Fifty-three failed knee replacements were revised using minimally constrained implants with smooth uncemented intramedullary stems and metal-backed tibial components. Polymethylmethacrylate was used only to replace lost bone near the surface of the implant. Excluding four knees which had serious postoperative complications, 91% had successful relief of pain, 84% had over 90 degrees of movement and 80% could walk for more than 30 minutes. Review of the radiographs showed that there were no progressive lucencies at the interface between bone and cement, and no subsidence of components or changes in alignment. At the uncemented stem-to-bone interface, thin white lines developed near the metal, and their significance is discussed. This revision technique is an effective treatment for aseptic failure of primary total knee arthroplasty.
Periprosthetic femoral fracture (PFF) is a potentially
devastating complication after total hip arthroplasty, with historically
high rates of complication and failure because of the technical
challenges of surgery, as well as the prevalence of advanced age
and comorbidity in the patients at risk. This study describes the short-term outcome after revision arthroplasty
using a modular, titanium, tapered, conical stem for PFF in a series
of 38 fractures in 37 patients. The mean age of the cohort was 77 years (47 to 96). A total of
27 patients had an American Society of Anesthesiologists grade of
at least 3. At a mean follow-up of 35 months (4 to 66) the mean
Oxford Hip Score (OHS) was 35 (15 to 48) and comorbidity was significantly
associated with a poorer OHS. All fractures united and no stem needed
to be revised. Three hips in three patients required further surgery
for infection, recurrent PFF and recurrent dislocation and three
other patients required closed manipulation for a single dislocation.
One stem subsided more than 5 mm but then stabilised and required
no further intervention. In this series, a modular, tapered, conical stem provided a versatile
reconstruction solution with a low rate of complications. Cite this article:
Migration of the acetabular component may give rise to oval-shaped bone defects in the acetabulum. The oblong implant is designed to fill these defects and achieve a stable cementless anchorage with no significant bone loss. We prospectively reviewed 133 oblong long oblique revision components at a mean follow-up of 9.74 years (0.6 to 14). All had been used in revisions for defects of type IIB to IIIB according to Paprosky. Aseptic loosening was the reason for revision in 11 cases (8.3%) and deep infection in seven (5.3%). The probability of implant survival over a 12-year follow-up estimated by the Kaplan-Meier method gave a survival rate of 0.85% respectively 0.90% when deep infection was excluded as the endpoint. Our study supports the use of these components in defects from IIB to IIIA. The main precondition for success is direct contact of more than half of the surface of the implant with the host acetabular bone.
Implantation of allograft bone is an integral part of revision surgery of the hip. One major concern with its use is the risk of transmission of infective agents. There are a number of methods of processing allograft bone in order to reduce this risk. One method requires washing the tissue using pulsed irrigation immediately before implantation. We report the incidence of deep bacterial infection in 138 patients (144 revision hip arthroplasties) who had undergone implantation of allograft bone. The bone used was fresh-frozen, non-irradiated and pulse-washed with normal saline before implantation. The deep infection rate at a minimum follow-up of one year was 0.7%. This method of processing appears to be associated with a very low risk of allograft-related bacterial infection.
We identified five (2.3%) fractures of the stem in a series of 219 revision procedures using a cementless, cylindrical, extensively porous-coated, distally-fixed femoral stem. Factors relating to the patients, the implant and the operations were compared with those with intact stems. Finite-element analysis was performed on two of the fractured implants. Factors associated with fracture of the stem were poor proximal bone support (type III–type IV; p = 0.001), a body mass index >
30; (p = 0.014), a smaller diameter of stem (<
13.5 mm; p = 0.007) and the use of an extended trochanteric osteotomy (ETO 4/5: p = 0.028). Finite-element analysis showed that the highest stresses on the stem occurred adjacent to the site of the fracture. The use of a strut graft wired over an extended trochanteric osteotomy in patients lacking proximal femoral cortical support decreased the stresses on the stem by 48%. We recommend the use of a strut allograft in conjunction with an extended trochanteric osteotomy in patients with poor proximal femoral bone stock.
We report the outcome at ten to 15 years of two-stage revision for hip infection in 99 patients using the Prostalac articulated hip spacer system. All the patients were contacted to determine their current functional and infection status using the Oxford-12, Short form-12, and Western Ontario and McMaster University Osteoarthritis Index questionnaires. A total of 11 of the 99 patients had a further infection, of whom seven responded to repeat surgery with no further sequelae. The mean interval between the stages was five months (1 to 36). We were able to review 48 living patients, with a mean age of 72 years (46 to 86), 34 (71%) of whom provided health-related quality-of-life outcome scores. The mean follow-up was 12 years (10 to 15). The long-term success rate was 89% and with additional surgery this rose to 96%. The mean global Western Ontario and McMaster University Osteoarthritis Index score was 80.6 ( Two-stage revision for hip infection using a Prostalac interim spacer offers a predictable and lasting solution for patients with this difficult problem.
