We biomechanically investigated whether the standard dynamic hip screw (DHS) or the DHS blade achieves better fixation in bone with regard to resistance to pushout, pullout and torsional stability. The experiments were undertaken in an artificial bone substrate in the form of polyurethane foam blocks with predefined mechanical properties. Pushout tests were also repeated in cadaveric femoral heads. The results showed that the DHS blade outperformed the DHS with regard to the two most important characteristics of implant fixation, namely resistance to pushout and rotational stability. We concluded that the DHS blade was the superior implant in this study.
Metallosis is a rare cause of failure after total knee replacement and has only previously been reported when there has been abnormal metal-on-metal contact. We describe 14 patients (15 knees) whose total knee replacement required revision for a new type of early failure caused by extensive metallosis. A modification of a cementless rotating platform implant, which had previously had excellent long-term survival, had been used in each case. The change was in the form of a new porous-beaded surface on the femoral component to induce cementless fixation, which had been used successfully in the fixation of acetabular and tibial components. This modification appeared to have resulted in metallosis due to abrasive two-body wear. The component has subsequently been recalled and is no longer in use. The presentation, investigation, and findings at revision are described and a possible aetiology and its implications are discussed.
Arthroscopy of the native hip is an established diagnostic and therapeutic procedure. Its application in the symptomatic replaced hip is still being explored. We describe the use of arthroscopy of the hip in 24 symptomatic patients following total hip replacement, resurfacing arthroplasty of the hip and partial resurfacing (study group), and compared it with arthroscopy of the native hip in 24 patients (control group). A diagnosis was made or confirmed at arthroscopy in 23 of the study group and a therapeutic arthroscopic intervention resulted in relief of symptoms in ten of these. In a further seven patients it led to revision hip replacement. In contrast, arthroscopy in the control group was diagnostic in all 24 patients and the resulting arthroscopic therapeutic intervention provided symptomatic relief in 21. The mean operative time in the study group (59.7 minutes (35 to 93)) was less than in the control group (71 minutes (40 to 100), p = 0.04) but the arthroscopic approach was more difficult in the arthroplasty group. We suggest that arthroscopy has a role in the management of patients with a symptomatic arthroplasty when other investigations have failed to provide a diagnosis.
We examined the effects of previous resection of the radial head and synovectomy on the outcome of subsequent total elbow arthroplasty in patients with rheumatoid arthritis. Fifteen elbows with a history of resection and synovectomy were compared with a control group of patients who had elbow arthroplasty with an implant of the same design. The mean age in both groups was 63 years. In the study group, resection of the radial head and synovectomy had been undertaken at a mean of 8.9 years before arthroplasty. The mean radiological follow-up for the 13 available patients in the study group was 5.89 years (0.3 to 11.0) and in the control group was 6.6 years (2.2 to 12.6). There were no revisions in either group. The mean Mayo elbow performance score improved from 29 to 96 in the study group, with similar improvement in the control group (28 to 87). The study group had excellent results in 13 elbows and good results in two. The control group had excellent results in seven and good results in six. Our experience indicates that previous resection of the radial head and synovectomy are not associated with an increased rate of revision following subsequent arthroplasty of the elbow. However, there was a higher rate of complication in the study group compared with the control group.
Injection or aspiration of the ankle may be performed through either an anteromedial or an anterolateral approach for diagnostic or therapeutic reasons. We evaluated the success of an intra-articular puncture in relation to its site in 76 ankles from 38 cadavers. Two orthopaedic surgical trainees each injected methylene blue dye into 18 of 38 ankles through an anterolateral approach and into 20 of 38 through an anteromedial. An arthrotomy was then performed to confirm the placement of the dye within the joint. Of the anteromedial injections 31 of 40 (77.5%, 95% confidence interval (CI) 64.6 to 90.4) were successful as were 31 of 36 (86.1%, 95% CI 74.8 to 97.4) anterolateral injections. In total 62 of 76 (81.6%, 95% CI 72.9 to 90.3) of the injections were intra-articular with a trend towards greater accuracy with the anterolateral approach, but this difference was not statistically significant (p = 0.25). In the case of trainee A, 16 of 20 anteromedial injections and 14 of 18 anterolateral punctures were intra-articular. Trainee B made successful intra-articular punctures in 15 of 20 anteromedial and 17 of 18 anterolateral approaches. There was no significant difference between them (p = 0.5 and p = 0.16 for the anteromedial and anterolateral approaches, respectively). These results were similar to those of other reported studies. Unintended peri-articular injection can cause complications and an unsuccessful aspiration can delay diagnosis. Placement of the needle may be aided by the use of ultrasonographic scanning or fluoroscopy which may be required in certain instances.
