Advertisement for orthosearch.org.uk
Results 381 - 400 of 464
Results per page:
The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 1 | Pages 28 - 37
1 Jan 2010
Jameson SS Langton DJ Nargol AVF

We present the early clinical and radiological results of Articular Surface Replacement (ASR) resurfacings in 214 hips (192 patients) with a mean follow-up of 43 months (30 to 57). The mean age of the patients was 56 years (28 to 74) and 85 hips (40%) were in 78 women.

The mean Harris hip score improved from 52 (11 to 81) to 95 (27 to 100) at two years and the mean University of California, Los Angeles activity score from 3.9 (1 to 10) to 7.4 (2 to 10) in the same period. Narrowing of the neck (to a maximum of 9%) was noted in 124 of 209 hips (60%). There were 12 revisions (5.6%) involving four (1.9%) early fractures of the femoral neck and two (0.9%) episodes of collapse of the femoral head secondary to avascular necrosis. Six patients (2.8%) had failure related to metal wear debris. The overall survival for our series was 93% (95% confidence interval 80 to 98) and 89% (95% confidence interval 82 to 96) for hips with acetabular components smaller than 56 mm in diameter.

The ASR implant has a lower diametrical clearance and a subhemispherical acetabular component when compared with other more frequently implanted metal-on-metal hip resurfacings. These changes may contribute to the higher failure rate than in other series, compared with other designs. Given our poor results with the small components we are no longer implanting the smaller size.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 7 | Pages 893 - 900
1 Jul 2007
Baker PN van der Meulen JH Lewsey J Gregg PJ

A postal questionnaire was sent to 10 000 patients more than one year after their total knee replacement (TKR). They were assessed using the Oxford knee score and were asked whether they were satisfied, unsure or unsatisfied with their TKR. The response rate was 87.4% (8231 of 9417 eligible questionnaires) and a total of 81.8% (6625 of 8095) of patients were satisfied. Multivariable regression modelling showed that patients with higher scores relating to the pain and function elements of the Oxford knee score had a lower level of satisfaction (p < 0.001), and that ongoing pain was a stronger predictor of this. Female gender and a primary diagnosis of osteoarthritis were found to be predictors of lower levels of patient satisfaction. Differences in the rate of satisfaction were also observed in relation to age, the American Society of Anesthesiologists grade and the type of prosthesis.

This study has provided data on the Oxford knee score and the expected levels of satisfaction at one year after TKR. The results should act as a benchmark of practice in the United Kingdom and provide a baseline for peer comparison between institutions.


The Bone & Joint Journal
Vol. 95-B, Issue 2 | Pages 145 - 145
1 Feb 2013
Villar RN


The Bone & Joint Journal
Vol. 95-B, Issue 12 | Pages 1587 - 1594
1 Dec 2013
Ibrahim MS Twaij H Giebaly DE Nizam I Haddad FS

The outcome after total hip replacement has improved with the development of surgical techniques, better pain management and the introduction of enhanced recovery pathways. These pathways require a multidisciplinary team to manage pre-operative education, multimodal pain control and accelerated rehabilitation. The current economic climate and restricted budgets favour brief hospitalisation while minimising costs. This has put considerable pressure on hospitals to combine excellent results, early functional recovery and shorter admissions.

In this review we present an evidence-based summary of some common interventions and methods, including pre-operative patient education, pre-emptive analgesia, local infiltration analgesia, pre-operative nutrition, the use of pulsed electromagnetic fields, peri-operative rehabilitation, wound dressings, different surgical techniques, minimally invasive surgery and fast-track joint replacement units.

Cite this article: Bone Joint J 2013;95-B:1587–94.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 1 | Pages 91 - 95
1 Jan 2011
Jensen CD Steval A Partington PF Reed MR Muller SD

Rivaroxaban has been recommended for routine use as a thromboprophylactic agent in patients undergoing lower-limb arthroplasty. However, trials supporting its use have not fully evaluated the risks of wound complications. This study of 1048 total hip/knee replacements records the rates of return to theatre and infection before and after the change from a low molecular weight heparin (tinzaparin) to rivaroxaban as the agent of chemical thromboprophylaxis in patients undergoing lower-limb arthroplasty. During a period of 13 months, 489 consecutive patients undergoing lower-limb arthroplasty received tinzaparin and the next 559 consecutive patients received rivaroxaban as thromboprophylaxis.

