1 Mar 2011
Matthies A Underwood R Cann P Ilo K Nawaz Z Skinner J Hart AJ

This study compared component wear rates and pre-revision blood metal ions levels in two groups of failed metal-on-metal hip arthroplasties: hip resurfacing and modular total hip replacement (THR).

There was no significant difference in the median rate of linear wear between the groups for both acetabular (p = 0.4633) and femoral (p = 0.0872) components. There was also no significant difference in the median linear wear rates when failed hip resurfacing and modular THR hips of the same type (ASR and Birmingham hip resurfacing (BHR)) were compared.

Unlike other studies of well-functioning hips, there was no significant difference in pre-revision blood metal ion levels between hip resurfacing and modular THR.

Edge loading was common in both groups, but more common in the resurfacing group (67%) than in the modular group (57%). However, this was not significant (p = 0.3479). We attribute this difference to retention of the neck in resurfacing of the hip, leading to impingement-type edge loading. This was supported by visual evidence of impingement on the femur.

These findings show that failed metal-on-metal hip resurfacing and modular THRs have similar component wear rates and are both associated with raised pre-revision blood levels of metal ions.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 8 | Pages 1126 - 1134
1 Aug 2012
Granchi D Cenni E Giunti A Baldini N

We report a systematic review and meta-analysis of the peer-reviewed literature focusing on metal sensitivity testing in patients undergoing total joint replacement (TJR). Our purpose was to assess the risk of developing metal hypersensitivity post-operatively and its relationship with outcome and to investigate the advantages of performing hypersensitivity testing.

We undertook a comprehensive search of the citations quoted in PubMed and EMBASE: 22 articles (comprising 3634 patients) met the inclusion criteria. The frequency of positive tests increased after TJR, especially in patients with implant failure or a metal-on-metal coupling. The probability of developing a metal allergy was higher post-operatively (odds ratio (OR) 1.52 (95% confidence interval (CI) 1.06 to 2.31)), and the risk was further increased when failed implants were compared with stable TJRs (OR 2.76 (95% CI 1.14 to 6.70)).

Hypersensitivity testing was not able to discriminate between stable and failed TJRs, as its predictive value was not statistically proven. However, it is generally thought that hypersensitivity testing should be performed in patients with a history of metal allergy and in failed TJRs, especially with metal-on-metal implants and when the cause of the loosening is doubtful.


Bone & Joint 360
Vol. 4, Issue 2 | Pages 2 - 6
1 Apr 2015
Lever CJ Robinson AHN

Ankle replacements have improved significantly since the first reported attempt at resurfacing of the talar dome in 1962. We are now at a stage where ankle replacement offers a viable option in the treatment of end-stage ankle arthritis. As the procedure becomes more successful, it is important to reflect and review the current surgical outcomes. This allows us to guide our patients in the treatment of end-stage ankle arthritis. What is the better surgical treatment – arthrodesis or replacement?


The Bone & Joint Journal
Vol. 95-B, Issue 11 | Pages 1464 - 1473
1 Nov 2013
Vendittoli P Rivière C Roy AG Barry J Lusignan D Lavigne M

A total of 219 hips in 192 patients aged between 18 and 65 years were randomised to 28-mm metal-on-metal uncemented total hip replacements (THRs, 107 hips) or hybrid hip resurfacing (HR, 112 hips). At a mean follow-up of eight years (6.6 to 9.3) there was no significant difference between the THR and HR groups regarding rate of revision (4.0% (4 of 99) vs 5.8% (6 of 104), p = 0.569) or re-operation rates without revision (5.1% (5 of 99) vs 2.9% (3 of 104), p = 0.428). In the THR group one recurrent dislocation, two late deep infections and one peri-prosthetic fracture required revision, whereas in the HR group five patients underwent revision for femoral head loosening and one for adverse reaction to metal debris. The mean University of California, Los Angeles activity scores were significantly higher in HR (7.5 (sd 1.7) vs 6.9 (sd 1.7), p = 0.035), but similar mean Western Ontario and McMaster Universities Osteoarthritis Index scores were obtained (5.8 (sd 9.5) in HR vs 5.1 (sd 8.9) in THR, p = 0.615) at the last follow-up. Osteolysis was found in 30 of 81 THR patients (37.4%), mostly in the proximal femur, compared with two of 83 HR patients (2.4%) (p < 0.001). At five years the mean metal ion levels were < 2.5 μg/l for cobalt and chromium in both groups; only titanium was significantly higher in the HR group (p = 0.001). Although revision rates and functional scores were similar in both groups at mid-term, long-term survival analysis is necessary to determine whether one procedure is more advantageous than the other.

