The Journal of Bone & Joint Surgery British Volume
Vol. 75-B, Issue 4 | Pages 624 - 629
1 Jul 1993
Jacobsson S Djerf K Gillquist J Hammerby S Ivarsson I

We compared two cementless femoral components, the 'isoelastic' Butel stem and the more rigid PCA design, in a randomised, prospective study of 56 patients with a mean follow-up of 4 years (2 to 5). Patients were matched in 28 pairs, and one of each pair was treated with each femoral component. The isoelastic stem gave fewer signs of stress shielding radiologically, but both the Harris hip score and the visual analogue pain scale showed significantly inferior results at the two-year and three-year follow-up. The overall failure rates for the femoral components were 43% for the Butel and 11% for the PCA. These results contrast with those of earlier experimental and clinical studies, in which isoelastic properties appeared to be advantageous. This review emphasises the importance of controlled studies before an altered or new design of prosthesis is released on to the market for general use


The Bone & Joint Journal
Vol. 102-B, Issue 6 Supple A | Pages 107 - 115
1 Jun 2020
Tetreault MW Perry KI Pagnano MW Hanssen AD Abdel MP

Aims

Metaphyseal fixation during revision total knee arthroplasty (TKA) is important, but potentially difficult when using historical designs of cone. Material and manufacturing innovations have improved the size and shape of the cones which are available, and simplified the required bone preparation. In a large series, we assessed the implant survivorship, radiological results, and clinical outcomes of new porous 3D-printed titanium metaphyseal cones featuring a reamer-based system.

Methods

We reviewed 142 revision TKAs in 139 patients using 202 cones (134 tibial, 68 femoral) which were undertaken between 2015 and 2016. A total of 60 involved tibial and femoral cones. Most cones (149 of 202; 74%) were used for Type 2B or 3 bone loss. The mean age of the patients was 66 years (44 to 88), and 76 (55 %) were female. The mean body mass index (BMI) was 34 kg/m2 (18 to 60). The patients had a mean of 2.4 (1 to 8) previous operations on the knee, and 68 (48%) had a history of prosthetic infection. The mean follow-up was 2.4 years (2 to 3.6).


Bone & Joint 360
Vol. 8, Issue 5 | Pages 33 - 35
1 Oct 2019


The Bone & Joint Journal
Vol. 101-B, Issue 7_Supple_C | Pages 48 - 54
1 Jul 2019
Kahlenberg CA Lyman S Joseph AD Chiu Y Padgett DE

Aims

The outcomes of total knee arthroplasty (TKA) depend on many factors. The impact of implant design on patient-reported outcomes is unknown. Our goal was to evaluate the patient-reported outcomes and satisfaction after primary TKA in patients with osteoarthritis undergoing primary TKA using five different brands of posterior-stabilized implant.

Patients and Methods

Using our institutional registry, we identified 4135 patients who underwent TKA using one of the five most common brands of implant. These included Biomet Vanguard (Zimmer Biomet, Warsaw, Indiana) in 211 patients, DePuy/Johnson & Johnson Sigma (DePuy Synthes, Raynham, Massachusetts) in 222, Exactech Optetrak Logic (Exactech, Gainesville, Florida) in 1508, Smith & Nephew Genesis II (Smith & Nephew, London, United Kingdom) in 1415, and Zimmer NexGen (Zimmer Biomet) in 779 patients. Patients were evaluated preoperatively using the Knee Injury and Osteoarthritis Outcome Score (KOOS), Lower Extremity Activity Scale (LEAS), and 12-Item Short-Form Health Survey questionnaire (SF-12). Demographics including age, body mass index, Charlson Comorbidity Index, American Society of Anethesiologists status, sex, and smoking status were collected. Postoperatively, two-year KOOS, LEAS, SF-12, and satisfaction scores were compared between groups.


The Bone & Joint Journal
Vol. 101-B, Issue 2 | Pages 178 - 188
1 Feb 2019
Chaudhary MM Lakhani PH

Aims

Double-level lengthening, bone transport, and bifocal compression-distraction are commonly undertaken using Ilizarov or other fixators. We performed double-level fixator-assisted nailing, mainly for the correction of deformity and lengthening in the same segment, using a straight intramedullary nail to reduce the time in a fixator.

Patients and Methods

A total of 23 patients underwent this surgery, involving 27 segments (23 femora and four tibiae), over a period of ten years. The most common indication was polio in ten segments and rickets in eight; 20 nails were inserted retrograde and seven antegrade. A total of 15 lengthenings were performed in 11 femora and four tibiae, and 12 double-level corrections of deformity without lengthening were performed in the femur. The mean follow-up was 4.9 years (1.1 to 11.4). Four patients with polio had tibial lengthening with arthrodesis of the ankle. We compared the length of time in a fixator and the external fixation index (EFI) with a control group of 27 patients (27 segments) who had double-level procedures with external fixation. The groups were matched for the gain in length, age, and level of difficulty score.


