Advertisement for orthosearch.org.uk
Results 281 - 300 of 569
Results per page:
Bone & Joint Research
Vol. 8, Issue 6 | Pages 253 - 254
1 Jun 2019
de Steiger R


The Bone & Joint Journal
Vol. 101-B, Issue 7 | Pages 787 - 792
1 Jul 2019
Goto K Kuroda Y Kawai T Kawanabe K Matsuda S

Aims

In the 1990s, a bioactive bone cement (BABC) containing apatite-wollastonite glass-ceramic (AW-GC) powder and bisphenol-a-glycidyl methacrylate resin was developed at our hospital. In 1996, we used BABC to fix the acetabular component in primary total hip arthroplasty (THA) in 20 patients as part of a clinical trial. The purpose of this study was to investigate the long-term results of primary THA using BABC.

Patients and Methods

A total of 20 patients (three men and 17 women) with a mean age of 57.4 years (40 to 71), a mean body weight of 52.3 kg (39 to 64), and a mean body mass index (BMI) of 23.0 kg/m2 (19.8 to 28.6) were evaluated clinically and radiologically. Survival analyses were undertaken, and wear analyses were carried out using a computer-aided method.


Bone & Joint Research
Vol. 8, Issue 6 | Pages 246 - 252
1 Jun 2019
Liddle A Webb M Clement N Green S Liddle J German M Holland J

Objectives

Previous studies have evidenced cement-in-cement techniques as reliable in revision arthroplasty. Commonly, the original cement mantle is reshaped, aiding accurate placement of the new stem. Ultrasonic devices selectively remove cement, preserve host bone, and have lower cortical perforation rates than other techniques. As far as the authors are aware, the impact of ultrasonic devices on final cement-in-cement bonds has not been investigated. This study assessed the impact of cement removal using the Orthosonics System for Cemented Arthroplasty Revision (OSCAR; Orthosonics) on final cement-in-cement bonds.

Methods

A total of 24 specimens were manufactured by pouring cement (Simplex P Bone Cement; Stryker) into stainless steel moulds, with a central rod polished to Stryker Exeter V40 specifications. After cement curing, the rods were removed and eight specimens were allocated to each of three internal surface preparation groups: 1) burr; 2) OSCAR; and 3) no treatment. Internal holes were recemented, and each specimen was cut into 5 mm discs. Shear testing of discs was completed by a technician blinded to the original grouping, recording ultimate shear strengths. Scanning electron microscopy (SEM) was completed, inspecting surfaces of shear-tested specimens.


The Journal of Bone & Joint Surgery British Volume
Vol. 79-B, Issue 4 | Pages 583 - 589
1 Jul 1997
Kobayashi A Donnelly WJ Scott G Freeman MAR

We reviewed a consecutive series of 527 uninfected hip replacements in patients resident in the UK which had been implanted from 1981 to 1993. All had the same basic design of femoral prosthesis, but four fixation techniques had been used: two press-fit, one HA-coated and one cemented. Review and radiography were planned prospectively. For assessment the components were retrospectively placed into two groups: those which had failed from two years onwards by aseptic femoral loosening and those in which the femoral component had survived without revision or recommendation for revision. All available radiographs in both groups were measured to determine vertical migration and examined by two observers to agree the presence of radiolucent lines (RLLs), lytic lesions, resorption of the neck, proximal osteopenia and distal intramedullary and distal subperiosteal formation of new bone. We then related the presence or absence of these features and the rate of migration at two years to the outcome with regard to aseptic loosening and determined the predictive value of each of these variables. Migration of ≥2 mm at two years, the presence of an RLL of 2 mm occupying one-third of any one zone, and subperiosteal formation of new bone at the tip of the stem were predictors of aseptic loosening after two years. There were too few lytic lesions to assess at two years, but at five years a lytic lesion ≥2 mm also predicted failure. We discuss the use of these variables as predictors of femoral aseptic loosening for groups of hips and for individual hips. We conclude that if a group of about 50 total hip replacements, perhaps with a new design of femoral stem, were studied in this way at two years, a mean migration of < 0.4 mm and an incidence of < 10% of RLLs of 2 mm in any one zone would predict 95% survival at ten years. For an individual prosthesis, migration of < 2 mm and the absence of an RLL of ≤2 mm at two years predict a 6% chance of revision over approximately ten years. If either 2 mm of migration or an RLL of 2 mm is present, the chances of revision rise to 27%, and if both radiological signs are present they are 50%. If at five years a lytic lesion has developed, whatever the situation at two years, there is approximately a 50% chance of failure in the following five years. Our findings suggest that replacements using a limited number of any new design of femoral prosthesis should be screened radiologically at two years before they are generally introduced. We also suggest that radiographs of individual patients at two years and perhaps at five years should be studied to help to decide whether or not the patient should remain under close review or be discharged from specialist follow-up


The Bone & Joint Journal
Vol. 101-B, Issue 6_Supple_B | Pages 97 - 103
1 Jun 2019
Novikov D Mercuri JJ Schwarzkopf R Long WJ Bosco III JA Vigdorchik JM

Aims

Studying the indications for revision total hip arthroplasty (THA) may enable surgeons to change their practice during the initial procedure, thereby reducing the need for revision surgery. The aim of this study was to identify and describe the potentially avoidable indications for revision THA within five years of the initial procedure.