Cite this article: Bone Joint J 2013;95-B:1464–73.


The Bone & Joint Journal
Vol. 97-B, Issue 4 | Pages 463 - 472
1 Apr 2015
Panagiotidou A Meswania J Osman K Bolland B Latham J Skinner J Haddad FS Hart A Blunn G

The aim of this study was to assess the effect of frictional torque and bending moment on fretting corrosion at the taper interface of a modular femoral component and to investigate whether different combinations of material also had an effect. The combinations we examined were 1) cobalt–chromium (CoCr) heads on CoCr stems 2) CoCr heads on titanium alloy (Ti) stems and 3) ceramic heads on CoCr stems.

In test 1 increasing torque was imposed by offsetting the stem in the anteroposterior plane in increments of 0 mm, 4 mm, 6 mm and 8 mm when the torque generated was equivalent to 0 Nm, 9 Nm, 14 Nm and 18 Nm.

In test 2 we investigated the effect of increasing the bending moment by offsetting the application of axial load from the midline in the mediolateral plane. Increments of offset equivalent to head + 0 mm, head + 7 mm and head + 14 mm were used.

Significantly higher currents and amplitudes were seen with increasing torque for all combinations of material. However, Ti stems showed the highest corrosion currents. Increased bending moments associated with using larger offset heads produced more corrosion: Ti stems generally performed worse than CoCr stems. Using ceramic heads did not prevent corrosion, but reduced it significantly in all loading configurations.

Cite this article: Bone Joint J 2015;97-B:463–72.


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 11 | Pages 1501 - 1508
1 Nov 2010
Donell ST Darrah C Nolan JF Wimhurst J Toms A Barker THW Case CP Tucker JK

Metal-on-metal total hip replacement has been targeted at younger patients with anticipated long-term survival, but the effect of the production of metal ions is a concern because of their possible toxicity to cells. We have reviewed the results of the use of the Ultima hybrid metal-on-metal total hip replacement, with a cemented polished tapered femoral component with a 28 mm diameter and a cobalt-chrome (CoCr) modular head, articulating with a 28 mm CoCr acetabular bearing surface secured in a titanium alloy uncemented shell.

Between 1997 and 2004, 545 patients with 652 affected hips underwent replacement using this system. Up to 31 January 2008, 90 (13.8%) hips in 82 patients had been revised. Pain was the sole reason for revision in 44 hips (48.9%) of which 35 had normal plain radiographs. Peri-prosthetic fractures occurred in 17 hips (18.9%) with early dislocation in three (3.3%) and late dislocation in 16 (17.8%). Infection was found in nine hips (10.0%).

At operation, a range of changes was noted including cavities containing cloudy fluid under pressure, necrotic soft tissues with avulsed tendons and denuded osteonecrotic upper femora. Corrosion was frequently observed on the retrieved cemented part of the femoral component. Typically, the peri-operative findings confirmed those found on pre-operative metal artefact reduction sequence MRI and histological examination showed severe necrosis.

Metal artefact reduction sequence MRI proved to be useful when investigating these patients with pain in the absence of adverse plain radiological features.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 11_Supple_A | Pages 78 - 81
1 Nov 2012
Benjamin-Laing H Haddad FS

In this paper, we will consider the current role of metal-on-metal bearings by looking at three subtypes of MoM hip arthroplasty separately: Hip resurfacing, large head (> 36 mm) MoM THA and MoM THA with traditional femoral head sizes.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 11_Supple_A | Pages 157 - 166
1 Nov 2012
Rosenberg AG Berend ME Berry DJ Della Valle CJ MacDonald SJ Minas T

This conversation represents an attempt by several arthroplasty surgeons to critique several abstracts presented over the last year as well as to use them as a jumping off point for trying to figure out where they fit in into our current understanding of multiple issues in modern hip and knee arthroplasty.