Between September 1993 and September 1996, we performed 34 Kudo 5 total elbow replacements in 31 rheumatoid patients. All 22 surviving patients were reviewed at a mean of 11.9 years (10 to 14). Their mean age was 56 years (37 to 78) at the time of operation. All had Larsen grade IV or V rheumatoid changes on X-ray. Nine (three bilateral replacements and six unilateral) had died from unrelated causes. One who had died before ten years underwent revision for dislocation. Of the 22 total elbow replacements reviewed six had required revision, four for aseptic loosening (one humeral and three ulnar) and two for infection. Post-operatively, one patient had neuropraxia of the ulnar nerve and one of the radial nerve. Two patients had valgus tilting of the ulnar component. With revision as the endpoint, the mean survival time for the prosthesis was 11.3 years (95% confidence interval (10 to 13) and the estimated survival of the prosthesis at 12 years according to Kaplan-Meier survival analysis was 74% (95% confidence interval 0.53 to 0.91). Of the 16 surviving implants, ten were free from pain, four had mild pain and two moderate. The mean arc of flexion/extension of the elbow was 106° (65° to 130°) with pronation/supination of 90° (30° to 150°) with the joint at 90° of flexion. The mean Mayo elbow performance score was 82 (60 to 100) with five excellent, ten good and one fair result. Good long-term results can be expected using the Kudo 5 total elbow replacement in patients with rheumatoid disease, with a low incidence of loosening of the components.
We present the case of an 83-year-old man who developed quadriparesis and respiratory embarrassment following osteomyelitis at the occipito-atlantoaxial junction. He had developed an abscess at this site after an earlier urinary infection with methicillin-resistant staphylococcus aureus. Stabilisation of the neck and antibiotic therapy led to an almost complete neurological recovery without recourse to anterior surgery.
We reviewed 123 second-generation uncemented total hip replacements performed on 115 patients by a single surgeon between 1993 and 1994. The acetabular component used in all cases was a fully porous-coated threaded hemispheric titanium shell (T-Tap ST) with a calcium ion stearate-free, isostatically compression-moulded polyethylene liner. The titanium femoral component used was a Taperloc with a reduced distal stem. No patient was lost to follow-up. Complete clinical and radiological follow-up was obtained for all 123 hips at a mean of 14 years (12 to 16). One femoral component was revised after a fracture, and three acetabular components for aseptic loosening. No additional femoral or acetabular components were judged loose by radiological criteria. Mild proximal femoral osteolysis was identified in two hips and minor acetabular osteolysis was present in four. The mean rate of penetration of the femoral head was 0.036 mm/year (0.000 to 0.227). These findings suggest that refinements in component design may be associated with excellent long-term fixation in cementless primary total hip replacement.
We report the survival, radiological and functional outcomes of a single surgeon series of his first 144 consecutive Birmingham hip resurfacing procedures (130 patients) at a minimum of ten years. There were ten revisions during this time. Although no patients were lost to follow-up some did not complete the scoring assessment or undergo radiological assessment at ten years. The ten-year survival for male patients was 98.0% (95% confidence interval 95.2 to 100). The ten-year survival for the total cohort with aseptic revision as the endpoint was 95.5% (95% confidence interval 91.8 to 99.0) and including revisions for sepsis was 93.5% (95% confidence interval 89.2 to 97.6). The median modified Oxford hip score at ten years was 4.2% (interquartile range 0 to 19) and the median University of California, Los Angeles score was 7.0 (interquartile range 5.0 to 8.0). This study confirms the midterm reports that metal-on-metal hip resurfacing using the Birmingham Hip provides a durable alternative to total hip replacement, particularly in younger male patients wishing to maintain a high level of function, with low risk of revision for at least ten years.
The aim of this study was to establish a classification system for the acromioclavicular joint using cadaveric dissection and radiological analyses of both reformatted computed tomographic scans and conventional radiographs centred on the joint. This classification should be useful for planning arthroscopic procedures or introducing a needle and in prospective studies of biomechanical stresses across the joint which may be associated with the development of joint pathology. We have demonstrated three main three-dimensional morphological groups namely flat, oblique and curved, on both cadaveric examination and radiological assessment. These groups were recognised in both the coronal and axial planes and were independent of age.