Patients and Methods

A retrospective review of 117 patients (73 women, 44 men; mean age 61.5 years (27 to 88)) who met the inclusion criteria was conducted. Three adult reconstruction surgeons independently reviewed the radiographs and medical records, and they classified the revision THAs into two categories: potentially avoidable and unavoidable. Baseline demographics, perioperative details, and quality outcomes up to the last follow-up were recorded.


The Bone & Joint Journal
Vol. 101-B, Issue 11 | Pages 1438 - 1446
1 Nov 2019
Kong X Chai W Chen J Yan C Shi L Wang Y

Aims

This study aimed to explore whether intraoperative nerve monitoring can identify risk factors and reduce the incidence of nerve injury in patients with high-riding developmental dysplasia.

Patients and Methods

We conducted a historical controlled study of patients with unilateral Crowe IV developmental dysplasia of the hip (DDH). Between October 2016 and October 2017, intraoperative nerve monitoring of the femoral and sciatic nerves was applied in total hip arthroplasty (THA). A neuromonitoring technician was employed to monitor nerve function and inform the surgeon of ongoing changes in a timely manner. Patients who did not have intraoperative nerve monitoring between September 2015 and October 2016 were selected as the control group. All the surgeries were performed by one surgeon. Demographics and clinical data were analyzed. A total of 35 patients in the monitoring group (ten male, 25 female; mean age 37.1 years (20 to 46)) and 56 patients in the control group (13 male, 43 female; mean age 37.9 years (23 to 52)) were enrolled. The mean follow-up of all patients was 13.1 months (10 to 15).


The Bone & Joint Journal
Vol. 100-B, Issue 10 | Pages 1303 - 1309
1 Oct 2018
Nodzo SR Chang C Carroll KM Barlow BT Banks SA Padgett DE Mayman DJ Jerabek SA

Aims

The aim of this study was to evaluate the accuracy of implant placement when using robotic assistance during total hip arthroplasty (THA).

Patients and Methods

A total of 20 patients underwent a planned THA using preoperative CT scans and robotic-assisted software. There were nine men and 11 women (n = 20 hips) with a mean age of 60.8 years (sd 6.0). Pelvic and femoral bone models were constructed by segmenting both preoperative and postoperative CT scan images. The preoperative anatomical landmarks using the robotic-assisted system were matched to the postoperative 3D reconstructions of the pelvis. Acetabular and femoral component positions as measured intraoperatively and postoperatively were evaluated and compared.


Bone & Joint 360
Vol. 7, Issue 6 | Pages 12 - 15
1 Dec 2018


The Bone & Joint Journal
Vol. 100-B, Issue 8 | Pages 1010 - 1017
1 Aug 2018
Jain S Magra M Dube B Veysi VT Whitwell GS Aderinto JB Emerton ME Stone MH Pandit HG

Aims

This study aimed to evaluate implant survival of reverse hybrid total hip arthroplasty (THA) at medium-term follow-up.

Patients and Methods

A consecutive series of 1082 THAs in 982 patients with mean follow-up of 7.9 years (5 to 11.3) is presented. Mean age was 69.2 years (21 to 94). Of these, 194 (17.9%) were in patients under 60 years, 663 (61.3%) in female patients and 348 (32.2%) performed by a trainee. Head size was 28 mm in 953 hips (88.1%) or 32 mm in 129 hips (11.9%). Survival analysis was performed and subgroups compared using log rank tests.


Bone & Joint 360
Vol. 8, Issue 4 | Pages 16 - 19
1 Aug 2019


Bone & Joint 360
Vol. 8, Issue 3 | Pages 11 - 13
1 Jun 2019


The Bone & Joint Journal
Vol. 101-B, Issue 7 | Pages 793 - 799
1 Jul 2019
Ugland TO Haugeberg G Svenningsen S Ugland SH Berg ØH Pripp AH Nordsletten L

Aims

The aim of this randomized trial was to compare the functional outcome of two different surgical approaches to the hip in patients with a femoral neck fracture treated with a hemiarthroplasty.