Nine patients in the control (tinzaparin) group (1.8%, 95% confidence interval 0.9 to 3.5) returned to theatre with wound complications within 30 days, compared with 22 patients in the rivaroxaban group (3.94%, 95% confidence interval 2.6 to 5.9). This increase was statistically significant (p = 0.046). The proportion of patients who returned to theatre and became infected remained similar (p = 0.10).

Our study demonstrates the need for further randomised controlled clinical trials to be conducted to assess the safety and efficacy of rivaroxaban in clinical practice, focusing on the surgical complications as well as the potential prevention of venous thromboembolism.


The Bone & Joint Journal
Vol. 95-B, Issue 8 | Pages 1040 - 1044
1 Aug 2013
Kazi HA Perera JR Gillott E Carroll FA Briggs TWR

We prospectively assessed the efficacy of a ceramic-on-metal (CoM) hip bearing with uncemented acetabular and femoral components in which cobalt­–chrome acetabular liners and alumina ceramic heads were used.

The cohort comprised 94 total hip replacements (THRs) in 83 patients (38 women and 45 men) with a mean age of 58 years (42 to 70). Minimum follow-up was two years. All patients had pre- and post-operative assessment using the Western Ontario and McMaster Universities osteoarthritis index (WOMAC), Oxford hip score and Short-Form 12 scores. All showed a statistically significant improvement from three months post-operatively onwards (all p < 0.001).

After two years whole blood metal ion levels were measured and chromosomal analysis was performed. The levels of all metal ions were elevated except vanadium. Levels of chromium, cobalt, molybdenum and titanium were significantly higher in patients who underwent bilateral THR compared with those undergoing unilateral THR (p < 0.001). Chromosomal analysis demonstrated both structural and aneuploidy mutations. There were significantly more breaks and losses than in the normal population (p < 0.001). There was no significant difference in chromosomal aberration between those undergoing unilateral and bilateral procedures (all analyses p ≥ 0.62).

The use of a CoM THR is effective clinically in the short-term, with no concerns, but the significance of high metal ion levels and chromosomal aberrations in the long-term remains unclear.

Cite this article: Bone Joint J 2013;95-B:1040–44.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 3 | Pages 307 - 314
1 Mar 2011
Matthies A Underwood R Cann P Ilo K Nawaz Z Skinner J Hart AJ

This study compared component wear rates and pre-revision blood metal ions levels in two groups of failed metal-on-metal hip arthroplasties: hip resurfacing and modular total hip replacement (THR).

There was no significant difference in the median rate of linear wear between the groups for both acetabular (p = 0.4633) and femoral (p = 0.0872) components. There was also no significant difference in the median linear wear rates when failed hip resurfacing and modular THR hips of the same type (ASR and Birmingham hip resurfacing (BHR)) were compared.

Unlike other studies of well-functioning hips, there was no significant difference in pre-revision blood metal ion levels between hip resurfacing and modular THR.

Edge loading was common in both groups, but more common in the resurfacing group (67%) than in the modular group (57%). However, this was not significant (p = 0.3479). We attribute this difference to retention of the neck in resurfacing of the hip, leading to impingement-type edge loading. This was supported by visual evidence of impingement on the femur.

These findings show that failed metal-on-metal hip resurfacing and modular THRs have similar component wear rates and are both associated with raised pre-revision blood levels of metal ions.


The Bone & Joint Journal
Vol. 95-B, Issue 5 | Pages 714 - 717
1 May 2013
Yates P Kellett C Huntley JS Whitwell D Reed MR Beadel G Snyckers C

In May 2012, in airports across the globe, seven orthopaedic surgeons bravely said goodbye to their loved ones, and slowly turned towards their respective aircraft. Filled with expectation and mild trepidation they stepped into the unknown… the ABC fellowship of 2012.