Dislocation is a common and well-studied complication after total hip replacement. However, subluxation, which we define as a clinically recognised episode of incomplete movement of the femoral head outside the acetabulum with spontaneous reduction, has not been studied previously. Out of a total of 2521 hip replacements performed over 12 years by one surgeon, 30 patients experienced subluxations which occurred in 31 arthroplasties. Data were collected prospectively with a minimum follow-up of two years. Subluxation occurred significantly more frequently after revision than after primary hip replacement, and resolved in 19 of 31 cases (61.3%). In six of the 31 hips (19.4%) the patient subsequently dislocated the affected hip, and in six hips (19.4%) intermittent subluxation continued. Four patients had a revision operation for instability, three for recurrent dislocation and one for recurrent subluxation. Clinical and radiological comparisons with a matched group of stable total hips showed no correlation with demographic or radiological parameters. Patients with subluxing hips reported significantly more concern that their hip would dislocate, more often changed their behaviour to prevent instability and had lower postoperative Harris hip scores than patients with stable replacements.
We compared the medium-term clinical and radiological results of hybrid total hip replacement (THR) with metal-on-metal Birmingham hip resurfacing (BHR) in two groups of 54 young patients matched for age, gender, body mass index and pre-operative levels of activity. The clinical outcome was assessed by the University of California, Los Angeles (UCLA) activity score, the Oxford Hip Score (OHS) and the EuroQol scores. Radiologically, all hips were assessed for migration and osteolysis, the hybrid THRs for polyethylene wear and the BHRs for a pedestal sign. The mean follow-up of the patients with a hybrid THR was ten years and for those with a BHR, nine years. Four patients with a hybrid THR and one with a BHR had died. In each group five were lost to follow-up. The revision rate of the hybrid THRs was 16.7% (9 of 54) and of the BHRs 9.3% (5 of 54) (p = 0.195). Radiographs of a further eight hybrid THRs demonstrated wear and osteolysis, and they await revision (p = 0.008). Of the unrevised BHRs 90% had radiological changes, of which approximately 50% had progressed over the previous four years. All hybrid THRs demonstrated linear polyethylene wear with a mean of 1.24 mm (0.06 to 3.03). The BHRs recorded superior OHS (p = 0.013), UCLA (p = 0.008), and EuroQol visual analogue scores (p = 0.009). After nine years, patients with BHRs remained more active and had a lower rate of revision than those with hybrid THRs. Both groups demonstrated progressive radiological changes at medium-term follow-up.
Lately, concerns have arisen following the use of large metal-on-metal bearings in hip replacements owing to reports of catastrophic soft-tissue reactions resulting in implant failure and associated complications. This review examines the literature and contemporary presentations on current clinical dilemmas in metal-on-metal hip replacement.
We report serum metal ion level data in patients with unilateral and bilateral hip resurfacing over a ten-year period. In these patients there is an increase in both cobalt and chromium levels above the accepted reference ranges during the first 18 months after operation. Metal ion levels remain elevated, but decline slowly for up to five years. However, the levels then appear to start rising again in some patients up to the ten-year mark. There was no significant difference in cobalt or chromium levels between men and women. These findings appear to differ from much of the current literature. The clinical significance of a raised metal ion level remains under investigation.
We implanted titanium and carbon fibre-reinforced plastic (CFRP) femoral prostheses of the same dimensions into five prosthetic femora. An abductor jig was attached and a 1 kN load applied. This was repeated with five control femora. Digital image correlation was used to give a detailed two-dimensional strain map of the medial cortex of the proximal femur. Both implants caused stress shielding around the calcar. Distally, the titanium implant showed stress shielding, whereas the CFRP prosthesis did not produce a strain pattern which was statistically different from the controls. There was a reduction in strain beyond the tip of both the implants. This investigation indicates that use of the CFRP stem should avoid stress shielding in total hip replacement.
