Mechanical wear and corrosion at the head-stem junction of total hip arthroplasties (THAs) (trunnionosis) have been implicated in their early revision, most commonly in metal-on-metal (MOM) hips. We can isolate the role of the head-stem junction as the predominant source of metal release by investigating non-MOM hips; this can help to identify clinically significant volumes of material loss and corrosion from these surfaces. In this study we examined a series of 94 retrieved metal-on-polyethylene (MOP) hips for evidence of corrosion and material loss at the taper junction using a well published visual grading method and an established roundness-measuring machine protocol. Hips were retrieved from 74 male and 20 female patients with a median age of 57 years (30 to 76) and a median time to revision of 215 months (2 to 324). The reasons for revision were loosening of both the acetabular component and the stem (n = 29), loosening of the acetabular component (n = 58) and infection (n = 7). No adverse tissue reactions were reported by the revision surgeons.Objectives
Methods
The aim of this study was to identify patient- and surgery-related
risk factors for sustaining an early periprosthetic fracture following
primary total hip arthroplasty (THA) performed using a double-tapered
cementless femoral component (Bi-Metric femoral stem; Biomet Inc.,
Warsaw, Indiana). A total of 1598 consecutive hips, in 1441 patients receiving
primary THA between January 2010 and June 2015, were retrospectively
identified. Level of pre-operative osteoarthritis, femoral Dorr
type and cortical index were recorded. Varus/valgus placement of
the stem and canal fill ratio were recorded post-operatively. Periprosthetic
fractures were identified and classified according to the Vancouver
classification. Regression analysis was performed to identify risk
factors for early periprosthetic fracture.Aims
Patients and Methods
Wound complications are reported in up to 10% hip and knee arthroplasties and there is a proven association between wound complications and deep prosthetic infections. In this randomised controlled trial (RCT) we explore the potential benefits of a portable, single use, incisional negative pressure wound therapy dressing (iNPWTd) on wound exudate, length of stay (LOS), wound complications, dressing changes and cost-effectiveness following total hip and knee arthroplasties. A total of 220 patients undergoing elective primary total hip and knee arthroplasties were recruited into in a non-blinded RCT. For the final analysis there were 102 patients in the study group and 107 in the control group.Objectives
Methods
The aim of this study was to establish what happens to patients
in the long term after endoprosthetic replacement for a primary
malignant tumour of bone. We conducted a retrospective analysis of a prospectively maintained
database to identify all patients who had undergone an endoprosthetic
replacement more than 25 years ago and who were still alive. Their
outcomes were investigated with reference to their complications and
need for further surgery. A total of 230 patients were identified.
Their mean age at diagnosis was 20.7 years (five to 62). The most
common diagnosis was osteosarcoma (132). The most common site was
the distal femur (102). Aims
Patients and Methods
This review summarises the evidence for the treatment
of displaced fractures of the femoral neck in elderly patients.
Results from randomised clinical trials and national register studies are
presented when available. The advantages of arthroplasty compared with internal fixation
are supported by several studies. A number of studies contribute
to the discussions of total hip arthroplasty (THA) The direct lateral approach reduces the risk of dislocation compared
with the posterior approach. Cemented implants lower the risk of
periprosthetic fracture and its subsequent morbidity and mortality.
As the risk of peri-operative death related to bone cement can be reduced
by adequate measures, cemented implants are recommended in fracture
cases. Take home message: There remains a great variation in the surgical
management of patients with a hip fracture, and an evidence-based
approach should improve the outcomes for this vulnerable patient
group. Cite this article:
The February 2015 Hip &
Pelvis Roundup360 looks at: Hip arthroplasty in Down syndrome; Bulk femoral autograft successful in acetabular reconstruction; Arthroplasty follow-up: is the internet the solution?; Total hip arthroplasty following acetabular fracture; Salvage arthroplasty following failed hip internal fixation; Bone banking sensible financially and clinically; Allogenic blood transfusion in arthroplasty.