The Bone & Joint Journal
Vol. 96-B, Issue 11_Supple_A | Pages 66 - 69
1 Nov 2014
Lombardi Jr AV Berend KR Adams JB

The common recommended treatment for infected total hip replacement is two-staged exchange including removal of all components. However, removal of well-fixed femoral stems can result in structural bone damage. We recently reported on an alternative treatment of partial two-stage exchange used in selected cases, in which a well-fixed femoral stem was left and only the acetabular component removed, the joint space was debrided thoroughly, an antibiotic-laden polymethylmethacrylate spacer was moulded using a bulb-type syringe and placed in the acetabulum, intravenous antibiotics were administered during the interval, and delayed re-implantation was performed. In 19 patients treated with this technique from January 2000 to January 2011, 89% were free of infection at a mean follow-up of four years (2 to 11). Since then, disposable silicone moulds have become available to fabricate spacers in separate femoral and head units. The head spacer mould, which incorporates various neck taper adapter options, greatly facilitates the technique of partial two-stage exchange. We report our early experience using disposable silicone head spacer moulds for partial two-stage exchange in seven patients with infected primary hip replacements.

Cite this article: Bone Joint J 2014;96-B(11 Suppl A):66–9


The Bone & Joint Journal
Vol. 96-B, Issue 11_Supple_A | Pages 87 - 92
1 Nov 2014
Kwong LM Nielsen ESN Ruiz DR Hsu AH Dines MD Mellano CM

A retrospective review was performed of patients undergoing primary cementless total knee replacement (TKR) using porous tantalum performed by a group of surgical trainees. Clinical and radiological follow-up involved 79 females and 26 males encompassing 115 knees. The mean age was 66.9 years (36 to 85). Mean follow-up was 7 years (2 to 11). Tibial and patellar components were porous tantalum monoblock implants, and femoral components were posterior stabilised (PS) in design with cobalt–chromium fibre mesh. Radiological assessments were made for implant positioning, alignment, radiolucencies, lysis, and loosening. There was 95.7% survival of implants. There was no radiological evidence of loosening and no osteolysis found. No revisions were performed for aseptic loosening. Average tibial component alignment was 1.4° of varus (4°of valgus to 9° varus), and 6.2° (3° anterior to 15° posterior) of posterior slope. Mean femoral component alignment was 6.6° (1° to 11°) of valgus. Mean tibiofemoral alignment was 5.6° of valgus (7° varus to 16° valgus). Patellar tilt was a mean of 2.4° lateral (5° medial to 28° lateral). Patient satisfaction with improvement in pain was 91%. Cementless TKR incorporating porous tantalum yielded good clinical and radiological outcomes at a mean of follow-up of seven-years.

Cite this article: Bone Joint J 2014;96-B(11 Suppl A):87–92.


Bone & Joint 360
Vol. 1, Issue 4 | Pages 12 - 15
1 Aug 2012

The August 2012 Knee Roundup360 looks at: meniscal defects and a polyurethane scaffold; which is best between a single or double bundle; OA of the knee; how to resolve anterior knee pain; whether yoga can be bad for your menisci; metal ions in the serum; whether ACI is any good; the ACL; whether hyaluronic acid delays collagen degradation; and hyaluronan and patellar tendinopathy.


The Bone & Joint Journal
Vol. 96-B, Issue 11 | Pages 1449 - 1454
1 Nov 2014
Imbuldeniya AM Walter WL Zicat BA Walter WK

We describe the clinical and radiological results of cementless primary total hip replacement (THR) in 25 patients (18 women and seven men; 30 THRs) with severe developmental dysplasia of the hip (DDH). Their mean age at surgery was 47 years (23 to 89). In all, 21 hips had Crowe type III dysplasia and nine had Crowe type IV. Cementless acetabular components with standard polyethylene liners were introduced as close to the level of the true acetabulum as possible. The modular cementless S-ROM femoral component was used with a low resection of the femoral neck.

A total of 21 patients (25 THRs) were available for review at a mean follow-up of 18.7 years (15.8 to 21.8). The mean modified Harris hip score improved from 46 points pre-operatively to 90 at final follow up (p < 0.001).