Patients and Methods

A total of 150 patients who were treated between February 2014 and July 2017 were included. Patients were allocated to undergo hemiarthroplasty using either an anterolateral or a direct lateral approach, and were followed for 12 months. The mean age of the patients was 81 years (69 to 90), and 109 were women (73%). Functional outcome measures, assessed by a physiotherapist blinded to allocation, and patient-reported outcome measures (PROMs) were collected postoperatively at three and 12 months.


The Bone & Joint Journal
Vol. 101-B, Issue 6_Supple_B | Pages 91 - 96
1 Jun 2019
Smith A Denehy K Ong KL Lau E Hagan D Malkani A

Aims

Cephalomedullary nails (CMNs) are commonly used for the treatment of intertrochanteric hip fractures. Total hip arthroplasty (THA) may be used as a salvage procedure when fixation fails in these patients. The aim of this study was to analyze the complications of THA following failed intertrochanteric hip fracture fixation using a CMN.

Patients and Methods

Patients who underwent THA were identified from the 5% subset of Medicare Parts A/B between 2002 and 2015. A subgroup involving those with an intertrochanteric fracture that was treated using a CMN during the previous five years was identified and compared with the remaining patients who underwent THA. The length of stay (LOS) was compared using both univariate and multivariate analysis. The incidence of infection, dislocation, revision, and re-admission was compared between the two groups, using multivariate analysis adjusted for demographic, hospital, and clinical factors.


The Bone & Joint Journal
Vol. 101-B, Issue 6_Supple_B | Pages 45 - 50
1 Jun 2019
Schloemann DT Edelstein AI Barrack RL

Aims

The aims of this study were to determine the change in pelvic sagittal alignment before, during, and after total hip arthroplasty (THA) undertaken with the patient in the lateral decubitus position, and to determine the impact of these changes on acetabular component position.

Patients and Methods

We retrospectively compared the radiological pelvic ratio among 91 patients undergoing THA. In total, 41 patients (46%) were female. The mean age was 61.6 years (sd 10.7) and the mean body mass index (BMI) was 20.0 kg/m2 (sd 5.5). Anteroposterior radiographs were obtained: in the standing position preoperatively and at six weeks postoperatively; in the lateral decubitus position after trial reduction intraoperatively; and in the supine position in the post-anaesthesia care unit. Pelvic ratio was defined as the ratio between the vertical distance from the inferior aspect of the sacroiliac (SI) joints to the superior pubic symphysis and the horizontal distance between the inferior aspect of the SI joints. Changes in the apparent component position based on changes in pelvic ratio were determined, with a change of > 5° considered clinically significant. Analyses were performed using Wilcoxon’s signed-rank test, with p < 0.05 considered significant.


The Bone & Joint Journal
Vol. 101-B, Issue 1_Supple_A | Pages 11 - 18
1 Jan 2019
Kayani B Konan S Thakrar RR Huq SS Haddad FS

Objectives

The primary objective of this study was to compare accuracy in restoring the native centre of hip rotation in patients undergoing conventional manual total hip arthroplasty (THA) versus robotic-arm assisted THA. Secondary objectives were to determine differences between these treatment techniques for THA in achieving the planned combined offset, component inclination, component version, and leg-length correction.

Materials and Methods

This prospective cohort study included 50 patients undergoing conventional manual THA and 25 patients receiving robotic-arm assisted THA. Patients undergoing conventional manual THA and robotic-arm assisted THA were well matched for age (mean age, 69.4 years (sd 5.2) vs 67.5 years (sd 5.8) (p = 0.25); body mass index (27.4 kg/m2 (sd 2.1) vs 26.9 kg/m2 (sd 2.2); p = 0.39); and laterality of surgery (right = 28, left = 22 vs right = 12, left = 13; p = 0.78). All operative procedures were undertaken by a single surgeon using the posterior approach. Two independent blinded observers recorded all radiological outcomes of interest using plain radiographs.


The Bone & Joint Journal
Vol. 101-B, Issue 6_Supple_B | Pages 31 - 36
1 Jun 2019
Nam D Nunley RM Clohisy JC Lombardi AV Berend KR Barrack RL

Aims

Whether patient-reported pain differs among surgical approaches in total hip arthroplasty (THA) remains unclear. This study’s purposes were to determine differences in pain based on surgical approach (direct anterior (DA) vs posterolateral (PL)) and PL approach incision length.

Patients and Methods

This was a retrospective investigation from two centres and seven surgeons (three DA, three PL, one both) of primary THAs. PL patients were categorized for incision length (6 cm to 8 cm, 8 cm to 12 cm, 12 cm to 15 cm). All patients had cementless femoral and acetabular fixation, at least one year’s follow-up, and well-fixed components. Patients completed a pain-drawing questionnaire identifying the location and intensity of pain on an anatomical diagram. Power analysis indicated 800 patients in each cohort for adequate power to detect a 4% difference in pain (alpha = 0.05, beta = 0.80).


