Conventional cemented acetabular components are
reported to have a high rate of failure when implanted into previously
irradiated bone. We recommend the use of a cemented reconstruction
with the addition of an acetabular reinforcement cross to improve
fixation. We reviewed a cohort of 45 patients (49 hips) who had undergone
irradiation of the pelvis and a cemented total hip arthroplasty
(THA) with an acetabular reinforcement cross. All hips had received
a minimum dose of 30 Gray (Gy) to treat a primary nearby tumour
or metastasis. The median dose of radiation was 50 Gy (Q1 to Q3:
45 to 60; mean: 49.57, 32 to 72). The mean follow-up after THA was 51 months (17 to 137). The cumulative
probability of revision of the acetabular component for a mechanical
reason was 0% (0 to 0%) at 24 months, 2.9% (0.2 to 13.3%) at 60
months and 2.9% (0.2% to 13.3%) at 120 months, respectively. One
hip was revised for mechanical failure and three for infection. Cemented acetabular components with a reinforcement cross provide
good medium-term fixation after pelvic irradiation. These patients
are at a higher risk of developing infection of their THA. Cite this article:
In light of the growing number of elderly osteopenic
patients with distal humeral fractures, we discuss the history of
their management and current trends. Under most circumstances operative
fixation and early mobilisation is the treatment of choice, as it
gives the best results. The relative indications for and results
of total elbow replacement
The purpose of this study was to describe the
radiological characteristics of a previously unreported finding: posterior
iliac offset at the sacroiliac joint and to assess its association
with pelvic instability as measured by initial displacement and
early implant loosening or failure. Radiographs from 42 consecutive
patients with a mean age of 42 years (18 to 77; 38 men, four women)
and mean follow-up of 38 months (3 to 96) with Anteroposterior Compression
II injuries, were retrospectively reviewed. Standardised measurements
were recorded for the extent of any diastasis of the pubic symphysis,
widening of the sacroiliac joint, static vertical ramus offset and
a novel measurement (posterior offset of the ilium at the sacroiliac
joint identified on axial CT scan). Pelvic fractures with posterior
iliac offset exhibited greater levels of initial displacement of
the anterior pelvis (anterior sacroiliac widening, pubic symphysis
diastasis and static vertical ramus offset, p <
0.001,0.034 and
0.028, respectively). Pelvic fractures with posterior ilium offset
also demonstrated higher rates of implant loosening regardless of
fixation method (p = 0.05). Posterior offset of the ilium was found
to be a reliable and reproducible measurement with substantial inter-observer
agreement (kappa = 0.70). Posterior offset of the ilium on axial
CT scan is associated with greater levels of initial pelvic displacement
and early implant loosening. Cite this article:
In this paper, we will consider the current role
of metal-on-metal bearings by looking at three subtypes of MoM hip
arthroplasty separately: Hip resurfacing, large head (>
36 mm) MoM
THA and MoM THA with traditional femoral head sizes.
The October 2014 Wrist &
Hand Roundup360 looks at: pulsed electromagnetic field of no use in acute scaphoid fractures; proximal interphalangeal joint replacement: one at a time or both at once; trapeziometacarpal arthrodesis in the young patient; Tamoxifen and Dupytren’s disease; and endoscopic or open for de Quervain’s syndrome?
Peri-acetabular tumour resections and their subsequent
reconstruction are among the most challenging procedures in orthopaedic
oncology. Despite the fact that a number of different pelvic endoprostheses
have been introduced, rates of complication remain high and long-term
results are mostly lacking. In this retrospective study, we aimed to evaluate the outcome
of reconstructing a peri-acetabular defect with a pedestal cup endoprosthesis
after a type 2 or type 2/3 internal hemipelvectomy. A total of 19 patients (11M:8F) with a mean age of 48 years (14
to 72) were included, most of whom had been treated for a primary
bone tumour (n = 16) between 2003 and 2009. After a mean follow-up
of 39 months (28 days to 8.7 years) seven patients had died. After
a mean follow-up of 7.9 years (4.3 to 10.5), 12 patients were alive,
of whom 11 were disease-free. Complications occurred in 15 patients.
Three had recurrent dislocations and three experienced aseptic loosening.