A total of 15 patients (17 THRs; 57%) underwent revision of the acetabular component at a mean of 14.6 years (7 to 20.8), all for osteolysis. Two patients (two THRs) had symptomatic loosening. No patient underwent femoral revision. Survival with revision of either component for any indication was 81% at 15 years (95% CI 60.1 to 92.3), with 21 patients at risk.

This technique may reduce the need for femoral osteotomy in severe DDH, while providing a good long-term functional result.

Cite this article: Bone Joint J 2014;96-B:1449–54.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 11 | Pages 1475 - 1481
1 Nov 2012
Berton C Puskas GJ Christofilopoulos P Stern R Hoffmeyer P Lübbeke A

There are no recent studies comparing cable with wire for the fixation of osteotomies or fractures in total hip replacement (THR). Our objective was to evaluate the five-year clinical and radiological outcomes and complication rates of the two techniques. We undertook a review including all primary and revision THRs performed in one hospital between 1996 and 2005 using cable or wire fixation. Clinical and radiological evaluation was performed five years post-operatively. Cables were used in 51 THRs and wires in 126, and of these, 36 THRs with cable (71%) and 101 with wire (80%) were evaluated at follow-up. The five-year radiographs available for 33 cable and 91 wire THRs revealed rates of breakage of fixation of 12 of 33 (36%) and 42 of 91 (46%), respectively. With cable there was a significantly higher risk of metal debris (68% vs 9%; adjusted relative risk (RR) 6.6; 95% confidence interval (CI) 3.0 to 14.1), nonunion (36% vs 21%; adjusted RR 2.0; 95% CI 1.0 to 3.9) and osteolysis around the material, acetabulum or femur (61% vs 19%; adjusted RR 3.9; 95% CI 2.3 to 6.5). Cable breakage increased the risk of osteolysis to 83%. There was a trend towards foreign-body reaction and increased infection with cables. Clinical results did not differ between the groups.

In conclusion, we found a higher incidence of complications and a trend towards increased infection and foreign-body reaction with the use of cables.


The Bone & Joint Journal
Vol. 96-B, Issue 10 | Pages 1349 - 1354
1 Oct 2014
Conway J Mansour J Kotze K Specht S Shabtai L

The treatment of infected nonunions is difficult. Antibiotic cement-coated (ACC) rods provide stability as well as delivering antibiotics. We conducted a review of 110 infected nonunions treated with ACC rods. Patients were divided into two groups: group A (67 patients) with an infected arthrodesis, and group B (43 patients) with an infected nonunion in a long bone. In group A, infected arthrodesis, the success rate after the first procedure was 38/67 (57%), 29/67 (43%) required further surgery for either control of infection or non-union. At last follow-up, five patients required amputation, representing a limb salvage rate of 62/67 (93%) overall. In all, 29/67 (43%) presented with a bone defect with a mean size of 6.78 cm (2 to 25). Of those with a bone defect, 13/29 (45%) required further surgery and had a mean size of defect of 7.2 cm (3.5 to 25). The cultures were negative in 17/67 (26%) and the most common organism cultured was methicillin-resistant staphylococcus aureus (MRSA) (23/67, (35%)). In group B, long bones nonunion, the success rate after the first procedure was 26/43 (60%), 17/43 (40%) required further surgery for either control of infection or nonunion. The limb salvage rate at last follow-up was 43/43 (100%). A total of 22/43 (51%) had bone defect with a mean size of 4.7 cm (1.5 to 11.5). Of those patients with a bone defect, 93% required further surgery with a mean size of defect of 5.4 cm (3 to 8.5). The cultures were negative in 10/43 (24%) and the most common organism cultured was MRSA, 15/43 (35%). ACC rods are an effective form of treatment for an infected nonunion, with an acceptable rate of complications.

Cite this article: Bone Joint J 2014; 96-B:1349–54


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 11 | Pages 1545 - 1546
1 Nov 2006
DANIEL J ZIAEE H PRADHAN C McMINN DJW


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 2 | Pages 281 - 286
1 Feb 2012
Chana R Esposito C Campbell PA Walter WK Walter WL

We report a case of a male patient presenting with bilateral painful but apparently well-positioned and -fixed large-diameter metal-on-metal hip replacements four years post-operatively. Multiple imaging modes revealed a thick-walled, cystic expansile mass in communication with the hip joint (a pseudotumour). Implant retrieval analysis and tissue culture eliminated high bearing wear or infection as causes for the soft-tissue reaction, but noted marked corrosion of the modular neck taper adaptor and corrosion products in the tissues. Therefore, we believe corrosion products from the taper caused by mismatch of the implant components led to pseudotumour formation requiring revision.


