We present the ten-year data of a cohort of patients, aged between
18 and 65 years (mean age 52.7 years; 19 to 64), who underwent total
hip arthroplasty. Patients were randomised to be treated with a
cobalt-chrome (CoCr) femoral head with an ultra-high molecular weight
polyethylene (UHMWPE), highly cross-linked polyethylene (XLPE) or
ceramic-on-ceramic (CoC) bearing surface. A total of 102 hips (91 patients) were randomised into the three
groups. At ten years, 97 hips were available for radiological and
functional follow-up. Two hips (two patients) had been revised (one
with deep infection and one for periprosthetic fracture) and three
were lost to follow-up. Radiological analysis was performed using
a validated digital assessment programme to give linear, directional
and volumetric wear of the two polyethylene groups.Aims
Patients and Methods
Taper junctions between modular hip arthroplasty femoral heads and stems fail by wear or corrosion which can be caused by relative motion at their interface. Increasing the assembly force can reduce relative motion and corrosion but may also damage surrounding tissues. The purpose of this study was to determine the effects of increasing the impaction energy and the stiffness of the impactor tool on the stability of the taper junction and on the forces transmitted through the patient’s surrounding tissues. A commercially available impaction tool was modified to assemble components in the laboratory using impactor tips with varying stiffness at different applied energy levels. Springs were mounted below the modular components to represent the patient. The pull-off force of the head from the stem was measured to assess stability, and the displacement of the springs was measured to assess the force transmitted to the patient’s tissues.Objectives
Methods
We present a case series of ten metal-on-polyethylene total hip
arthroplasties (MoP THAs) with delayed dislocation associated with
unrecognised adverse local tissue reaction due to corrosion at the
trunnion and pseudotumour formation. The diagnosis was not suspected in nine of the ten patients (six
female/four male; mean age 66 years), despite treatment in a specialist
unit (mean time from index surgery to revision was 58 months, 36
to 84). It was identified at revision surgery and subsequently confirmed
by histological examination of resected tissue. Pre-operative assessment
and culture results ruled out infection. A variety of treatment
strategies were used, including resection of the pseudotumour and
efforts to avoid recurrent dislocation. Aims
Methods
The widely used and well-proven Palacos R (a.k.a. Refobacin Palacos
R) bone cement is no longer commercially available and was superseded
by Refobacin bone cement R and Palacos R + G in 2005. However, the
performance of these newly introduced bone cements have not been
tested in a phased evidence-based manner, including roentgen stereophotogrammetric
analysis (RSA). In this blinded, randomised, clinical RSA study, the migration
of the Stanmore femoral component was compared between Refobacin
bone cement R and Palacos R + G in 62 consecutive total hip arthroplasties.
The primary outcome measure was femoral component migration measured
using RSA and secondary outcomes were Harris hip score (HHS), Hip
disability and Osteoarthritis Outcome Score (HOOS), EuroQol 5D (EQ-5D)
and Short Form 36 (SF-36).Aims
Patients and Methods
The long-term results of grafting with hydroxyapatite granules for acetabular deficiency in revision total hip replacement are not well known. We have evaluated the results of revision using a modular cup with hydroxyapatite grafting for Paprosky type 2 and 3 acetabular defects at a minimum of ten years’ follow-up. We retrospectively reviewed 49 acetabular revisions at a mean of 135 months (120 to 178). There was one type 2B, ten 2C, 28 3A and ten 3B hips. With loosening as the endpoint, the survival rate was 74.2% (95% confidence interval 58.3 to 90.1). Radiologically, four of the type 3A hips (14%) and six of the type 3B hips (60%) showed aseptic loosening with collapse of the hydroxyapatite layer, whereas no loosening occurred in type 2 hips. There was consolidation of the hydroxyapatite layer in 33 hips (66%). Loosening was detected in nine of 29 hips (31%) without cement and in one of 20 hips (5%) with cement (p = 0.03, Fisher’s exact probability test). The linear wear and annual wear rate did not correlate with loosening. These results suggest that the long-term results of hydroxyapatite grafting with cement for type 2 and 3A hips are encouraging.