There were no mechanical failures. Infection occurred in nine patients,
six of whom required removal of the prosthesis. Two patients underwent
hindquarter amputation for local recurrence. The implant survival rate at five years was 50% for all reasons,
and 61% for non-oncological reasons. The mean Musculoskeletal Tumor
Society score at final follow-up was 49% (13 to 87). Based on these poor results, we advise caution if using the pedestal
cup for reconstruction of a peri-acetabular tumour resection. Cite this article:
Obesity is a global epidemic of 2.1 billion people and a well known cause of osteoarthritis. Joint replacement in the obese attracts more complications, poorer outcomes and higher revision rates. It is a reversible condition and the fundamental principles of dealing with reversible medical conditions prior to elective total joint replacement should apply to obesity. The dilemma for orthopaedic surgeons is when to offer surgery in the face of a reversible condition, which if treated may obviate joint replacement and reduce the risk and severity of obesity related disease in both the medical arena and the field of orthopaedics.
We compared the clinical, radiological and quality-of-life
outcomes between hybrid and total pedicle screw instrumentation
in patients undergoing surgery for neuromuscular scoliosis. Total pedicle screw instrumentation provided shorter operating
times, less blood loss and better correction of the major curve
compared with hybrid constructs in patients undergoing surgery for
neuromuscular scoliosis.
We compared the clinical and radiographic results
of total ankle replacement (TAR) performed in non-diabetic and diabetic
patients. We identified 173 patients who underwent unilateral TAR
between 2004 and 2011 with a minimum of two years’ follow-up. There
were 88 male (50.9%) and 85 female (49.1%) patients with a mean
age of 66 years ( The mean AOS and AOFAS scores were significantly better in the
non-diabetic group (p = 0.018 and p = 0.038, respectively). In all,
nine TARs (21%) in the diabetic group had clinical failure at a
mean follow-up of five years (24 to 109), which was significantly
higher than the rate of failure of 15 (11.6%) in the non-diabetic
group (p = 0.004). The uncontrolled diabetic subgroup had a significantly
poorer outcome than the non-diabetic group (p = 0.02), and a higher
rate of delayed wound healing. The incidence of early-onset osteolysis was higher in the diabetic
group than in the non-diabetic group (p = 0.02). These results suggest
that diabetes mellitus, especially with poor glycaemic control,
negatively affects the short- to mid-term outcome after TAR. Cite this article:
The aim of this study was to determine whether
an osteoplasty of the femoral neck performed at the same time as an
intertrochanteric Imhäuser osteotomy led to an improved functional
outcome or increased morbidity. A total of 20 hips in 19 patients
(12 left, 8 right, 13 male, 6 female), who underwent an Imhäuser
intertrochanteric osteotomy following a slipped capital femoral
epiphysis were assessed over an eight-year period. A total of 13
hips in 13 patients had an osteoplasty of the femoral neck at the
same time. The remaining six patients (seven hips) had intertrochanteric
osteotomy alone. The mean age was 15.3 years (13 to 20) with a mean
follow-up of 57.8 months (15 to 117); 19 of the slips were severe
(Southwick grade III) and one was moderate (grade II), with a mean
slip angle of 65.3° (50° to 80°); 17 of the slips were stable and
three unstable at initial presentation. The mean Non-Arthritic Hip Scores
(NAHS) in patients who underwent osteoplasty was 91.7 (76.3 to 100)
and the mean NAHS in patients who did not undergo osteoplasty was
76.6 (41.3 to 100) (p = 0.056). Two patients required a subsequent
arthroplasty and neither of these patients had an osteoplasty. No
hips developed osteonecrosis or chondrolysis, and there was no increase
in complications related to the osteoplasty. We recommend that for
patients with a slipped upper femoral epiphysis undergoing an intertrochanteric
osteotomy, the addition of an osteoplasty of the femoral neck should
be considered. Cite this article:
The June 2014 Hip &
Pelvis Roundup360 looks at: Modular femoral necks: early signs are not good; is corrosion to blame for modular neck failures; metal-on-metal is not quite a closed book; no excess failures in fixation of displaced femoral neck fractures; noise no problem in hip replacement; heterotopic ossification after hip arthroscopy: are NSAIDs the answer?; thrombotic and bleeding events surprisingly low in total joint replacement; and the elephant in the room: complications and surgical volume.