The amount of anteroposterior laxity required for a good range of movement and knee function in a cruciate-retaining total knee replacement (TKR) continues to be debated. We undertook a retrospective study to evaluate the effects of anteroposterior laxity on the range of movement and knee function in 55 patients following the e-motion cruciate-retaining TKR with a minimum follow-up of two years. The knees were divided into stable (anteroposterior translation, ≤ 10 mm, 38 patients) and unstable (anteroposterior translation, >
10 mm, 17) groups based on the anteroposterior laxity, measured using stress radiographs. We compared the Hospital for Special Surgery (HSS) scores, the Western Ontario MacMasters University Osteoarthritis (WOMAC) index, weight-bearing flexion, non-weight-bearing flexion and the reduction of flexion under weight-bearing There were no differences between the stable and unstable groups with regard to the mean HHS and WOMAC total scores, as well as weight-bearing and non-weight-bearing flexion (p = 0.277, p = 0.082, p = 0.095 and p = 0.646, respectively). However, the stable group had a better WOMAC function score and less delta flexion than the unstable group (p = 0.011 and p = 0.005, respectively). Our results suggest that stable knees with laxity ≤ 10 mm have a good functional outcome and less reduction of flexion under weight-bearing conditions than unstable knees with laxity >
10 mm following an e-motion cruciate-retaining TKR.
The purposes of this study were to define the range of laxity of the interosseous ligaments in cadaveric wrists and to determine whether this correlated with age, the morphology of the lunate, the scapholunate (SL) gap or the SL angle. We evaluated 83 fresh-frozen cadaveric wrists and recorded the SL gap and SL angle. Standard arthroscopy of the wrist was then performed and the grades of laxity of the scapholunate interosseous ligament (SLIL) and the lunotriquetral interosseous ligament (LTIL) and the morphology of the lunate were recorded. Arthroscopic evaluation of the SLIL revealed four (5%) grade I specimens, 28 (34%) grade II, 40 (48%) grade III and 11 (13%) grade IV. Evaluation of the LTIL showed 17 (20%) grade I specimens, 40 (48%) grade II, 28 (30%) grade III and one (1%) grade IV. On both bivariate and multivariate analysis, the grade of both the SLIL and LTIL increased with age, but decreased with female gender. The grades of SLIL or LTIL did not correlate with the morphology of the lunate, the SL gap or the SL angle. The physiological range of laxity at the SL and lunotriquetral joints is wider than originally described. The intercarpal ligaments demonstrate an age-related progression of laxity of the SL and lunotriquetral joints. There is no correlation between the grades of laxity of the SLIL or LTIL and the morphology of the lunate, the SL gap or the SL grade. Based on our results, we believe that the Geissler classification has a role in describing intercarpal laxity, but if used alone it cannot adequately diagnose pathological instability. We suggest a modified classification with a mechanism that may distinguish physiological laxity from pathological instability.
We sought to establish the incidence of joint failure secondary to adverse reaction to metal debris (ARMD) following metal-on-metal hip resurfacing in a large, three surgeon, multicentre study involving 4226 hips with a follow-up of 10 to 142 months. Three implants were studied: the Articular Surface Replacement; the Birmingham Hip Resurfacing; and the Conserve Plus. Retrieved implants underwent analysis using a co-ordinate measuring machine to determine volumetric wear. There were 58 failures associated with ARMD. The median chromium and cobalt concentrations in the failed group were significantly higher than in the control group (p <
0.001). Survival analysis showed a failure rate in the patients with Articular Surface Replacement of 9.8% at five years, compared with <
1% at five years for the Conserve Plus and 1.5% at ten years for the Birmingham Hip Resurfacing. Two ARMD patients had relatively low wear of the retrieved components. Increased wear from the metal-on-metal bearing surface was associated with an increased rate of failure secondary to ARMD. However, the extent of tissue destruction at revision surgery did not appear to be dose-related to the volumetric wear.
In a study on ten fresh human cadavers we examined the change in the height of the intervertebral disc space, the angle of lordosis and the geometry of the facet joints after insertion of intervertebral total disc replacements. SB III Charité prostheses were inserted at L3-4, L4-5, and L5-S1. The changes studied were measured using computer navigation sofware applied to CT scans before and after instrumentation. After disc replacement the mean lumbar disc height was doubled (p <
0.001). The mean angle of lordosis and the facet joint space increased by a statistically significant extent (p <
0.005 and p = 0.006, respectively). By contrast, the mean facet joint overlap was significantly reduced (p <
0.001). Our study indicates that the increase in the intervertebral disc height after disc replacement changes the geometry at the facet joints. This may have clinical relevance.
This paper outlines the recent development of an exchange Travelling Fellowship scheme between the British and American Orthopaedic Research Societies.