We compared the incidence of pseudotumours after
large head metal-on-metal (MoM) total hip arthroplasty (THA) with
that after conventional metal-on-polyethylene (MoP) THA and assessed
the predisposing factors to pseudotumour formation. From a previous randomised controlled trial which compared large
head (38 mm to 60 mm) cementless MoM THA with conventional head
(28 mm) cementless MoP THA, 93 patients (96 THAs: 41 MoM (21 males,
20 females, mean age of 64 years, standard deviation ( The incidence of pseudotumours, measured using a standardised
CT protocol was 22 (53.7%) after MoM THA and 12 (21.8%) after MoP
THA. Women with a MoM THA were more likely to develop a pseudotumour
than those with a MoP THA (15 Contrary to popular belief, pseudotumours occur frequently around
MoP THAs. Women with a MoM THA and an elevated cobalt level are
at greatest risk. In this study, pseudotumours had no effect on
the functional outcome after either large head MoM or conventional
MoP THA. Cite this article:
We reviewed the results of 84 total hip replacements performed with a short metaphyseal-fitting anatomical cementless femoral component in 84 unselected consecutive patients with a mean age of 78.9 years (70 to 88). The mean follow-up was 4.6 years (4 to 5). The mean pre-operative Harris hip score was 26 points (0 to 56), which improved to 89 (61 to 100) at the final follow-up. No patient had thigh pain. The mean pre-operative Western Ontario and McMaster Universities osteoarthritis index score was 61 points (48 to 75), which improved to 21 (6 to 46). The mean University of California, Los Angeles activity score was 5.5 points (3 to 7) at the final follow-up. Osseointegration was seen in all femoral and acetabular components. All hips had grade 1 stress shielding of the proximal femur. No acetabular or femoral osteolysis was identified. These results demonstrate that a short metaphyseal-fitting femoral component achieves optimal fixation without diaphyseal anchorage in elderly patients.
A modular femoral head–neck junction has practical
advantages in total hip replacement. Taper fretting and corrosion
have so far been an infrequent cause of revision. The role of design
and manufacturing variables continues to be debated. Over the past
decade several changes in technology and clinical practice might
result in an increase in clinically significant taper fretting and
corrosion. Those factors include an increased usage of large diameter
(36 mm) heads, reduced femoral neck and taper dimensions, greater
variability in taper assembly with smaller incision surgery, and
higher taper stresses due to increased patient weight and/or physical
activity. Additional studies are needed to determine the role of
taper assembly compared with design, manufacturing and other implant
variables. Cite this article:
We studied prospectively the long-term results of the Charnley Elite-Plus femoral stem in 184 consecutive young patients (194 hips). There were 130 men and 54 women with a mean age of 49.1 years (21 to 60). The predominant diagnosis was osteonecrosis of the femoral head (63.6%, 117 patients). Clinical and radiological evaluation was undertaken at each follow-up. The mean follow-up was 11.2 years (10 to 12). The mean pre-operative Harris hip score was 43.4 (12 to 49) which improved to 91 (59 to 100) at the final follow-up. The survival of the femoral stem at 12 years was 99% with revision as the end-point. The mean annual linear wear of the polyethylene liner was 0.17 mm (0.13 to 0.22). The prevalence of acetabular osteolysis was 10.8% (21 hips) and osteolysis of the calcar femorale 12.9% (25 hips). A third-generation cementing technique, accurate alignment of the stem and the use of a 22 mm zirconia head were important factors in the prevention of aseptic loosening of the Elite Plus femoral stem in these high-risk young patients.
The aim of this prospective randomised study
was to compare the clinical and radiological results of a cemented
all-polyethylene Ultima acetabular component with those of a cementless
porous-coated acetabular component (PFC) following total hip replacement
(THR). A total of 287 patients received either a polyethylene acetabular
component (group A) or a cobalt–chromium porous-coated component
(group B) with an identical cemented femoral component and 28 mm
cobalt-chromium head, thus making it the largest study of its type.