The Bone & Joint Journal
Vol. 95-B, Issue 11 | Pages 1582 - 1582
1 Nov 2013
Haddad FS

Kazi HA, Perera JR, Gillott E, Carroll FA, Briggs TW. A prospective study of a ceramic-on-metal bearing in total hip arthroplasty: Clinical results, metal ion levels and chromosome analysis at two years. Bone Joint J 2013;95-B:1040-1044.


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 3 | Pages 402 - 404
1 Mar 2006
Boardman DR Middleton FR Kavanagh TG

As metal-on-metal arthroplasty becomes more widespread, concerns are being raised about the potential dangers of metal particulate debris. We present the case of a benign psoas mass secondary to the presence of such particles. The mass was excised and the hip resurfacing subsequently revised to a total hip replacement.


The Bone & Joint Journal
Vol. 96-B, Issue 10 | Pages 1359 - 1365
1 Oct 2014
Large R Tambe A Cresswell T Espag M Clark DI

Medium-term results of the Discovery elbow replacement are presented. We reviewed 51 consecutive primary Discovery total elbow replacements (TERs) implanted in 48 patients. The mean age of the patients was 69.2 years (49 to 92), there were 19 males and 32 females (37%:63%) The mean follow-up was 40.6 months (24 to 69). A total of six patients were lost to follow-up. Statistically significant improvements in range movement and Oxford Elbow Score were found (p < 0.001). Radiolucent lines were much more common in, and aseptic loosening was exclusive to, the humeral component. Kaplan–Meier survivorship at five years was 92.2% (95% CI 74.5% to 96.4%) for aseptic loosening. In four TERs, periprosthetic infection occurred resulting in failure. A statistically significant association between infection and increased BMI was found (p = 0.0268). Triceps failure was more frequent after the Mayo surgical approach and TER performed after previous trauma surgery. No failures of the implant were noted.

Our comparison shows that the Discovery has early clinical results that are similar to other semi-constrained TERs. We found continued radiological surveillance with particular focus on humeral lucency is warranted and has not previously been reported. Despite advances in the design of total elbow replacement prostheses, rates of complication remain high.

Cite this article: Bone Joint J 2014;96-B:1359–65


The Bone & Joint Journal
Vol. 96-B, Issue 6 | Pages 730 - 736
1 Jun 2014
Imbuldeniya AM Walter WK Zicat BA Walter WL

We report on 397 consecutive revision total hip replacements in 371 patients with a mean clinical and radiological follow-up of 12.9 years (10 to 17.7). The mean age at surgery was 69 years (37 to 93). A total of 28 patients (8%) underwent further revision, including 16 (4%) femoral components. In all 223 patients (56%, 233 hips) died without further revision and 20 patients (5%, 20 hips) were lost to follow-up. Of the remaining patients, 209 (221 hips) were available for clinical assessment and 194 (205 hips) for radiological review at mean follow-up of 12.9 years (10 to 17.7).

The mean Harris Hip Score improved from 58.7 (11 to 92) points to 80.7 (21 to 100) (p <  0.001) and the mean Merle d’Aubigné and Postel hip scores at final follow-up were 4.9 (2 to 6), 4.5 (2 to 6) and 4.3 (2 to 6), respectively for pain, mobility and function. Radiographs showed no lucencies around 186 (90.7%) femoral stems with stable bony ingrowth seen in 199 stems (97%). The survival of the S-ROM femoral stem at 15 years with revision for any reason as the endpoint was 90.5% (95% confidence interval (CI) 85.7 to 93.8) and with revision for aseptic loosening as the endpoint 99.3% (95% CI 97.2 to 99.8).

We have shown excellent long-term survivorship and good clinical outcome of a cementless hydroxyapatite proximally-coated modular femoral stem in revision hip surgery.

Cite this article: Bone Joint J 2014;96-B:730–6.