The ‘jumbo’ acetabular component is now commonly
used in acetabular revision surgery where there is extensive bone
loss. It offers high surface contact, permits weight bearing over
a large area of the pelvis, the need for bone grafting is reduced
and it is usually possible to restore centre of rotation of the
hip. Disadvantages of its use include a technique in which bone
structure may not be restored, a risk of excessive posterior bone
loss during reaming, an obligation to employ screw fixation, limited
bone ingrowth with late failure and high hip centre, leading to increased
risk of dislocation. Contraindications include unaddressed pelvic
dissociation, inability to implant the component with a rim fit,
and an inability to achieve screw fixation. Use in acetabulae with
<
50% bone stock has also been questioned. Published results
have been encouraging in the first decade, with late failures predominantly because
of polyethylene wear and aseptic loosening. Dislocation is the most
common complication of jumbo acetabular revisions, with an incidence
of approximately 10%, and often mandates revision. Based on published results,
a hemispherical component with an enhanced porous coating, highly
cross-linked polyethylene, and a large femoral head appears to represent
the optimum tribology for jumbo acetabular revisions. Cite this article:
There are many guidelines that help direct the management of
patients with metal-on-metal (MOM) hip arthroplasties. We have undertaken
a study to compare the management of patients with MOM hip arthroplasties in
different countries. Six international tertiary referral orthopaedic centres were
invited to participate by organising a multi-disciplinary team (MDT)
meeting, consisting of two or more revision hip arthroplasty surgeons
and a musculoskeletal radiologist. A full clinical dataset including
history, blood tests and imaging for ten patients was sent to each
unit, for discussion and treatment planning. Differences in the
interpretation of findings, management decisions and rationale for
decisions were compared using quantitative and qualitative methods.Aims
Methods
T-cells are considered to play an important role in the inflammatory response causing arthroplasty failure. The study objectives were to investigate the composition and distribution of CD4+ T-cell phenotypes in the peripheral blood (PB) and synovial fluid (SF) of patients undergoing revision surgery for failed metal-on-metal (MoM) and metal-on-polyethylene (MoP) hip arthroplasties, and in patients awaiting total hip arthroplasty. In this prospective case-control study, PB and SF were obtained from 22 patients (23 hips) undergoing revision of MoM (n = 14) and MoP (n = 9) hip arthroplasties, with eight controls provided from primary hip osteoarthritis cases awaiting arthroplasty. Lymphocyte subtypes in samples were analysed using flow cytometry.Objectives
Methods
Periprosthetic joint infection (PJI) is a devastating
complication for patients and results in greatly increased costs
of care for both healthcare providers and patients. More than 15
500 revision hip and knee procedures were recorded in England, Wales
and Northern Ireland in 2013, with infection accounting for 13%
of revision hip and 23% of revision knee procedures. We report our experience of using antibiotic eluting absorbable
calcium sulphate beads in 15 patients (eight men and seven women
with a mean age of 64.8 years; 41 to 83) as part of a treatment
protocol for PJI in revision arthroplasty. The mean follow-up was 16 months (12 to 22). We report the outcomes
and complications, highlighting the risk of hypercalcaemia which
occurred in three patients. We recommend that serum levels of calcium be routinely sought
following the implantation of absorbable calcium sulphate beads
in orthopaedic surgery. Cite this article:
Infection is a leading indication for revision
arthroplasty. Established criteria used to diagnose prosthetic joint infection
(PJI) include a range of laboratory tests. Leucocyte esterase (LE)
is widely used on a colorimetric reagent strip for the diagnosis
of urinary tract infections. This inexpensive test may be used for
the diagnosis or exclusion of PJI. Aspirates from 30 total hip arthroplasties
(THAs) and 79 knee arthroplasties (KA) were analysed for LE activity. Semi-quantitative
reagent strip readings of 15, 70, 125 and 500 white blood cells
(WBC) were validated against a manual synovial white cell count
(WCC). A receiver operating characteristic (ROC) curve was constructed
to determine the optimal cut-off point for the semi-quantitative
results. Based on established criteria, six THAs and 15 KAs were
classified as infected. The optimal cut-off point for the diagnosis