Patients were evaluated radiologically and clinically using the
Harris hip score (HHS). Group A comprised 183 patients (73 male,
110 female) with a mean age of
71.3 years (55 to 89). Group B comprised 104 patients (48 male,
56 female) with a mean age of 69.8 years (56 to 89). A total of
16 patients (13 in Group A, three in Group B) did not have post-operative
data for analysis. The mean follow-up in group A was 7.52 years
(0.4 to 15.0) and in Group B 7.87 years (0.5 to 14.0). At final follow-up the mean HHS was similar between groups A
and B (74.5 (25 to 100) and 78.0 (37 to 100), respectively; p =
0.068). The total number of revisions for any cause was 28, 17 of
which were in group A and 11 in group B. The ten-year survivorship
was 86.8% (95% confidence interval (CI) 78.4 to 92.1) and 89.2%
(95% CI 78.3 to 94.8) for groups A and B, respectively (log-rank
p-value = 0.938). A total of 20 cemented and two cementless acetabular
components had evidence of acetabular radiolucencies or acetabular
component migration at last follow-up (p = 0.001). These results indicate that patients with a cemented all-polyethylene
and cementless porous-coated polyethylene lined acetabular component
have similar long-term clinical outcomes.
Metallosis after shoulder replacement has not previously been described in the literature. We report a patient who developed extensive metallosis after implantation of an uncemented Nottingham shoulder replacement. He underwent a revision procedure. Examination of the retrieved prosthesis showed that the titanium porous coating was separating from the humeral stem and becoming embedded in the ultra-high-molecular-weight polyethylene glenoid component, resulting in abrasive wear of the humeral component. There was metallosis despite exchange of the modular humeral head. Both components had to be exchanged to resolve the problem.
In Africa the amount of joint replacement surgery is increasing, but the indications for operation and the age of the patients are considerably different from those in the developed world. New centres with variable standards of care and training of the surgeons are performing these procedures and it is important that a proper audit of this work is undertaken. In Malawi, we have pioneered a Registry which includes all joint replacements that have been carried out in the country. The data gathered include the age, gender, indication for operation, the prosthesis used, the surgical approach, the use of bone graft, the type of cement, pressurising systems and the thromboprophylaxis used. All patients have their clinical scores recorded pre-operatively and then after three and six months and at one year. Before operation all patients are counselled and on consent their HIV status is established allowing analysis of the effect of HIV on successful joint replacement. To date, 73 total hip replacements (THRs) have been carried out in 58 patients by four surgeons in four different hospitals. The most common indications for THR were avascular necrosis (35 hips) and osteoarthritis (22 hips). The information concerning 20 total knee replacements has also been added to the Registry.
We report the case of a 12-year-old boy with polyostotic fibrous dysplasia of the upper femur in whom a massive customised polyethylene prosthesis functioned successfully for more than fifty years.
In 2004 we described the ten-year prospective results of 38 total hip replacements using the Furlong hydroxyapatite-ceramic-coated femoral component in 35 patients <
50 years old. We have now reviewed the surviving 35 arthroplasties in 33 patients at a mean of 16 years (10.3 to 19.9). The mean age of the surviving patients at the time of operation was 41.3 years (26.0 to 49.0). Of these, eight have undergone revision of their acetabular component for aseptic loosening. None of the femoral components has had revision for aseptic loosening giving a survival rate of 100% at 16 years (95% confidence interval 89% to 100%). The Furlong hydroxyapatite-ceramic-coated femoral component gives excellent long-term survival in young and active patients.