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 9 | Pages 1253 - 1258
1 Sep 2010
Scott CEH Howie CR MacDonald D Biant LC

Up to 20% of patients are not satisfied with the outcome following total knee replacement (TKR). This study investigated the pre- and post-operative predictors of dissatisfaction in a large cohort of patients undergoing TKR. We assessed 1217 consecutive patients between 2006 and 2008 both before operation and six months after, using the Short-form (SF)-12 health questionnaire and the Oxford Knee Score. Detailed information concerning comorbidity was also gathered. Satisfaction was measured at one year when 18.6% (226 of 1217) of patients were unsure or dissatisfied with their replacement and 81.4% (911 of 1217) were satisfied or very satisfied. Multivariate regression analysis was performed to identify independent predictors of dissatisfaction. Significant (p < 0.001) predictors at one year included the pre-operative SF-12 mental component score, depression and pain in other joints, the six-month SF-12 score and poorer improvement in the pain element of the Oxford Knee Score.

Patient expectations were highly correlated with satisfaction. Satisfaction following TKR is multifactorial. Managing the expectations and mental health of the patients may reduce dissatisfaction. However, the most significant predictor of dissatisfaction is a painful total knee replacement.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 11 | Pages 1455 - 1456
1 Nov 2012
Oussedik S Gould K Stockley I Haddad FS

Peri-prosthetic infection remains a leading cause of revision surgery. Recent publications from the American Musculoskeletal Infection Society have sought to establish a definition of peri-prosthetic infection based on clinical findings and laboratory investigations. The limitations of their approach are discussed and an alternative definition is proposed, which it is felt may better reflect the uncertainties encountered in clinical practice.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 1 | Pages 1 - 4
1 Jan 2007
Robertsson O

This article considers the establishment, purpose and conduct of knee arthroplasty registers using the Swedish register as an example. The methods of collection of appropriate data, the cost, and the ways in which this information may be used are considered.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 11_Supple_A | Pages 52 - 54
1 Nov 2012
Rodriguez JA Rathod PA

Large femoral heads have been used with increasing frequency over the last decade. The prime reason is likely the effect of large heads on stability. The larger head neck ratio, combined with the increased jump distance of larger heads result in a greater arc of impingement free motion, and greater resistance to dislocation in a provocative position. Multiple studies have demonstrated clear clinical efficacy in diminishing dislocation rates with the use of large femoral heads. With crosslinked polyethylene, wear has been shown to be equivalent between larger and smaller heads. However, the stability advantages of increasing diameter beyond 38 mm have not been clearly demonstrated. More importantly, recent data implicates large heads in the increasing prevalence of groin pain and psoas impingement. There are clear benefits with larger femoral head diameters, but the advantages of diameters beyond 38 mm have not yet been demonstrated clinically.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 3 | Pages 298 - 306
1 Mar 2011
van der Weegen W Hoekstra HJ Sijbesma T Bos E Schemitsch EH Poolman RW

We systematically reviewed the peer-reviewed literature to relate the survival of hybrid metal-on-metal hip resurfacing arthroplasty devices to a National Institute of Clinical Excellence (NICE) benchmark for choosing a primary total hip replacement, which is a survival rate of 90% at a follow-up of ten years.

A total of 29 articles (10 621 resurfaced hips) met the inclusion criteria. The mean follow-up ranged from 0.6 to 10.5 years and the survival of the implant ranged from 84% to 100%. Of the 10 621 hips, 370 were revised (3.5%), with aseptic loosening as the most frequent mode of failure.

None of the hip resurfacing arthroplasty implants used to date met the full ten-year NICE benchmark of survival. A total of 13 studies showed satisfactory survival compared with the three-year NICE benchmark.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 5 | Pages 567 - 573
1 May 2007
Keegan GM Learmonth ID Case CP

The long-term effects of metal-on-metal arthroplasty are currently under scrutiny because of the potential biological effects of metal wear debris. This review summarises data describing the release, dissemination, uptake, biological activity, and potential toxicity of metal wear debris released from alloys currently used in modern orthopaedics. The introduction of risk assessment for the evaluation of metal alloys and their use in arthroplasty patients is discussed and this should include potential harmful effects on immunity, reproduction, the kidney, developmental toxicity, the nervous system and carcinogenesis.


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 1 | Pages 111 - 115
1 Jan 2006
Jain N Willett KM

In order to assess the efficacy of inspection and accreditation by the Specialist Advisory Committee for higher surgical training in orthopaedic surgery and trauma, seven training regions with 109 hospitals and 433 Specialist Registrars were studied over a period of two years.

