal-on-metal articulations which are loose.


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 4 | Pages 496 - 500
1 Apr 2005
Mileti J Sperling JW Cofield RH Harrington JR Hoskin TL

There are theoretical and practical advantages to modular rather than monoblock designs of prostheses for shoulder arthroplasty, but there are no reported studies which specifically compare the clinical and radiological results of their use. We have compared the results of unconstrained total shoulder arthroplasty for osteoarthritis using both types of implant. The monoblock design was used between 1992 and 1995 and the modular design after 1995. Both had cemented all-polyethylene glenoids, the monoblock with matched and the modular with mismatched radii of curvature. There were 34 consecutive shoulders in each group with a mean follow-up of 6.1 years in the first and 5.2 years in the second.

There were no significant differences in improvement of pain scores, active elevation, external rotation, internal rotation, patient satisfaction, or the Neer ratings between the two groups. Two of 28 glenoid components in the first group and six of 30 in the second met the criteria for being radiologically at risk for loosening (p = 0.25). There were no significant differences in clinical outcome or radiological changes between the first- and second-generation designs of implant for shoulder arthroplasty.