The original forged Müller straight stem (CoNiCr) has shown excellent ten- to 15-year results. We undertook a long-term survival analysis with special emphasis on radiological changes within a 20-year period of follow-up. In all, 165 primary total hip replacements, undertaken between July 1984 and June 1987 were followed prospectively. Clinical follow-up included a standardised clinical examination, and radiological assessment was based on a standardised anteroposterior radiograph of the pelvis, which was studied for the presence of osteolysis, debonding and cortical atrophy. Survival of the stem with revision for any reason was 81% (95% confidence interval (CI), 76 to 86) at 20 years and for aseptic loosening 87% (95% CI, 82 to 90). At the 20-year follow-up, 15 of the surviving 36 stems showed no radiological changes. Debonding (p = 0.005), osteolysis (p = 0.003) and linear polyethylene wear (p = 0.016) were associated with aseptic loosening, whereas cortical atrophy was not associated with failure (p = 0.008). The 20-year results of the Müller straight stem are comparable to those of other successful cemented systems with similar follow-up. Radiological changes are frequently observed, but with a low incidence of progression, and rarely result in revision. Cortical atrophy appears to be an effect of ageing and not a sign of loosening of the femoral component.
Bone surface strains were measured in cadaver femora during loading prior to and after resurfacing of the hip and total hip replacement using an uncemented, tapered femoral component. In vitro loading simulated the single-leg stance phase during walking. Strains were measured on the medial and the lateral sides of the proximal aspect and the mid-diaphysis of the femur. Bone surface strains following femoral resurfacing were similar to those in the native femur, except for proximal shear strains, which were significantly less than those in the native femur. Proximomedial strains following total hip replacement were significantly less than those in the native and the resurfaced femur. These results are consistent with previous clinical evidence of bone loss after total hip replacement, and provide support for claims of bone preservation after resurfacing arthroplasty of the hip.
Between June 1991 and January 1995, 42 hydroxyapatite-coated CAD-CAM femoral components were inserted in 25 patients with inflammatory polyarthropathy, 21 of whom had juvenile idiopathic arthritis. Their mean age was 21 years (11 to 35). All the patients were reviewed clinically and radiologically at one, three and five years. At the final review at a mean of 11.2 years (8 to 13) 37 hips in 23 patients were available for assessment. A total of four femoral components (9.5%) had failed, of which two were radiologically loose and two were revised. The four failed components were in patients aged 16 years or less at the time of surgery. Hydroxyapatite-coated customised femoral components give excellent medium- to long-term results in skeletally-mature young adults with inflammatory polyarthropathy. Patients aged less than 16 years at the time of surgery have a risk of 28.5% of failure of the femoral component at approximately ten years.
We describe the results at five years of a prospective study of a new tri-tapered polished, cannulated, cemented femoral stem implanted in 51 patients (54 hips) with osteoarthritis. The mean age and body mass index of the patients was 74 years and 27.9, respectively. Using the anterolateral approach, half of the stems were implanted by a consultant orthopaedic surgeon and half by six different registrars. There were three withdrawals from the study because of psychiatric illness, a deep infection and a recurrent dislocation. Five deaths occurred prior to five-year follow-up and one patient withdrew from clinical review. In the remaining 51 hips the mean pre-operative Oxford hip score was 47 points which decreased to 19 points at five years (45 hips). Of the stems 49 (98%) were implanted within 1° of neutral in the femoral canal. The mean migration of the stem at five years was 1.9 mm and the survivorship for aseptic loosening was 100%. There was no significant difference in outcome between the consultant and registrar groups. At five years, the results were comparable with those of other polished, tapered, cemented stems. Long-term surveillance continues.
We carried out a clinical and radiological review of 103 cementless primary hip arthroplasties with a tapered rectangular grit-blasted titanium press-fit femoral component and a threaded conical titanium acetabular component at a mean follow-up of 14.4 years (10.2 to 17.1). The mean Harris hip score at the last follow-up was 89.2 (32 to 100). No early loosening and no fracture of the implant were found. One patient needed revision surgery because of a late deep infection. In 11 hips (10.7%), the reason for revision was progressive wear of the polyethylene liner. Exchange of the acetabular component because of aseptic loosening without detectable liner wear was carried out in three hips (2.9%). After 15 years the survivorship with aseptic loosening as the definition for failure was 95.6% for the acetabular component and 100% for the femoral component.