There were initial deficiencies in a mean of 14.8% of required standards (10.3% to 19.2%). This improved following completion of the inspection, with a mean residual deficiency in 8.9% (6.5% to 12.7%.) Overall, 84% of standards were checked, 68% of the units improved and training was withdrawn in 4%.

Most units (97%) were deficient on initial assessment. Moderately good rectification was achieved but the process of follow-up and collection of data require improvement. There is an imbalance between the setting of standards and their implementation. Any major revision of the process of accreditation by the new Post-graduate Medical Education and Training Board should recognise the importance of assessment of training by direct inspection on site, of the relationship between service and training, and the advantage of defining mandatory and developmental standards.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 4 | Pages 540 - 544
1 Apr 2009
Robinson PM O’Meara MJ

The Thomas splint is a common piece of equipment in emergency departments and orthopaedic units in hospitals worldwide. Its basic design has changed little since its first description by Hugh Owen Thomas was published in 1875. We have reviewed the origins of the Thomas splint as a means of correction of deformities of the lower limb and its evolution through both World Wars into an essential item for the management of trauma of the lower limb.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 11_Supple_A | Pages 78 - 81
1 Nov 2012
Benjamin-Laing H Haddad FS

In this paper, we will consider the current role of metal-on-metal bearings by looking at three subtypes of MoM hip arthroplasty separately: Hip resurfacing, large head (> 36 mm) MoM THA and MoM THA with traditional femoral head sizes.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 11_Supple_A | Pages 14 - 18
1 Nov 2012
Lombardi, Jr AV Barrack RL Berend KR Cuckler JM Jacobs JJ Mont MA Schmalzried TP

Since 1996 more than one million metal-on-metal articulations have been implanted worldwide. Adverse reactions to metal debris are escalating. Here we present an algorithmic approach to patient management. The general approach to all arthroplasty patients returning for follow-up begins with a detailed history, querying for pain, discomfort or compromise of function. Symptomatic patients should be evaluated for intra-articular and extra-articular causes of pain. In large head MoM arthroplasty, aseptic loosening may be the source of pain and is frequently difficult to diagnose. Sepsis should be ruled out as a source of pain. Plain radiographs are evaluated to rule out loosening and osteolysis, and assess component position. Laboratory evaluation commences with erythrocyte sedimentation rate and C-reactive protein, which may be elevated. Serum metal ions should be assessed by an approved facility. Aspiration, with manual cell count and culture/sensitivity should be performed, with cloudy to creamy fluid with predominance of monocytes often indicative of failure. Imaging should include ultrasound or metal artifact reduction sequence MRI, specifically evaluating for fluid collections and/or masses about the hip. If adverse reaction to metal debris is suspected then revision to metal or ceramic-on-polyethylene is indicated and can be successful. Delay may be associated with extensive soft-tissue damage and hence poor clinical outcome.


Bone & Joint 360
Vol. 1, Issue 1 | Pages 28 - 28
1 Feb 2012
Aprato A


The Bone & Joint Journal
Vol. 95-B, Issue 1 | Pages 122 - 126
1 Jan 2013
McWilliams AB Douglas SL Redmond AC Grainger AJ O’Connor PJ Stewart TD Stone MH

The results of hip and knee replacement surgery are generally regarded as positive for patients. Nonetheless, they are both major operations and have recognised complications. We present a review of relevant claims made to the National Health Service Litigation Authority. Between 1995 and 2010 there were 1004 claims to a value of £41.5 million following hip replacement surgery and 523 claims to a value of £21 million for knee replacement. The most common complaint after hip surgery was related to residual neurological deficit, whereas after knee replacement it was related to infection. Vascular complications resulted in the highest costs per case in each group.

Although there has been a large increase in the number of operations performed, there has not been a corresponding relative increase in litigation. The reasons for litigation have remained largely unchanged over time after hip replacement. In the case of knee replacement, although there has been a reduction in claims for infection, there has been an increase in claims for technical errors. There has also been a rise in claims for non-specified dissatisfaction. This information is of value to surgeons and can be used to minimise the potential mismatch between patient expectation, informed consent and outcome.

Cite this article: Bone Joint J 2013;95-B:122–6.