Nanotechnology is the study, production and controlled
manipulation of materials with a grain size <
100 nm. At this
level, the laws of classical mechanics fall away and those of quantum
mechanics take over, resulting in unique behaviour of matter in
terms of melting point, conductivity and reactivity. Additionally,
and likely more significant, as grain size decreases, the ratio
of surface area to volume drastically increases, allowing for greater interaction
between implants and the surrounding cellular environment. This
favourable increase in surface area plays an important role in mesenchymal
cell differentiation and ultimately bone–implant interactions. Basic science and translational research have revealed important
potential applications for nanotechnology in orthopaedic surgery,
particularly with regard to improving the interaction between implants
and host bone. Nanophase materials more closely match the architecture
of native trabecular bone, thereby greatly improving the osseo-integration
of orthopaedic implants. Nanophase-coated prostheses can also reduce
bacterial adhesion more than conventionally surfaced prostheses.
Nanophase selenium has shown great promise when used for tumour
reconstructions, as has nanophase silver in the management of traumatic
wounds. Nanophase silver may significantly improve healing of peripheral
nerve injuries, and nanophase gold has powerful anti-inflammatory
effects on tendon inflammation. Considerable advances must be made in our understanding of the
potential health risks of production, implantation and wear patterns
of nanophase devices before they are approved for clinical use.
Their potential, however, is considerable, and is likely to benefit
us all in the future. Cite this article:
We report on 397 consecutive revision total hip
replacements in 371 patients with a mean clinical and radiological follow-up
of 12.9 years (10 to 17.7). The mean age at surgery was 69 years
(37 to 93). A total of 28 patients (8%) underwent further revision,
including 16 (4%) femoral components. In all 223 patients (56%,
233 hips) died without further revision and 20 patients (5%, 20
hips) were lost to follow-up. Of the remaining patients, 209 (221
hips) were available for clinical assessment and 194 (205 hips)
for radiological review at mean follow-up of 12.9 years (10 to 17.7). The mean Harris Hip Score improved from 58.7 (11 to 92) points
to 80.7 (21 to 100) (p <
0.001) and the mean Merle d’Aubigné and
Postel hip scores at final follow-up were 4.9 (2 to 6), 4.5 (2 to
6) and 4.3 (2 to 6), respectively for pain, mobility and function.
Radiographs showed no lucencies around 186 (90.7%) femoral stems
with stable bony ingrowth seen in 199 stems (97%). The survival
of the S-ROM femoral stem at 15 years with revision for any reason as
the endpoint was 90.5% (95% confidence interval (CI) 85.7 to 93.8)
and with revision for aseptic loosening as the endpoint 99.3% (95%
CI 97.2 to 99.8). We have shown excellent long-term survivorship and good clinical
outcome of a cementless hydroxyapatite proximally-coated modular
femoral stem in revision hip surgery. Cite this article:
We compared the outcome of patients treated for an intertrochanteric fracture of the femoral neck with a locked, long intramedullary nail with those treated with a dynamic hip screw (DHS) in a prospective randomised study. Each patient who presented with an extra-capsular hip fracture was randomised to operative stabilisation with either a long intramedullary Holland nail or a DHS. We treated 92 patients with a Holland nail and 98 with a DHS. Pre-operative variables included the Mini Mental test score, patient mobility, fracture pattern and American Society of Anesthesiologists grading. Peri-operative variables were anaesthetic time, operating time, radiation time and blood loss. Post-operative variables were time to mobilising with a frame, wound infection, time to discharge, time to fracture union, and mortality. We found no significant difference in the pre-operative variables. The mean anaesthetic and operation times were shorter in the DHS group than in the Holland nail group (29.7 We conclude that the DHS can be implanted more quickly and with less exposure to radiation than the Holland nail. However, the resultant blood loss and need for transfusion is greater. The Holland nail allows patients to mobilise faster and to a greater extent. We have therefore adopted the Holland nail as our preferred method of treating intertrochanteric fractures of the hip.
This prospective study compares the outcome of
157 hydroxyapatite (HA)-coated tibial components with 164 cemented
components in the ROCC Rotating Platform total knee replacement
in 291 patients. The mean follow-up was 7.6 years (5.2 to 11). There
were two revisions for loosening: one for an HA-coated and one for
a cemented tibial component. Radiological evaluation demonstrated
no radiolucent lines with the HA-coated femoral components. A total
of three HA-coated tibial components exhibited radiolucent lines
at three months post-operatively and these disappeared after three
further months of protected weight-bearing. With HA-coated components
the operating time was shorter (p <
0.006) and the radiological
assessment of the tibial interface was more stable (p <
0.01).