The Bone & Joint Journal
Vol. 101-B, Issue 7 | Pages 768 - 778
1 Jul 2019
Galea VP Rojanasopondist P Ingelsrud LH Rubash HE Bragdon C Huddleston III JI Malchau H Troelsen A

Aims

The primary aim of this study was to quantify the improvement in patient-reported outcome measures (PROMs) following total hip arthroplasty (THA), as well as the extent of any deterioration through the seven-year follow-up. The secondary aim was to identify predictors of PROM improvement and deterioration.

Patients and Methods

A total of 976 patients were enrolled into a prospective, international, multicentre study. Patients completed a battery of PROMs prior to THA, at three months post-THA, and at one, three, five, and seven-years post-THA. The Harris Hip Score (HHS), the 36-Item Short-Form Health Survey (SF-36) Physical Component Summary (PCS), the SF-36 Mental Component Summary (MCS), and the EuroQol five-dimension three-level (EQ-5D) index were the primary outcomes. Longitudinal changes in each PROM were investigated by piece-wise linear mixed effects models. Clinically significant deterioration was defined for each patient as a decrease of one half of a standard deviation (group baseline).


The Bone & Joint Journal
Vol. 101-B, Issue 6 | Pages 646 - 651
1 Jun 2019
Aggarwal VK Elbuluk A Dundon J Herrero C Hernandez C Vigdorchik JM Schwarzkopf R Iorio R Long WJ

Aims

A variety of surgical approaches are used for total hip arthroplasty (THA), all with reported advantages and disadvantages. A number of common complications can occur following THA regardless of the approach used. The purpose of this study was to compare five commonly used surgical approaches with respect to the incidence of surgery-related complications.

Patients and Methods

The electronic medical records of all patients who underwent primary elective THA at a single large-volume arthroplasty centre, between 2011 and 2016, with at least two years of follow-up, were reviewed. After exclusion, 3574 consecutive patients were included in the study. There were 1571 men (44.0%) and 2003 women (56.0%). Their mean age and body mass index (BMI) was 63.0 years (sd 11.8) and 29.1 kg/m2 (sd 6.1), respectively. Data gathered included the age of the patient, BMI, the American Society of Anesthesiologists (ASA) score, estimated blood loss (EBL), length of stay (LOS), operating time, the presence of intra- or postoperative complications, type of complication, and the surgical approach. The approaches used during the study were posterior, anterior, direct lateral, anterolateral, and the northern approach. The complications that were recorded included prolonged wound drainage without infection, superficial infection, deep infection, dislocation, aseptic loosening, and periprosthetic fracture. Finally, the need for re-operation was recorded. Means were compared using analysis of variance (ANOVA) and Student’s t-tests where appropriate and proportions were compared using the chi-squared test.


The Bone & Joint Journal
Vol. 100-B, Issue 10 | Pages 1310 - 1319
1 Oct 2018
Langton DJ Wells SR Joyce TJ Bowsher JG Deehan D Green S Nargol AVF Holland JP

Aims

There are limited published data detailing the volumetric material loss from tapers of conventional metal-on-polyethylene (MoP) total hip arthroplasties (THAs). Our aim was to address this by comparing the taper wear rates measured in an explanted cohort of the widely used Exeter THA with those measured in a group of metal-on-metal (MoM) THAs.

Patients and Methods

We examined an existing retrieval database to identify all Exeter V40 and Universal MoP THAs. Volumetric wear analysis of the taper surfaces was conducted using previously validated methodology. These values were compared with those obtained from a series of MoM THAs using non-parametric statistical methodology. A number of patient and device variables were accounted for using multiple regression modelling.


The Bone & Joint Journal
Vol. 100-B, Issue 12 | Pages 1559 - 1564
1 Dec 2018
McConnell JS Saunders PRJ Young SK

Aims

Cementless femoral stems must be correctly sized and well-seated to obtain satisfactory biological fixation. The change in sound that occurs during impaction of the femoral broach is said to indicate good fit, but this has not been widely studied. We set out to find whether the presence or absence of these sound changes could predict correct sizing.

Patients and Methods

We recorded the sound generated during femoral broaching for 105 cementless total hip arthroplasties using the Corail stem. Four cases were excluded, leaving 101 recordings for analysis. There were 36 male patients and 65 female patients, with a mean age of 69.9 years (sd 12.3) and median body mass index (BMI) of 29 kg/m2 (interquartile range (IQR) 26 to 32). The recordings were analyzed to identify the frequencies of the sounds produced during impaction of the femoral broach.