It may not be possible to undertake revision total hip arthroplasty
(THA) in the presence of massive loss of acetabular bone stock using
standard cementless hemispherical acetabular components and metal
augments, as satisfactory stability cannot always be achieved. We
aimed to study the outcome using a reconstruction cage and a porous
metal augment in these patients. A total of 22 acetabular revisions in 19 patients were performed
using a combination of a reconstruction cage and porous metal augments.
The augments were used in place of structural allografts. The mean
age of the patients at the time of surgery was 70 years (27 to 85)
and the mean follow-up was 39 months (27 to 58). The mean number
of previous THAs was 1.9 (1 to 3). All patients had segmental defects
involving more than 50% of the acetabulum and seven hips had an
associated pelvic discontinuity. Aims
Patients and Methods
We reviewed histologically the incidence and pathogenesis of the deposition of calcium pyrophosphate dihydrate (CPPD) crystals in the pseudocapsule, femoral and acetabular membranes and periprosthetic tissue at revision of 789 cases of failed total hip replacement. In 13, periprosthetic tissues were found to have deposits of CPPD crystals in areas of cartilaginous metaplasia; four also showed evidence of localised deposition of amyloid. None of the patients had a history of chondrocalcinosis in the hip or other joints. Cartilaginous metaplasia and other changes in periprosthetic tissues may predispose to the deposition of CPPD and associated localised amyloid.
We studied the histology of the cement-bone interface and the regenerated synovium in ten aseptically loosened arthroplasties in rheumatoid patients. In two patients we found rheumatoid nodules at the cement-bone interface and marked lymphoplasmacytic infiltration in another three. Failed joints in osteoarthritic patients did not show these changes, and it seems likely that the presence of abundant immunocompetent cells was due to the background disease. We speculate that rheumatoid inflammation contributed to the process of loosening.
Revision hip arthroplasty in patients with massive acetabular bone deficiency has generally given poor long-term results. We report the use of an 'anti-protrusio cage', secured to the ischium and ilium, which bridges areas of acetabular bone loss, provides support for the acetabular socket, and allows pelvic bone grafting in an environment protected from excessive stress. Forty-two failed hip arthroplasties with massive acetabular bone loss were revised with the Burch-Schneider anti-protrusio cage and evaluated after two to 11 years (mean five years). There was failure due to sepsis in five hips (12%) and aseptic loosening in five (12%); the remaining 32 hips (76%) showed no evidence of acetabular component failure or loosening.
Aims. To compare rates of serious adverse events in patients undergoing revision knee arthroplasty with consideration of the indication for revision (urgent versus elective indications), and compare these with primary arthroplasty and re-revision arthroplasty. Methods. Patients undergoing primary knee arthroplasty were identified in the national Hospital Episode Statistics (HES) between 1 April 1997 to 31 March 2017. Subsequent revision and re-revision arthroplasty procedures in the same patients and same knee were identified. The primary outcome was 90-day mortality and a logistic regression model was used to investigate factors associated with 90-day mortality and secondary adverse outcomes, including infection (undergoing surgery), pulmonary embolism, myocardial infarction, and stroke. Urgent indications for
Aims. Treatment outcomes for methicillin-resistant Staphylococcus aureus (MRSA) periprosthetic joint infection (PJI) using systemic vancomycin and antibacterial cement spacers during two-stage
Aims. This study aimed to investigate the estimated change in primary and
Aims. The aim of this study was to estimate the 90-day periprosthetic joint infection (PJI) rates following total knee arthroplasty (TKA) and total hip arthroplasty (THA) for osteoarthritis (OA). Methods. This was a data linkage study using the New South Wales (NSW) Admitted Patient Data Collection (APDC) and the Australian Orthopaedic Association National Joint Replacement Registry (AOANJRR), which collect data from all public and private hospitals in NSW, Australia. Patients who underwent a TKA or THA for OA between 1 January 2002 and 31 December 2017 were included. The main outcome measures were 90-day incidence rates of hospital readmission for:
Aims. The aim of this study was to determine whether fixation, as opposed to
Aims. The aim of this study was to evaluate the optimal deep tissue specimen sample number for histopathological analysis in the diagnosis of periprosthetic joint infection (PJI). Methods. In this retrospective diagnostic study, patients undergoing revision surgery after total hip or knee arthroplasty (n = 119) between January 2015 and July 2018 were included. Multiple specimens of the periprosthetic membrane and pseudocapsule were obtained for histopathological analysis at
Aims. The purpose of this study was to determine the association between prior sleeve gastrectomy in patients undergoing primary total hip and knee arthroplasty, and 90-day complications, incidence of
Aims. Removal of infected components and culture-directed antibiotics are important for the successful treatment of chronic periprosthetic joint infection (PJI). However, as many as 27% of chronic PJI patients yield negative culture results. Although culture negativity has been thought of as a contraindication to one-stage revision, data supporting this assertion are limited. The aim of our study was to report on the clinical outcomes for one-stage and two-stage exchange arthroplasty performed in patients with chronic culture-negative PJI. Methods. A total of 105 consecutive patients who underwent