Using revision for aseptic loosening of the tibial component as
the end point, the survival rates at nine years was identical for
both groups at 99.1%. Our results suggest that HA-coated components perform at least
as well as the same design with cemented components and compare
favourably with those of series describing cemented or porous-coated
knee replacements, suggesting that fixation of both components with
hydroxyapatite is a reliable option in primary total knee replacement.
The treatment of peri-prosthetic joint infection
(PJI) of the ankle is not standardised. It is not clear whether
an algorithm developed for hip and knee PJI can be used in the management
of PJI of the ankle. We evaluated the outcome, at two or more years
post-operatively, in 34 patients with PJI of the ankle, identified
from a cohort of 511 patients who had undergone total ankle replacement.
Their median age was 62.1 years (53.3 to 68.2), and 20 patients
were women. Infection was exogenous in 28 (82.4%) and haematogenous
in six (17.6%); 19 (55.9%) were acute infections and 15 (44.1%)
chronic. Staphylococci were the cause of 24 infections (70.6%).
Surgery with retention of one or both components was undertaken
in 21 patients (61.8%), both components were replaced in ten (29.4%),
and arthrodesis was undertaken in three (8.8%). An infection-free
outcome with satisfactory function of the ankle was obtained in
23 patients (67.6%). The best rate of cure followed the exchange
of both components (9/10, 90%). In the 21 patients in whom one or
both components were retained, four had a relapse of the same infecting organism
and three had an infection with another organism. Hence the rate
of cure was 66.7% (14 of 21). In these 21 patients, we compared
the treatment given to an algorithm developed for the treatment
of PJI of the knee and hip. In 17 (80.9%) patients, treatment was
not according to the algorithm. Most (11 of 17) had only one criterion against
retention of one or both components. In all, ten of 11 patients
with severe soft-tissue compromise as a single criterion had a relapse-free
survival. We propose that the treatment concept for PJI of the ankle
requires adaptation of the grading of quality of the soft tissues. Cite this article
We have assessed the different adhesive properties
of some of the most common bacteria associated with periprosthetic
joint infection on various types of ultra high molecular Weight
Polyethylene (UHMWPE). Quantitative Cite this article:
Peri- and sub-prosthetic fractures, or pathological fractures below an existing well-fixed femoral component, with or without an ipsilateral knee replacement, present a difficult surgical challenge. We describe a simple solution, in which a custom-made prosthesis with a cylindrical design is cemented proximally to the stem of an existing, well-fixed femoral component. This effectively treats the fracture without sacrificing the good hip. We describe five patients with a mean age of 73 years (60 to 81) and a mean follow-up of 47 months (6 to 108). The mean overlap of the prosthesis over the femoral component was 7.5 cm (5.5 to 10). There have been no mechanical failures, no new infections and no re-operations. We suggest that in highly selected cases, in which conventional fixation is not feasible, this technique offers a durable option and avoids the morbidity of a total femoral replacement.
The December 2013 Oncology Roundup360 looks at: Peri-articular resection fraught with complications; Navigated margins; Skeletal tumours and thromboembolism; Conditional survival in Ewing’s sarcoma; Reverse shoulders and tumour; For how long should we follow up sarcoma patients?; and already metastasised?
We examined the risk of thrombotic and major
bleeding events in patients undergoing total hip and knee replacement
(THR and TKR) treated with thromboprophylaxis, using nationwide
population-based databases. We identified 83 756 primary procedures
performed between 1997 and 2011. The outcomes were symptomatic venous thromboembolism
(VTE), myocardial infarction (MI), stroke, death and major bleeding
requiring hospitalisation within 90 days of surgery. A total of 1114 (1.3%) and 483 (0.6%) patients experienced VTE
and bleeding, respectively. The annual risk of VTE varied between
0.9% and 1.6%, and of bleeding between 0.4% and 0.8%. The risk of
VTE and bleeding was unchanged over a 15-year period. A total of
0.7% of patients died within 90 days, with a decrease from 1% in
1997 to 0.6% in 2011 (p <
0.001). A high level of comorbidity
and general anaesthesia were strong risk factors for both VTE and
bleeding, with no difference between THR and TKR patients. The risk
of both MI and stroke was 0.5%, which remained unchanged during
the study period. In this cohort study of patients undergoing THR and TKR patients
in routine clinical practice, approximately 3% experienced VTE,
MI, stroke or bleeding. These risks did not decline during the 15-year
study period, but the risk of dying fell substantially. Cite this article: