Metal-on-metal hip arthroplasty gained significant
favor in the first decade of the millennium. However, the past several
years have seen increasing reports of failure, pseudotumor and other
adverse reactions. This study presents the results of a single center’s
15-year experience with metal-on-metal total hip arthroplasty as
strong evidence that metal-on-metal is going, going, gone.
We reviewed the outcome of 69 uncemented, custom-made,
distal femoral endoprosthetic replacements performed in 69 patients
between 1994 and 2006. There were 31 women and 38 men with a mean
age at implantation of 16.5 years (5 to 37). All procedures were
performed for primary malignant bone tumours of the distal femur.
At a mean follow-up of 124.2 months (4 to 212), 53 patients were
alive, with one patient lost to follow-up. All nine implants (13.0%)
were revised due to aseptic loosening at a mean of 52 months (8
to 91); three implants (4.3%) were revised due to fracture of the
shaft of the prosthesis and three patients (4.3%) had a peri-prosthetic
fracture. Bone remodelling associated with periosteal cortical thinning
adjacent to the uncemented intramedullary stem was seen in 24 patients
but this did not predispose to failure. All aseptically loose implants
in this series were diagnosed to be loose within the first five
years. The results from this study suggest that custom-made uncemented
distal femur replacements have a higher rate of aseptic loosening
compared to published results for this design when used with cemented
fixation. Loosening of uncemented replacements occurs early indicating
that initial fixation of the implant is crucial. Cite this article:
Conventional uncemented femoral implants provide
dependable long-term fixation in patients with a wide range of functional
requirements. Yet challenges associated with proximal–distal femoral
dimensional mismatch, preservation of bone stock, and minimally
invasive approaches have led to exploration into alternative implant designs.
Short stem designs focusing on a stable metaphyseal fit have emerged
to address these issues in total hip replacement (THR). Uncemented
metaphyseal-engaging short stem implants are stable and are associated
with proximal bone remodeling closer to the metaphysis when compared
with conventional stems and they also have comparable clinical performances.
Short stem metaphyseal-engaging implants can meet the goals of a
successful THR, including tolerating a high level of patient function,
as well as durable fixation. Cite this article:
Some surgeons are reluctant to perform a reverse
total shoulder arthroplasty (RTSA) on both shoulders because of concerns
regarding difficulty with activities of daily living post-operatively
as a result of limited rotation of the shoulders. Nevertheless,
we hypothesised that outcomes and patient satisfaction following
bilateral RTSA would be comparable to those following unilateral
RTSA. A single-surgeon RTSA registry was reviewed for patients who underwent
bilateral staged RTSA with a minimum follow-up of two years. A unilateral
RTSA matched control was selected for each shoulder in those patients
undergoing bilateral procedures. The Constant–Murley score (CMS), American
Shoulder and Elbow Surgeons (ASES) score, Subjective Shoulder Values
(SSV), visual analogue scale (VAS) for pain, range of movement and
strength were measured pre- and post-operatively. The mean CMS,
ASES, SSV, VAS scores, strength and active forward elevation were
significantly improved
(all p <
0.01) following each operation in those undergoing bilateral
procedures. The mean active external rotation (p = 0.63 and p =
0.19) and internal rotation (p = 0.77 and p = 0.24) were not significantly
improved. The improvement in the mean ASES score after the first
RTSA was greater than the improvement in its control group (p =
0.0039). The improvement in the mean CMS, ASES scores and active
forward elevation was significantly less after the second RTSA than
in its control group (p = 0.0244, p = 0.0183, and p = 0.0280, respectively).