of PJI was 97 WBC. The closest semi-quantitative reading for a positive
result was 125 WBC, achieving a sensitivity of 81% and a specificity
of 93%. The positive and negative predictive values of the LE test
strip were 74% and 95% respectively. The LE reagent strip had a high specificity and negative predictive
value. A negative result may exclude PJI and negate the need for
further diagnostic tests. Cite this article:
The June 2015 Oncology Roundup360 looks at: Infection in megaprosthesis; Impressive results for mid femoral reconstruction; Revered teaching or old myth? Femoral neck protection in metastatic disease; Megaprosthesis about the knee; Malignant transformation in multiple hereditary exostoses; Fracture of intercalary bone allograft; Comorbidity and outcomes in sarcoma; A worrying turn? Use of denosumab for giant cell tumour of bone
An extensive review of the spinal and arthroplasty
literature was undertaken to evaluate the effectiveness of local
antibiotic irrigation during surgery. The efficacy of antibiotic
irrigation for the prevention of acute post-operative infection
after total joint arthroplasty was evaluated retrospectively in
2293 arthroplasties (1990 patients) between January 2004 and December
2013. The mean follow-up was 73 months (20 to 139). One surgeon
performed all the procedures with minimal post-operative infection. The intra-operative protocol included an irrigation solution
of normal saline with vancomycin 1000 mg/l and polymyxin 250 000
units/l at the rate of 2 l per hour. No patient required re-admission
for primary infection or further antibiotic treatment. Two morbidly obese
patients (two total hip arthroplasties) developed subcutaneous fat
necrosis requiring debridement and one was revised because the deep
capsular sutures were contaminated by the draining subcutaneous
haematoma. One patient who had undergone total knee arthroplasty
had unrecognised damage to the lateral superior geniculate artery
and developed a haematoma that became infected secondarily four
months after the surgery and underwent revision. The use of antibiotic irrigation during arthroplasty surgery
has been highly effective for the prevention of infection in the
author’s practice. However, it should be understood that any routine
prophylactic use of antibiotics may result in resistant organisms,
and the wise stewardship of the use of antibiotics is an important
part of surgical practice. Cite this article:
The June 2015 Knee Roundup360 looks at: Cruciate substituting
Ideal placement of the acetabular component remains
elusive both in terms of defining and achieving a target. Our aim
is to help restore original anatomy by using the transverse acetabular
ligament (TAL) to control the height, depth and version of the component.
In the normal hip the TAL and labrum extend beyond the equator of
the femoral head and therefore, if the definitive acetabular component
is positioned such that it is cradled by and just deep to the plane
of the TAL and labrum and is no more than 4mm larger than the original
femoral head, the centre of the hip should be restored. If the face
of the component is positioned parallel to the TAL and psoas groove
the patient specific version should be restored. We still use the
TAL for controlling version in the dysplastic hip because we believe
that the TAL and labrum compensate for any underlying bony abnormality. The TAL should not be used as an aid to inclination. Worldwide,
>
75% of surgeons operate with the patient in the lateral decubitus
position and we have shown that errors in post-operative radiographic
inclination (RI) of >
50° are generally caused by errors in patient positioning.
Consequently, great care needs to be taken when positioning the
patient. We also recommend 35° of apparent operative inclination
(AOI) during surgery, as opposed to the traditional 45°. Cite this article:
The February 2014 Knee Roundup360 looks at: whether sham surgery is as good as arthroscopic meniscectomy; distraction in knee osteoarthritis; whether trans-tibial tunnel placement increases the risk of graft failure in ACL surgery; whether joint replacements prevent cardiac events; the size of the pulmonary embolism problem; tranexamic acid and knee replacement haemostasis; matching the demand for knee replacement and follow-up; predicting the length of stay after knee replacement; and popliteal artery injury in TKR.
The December 2013 Hip &
Pelvis Roundup360 looks at: Enhanced recovery works; Acetabular placement; Exercise better than rest in osteoarthritis patients; if Birmingham hip resurfacing is immune from pseudotumour; HIV and arthroplasty; Labral tears revisited; Prophylactic surgery for FAI; and Ceramics and impaction grafting