Advertisement for orthosearch.org.uk
Results 361 - 380 of 1373
Results per page:
The Journal of Bone & Joint Surgery British Volume
Vol. 79-B, Issue 2 | Pages 322 - 326
1 Mar 1997
Katz RP Callaghan JJ Sullivan PM Johnston RC

We performed 83 consecutive cemented revision total hip arthroplasties in 77 patients between 1977 and 1983 using improved cementing techniques. One patient (two hips) was lost to follow-up. The remaining 76 patients (81 hips) had an average age at revision of 63.7 years (23 to 89). At the final follow-up 18 hips (22%) had had a reoperation, two (2.5%) for sepsis, three (4%) for dislocation and 13 (16%) for aseptic loosening. The incidence of rerevision for aseptic femoral loosening was 5.4% and for aseptic acetabular loosening 16%. These results confirm that cemented femoral revision is a durable option in revision hip surgery when improved cementing techniques are used, but that cemented acetabular revision is unsatisfactory


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 3 | Pages 310 - 313
1 Mar 2005
Buckley SC Stockley I Hamer AJ Kerry RM

We report the results of the revision of 123 acetabular components for aseptic loosening treated by impaction bone grafting using frozen, morsellised, irradiated femoral heads and cemented sockets. This is the first large series using this technique to be reported. A survivorship of 88% with revision as the end-point after a mean of five years is comparable with that of other series


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 7 | Pages 879 - 886
1 Jul 2007
Howie DW Wimhurst JA McGee MA Carbone TA Badaruddin BS

We investigated 219 revisions of total hip replacement (THR) in 211 patients using a collarless double-taper cemented femoral component. The mean age of the patients was 72 years (30 to 90). The 137 long and 82 standard length stems were analysed separately. The mean follow-up was six years (2 to 18), and no patient was lost to follow-up. Survival of the long stems to re-revision for aseptic loosening at nine years was 98% (95% confidence interval (CI) 94 to 100), and for the standard stems was 93% (95% CI 85 to 100). At five years, one long stem was definitely loose radiologically and one standard stem was probably loose. Pre-operative femoral bone deficiency did not influence the results for the long stems, and corrective femoral osteotomy was avoided, as were significant subsidence, major stress shielding and persistent thigh pain. Because of these reliable results, cemented long collarless double-taper femoral components are recommended for routine revision THR in older patients


The Journal of Bone & Joint Surgery British Volume
Vol. 86-B, Issue 5 | Pages 759 - 770
1 Jul 2004
Vermes C Chandrasekaran R Dobai JG Jacobs JJ Andersson GBJ An H Hallab NJ Galante JO Glant TT

Periprosthetic bone loss after total joint arthroplasty is a major clinical problem resulting in aseptic loosening of the implant. Among many cell types, osteoblasts play a crucial role in the development of peri-implant osteolysis. In this study, we tested the effects of calcitriol (1α,25-dihydroxy-vitamin-D. 3. ) and the bisphosphonate pamidronate on titanium-particle- and TNF-α-induced release of interleukin-6 and suppression of osteoblast-specific gene expressions in bone-marrow-derived stromal cells with an osteoblastic phenotype. We monitored the expression of procollagen α1[1], osteocalcin, osteonectin and alkaline phosphatase mRNAs by Northern blots and real-time reverse transcription and polymerase chain reaction analyses. The release of various cytokines was also analysed by ELISA. We found that calcitriol or pamidronate could only partially recover the altered functions of osteoblasts when added alone. Only a combination of these compounds restored all the tested functions of osteoblasts. The local delivery of these drugs may have therapeutic potential to prevent or to treat periprosthetic osteolysis and aseptic loosening of implants


The Bone & Joint Journal
Vol. 104-B, Issue 7 | Pages 801 - 810
1 Jul 2022
Krull P Steinbrück A Grimberg AW Melsheimer O Morlock M Perka C

Aims

Registry studies on modified acetabular polyethylene (PE) liner designs are limited. We investigated the influence of standard and modified PE acetabular liner designs on the revision rate for mechanical complications in primary cementless total hip arthroplasty (THA).

Methods

We analyzed 151,096 primary cementless THAs from the German Arthroplasty Registry (EPRD) between November 2012 and November 2020. Cumulative incidence of revision for mechanical complications for standard and four modified PE liners (lipped, offset, angulated/offset, and angulated) was determined using competing risk analysis at one and seven years. Confounders were investigated with a Cox proportional-hazards model.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 1 | Pages 24 - 30
1 Jan 2009
Landor I Vavrik P Jahoda D Pokorny D Tawa A Sosna A

Migration of the acetabular component may give rise to oval-shaped bone defects in the acetabulum. The oblong implant is designed to fill these defects and achieve a stable cementless anchorage with no significant bone loss. We prospectively reviewed 133 oblong long oblique revision components at a mean follow-up of 9.74 years (0.6 to 14). All had been used in revisions for defects of type IIB to IIIB according to Paprosky. Aseptic loosening was the reason for revision in 11 cases (8.3%) and deep infection in seven (5.3%). The probability of implant survival over a 12-year follow-up estimated by the Kaplan-Meier method gave a survival rate of 0.85% respectively 0.90% when deep infection was excluded as the endpoint. Our study supports the use of these components in defects from IIB to IIIA. The main precondition for success is direct contact of more than half of the surface of the implant with the host acetabular bone


The Journal of Bone & Joint Surgery British Volume
Vol. 80-B, Issue 1 | Pages 151 - 155
1 Jan 1998
Ahmed M Bergström J Lundblad H Gillespie WJ Kreicbergs A

We studied the presence of sensory nerves by immunohistochemistry in the interface membranes of hip prostheses after aseptic loosening. Substance P (SP), neurokinin A (NKA) and calcitonin gene-related peptide (CGRP) were analysed as was protein gene product (PGP) 9.5, a general marker for nerve fibres. We identified nerve fibres in all samples but differences in their density were found. SP- and NKA-positive fibres were predominantly non-vascular, forming varicose nerve terminals. CGRP-immunoreactive nerve fibres with varicose terminals were seen mostly close to blood vessels, but also as free nerve endings. Sensory neuropeptides participate not only in nociception but also stimulate immune cells to release cytokines. The presence of sensory nerves in the interface membrane may reflect a pathophysiological response contributing to the aseptic loosening of hip prostheses


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 6 | Pages 867 - 874
1 Jun 2010
Hanna SA Sewell MD Aston WJS Pollock RC Skinner JA Cannon SR Briggs TWR

Segmental resection of malignant bone disease in the femoral diaphysis with subsequent limb reconstruction is a major undertaking. This is a retrospective review of 23 patients who had undergone limb salvage by endoprosthetic replacement of the femoral diaphysis for a primary bone tumour between 1989 and 2005. There were 16 males and seven females, with a mean age of 41.3 years (10 to 68). The mean overall follow-up was for 97 months (3 to 240), and 120 months (42 to 240) for the living patients. The cumulative patient survival was 77% (95% confidence interval 63% to 95%) at ten years. Survival of the implant, with failure of the endoprosthesis as an endpoint, was 85% at five years and 68% (95% confidence interval 42% to 92%) at ten years. The revision rate was 22% and the overall rate of re-operation was 26%. Complications included deep infection (4%), breakage of the prosthesis (8%), periprosthetic fracture (4%), aseptic loosening (4%), local recurrence (4%) and metastases (17%). The 16 patients who retained their diaphyseal endoprosthesis had a mean Musculoskeletal Tumour Society score of 87% (67% to 93%). They were all able to comfortably perform most activities of daily living. Femoral diaphyseal endoprosthetic replacement is a viable option for reconstruction following segmental resection of malignant bone disease. It allows immediate weight-bearing, is associated with a good long-term functional outcome, has an acceptable complication and revision rate and, most importantly, does not appear to compromise patient survival


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 1 | Pages 12 - 15
1 Jan 2005
Robertson A Lavalette D Morgan S Angus PD

We present the clinical and radiographic outcome of 68 consecutive primary total hip replacements performed in 54 patients under the age of 55 years using a hydroxyapatite-coated femoral component and threaded cup with a modular ceramic head (JRI-Furlong). We reviewed 62 (91%) hips at a median follow-up of 8.8 years (5 to 13.8) after implantation; six (9%) were lost to follow-up. At review there had been four (6%) revisions but only one for aseptic loosening (acetabulum). Radiographic review of the remaining hips did not identify any evidence of femoral or acetabular loosening. The median Harris and Merle d’Aubigné and Postel hip scores were 95.9 (42.7 to 100) and 17 (3 to 18) respectively. The JRI-Furlong hip gives promising functional and radiographic results in young patients in the medium term


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 12 | Pages 1561 - 1565
1 Dec 2009
Blakey CM Eswaramoorthy VK Hamilton LC Biant LC Field RE

We report the minimum five-year follow-up of 352 primary total hip replacements using the uncemented hydroxyapatite-coated ANCA-Fit femoral component with a modular neck and head. The series comprised 319 patients (212 men, 107 women) with a mean age at operation of 64.4 years (28 to 97). The principal diagnosis was osteoarthritis. A total of 18 patients (21 hips) died before their follow-up at five years, nine patients (11 hips) were lost to follow-up, and four (four hips) declined further follow-up. Patient-reported outcomes have been recorded for 288 patients (316 hips). Their mean Oxford Hip Score improved significantly from 41 points (16 to 57) pre-operatively to 20 points (12 to 44) at five-year follow-up. Radiological assessment showed good bony stability in 98% of implants. There were two cases of aseptic loosening of the femoral component. There were no clinical or radiological complications related to modularity. In our series we did not see the high rate of intra-operative fracture previously reported for this implant. This medium-term follow-up study demonstrates that the clinical outcome of the ANCA-Fit femoral component is, to date, comparable with that of other metaphyseal loading femoral components


The Journal of Bone & Joint Surgery British Volume
Vol. 77-B, Issue 6 | Pages 962 - 966
1 Nov 1995
Shahgaldi B Heatley F Dewar A Corrin B

Tissue reaction to wear particles from metal implants may play a major role in the aseptic loosening of implants. We used electron microprobe elemental analysis to determine the chemical composition of wear particles embedded in the soft tissues around hip and knee implants from 11 patients at revision surgery for aseptic loosening. The implants were made of cobalt-chromium-molybdenum alloy or titanium-aluminium-vanadium alloy. Histological examination showed a widespread giant-cell reaction to the particles. Elemental analysis showed that the chemical composition of the particles was different from that of the implanted alloys: cobalt and titanium were reduced, often down to zero, whereas chromium and aluminium persisted. Our findings indicate that corrosion is continually changing the shape, size and chemical composition of the implanted alloy. This may alter the biochemical environment of the tissue surrounding an implant to favour bone resorption


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 12 | Pages 1641 - 1648
1 Dec 2012
Baker PN Jameson SS Deehan DJ Gregg PJ Porter M Tucker K

Current analysis of unicondylar knee replacements (UKRs) by national registries is based on the pooled results of medial and lateral implants. Consequently, little is known about the differential performance of medial and lateral replacements and the influence of each implant type within these pooled analyses. Using data from the National Joint Registry for England and Wales (NJR) we aimed to determine the proportion of UKRs implanted on the lateral side of the knee, and their survival and reason for failure compared with medial UKRs. By combining information on the side of operation with component details held on the NJR, we were able to determine implant laterality (medial versus lateral) for 32 847 of the 35 624 unicondylar replacements (92%) registered before December 2010. Of these, 2052 (6%) were inserted on the lateral side of the knee. The rates of survival at five years were 93.1% (95% confidence interval (CI) 92.7 to 93.5) for medial and 93.0% (95% CI 91.1 to 94.9) for lateral UKRs (p = 0.49). The rates of failure remained equivalent after adjusting for patient age, gender, American Society of Anesthesiologists (ASA) grade, indication for surgery and implant design using Cox’s proportional hazards method (hazard ratio for lateral relative to medial replacement = 0.88 (95% CI 0.69 to 1.13); p = 0.32). Aseptic loosening/lysis and unexplained pain were the main reasons for revision in both groups, although the reasons did vary depending on whether a mobile- or a fixed-bearing design was used. At a maximum of eight years the mid-term survival rates of medial and lateral UKRs are similar


The Journal of Bone & Joint Surgery British Volume
Vol. 72-B, Issue 5 | Pages 757 - 760
1 Sep 1990
Mulroy R Harris W

Improved cementing techniques have been shown to decrease the rate of aseptic loosening of femoral components of cemented total hip replacements at five to seven years. We now report our results in 105 hips in 93 patients at 10 to 12.7 years (mean 11.2). The improved techniques included use of a medullary plug, a cement gun, a doughy mix of Simplex P and a collared stem of chrome cobalt. Only three femoral components had definitely loosened, none were probably loose and 24 were graded as possibly loose. In contrast, the incidence of radiographic loosening on the acetabular side was 42%. Improved cementing techniques have produced a marked reduction in the rate of aseptic loosening of the femoral component, but the incidence of acetabular loosening is unchanged


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 4 | Pages 449 - 454
1 Apr 2007
Kim Y Kim J Yoon S

We studied prospectively the long-term results of the Charnley Elite-Plus femoral stem in 184 consecutive young patients (194 hips). There were 130 men and 54 women with a mean age of 49.1 years (21 to 60). The predominant diagnosis was osteonecrosis of the femoral head (63.6%, 117 patients). Clinical and radiological evaluation was undertaken at each follow-up. The mean follow-up was 11.2 years (10 to 12). The mean pre-operative Harris hip score was 43.4 (12 to 49) which improved to 91 (59 to 100) at the final follow-up. The survival of the femoral stem at 12 years was 99% with revision as the end-point. The mean annual linear wear of the polyethylene liner was 0.17 mm (0.13 to 0.22). The prevalence of acetabular osteolysis was 10.8% (21 hips) and osteolysis of the calcar femorale 12.9% (25 hips). A third-generation cementing technique, accurate alignment of the stem and the use of a 22 mm zirconia head were important factors in the prevention of aseptic loosening of the Elite Plus femoral stem in these high-risk young patients


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 11 | Pages 1513 - 1520
1 Nov 2009
Sewell MD Spiegelberg BGI Hanna SA Aston WJS Bartlett W Blunn GW David LA Cannon SR Briggs TWR

We undertook a retrospective review of 33 patients who underwent total femoral endoprosthetic replacement as limb salvage following excision of a malignant bone tumour. In 22 patients this was performed as a primary procedure following total femoral resection for malignant disease. Revision to a total femoral replacement was required in 11 patients following failed segmental endoprosthetic or allograft reconstruction. There were 33 patients with primary malignant tumours, and three had metastatic lesions. The mean age of the patients was 31 years (5 to 68). The mean follow-up was 4.2 years (9 months to 16.4 years). At five years the survival of the implants was 100%, with removal as the endpoint and 56% where the endpoint was another surgical intervention. At five years the patient survival was 32%. Complications included dislocation of the hip in six patients (18%), local recurrence in three (9%), peri-prosthetic fracture in two and infection in one. One patient subsequently developed pulmonary metastases. There were no cases of aseptic loosening or amputation. Four patients required a change of bushings. The mean Musculoskeletal Tumour Society functional outcome score was 67%, the mean Harris Hip Score was 70, and the mean Oxford Knee Score was 34. Total femoral endoprosthetic replacement can provide good functional outcome without compromising patient survival, and in selected cases provides an effective alternative to amputation


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 2 | Pages 170 - 175
1 Feb 2009
Young L Duckett S Dunn A

We describe the survivorship of the Exeter femoral component in a District General Hospital. Between 1994 and 1996, 230 Exeter Universal cemented femoral components were implanted in 215 patients who were reviewed at a mean of 11.2 years (10 to 13). We used one acetabular implant, the Elite Ogee component, in 218 of the 230 hips. During the period of this study 76 patients (79 hips) died. Of the remaining 139 patients (151 hips), 121 were able to attend for radiological analysis at a minimum of ten years. One patient was lost to follow-up. No femoral component was revised for aseptic loosening. Three hips were revised for deep infection and six acetabular components required revision, four for loosening and two for recurrent dislocation. Taking the ‘worst-case scenario’ including the one patient lost to follow-up, the overall survival rate was 94.4% at 13 years. Our results confirm excellent medium-term results for the Exeter Universal femoral component, implanted in a general setting. The excellent survival of this femoral component, when used in combination with the Ogee acetabular component, suggests that this is a successful pairing


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 7 | Pages 946 - 949
1 Jul 2005
Talwalkar SC Givissis PK Trail IA Nuttall D Stanley JK

We divided 309 patients with an inflammatory arthritis who had undergone primary elbow replacement using the Souter-Strathclyde implant into two groups according to their age. The mean follow-up in the older group (mean age 64 years) was 7.3 years while in the younger patients (mean age 42 years) it was 12 years. Survivorship for three different failure end-points (revision, revision because of aseptic loosening of the humeral component, and gross loosening of the humeral implant), was compared in both groups. Our findings showed that there was no significant difference in the incidence of loosening when young rheumatoid patients were compared with an older age group


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 12 | Pages 1574 - 1580
1 Dec 2007
Hallan G Lie SA Furnes O Engesaeter LB Vollset SE Havelin LI

Primary uncemented femoral stems reported to the Norwegian arthroplasty register between 1987 and 2005 were included in this prospective observational study. There were 11 516 hips (9679 patients) and 14 different designs of stem. Kaplan-Meier survival probabilities and Cox regression were used to analyse the data. With aseptic loosening as the end-point, all currently used designs performed excellently with survival of 96% to 100% at ten years. With the end-point as stem revision for any cause, the long-term results of the different designs varied from poor to excellent, with survival at 15 years ranging between 29% and 97%. Follow-up for longer than seven years was needed to identify some of the poorly-performing designs. There were differences between the stems; the Corail, used in 5456 hips, was the most frequently used stem with a survival of 97% at 15 years. Male gender was associated with an increased risk of revision of × 1.3 (95% confidence interval 1.05 to 1.52), but age and diagnosis had no influence on the results. Overall, modern uncemented femoral stems performed well. Moderate differences in survival between well-performing stems should be interpreted with caution since the differences may be caused by factors other than the stem itself


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 4 | Pages 465 - 470
1 Apr 2005
Reinartz P Mumme T Hermanns B Cremerius U Wirtz DC Schaefer WM Niethard F Buell U

Two major complications of hip replacement are loosening and infection. Reliable differentiation between these pathological processes is difficult since both may be accompanied by similar symptoms. Our aim was to assess the diagnostic ability of triple-phase bone scanning (TPBS) and positron-emission tomography (PET) to detect and differentiate these complications in patients with a hip arthroplasty. Both TPBS and PET were performed in 63 patients (92 prostheses). The radiotracer for PET imaging was . 18. F-fluorodeoxyglucose (FDG). Image interpretation was performed according to qualitative and quantitative criteria although the final diagnosis was based upon either surgical findings or clinical follow-up. The sensitivity, specificity and accuracy of PET was 0.94, 0.95 and 0.95 respectively, compared with 0.68, 0.76 and 0.74 for TPBS. We found that an image interpretation based exclusively upon quantitative criteria was inappropriate because of its low selectivity. The histological examination indicated that increased periprosthetic uptake of FDG in patients with aseptic loosening was caused by wear-induced polyethylene particles and the subsequent growth of aggressive granulomatous tissue


The Bone & Joint Journal
Vol. 97-B, Issue 5 | Pages 649 - 653
1 May 2015
Hawi N Kendoff D Citak M Gehrke T Haasper C

Knee arthrodesis is a potential salvage procedure for limb preservation after failure of total knee arthroplasty (TKA) due to infection. In this study, we evaluated the outcome of single-stage knee arthrodesis using an intramedullary cemented coupled nail without bone-on-bone fusion after failed and infected TKA with extensor mechanism deficiency. Between 2002 and 2012, 27 patients (ten female, 17 male; mean age 68.8 years; 52 to 87) were treated with septic single-stage exchange. Mean follow-up duration was 67.1months (24 to 143, n = 27) (minimum follow-up 24 months) and for patients with a minimum follow-up of five years 104.9 (65 to 143,; n = 13). A subjective patient evaluation (Short Form (SF)-36) was obtained, in addition to the Visual Analogue Scale (VAS). The mean VAS score was 1.44 (SD 1.48). At final follow-up, four patients had recurrent infections after arthrodesis (14.8%). Of these, three patients were treated with a one-stage arthrodesis nail exchange; one of the three patients had an aseptic loosening with a third single-stage exchange, and one patient underwent knee amputation for uncontrolled sepsis at 108 months. All patients, including the amputee, indicated that they would choose arthrodesis again. Data indicate that a single-stage knee arthrodesis offers an acceptable salvage procedure after failed and infected TKA. Cite this article: Bone Joint J 2015;97-B:649–53