Pain relief and function significantly improved after each RTSA
in those undergoing a bilateral procedure. Bilateral RTSA is thus a reasonable form of treatment for patients
with severe bilateral rotator cuff deficiency, although inferior
results may be seen after the second procedure compared with the
first. Cite this article:
We investigated the characteristics of patients
who achieved Japanese-style deep flexion (seiza-sitting) after total knee
replacement (TKR) and measured three-dimensional positioning and
the contact positions of the femoral and tibial components. Seiza-sitting
was achieved after surgery by 23 patients (29 knees) of a series
of 463 TKRs in 341 patients. Pre-operatively most of these patients
were capable of seiza-sitting, had a lower body mass index and a favourable
attitude towards the Japanese lifestyle (27 of 29 knees). According
to two-/three-dimensional image registration analysis in the seiza-sitting
position, flexion, varus and internal rotation angles of the tibial
component relative to the femoral component had means of 148° ( Cite this article:
As part of the national initiative to reduce
waiting times for joint replacement surgery in Wales, the Cardiff
and Vale NHS Trust referred 224 patients to the NHS Treatment Centre
in Weston-Super-Mare for total knee replacement (TKR). A total of
258 Kinemax TKRs were performed between November 2004 and August
2006. Of these, a total of 199 patients (232 TKRs, 90%) have been
followed up for five years. This cohort was compared with 258 consecutive
TKRs in 250 patients, performed at Cardiff and Vale Orthopaedic
Centre (CAVOC) over a similar time period. The five year cumulative
survival rate was 80.6% (95% confidence interval (CI) 74.0 to 86.0)
in the Weston-Super-Mare cohort and 95.0% (95% CI 90.2 to 98.2)
in the CAVOC cohort with revision for any reason as the endpoint.
The relative risk for revision at Weston-Super-Mare compared with
CAVOC was 3.88 (p <
0.001). For implants surviving five years,
the mean Oxford knee scores (OKS) and mean EuroQol (EQ-5D) scores
were similar (OKS: Weston-Super-Mare 29 (2 to 47) The results show a higher revision rate for patients operated
at Weston-Super-Mare Treatment Centre, with a reduction in functional
outcome and quality of life after revision. This further confirms
that patients moved from one area to another for joint replacement
surgery fare poorly.
The optimum cementing technique for the tibial
component in cemented primary total knee replacement (TKR) remains
controversial. The technique of cementing, the volume of cement
and the penetration are largely dependent on the operator, and hence
large variations can occur. Clinical, experimental and computational
studies have been performed, with conflicting results. Early implant
migration is an indication of loosening. Aseptic loosening is the
most common cause of failure in primary TKR and is the product of
several factors. Sufficient penetration of cement has been shown
to increase implant stability. This review discusses the relevant literature regarding all aspects
of the cementing of the tibial component at primary TKR. Cite this article:
Hip implant retrieval analysis is the most important
source of insight into the performance of new materials and designs
of hip arthroplasties. Even the most rigorous
This editorial considers the shortcomings of assessing outcome after joint replacement only by the survival of the implant.
Correct positioning and alignment of components during primary total knee replacement (TKR) is widely accepted to be an important predictor of patient satisfaction and implant durability. This retrospective study reports the effect of the post-operative mechanical axis of the lower limb in the coronal plane on implant survival following primary TKR. A total of 501 TKRs in 396 patients were divided into an aligned group with a neutral mechanical axis (± 3°) and a malaligned group where the mechanical axis deviated from neutral by >
3°. At 15 years’ follow-up, 33 of 458 (7.2%) TKRs were revised for aseptic loosening. Kaplan-Meier survival analysis showed a weak tendency towards improved survival with restoration of a neutral mechanical axis, but this did not reach statistical significance (p = 0.47). We found that the relationship between survival of a primary TKR and mechanical axis alignment is weaker than that described in a number of previous reports.
Ensuring correct rotation of the femoral component
is a challenging aspect of patellofemoral replacement surgery. Rotation
equal to the epicondylar axis or marginally more external rotation
is acceptable. Internal rotation is associated with poor outcomes.
This paper comprises two studies evaluating the use of the medial
malleolus as a landmark to guide rotation. We used 100 lower-leg anteroposterior radiographs to evaluate
the reliability of the medial malleolus as a landmark. Assessment
was made of the angle between the tibial shaft and a line from the
intramedullary rod entry site to the medial malleolus. The femoral
cut was made in ten cadaver knees using the inferior tip of the
medial malleolus as a landmark for rotation. Rotation of the cut
relative to the anatomical epicondylar axis was assessed using CT.
The study of radiographs found the position of the medial malleolus
relative to the tibial axis is consistent. Using the inferior tip
of the medial malleolus in the cadaver study produced a mean external
rotation of 1.6° (0.1° to 3.7°) from the anatomical epicondylar
axis. Using the inferior tip of the medial malleolus to guide the
femoral cutting jig avoids internal rotation and introduces an acceptable
amount of external rotation of the femoral component.
The purpose of this paper is to discuss the risk
factors, prevention strategies, classification, and treatment of
intra-operative femur fractures sustained during primary and revision
total hip arthroplasty.
Peri-prosthetic patellar fracture following resurfacing
as part of total knee replacement (TKR) is an infrequent yet challenging
complication. This case-control study was performed to identify
clinical, radiological and surgical factors that increase the risk
of developing a spontaneous patellar fracture after TKR. Patellar
fractures were identified in 74 patients (88 knees) from a series
of 7866 consecutive TKRs conducted between 1998 and 2009. After excluding
those with a previous history of extensor mechanism realignment
or a clear traumatic event, a metal-backed patella, any uncemented
component or subsequent infection, the remaining 64 fractures were
compared with a matched group of TKRs with an excellent outcome
defined by the Knee Society score. The mean age of patients with
a fracture was 70 years (51 to 81) at the time of TKR. Patellar
fractures were detected at a mean of 13.4 months (2 to 84) after
surgery. The incidence of patellar fracture was found to be strongly
associated with the number of previous knee operations, greater
pre-operative mechanical malalignment, smaller post-operative patellar
tendon length, thinner post-resection patellar thickness, and a
lower post-operative Insall-Salvati ratio. An understanding of the risk factors associated with spontaneous
patellar fracture following TKR provides a valuable insight into
prevention of this challenging complication.
This prospective randomised controlled double-blind
trial compared two types of PFC Sigma total knee replacement (TKR),
differing in three design features aimed at improving flexion. The
outcome of a standard fixed-bearing posterior cruciate ligament-preserving
design (FB-S) was compared with that of a high-flexion rotating-platform
posterior-stabilised design (RP-F) at one year after TKR. The study group of 77 patients with osteoarthritis of the knee
comprised 37 men and 40 women, with a mean age of 69 years (44.9
to 84.9). The patients were randomly allocated either to the FB-S
or the RP-F group and assessed pre-operatively and at one year post-operatively.
The mean post-operative non-weight-bearing flexion was 107° (95%
confidence interval (CI) 104° to 110°)) for the FB-S group and 113°
(95% CI 109° to 117°) for the RP-F group, and this difference was
statistically significant (p = 0.032). However, weight-bearing range
of movement during both level walking and ascending a slope as measured
during flexible electrogoniometry was a mean of 4° lower in the RP-F
group than in the FB-S group, with 58° (95% CI 56° to 60°) Although the RP-F group achieved higher non-weight-bearing knee
flexion, patients in this group did not use this during activities
of daily living and reported more pain one year after surgery
This was a safety study where the hypothesis was that the newer-design CPCS femoral stem would demonstrate similar early clinical results and micromovement to the well-established Exeter stem. Both are collarless, tapered, polished cemented stems, the only difference being a slight lateral to medial taper with the CPCS stem. A total of 34 patients were enrolled in a single-blinded randomised controlled trial in which 17 patients received a dedicated radiostereometric CPCS stem and 17 a radiostereometric Exeter stem. No difference was found in any of the outcome measures pre-operatively or post-operatively between groups. At two years, the mean subsidence for the CPCS stem was nearly half that seen for the Exeter stem (0.77 mm (−0.943 to 1.77) and 1.25 mm (0.719 to 1.625), respectively; p = 0.032). In contrast, the mean internal rotation of the CPCS stem was approximately twice that of the Exeter (1.61° (−1.07° to 4.33°) and 0.59° (0.97° to 1.64°), respectively; p = 0.048). Other migration patterns were not significantly different between the stems. The subtle differences in designs may explain the different patterns of migration. Comparable migration with the Exeter stem suggests that the CPCS design will perform well in the long term.
The aim of this retrospective cohort study was
to identify any difference in femoral offset as measured on pre-operative
anteroposterior (AP) radiographs of the pelvis, AP radiographs of
the hip and corresponding CT scans in a consecutive series of 100
patients with primary end-stage osteoarthritis of the hip (43 men
and 57 women with a mean age of 61 years (45 to 74) and a mean body
mass index of 28 kg/m2 (20 to 45)). Patients were positioned according to a standardised protocol
to achieve reproducible projection and all images were calibrated.
Inter- and intra-observer reliability was evaluated and agreement
between methods was assessed using Bland-Altman plots. In the entire cohort, the mean femoral offset was 39.0 mm (95%
confidence interval (CI) 37.4 to 40.6) on radiographs of the pelvis,
44.0 mm (95% CI 42.4 to 45.6) on radiographs of the hip and 44.7
mm (95% CI 43.5 to 45.9) on CT scans. AP radiographs of the pelvis
underestimated femoral offset by 13% when compared with CT (p <
0.001).
No difference in mean femoral offset was seen between AP radiographs
of the hip and CT (p = 0.191). Our results suggest that femoral offset is significantly underestimated
on AP radiographs of the pelvis but can be reliably and accurately
assessed on AP radiographs of the hip in patients with primary end-stage
hip osteoarthritis. We, therefore, recommend that additional AP radiographs of the
hip are obtained routinely for the pre-operative assessment of femoral
offset when templating before total hip replacement.
Increasing numbers of posterior lumbar fusions
are being performed. The purpose of this study was to identify trends
in demographics, mortality and major complications in patients undergoing
primary posterior lumbar fusion. We accessed data collected for
the Nationwide Inpatient Sample for each year between 1998 and 2008
and analysed trends in the number of lumbar fusions, mean patient
age, comorbidity burden, length of hospital stay, discharge status,
major peri-operative complications and mortality. An estimated 1 288 496
primary posterior lumbar fusion operations were performed between
1998 and 2008 in the United States. The total number of procedures,
mean patient age and comorbidity burden increased over time. Hospital
length of stay decreased, although the in-hospital mortality (adjusted
and unadjusted for changes in length of hospital stay) remained
stable. However, a significant increase was observed in peri-operative
septic, pulmonary and cardiac complications. Although in-hospital mortality
rates did not change over time in the setting of increases in mean
patient age and comorbidity burden, some major peri-operative complications
increased. These trends highlight the need for appropriate peri-operative services
to optimise outcomes in an increasingly morbid and older population
of patients undergoing lumbar fusion.
We report the long-term results of revision total
hip replacement using femoral impaction allografting with both uncemented
and cemented Freeman femoral components. A standard design of component
was used in both groups, with additional proximal hydroxyapatite
coating in the uncemented group. A total of 33 hips in 30 patients received
an uncemented component and 31 hips in 30 patients a cemented component.
The mean follow-up was 9.8 years (2 to 17) in the uncemented group
and 6.2 years (1 to 11) in the cemented group. Revision procedures
(for all causes) were required in four patients (four hips) in the
uncemented group and in five patients (five hips) in the cemented
group. Harris hip scores improved significantly in both groups and
were maintained independently of the extent of any migration of
the femoral component within the graft or graft–cement mantle.
Blood metal ions have been widely used to investigate
metal-on-metal hip replacements, but their ability to discriminate
between well-functioning and failed hips is not known. The Medicines
and Healthcare products Regulatory Agency (MHRA) has suggested a
cut-off level of 7 parts per billion (ppb). We performed a pair-matched, case-control study to investigate
the sensitivity and specificity of blood metal ion levels for diagnosing
failure in 176 patients with a unilateral metal-on-metal hip replacement.
We recruited 88 cases with a pre-revision, unexplained failed hip
and an equal number of matching controls with a well-functioning
hip. We investigated the 7 ppb cut-off level for the maximum of
cobalt or chromium and determined optimal mathematical cut-off levels
from receiver-operating characteristic curves. The 7 ppb cut-off level for the maximum of cobalt or chromium
had a specificity of 89% and sensitivity 52% for detecting a pre-operative
unexplained failed metal on metal hip replacement. The optimal cut-off
level for the maximum of cobalt or chromium was 4.97 ppb and had
sensitivity 63% and specificity 86%. Blood metal ions had good discriminant ability to separate failed
from well-functioning hip replacements. The MHRA cut-off level of
7 ppb provides a specific test but has poor sensitivity.