The Journal of Bone & Joint Surgery British Volume
Vol. 78-B, Issue 2 | Pages 226 - 228
1 Mar 1996
Feller JA Bartlett RJ Lang DM

We studied 40 patients in whom the patella was not severely deformed and who were undergoing primary total knee arthroplasty (TKA) for osteoarthritis by one surgeon using one type of prosthesis. They were randomly allocated either to have the patella retained or resurfaced with a cemented, all-polyethylene component regardless of the state of the patellar articular cartilage. Apart from removal of osteophytes, no surgery was undertaken on the retained patellae. All 38 surviving patients were evaluated at three years using the HSS knee score and a new, specifically designed Patellar score (maximum score of 30). No TKA was revised, but two patients in the resurfacing group had a further unrelated procedure. The mean HSS and Patellar scores at follow-up were 89 and 28 in the patellar retention group and 83 and 26 in the patellar resurfacing group. Statistically significant lower scores for both were recorded in women and in heavier patients. Stair-climbing ability was significantly better in the retention group. Although there were no complications related to patellar resurfacing, in the medium term we did not find any significant benefit from resurfacing the patella during TKA for osteoarthritis if it was not severely deformed


The Journal of Bone & Joint Surgery British Volume
Vol. 77-B, Issue 4 | Pages 528 - 533
1 Jul 1995
Gillespie W Pekarsky B O'Connell D

Cost is a factor in the choice of prosthetic components in joint replacement. For a given performance, the least expensive components are the most cost-effective. When evaluating a new prosthesis with an unknown outcome, the use of an economic model allows estimation of potential cost-effectiveness. We used published data for the survival of cemented total hip replacements from Sweden, and cost and demographic information from New South Wales, Australia, in such a model. In young active total hip recipients a new prosthetic design which offered a 90% improvement in survivorship over 15 years and a 15% reduction in the cost of revision surgery, could be sold at a price of 2 to 2.5 times that of conventional cemented components such as the Charnley Low Friction Arthroplasty and still be cost-effective. Using more likely estimates of the improved performance of new technology, however, the upper limit of cost-effectiveness is an increase of 1.5 to 1. Only a very small increase in the cost of a prosthesis could ever be justified for older patients of either sex. Most of the potential benefits of a better level of survivorship appear towards the end of the 15-year period. The results of modelling may be incorporated in clinical trial design. Given the known performance of some well-established and relatively inexpensive designs of prostheses, very large randomised studies would be required to prove an improvement in performance


Bone & Joint 360
Vol. 7, Issue 4 | Pages 25 - 28
1 Aug 2018


The Journal of Bone & Joint Surgery British Volume
Vol. 77-B, Issue 5 | Pages 705 - 714
1 Sep 1995
Walker P Mai S Cobb A Bentley G Hua J

We report the theoretical basis of a method to measure axial migration of femoral components of total hip replacements (THR). The use of the top of the greater trochanter and a lateral point on the collar of the stem, allowing for variations of up to 10 degrees rotation of the femur in any direction between successive radiographs, gave a maximum error of 0.37 mm. At a more realistic 5 degrees rotational variation, the error was only 0.13 mm. These data were confirmed in an experimental study using digitisation of points and special software. We also showed that the centre of the femoral head, the stem tip, and the lesser trochanter provided less accurate landmarks. In a second study we digitised a series of radiographs of 51 Charnley and 57 Stanmore THRs; the mean migration rates were found to be identical. We then studied 46 successful stems with a minimum follow-up of eight years and 46 stems which had failed by aseptic loosening at different times. At two years, the successful stems had migrated by a mean of 1.45 +/- 0.68 mm, but the failed cases had a mean migration of 4.32 +/- 2.58 mm (p < 0.0001). Of the successful cases 76% had migrated less than 2 mm, while in the failed group 84% had migrated more than 2 mm. For any particular case migration of more than 2.6 mm at two years had only a 5% chance of continuing success and would therefore merit special follow-up. Only 24% of the eventually successful stems showed migration at the stem-cement interface, but this had happened in every failed stem. We conclude that it would be possible to evaluate a new cemented design of femoral stem over a two-year period by the use of our method and to compare its performance against the reported known standard of the Charnley and Stanmore designs


The Bone & Joint Journal
Vol. 100-B, Issue 1 | Pages 11 - 19
1 Jan 2018
Darrith B Courtney PM Della Valle CJ

Aims

Instability remains a challenging problem in both primary and revision total hip arthroplasty (THA). Dual mobility components confer increased stability, but there are concerns about the unique complications associated with these designs, as well as the long-term survivorship.

Materials and Methods

We performed a systematic review of all English language articles dealing with dual mobility THAs published between 2007 and 2016 in the MEDLINE and Embase electronic databases. A total of 54 articles met inclusion criteria for the final analysis of primary and revision dual mobility THAs and dual mobility THAs used in the treatment of fractures of the femoral neck. We analysed the survivorship and rates of aseptic loosening and of intraprosthetic and extra-articular dislocation.


Bone & Joint Research
Vol. 6, Issue 5 | Pages 270 - 276
1 May 2017
Gosiewski JD Holsgrove TP Gill HS

Objectives

Fractures of the proximal femur are a common clinical problem, and a number of orthopaedic devices are available for the treatment of such fractures. The objective of this study was to assess the rotational stability, a common failure predictor, of three different rotational control design philosophies: a screw, a helical blade and a deployable crucifix.

Methods

Devices were compared in terms of the mechanical work (W) required to rotate the implant by 6° in a bone substitute material. The substitute material used was Sawbones polyurethane foam of three different densities (0.08 g/cm3, 0.16 g/cm3 and 0.24 g/cm3). Each torsion test comprised a steady ramp of 1°/minute up to an angular displacement of 10°.


The Bone & Joint Journal
Vol. 99-B, Issue 6 | Pages 749 - 758
1 Jun 2017
García-Rey E Cruz-Pardos A García-Cimbrelo E

Aims

To determine the effect of a change in design of a cementless ceramic acetabular component in fixation and clinical outcome after total hip arthroplasty

Patients and Methods

We compared 342 hips (302 patients) operated between 1999 and 2005 with a relatively smooth hydroxyapatite coated acetabular component (group 1), and 337 hips (310 patients) operated between 2006 and 2011 using a similar acetabular component with a macrotexture on the entire outer surface of the component (group 2). The mean age of the patients was 53.5 (14 to 70) in group 1 and 53.0 (15 to 70) in group 2. The mean follow-up was 12.7 years (10 to 17) for group 1 and 7.2 years (4 to 10) for group 2.


Bone & Joint Research
Vol. 5, Issue 6 | Pages 215 - 217
1 Jun 2016
Pijls BG Nelissen RGHH


Bone & Joint Research
Vol. 6, Issue 5 | Pages 345 - 350
1 May 2017
Di Laura A Hothi H Henckel J Swiatkowska I Liow MHL Kwon Y Skinner JA Hart AJ

Objectives

The use of ceramic femoral heads in total hip arthroplasty (THA) has increased due to their proven low bearing wear characteristics. Ceramic femoral heads are also thought to reduce wear and corrosion at the head-stem junction with titanium (Ti) stems when compared with metal heads. We sought to evaluate taper damage of ceramic compared with metal heads when paired with cobalt chromium (CoCr) alloy stems in a single stem design.

Methods

This retrieval study involved 48 total hip arthroplasties (THAs) with CoCr V40 trunnions paired with either CoCr (n = 21) or ceramic (n = 27) heads. The taper junction of all hips was evaluated for fretting/corrosion damage and volumetric material loss using a roundness-measuring machine. We used linear regression analysis to investigate taper damage differences after adjusting for potential confounding variables.


The Bone & Joint Journal
Vol. 98-B, Issue 10_Supple_B | Pages 28 - 33
1 Oct 2016
Lum ZC Lombardi AV Hurst JM Morris MJ Adams JB Berend KR

Aims

Since redesign of the Oxford phase III mobile-bearing unicompartmental knee arthroplasty (UKA) femoral component to a twin-peg design, there has not been a direct comparison to total knee arthroplasty (TKA). Thus, we explored differences between the two cohorts.

Patients and Methods

A total of 168 patients (201 knees) underwent medial UKA with the Oxford Partial Knee Twin-Peg. These patients were compared with a randomly selected group of 177 patients (189 knees) with primary Vanguard TKA. Patient demographics, Knee Society (KS) scores and range of movement (ROM) were compared between the two cohorts. Additionally, revision, re-operation and manipulation under anaesthesia rates were analysed.


Bone & Joint 360
Vol. 5, Issue 1 | Pages 14 - 16
1 Feb 2016


The Bone & Joint Journal
Vol. 97-B, Issue 10_Supple_A | Pages 9 - 15
1 Oct 2015
Parratte S Ollivier M Lunebourg A Abdel MP Argenson J

Partial knee arthroplasty (PKA), either medial or lateral unicompartmental knee artroplasty (UKA) or patellofemoral arthroplasty (PFA) are a good option in suitable patients and have the advantages of reduced operative trauma, preservation of both cruciate ligaments and bone stock, and restoration of normal kinematics within the knee joint. However, questions remain concerning long-term survival. The goal of this review article was to present the long-term results of medial and lateral UKA, PFA and combined compartmental arthroplasty for multicompartmental disease. Medium- and long-term studies suggest reasonable outcomes at ten years with survival greater than 95% in UKA performed for medial osteoarthritis or osteonecrosis, and similarly for lateral UKA, particularly when fixed-bearing implants are used. Disappointing long-term outcomes have been observed with the first generation of patellofemoral implants, as well as early Bi-Uni (ie, combined medial and lateral UKA) or Bicompartmental (combined UKA and PFA) implants due to design and fixation issues. Promising short- and med-term results with the newer generations of PFAs and bicompartmental arthroplasties will require long-term confirmation.

Cite this article: Bone Joint J 2015;97-B(10 Suppl A):9–15.


The Bone & Joint Journal
Vol. 97-B, Issue 1 | Pages 64 - 70
1 Jan 2015
Hamilton DF Burnett R Patton JT Howie CR Moran M Simpson AHRW Gaston P

Total knee arthroplasty (TKA) is an established and successful procedure. However, the design of prostheses continues to be modified in an attempt to optimise the functional outcome of the patient.

The aim of this study was to determine if patient outcome after TKA was influenced by the design of the prosthesis used.

A total of 212 patients (mean age 69; 43 to 92; 131 female (62%), 81 male (32%)) were enrolled in a single centre double-blind trial and randomised to receive either a Kinemax (group 1) or a Triathlon (group 2) TKA.

Patients were assessed pre-operatively, at six weeks, six months, one year and three years after surgery. The outcome assessments used were the Oxford Knee Score; range of movement; pain numerical rating scales; lower limb power output; timed functional assessment battery and a satisfaction survey. Data were assessed incorporating change over all assessment time points, using repeated measures analysis of variance longitudinal mixed models. Implant group 2 showed a significantly greater range of movement (p = 0.009), greater lower limb power output (p = 0.026) and reduced report of ‘worst daily pain’ (p = 0.003) over the three years of follow-up. Differences in Oxford Knee Score (p = 0.09), report of ‘average daily pain’ (p = 0.57) and timed functional performance tasks (p = 0.23) did not reach statistical significance. Satisfaction with outcome was significantly better in group 2 (p = 0.001).

These results suggest that patient outcome after TKA can be influenced by the prosthesis used.

Cite this article: Bone Joint J 2015;97-B:64–70.


Bone & Joint 360
Vol. 2, Issue 3 | Pages 6 - 14
1 Jun 2013
Wallace WA

In the UK we have many surgeon inventors – surgeons who innovate and create new ways of doing things, who invent operations, who design new instruments to facilitate surgery or design new implants for using in patients. However truly successful surgeon inventors are a rare breed and they need to develop additional knowledge and skills during their career in order to push forward their devices and innovations. This article reviews my own experiences as a surgeon inventor and the highs and lows over the whole of my surgical career


The Bone & Joint Journal
Vol. 96-B, Issue 1 | Pages 114 - 121
1 Jan 2014
Pekmezci M McDonald E Buckley J Kandemir U

We investigated a new intramedullary locking nail that allows the distal interlocking screws to be locked to the nail. We compared fixation using this new implant with fixation using either a conventional nail or a locking plate in a laboratory simulation of an osteoporotic fracture of the distal femur. A total of 15 human cadaver femora were used to simulate an AO 33-A3 fracture pattern. Paired specimens compared fixation using either a locking or non-locking retrograde nail, and using either a locking retrograde nail or a locking plate. The constructs underwent cyclical loading to simulate single-leg stance up to 125 000 cycles. Axial and torsional stiffness and displacement, cycles to failure and modes of failure were recorded for each specimen. When compared with locking plate constructs, locking nail constructs had significantly longer mean fatigue life (75 800 cycles (sd 33 900) vs 12 800 cycles (sd 6100); p = 0.007) and mean axial stiffness (220 N/mm (sd 80) vs 70 N/mm (sd 18); p = 0.005), but lower mean torsional stiffness (2.5 Nm/° (sd 0.9) vs 5.1 Nm/° (sd 1.5); p = 0.008). In addition, in the nail group the mode of failure was either cut-out of the distal screws or breakage of nails, and in the locking plate group breakage of the plate was always the mode of failure. Locking nail constructs had significantly longer mean fatigue life than non-locking nail constructs (78 900 cycles (sd 25 600) vs 52 400 cycles (sd 22 500); p = 0.04).

The new locking retrograde femoral nail showed better stiffness and fatigue life than locking plates, and superior fatigue life to non-locking nails, which may be advantageous in elderly patients.

Cite this article: Bone Joint J 2014;96-B:114–21.